PINPOINT Endoscopic Fluorescence Imaging System

K150956 · Novadaq Technologies, Inc. · GCJ · May 23, 2016 · Gastroenterology, Urology

Device Facts

Record IDK150956
Device NamePINPOINT Endoscopic Fluorescence Imaging System
ApplicantNovadaq Technologies, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateMay 23, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K150956 · May 23, 2016PINPOINT Endoscopic Fluorescence Imaging SystemNovadaq Technologies, Inc.Published clinical literature (Journal of Gastrointestinal Surgery, Surgical Innovation, Int'l J Surg Case Reports); Post-market surveillance clinical adverse event informationThe sponsor conducted a clinical evaluation using literature reviews and post-market surveillance data to demonstrate the continued safety and performance of the PINPOINT System and to support expanded indications for use.Literature review; Post-market surveillance; Case reports; Clinical evaluation report

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Clinical literature search and post-market surveillance evaluation; Literature review and post-market surveillance analysisPatients undergoing minimally invasive surgery using the PINPOINT SystemNot applicable for this studyClinical safety and performance

Indications for Use

The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Device Story

PINPOINT Endoscopic Fluorescence Imaging System (PC9000) provides real-time visible and near-infrared (NIR) fluorescence imaging during minimally invasive surgery. System components include endoscopic video processor/illuminator (VPI), surgical laparoscopes (VIS and NIR optimized), camera head, and light guide cable. Operation involves illuminating surgical site with visible light for standard imaging; for NIR fluorescence, patient is injected with indocyanine green (ICG) imaging agent. VPI provides NIR excitation light; camera captures fluorescence response; processed images displayed on HD video monitor. Used in hospital operating rooms by surgeons. Output assists in visual assessment of vessels, blood flow, tissue perfusion, and extra-hepatic bile ducts. Complements standard white light and intraoperative cholangiography. Benefits include enhanced intraoperative visualization of tissue perfusion and biliary anatomy.

Clinical Evidence

Clinical evaluation based on literature search and post-market surveillance. Evidence includes published clinical studies and case reports in journals (e.g., Journal of Gastrointestinal Surgery, Surgical Innovation). Porcine animal model testing validated in vivo fluorescence imaging capability. No new safety or effectiveness concerns identified.

Technological Characteristics

System includes VPI, laparoscopes (models SC9100, SC9104, SC9130, SC9131, SC9134, SC9144), camera head, and light guide. Conforms to IEC 60601-1:2012 (safety), IEC 60601-1-2:2007 (EMC), IEC 60601-2-18:2009 (endoscopic equipment), and IEC 60825:2007 (Class 3R laser device). Imaging relies on ICG fluorescence under NIR excitation.

Indications for Use

Indicated for patients undergoing minimally invasive surgery requiring real-time endoscopic visible and near-infrared fluorescence imaging for visual assessment of vessels, blood flow, tissue perfusion, and extra-hepatic bile ducts. Not for standalone biliary duct visualization.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal are three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 23, 2016 Novadaq Technologies Inc. Ms. Jen Pendlebury Director of Regulatory Affairs 8329 Eastlake Drive Unit 101 Burnaby, British Columbia Canada V5A 4W2 Re: K150956 Trade/Device Name: Pinpoint Endoscopic Fluorescence Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, IZI Dated: May 13, 2016 Received: May 16, 2016 Dear Ms. Pendlebury: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Jennifer R. Stevenson -A For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150956 #### Device Name PINPOINT Endoscopic Fluorescence Imaging System (PC9000) Indications for Use (Describe) The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 5 - 510(k) Summary This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR § 807.92. | Trade Name: | PINPOINT Endoscopic Fluorescence Imaging System | |-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Model Number: | PC9000 | | Common Name: | Endoscope Video Imaging System | | Classification: | 21 CFR § 876.1500 | | Classification Name: | Laparoscope, General & Plastic Surgery; Angiographic X-ray<br>System | | Product Code: | GCJ; IZI | | Classification: | Class II | | Manufacturer: | Novadaq Technologies Inc.<br>8329 Eastlake Drive, Unit 101<br>Burnaby, British Columbia<br>Canada V5A 4W2 | | Contact Name: | Jen Pendlebury<br>Director of Regulatory Affairs<br>Tel: (905) 629-3822 ext. 205<br>Fax: (905) 249-0656 | | Date 510(k) Summary Prepared: | May 12, 2016 | | Predicate Devices: | SPY Scope Intra-Operative Imaging System/ PINPOINT<br>(K091515)<br>(Novadaq Technologies Inc.)<br><br>Da Vinci Firefly Imaging System (K141077) (Intuitive Surgical<br>Inc.)<br><br>Stryker Infrared Light Source and SafeLight Cable (K142310)<br>(Stryker Endoscopy) | # Device Description: The PINPOINT Endoscopic Fluorescence Imaging System (PINPOINT, PINPOINT System) is comprised of an endoscopic video processor/ illuminator (VPI) which is capable of providing visible and near-infrared illumination to a surgical laparoscope, surgical laparoscopes optimized for visible (VIS) and near-infrared (NIR) illumination and imaging, a camera head that is also optimized for visible and near-infrared imaging, and a flexible light guide cable. The following {4}------------------------------------------------ laparoscope models are included as part of the PINPOINT System: SC9100, SC9104, SC9130, SC9131, SC9134 and SC9144. These are the major components of the PINPOINT System. During surgical procedures, PINPOINT may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the illuminator and the resulting reflected light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein. To provide NIR fluorescence imaging, PINPOINT is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the VPI, and the fluorescence response is then imaged with the camera, processed and displayed on an HD video monitor. When used with a VIS/NIR laparoscope, PINPOINT can operate to provide illumination and imaging of both visible light and ICG fluorescence images simultaneously. # Proposed Indications for Use of the PINPOINT System: The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. # Summary of Technological Characteristics of the PINPOINT System and Predicate Devices Based on the technological characteristics and fundamental scientific premise, the PINPOINT System (PC9000) has been determined to be substantially equivalent to the predicate devices. the SPY Intra-Operative Imaging System/PINPOINT (K091515), as well as the da Vinci Firefly Imaging System (K141077) and the Stryker IRF Light Source and SafeLight Cable (K142310). PINPOINT is substantially equivalent to the SPY Scope Intra-Operative Imaging System (SC8000), as both systems use the same technology and mode of imaging to provide real-time endoscopic visible and NIR fluorescence imaging during minimally invasive surgical procedures. PINPOINT is also substantially equivalent to the da Vinci Firefly Imaging System (K141077) and to the Stryker IRF Light Source and SafeLight Cable (K142310). Both systems and PINPOINT utilize very similar technology and rely on the same scientific premise – they are endoscopic video imaging systems which are used for high definition visible light and NIR fluorescence imaging during minimally invasive surgery. All of these systems utilize the same mode of imaging – visible and near infrared fluorescence imaging, with ICG as the imaging agent, used in the hospital operating room. All of these systems have the same integral components – a light source console and a light cable for outputting light, a camera control unit for processing NIR and VIS light images, a coupler attached to the laparoscope and a camera head, and a laparoscope for VIS and NIR light illumination and imaging. {5}------------------------------------------------ #### Performance Data: ### Non-Clinical Performance Testing of the PINPOINT System The PINPOINT System was designed and developed by Novadag Technologies Inc. (NOVADAQ), in accordance with the applicable requirements and standards to establish performance and safety of the device. Device safety and performance were verified by tests conducted by NOVADAQ and accredited third party laboratories. The PINPOINT System was tested in accordance with IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance. IEC 60601-1-2:2007 Medical Electrical Equipment – Part 1-2: General requirements for safety – Collateral standard: Electromagnetic compatibility – Requirements and testing was also conducted on the PINPOINT System and test results showed that PINPOINT conforms to the applicable requirements. PINPOINT was also tested in accordance with IEC 60601-2-18:2009 Medical Electrical Equipment – Part 2: Particular requirements for the basic safety and essential performance of endoscopic equipment. and has demonstrated conformance to the standard. Conformance of PINPOINT with IEC 60825:2007 Safety of laser products - Part 1: Equipment classification and requirements was assessed by Underwriters Laboratories Inc. (UL) and showed that PINPOINT is a Class 3R laser device with internal maximum Class 4 laser radiation. Animal testing using a porcine model validated the in vivo fluorescence imaging capability of the PINPOINT System. #### Clinical Performance Data A clinical evaluation based on a clinical literature search and reported clinical adverse event information from post-market surveillance was considered appropriate to evaluate the performance and safety of the PINPOINT System. It was concluded to objectively verify the continued clinical safety and performance of NOVADAQ'S Fluorescence Imaging devices, including the PINPOINT System. Data presented in the clinical evaluation report provides evidence to support the safety and performance claims for the PINPOINT System. Additionally, results from a clinical study and case reports utilizing the PINPOINT System and published in the Journal of Gastrointestinal Surgerv and in Surgical Innovation and Int! J Sur Case Reports, respectively, together with findings from the most recent literature search are enclosed to support the expanded Indications for Use of PINPOINT (see section 20, Performance Testing - Clinical, for the summary of the relevant clinical literature reports). Findings demonstrated that no new concerns related to the safety and effectiveness of the device have been introduced. #### Conclusions It has been demonstrated in this 510(k) submission that the PINPOINT System, with the expanded Indications for Use, is substantially equivalent to the predicate devices discussed herein and there are no issues related to its safety or effectiveness. The additional performance data and scientific literature from a recent literature search supports the expanded Indications of the device, and further validates the well-established safety profile of NOVADAQ'S imaging systems.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...