K150225 · Neometrics, Inc. · OCY · Mar 16, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K150225
Device Name
NovaGold High Performance Guidewire
Applicant
Neometrics, Inc.
Product Code
OCY · Gastroenterology, Urology
Decision Date
Mar 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
The NovaGold Guidewire is intended for use in selective cannulation of the billary ducts including the common bile, pancreatic, cystic, right and left hepatic ducts, and to aid in the placement of diagnostic devices during endoscopic procedures.
Device Story
NovaGold High Performance Guidewire is a steerable, metallic guidewire used during endoscopic procedures. Device consists of a nitinol alloy core with a PTFE polymer shaft coating and a hydrophilic coating on the distal portion; features a radiopaque, floppy tip. Operated manually by a physician to navigate biliary and pancreatic ducts; facilitates placement of diagnostic and therapeutic devices. Provides mechanical guidance within the ductal system. No electronic components or software.
Clinical Evidence
Bench testing only. Performance verified through fracture resistance, flex resistance, tensile strength, torqueability, torque strength, tip flexibility, and distal fatigue resistance testing.
Technological Characteristics
Steerable nitinol alloy core wire; PTFE polymer shaft coating; hydrophilic distal coating; radiopaque floppy tip. Nominal diameter 0.018"; lengths 260 and 480 cm. Manual operation. No energy source. No software.
Indications for Use
Indicated for patients undergoing endoscopic procedures requiring selective cannulation of biliary and pancreatic ducts, including common bile, cystic, and hepatic ducts, and placement of diagnostic devices.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 16, 2015
NeoMetrics, Inc. Dave Liebl President and Chief Technology Officer 2605 Fernbrook Lane North, Suite J Plymouth, MN 55447
Re: K150225
> Trade/Device Name: NovaGold™ High Performance Guidewire Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCY Dated: January 30, 2015 Received: February 2, 2015
Dear Dave Liebl,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150225
Device Name NovaGold™ High Performance Guidewire
Indications for Use (Describe)
The NovaGold Guidewire is intended for use in selective cannulation of the billary ducts including the common bile, pancreatic, cystic, right and left hepatic ducts, and to aid in the placement of diagnostic devices during endoscopic procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 6.0 510(k) Summary
| Submitter: | NeoMetrics, Inc.<br>2605 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447 |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dave Liebl, President and Chief Technology Officer<br>2605 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447 |
| Date Prepared: | January 30, 2015 |
| Trade Name: | NovaGold™ High Performance Guidewire |
| Classification: | Class II<br>Regulation Number: 21 CFR 876.1500. |
| Product Code: | OCY |
| Predicate Device: | The subject device is substantially equivalent to K133076;<br>NovaGold High Perfomance Guidewire manufactured by<br>NeoMetrics, Inc. |
| Device Description: | The NovaGold Guidewire is constructed from a steerable,<br>metallic core with a PTFE polymer coating over the shaft. A<br>hydrophilic coating is applied over the distal portion of the<br>device. The guidewire has a radiopaque, floppy tip. |
| Indication for Use: | The NovaGold Guidewire is intended for use in selective<br>cannulation of the biliary ducts including the common bile,<br>pancreatic, cystic, right and left hepatic ducts, and to aid in the<br>placement of diagnostic and therapeutic devices during<br>endoscopic procedures. |
| Principle of<br>Operation: | The NovaGold Guidewire is manually inserted and advanced to<br>the target region. |
| Functional and Safety<br>Testing: | To verify that device design met functional and performance<br>requirements, representative samples of the device underwent<br>bench testing in accordance to applicable standards and<br>guidances. |
| | These data provides an acceptable assurance of the safety and<br>effectiveness of the NovaGold guidewire and demonstrated the<br>device is equivalent to the predicate. |
| Comparative<br>Technology<br>Characteristics | A comparison of the characteristics of the proposed device and<br>the predicate device shows the NovaGold guidewire to have the<br>same technological characteristics to the predicate which has<br>received 510(k) clearance.<br>• Identical intended use<br>• Identical operating principle<br>• Identical packaging and sterilization process<br>• Identical overall design, materials of construction, and<br>technology<br>Identical:<br>• Nominal diameter: 0.018”<br>• Guidewire lengths: 260 and 480 cm<br>• Nitinol alloy core wires<br>• Distal radiopaque tip<br>• Lubricious coatings |
| Non-Clinical Tests<br>Submitted | The following tests were performed to support NovaGold's<br>substantial equivalence. |
| | • Fracture resistance<br>• Flex resistance<br>• Tensile strength<br>• Torqueability<br>• Torque Strength<br>• Tip flexibility<br>• Distal fatigue resistance |
| Conclusion: | NeoMetrics Inc. considers the NovaGold guidewire to be<br>equivalent to the predicate device. This conclusion is based upon<br>the fact that device has an equivalent intended use, and there are<br>no differences that raise new types of questions of safety and<br>effectiveness. |
{4}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.