K181887 · Advanced Medical Design Co., Ltd. · OCT · Jan 22, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K181887
Device Name
AMD Anti-Fog Solution
Applicant
Advanced Medical Design Co., Ltd.
Product Code
OCT · Gastroenterology, Urology
Decision Date
Jan 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the distal lenses of the endoscopic/laparoscopic instruments during the surgery. The product is intended for prescription use only.
Device Story
AMD Anti-Fog Solution is a sterile, single-use laparoscopic accessory; consists of aqueous solution (surfactant, isopropyl alcohol, water) in 6 mL dispenser bottle and polyurethane foam pad with radiopaque ribbon. Used in hospital OR by surgeons/staff; solution applied to endoscopic lens surface via foam pad prior to and during surgery. Mechanism: surfactant reduces water surface tension, creating non-scattering film (wetting) instead of droplets on lens. Prevents condensation, maintaining clear visualization during minimally invasive procedures; benefits patient by ensuring continuous surgical field visibility.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity) passed. Sterilization validated per ISO 11137-2 (VDmax 25, SAL 10^-6). Shelf life (5 years) validated per ISO 11607-1 (package integrity, seal strength, sterility, and fog resistance). Performance testing confirmed effectiveness equivalent to predicate.
Technological Characteristics
Aqueous solution (surfactant ≤2%, isopropyl alcohol ≤2%, water ≥96%); polyurethane foam pad with radiopaque ribbon. Dimensions: 6 mL bottle. Energy source: none. Sterilization: Gamma-radiation (10^-6 SAL). Packaging: Tyvek pouch. Standards: ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, USP <151>, ISO 11137-2, ISO 11737-1/2, ISO 11607-1, ASTM F1929-15, ASTM F88/F88M-15.
Indications for Use
Indicated for patients undergoing minimally invasive laparoscopic surgery to prevent condensation on distal lenses of endoscopic/laparoscopic instruments during procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 22, 2019
Advanced Medical Design Co., Ltd Silvia Yeh Regulatory Affairs 4-5F, No 29, Wuquan 5th Rd., Wugu Dist. New Taipei City, 24888 Taiwan
Re: K181887 Trade/Device Name: AMD Anti-Fog Solution Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OCT Dated: July 10, 2018 Received: December 12, 2018
Dear Silvia Yeh:
We reviewed your Section 510(k) premarket notification of intent to market the device referenced above and determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmb/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark R. Kreitz -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181887
Device Name AMD Anti-Fog Solution
#### Indications for Use (Describe)
AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the endoscopic/ laparoscopic instruments during the surgery. The product is intended for prescription use only.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for AMD Global Solutions for Medical Industry. The logo features the letters "AMD" in large, bold, blue font, with a globe graphic incorporated into the "A". Below the letters is the text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" in a smaller font. The logo is clean and professional, suggesting a company that is reliable and innovative.
I. SUBMITTER
# 510(k) Summary
| Company Name: | Advanced Medical Design Co., Ltd. |
|------------------|----------------------------------------------------------------------------------------------------------|
| Mailing Address: | 4F~5F, No29, Wuquan 5th Rd., Wugu Dist., New Taipei city, 248 Taiwan<br>R.O.C. |
| Phone: | +886-2-2290-2627 |
| Fax: | +886-2-2298-1789 |
| Contact Person: | Silvia Yeh<br>Regulatory Affair<br>Phone: +886-2-2290-2627 Ext. 511<br>E-mail: silvia.yeh@amdintl.com.tw |
| Date Prepared: | July 10, 2018 |
#### II. DEVICE INFORMATION
| Name of Device: | AMD Anti-Fog Solution |
|-----------------------|---------------------------------------------|
| Common or Usual Name: | Endoscope Anti-fogging Device |
| Classification Name: | Endoscope and accessories (21 CFR 876.1500) |
| Regulatory Class: | II |
| Product Code: | OCT |
### III. PREDICATE DEVICE
| Trade/Device Name | Clear-It Anti-Fog Solution |
|----------------------|-------------------------------|
| Common or Usual Name | Endoscope Anti-fogging Device |
| 510(k) Number | K022826 |
| Product Code | OCT |
| Submitter | Preservation Solution, Inc. |
## IV. DEVICE DESCRIPTION
The AMD Anti-Fog Solution is a single-use, sterile, and biocompatible laparoscopic accessory device. The product is packaged by the Tyvek" Pouch which includes one clear plastic screw top dispenser bottle with the volume of 6 mL and one polyurethane (PU) foam pad with x-ray detectable radiopaque
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for AMD Global Solutions for Medical Industry. The logo is blue and white. The logo has a globe on the left side of the word AMD. The text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is below the logo.
ribbon. The solution is a clear/colorless, odorless, and aqueous solution which comprises of surfactant (≤2%), isopropyl alcohol (≤2%), and water (≥96%). The polyurethane (PU) foam pad is an adhesivebacked sponge with radiopaque ribbon that can be used to introduce the solution into the surface of the lenses.
The device is intended to be used prior and during minimally invasive surgical operation, to prevent condensation of the distal lenses of the endoscopic instruments during the clinical operation. AMD Anti-Fog Solution is comprised of a surfactant that provides the same physical characteristics as the predicate device. The mechanism of anti-fogging function of AMD Anti-Fog Solution is to reduce the surface tension of the water, resulting in a non-scattering film of water instead of single droplets, on the lenses once the product is smeared on the lenses. Therefore, the non-scattering film of water was formed on the lenses instead of the single water droplets. This phenomenon is called "wetting".
## V. INDICATIONS FOR USE
AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the distal lenses of the endoscopic/laparoscopic instruments during the surgery. The product is intended for prescription use only.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
AMD Anti-Fog Solution is designed to prevent condensation of the distal lenses, that is the same as the predicate device (K number: K022826). A comparison of the device features, technological characteristics, intended use, and other information demonstrates that AMD Anti-Fog Solution is substantially equivalent to the predicate device. The substantially equivalent comparison table was summarized in Table 5.1.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AMD Global Solutions for the Medical Industry. The logo is in blue and features a globe on the left side of the letters AMD. The text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is written in a smaller font below the letters AMD.
| Device Name | Proposed Device | Predicate Device | Comment |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the distal lenses of the endoscopic/laparoscopic instruments during the surgery. The product is intended for prescription use only. | Clear-It Anti-Fog Solution (K022826)<br>Clear-It Anti-Fog Solution is indicated for use in the sterile surgical arena to eliminate condensation from endoscopic lenses, goggles and other devices that are likely to fog. | Same |
| Reusable / Disposable | Disposable | Disposable | Same |
| Target Patient Population | The patient who treated by minimum invasive laparoscopic surgery. | The patient who treated the surgery in the sterile surgical arena. | Similar. The patient population of the proposed device is for minimum invasive laparoscopic surgery, the predicate device included. |
| Target User Population | Prescription use only | Prescription use only | Same |
| Where Used | Hospital O.R. room | Hospital O.R. room | Same |
| Contraindications | There are no known contraindications to the patient. | N/A | Same |
Page 3 of 7
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for AMD Global Solutions for the Medical Industry. The logo is in blue and features a globe on the left side of the letters AMD. The text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is located below the letters AMD.
| Table 5.1. The Substantially Equivalent Comparison Table | | | |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Device<br>Name | Proposed Device | Predicate Device<br>(K022826) | Comment |
| Elements | - Surfactant (≤2%)<br>- Isopropyl alcohol (≤2%)<br>- Water (≥96%)<br>b. Sponge<br>- Polyurethane (PU) foam | - Water<br>- Isopropyl alcohol<br>- Sodium Alcohol Ether<br>Sulfate<br>- Ammonium<br>Dodecylbenzene<br>b. Sponge<br>- Polyurethane (PU) foam | biological evaluation<br>results of the<br>proposed device<br>meet the<br>requirements of the<br>applicable<br>standards. |
| Method of<br>Introduction | Wiping distal end of the lens<br>by sponge with solution. | Wiping distal end of the lens<br>by sponge with solution. | Same |
| Compatibility<br>With Other<br>Devices | Laparoscope/Endoscope:<br>5 mm, 10 mm and all other<br>sizes | Laparoscope/Endoscope:<br>5 mm, 10 mm and all other<br>sizes | Same |
| Performance | Preventing condensation of<br>the distal lenses of the<br>endoscopic/laparoscopic<br>instruments during the<br>clinical operation. | Preventing condensation of<br>the distal lenses of the<br>endoscopic/laparoscopic<br>instruments during the<br>clinical operation. | Same |
| Safety | The following<br>biocompatibility tests were<br>conducted.<br>- Cytotoxicity Test<br>- Sensitization Test<br>- Intracutaneous Reactivity<br>Test<br>- Acute Systemic Toxicity Test<br>- Material-Mediated<br>Pyrogenicity Test<br>All of the tests passed the<br>requirements as indicated in<br>the applicable standards | The cytotoxicity test was<br>conducted. | Same. We<br>performed all of the<br>biocompatibility<br>tests that are<br>indicated in the FDA<br>recognized<br>standards. |
# Table 5,1, The Substantially Equivalent Comparison Table
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for AMD Global Solutions for the Medical Industry. The logo is in blue and features a globe on the left side. The letters AMD are in large, bold font and are the main focus of the logo. Below the letters, the words "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" are written in a smaller font.
| | Proposed Device | Predicate Device | Comment |
|-------------------------|---------------------------------------------|---------------------------------------------|---------|
| Device<br>Name | AMD Anti-Fog Solution | Clear-It Anti-Fog Solution<br>(K022826) | |
| Sterile<br>Barrier | Single Layer Of Tyvek® Pouch | Single Layer Of Tyvek® Pouch | Same |
| Sterilization<br>Method | Gamma-radiation Sterilization<br>(10-6 SAL) | Gamma-radiation Sterilization<br>(10-6 SAL) | Same |
| Energy<br>Source | No energy source | No energy source | Same |
Table 5.1. The Substantially Equivalent Comparison Table
Through the substantially equivalent comparison table, the differences do not raise any different issues on the safety or effectiveness of the product.
## VII. PERFORMANCE DATA
A series of the studies were performed to evaluate the safety and effectiveness of AMD Anti-Fog Solution. The following test results were provided to confirm the product is safe and effective as indicated.
### A. Biocompatibility Testing
The biocompatibility test was evaluated per the FDA recognized consensus standard named "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued on June 16, 2016, the biocompatibility tests include the following items since the product is classified as the classification.
| Nature of Body Contact | | | Contact Duration |
|-------------------------------|--------------------|-----------------|------------------|
| Category | Contact | | |
| External communicating device | Tissue/bone/dentin | Limited (≤24 h) | |
| No. | Test Name | Applicable Standards | Comment |
|-----|--------------------|----------------------------------------------------------------------------------------------------------------------|---------|
| 1 | Cytotoxicity Test | ISO 10993-5:2009 Biological Evaluation Of Medical Devices<br>- Part 5: Tests For In Vitro Cytotoxicity | Pass |
| 2 | Sensitization Test | ISO 10993-10:2010 Biological Evaluation Of Medical<br>Devices - Part 10: Tests For Irritation And Skin Sensitization | Pass |
| 3 | Intracutaneous | ISO 10993-10:2010 Biological Evaluation Of Medical | Pass |
Page 5 of 7
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for AMD Global Solutions for Medical Industry. The logo is in blue and features a globe on the left side. The text "AMD" is in large, bold letters, and the text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is in smaller letters below the logo.
| No. | Test Name | Applicable Standards | Comment |
|-----|-----------------------------------|------------------------------------------------------------------------------------------------------|---------|
| | Reactivity Test | Devices - Part 10: Tests For Irritation And Skin Sensitization | |
| 4 | Acute Systemic<br>Toxicity Test | ISO 10993-11:2017 Biological Evaluation Of Medical<br>Devices - Part 11: Tests For Systemic Toxicity | Pass |
| 5 | Material-Mediated<br>Pyrogenicity | USP <151> Pyrogen Test | Pass |
All of the test studies listed above showed that AMD Anti-Fog Solution did not raise any safety issues and is biocompatible.
# B. Sterilization Validation and Shelf Life Study
The product is designed to perform gamma-radiation sterilization prior to place into the market, therefore the following studies should be evaluated by the applicable standards/guidance.
# a. Gamma-radiation Sterilization Validation Study
The gamma-radiation sterilization study was performed per the requirements of the FDA recognized consensus standards listed below.
- ISO 11137-2:2013 Sterilization of Health Care Products Radiation Part 2: Establishing the Sterilization Dose
- ISO 11737-1:2006 Sterilization of Medical Devices Microbiological Methods Part 1: Determination of A Population of Microorganisms on Products
- ISO 11737-2:2009 Sterilization of Medical Devices Microbiological Methods Part 2: Tests of Sterility Performed in The Definition, Validation And Maintenance Of A Sterilization Process
The method used for gamma-radiation sterilization study was VDmax 25 method, since the product bioburden was less than 1000 CFU/sample. The test reports showed that the product can become sterile when the routine sterilization parameter was controlled at the dose of no less than 25 kGy, which meets the regulatory requirement of sterile condition (SAL < 10t).
# b. Product Aging Validation Study (Shelf Life Study)
The product aging validation study was performed for 5 years per the FDA recognized consensus standards "ISO 11607-1:2006 Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems And Packaging Systems" to determine the shelf life of the product, since the product is supplied in the sterile status.
The Aging Validation Study included the following test studies. Since the shelf life of the product is proposed to be stored for 5 years, the aging validation study is
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for AMD Global Solutions for the Medical Industry. The logo is blue and white. The logo has a globe on the left side and the letters AMD on the right side. Below the logo is the text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY".
performed which included the following test items.
| No. | Test Item | Applicable Standards/Guidance | Comment |
|-----|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 1 | Package Integrity Test<br>(Dye penetration Test) | ASTM F1929-15 Standard Test Method For Detecting<br>Seal Leaks In Porous Medical Packaging By Dye<br>Penetration | Pass |
| 2 | Seal Peel Strength Test | ASTM F88/F88M-15 Standard Test Method For Seal<br>Strength Of Flexible Barrier Materials | Pass |
| 3 | Product Sterility Test | ISO 11737-2:2009 Sterilization Of Medical Devices -<br>Microbiological Methods - Part 2: Tests Of Sterility<br>Performed In The Definition, Validation And<br>Maintenance Of A Sterilization Process | Pass |
| 4 | Product Stability Studies<br>(Fog Resistance Test) | N/A (followed by the internal testing protocol) | Pass |
All of the aging test data showed that the product can be safe and effective during its predetermined shelf life. Hence, the sterile assurance level and the functional specification of the product meet the firm's definition and regulatory requirement, the shelf life of the product is 5 years.
### C. Product Performance Test
The test result of the fog resistance test shows that AMD Anti-Fog Solution is as effective as the predicate device.
### VIII. CONCLUSIONS
Based on the previous data and comparison to the predicate device, AMD Anti-Fog Solution is as safe and effective as the predicate device and do not raise any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.