K163257 · Sio2 Nanotech, LLC · OCT · May 24, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K163257
Device Name
VitreOx Anti-fog Solution
Applicant
Sio2 Nanotech, LLC
Product Code
OCT · Gastroenterology, Urology
Decision Date
May 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
VitreOx® is a temporary anti-fog coating and therein defogs devices such as a laparoscopic accessory intended to facilitate intraoperative defogging of the laparoscope lens thereby maintaining visualization of the surgical site.
Device Story
VitreOx is a sterile, aqueous, alcohol-free, anti-fog solution; applied to endoscope/laparoscope lenses via syringe, dropper bottle, or sterile applicator. Mechanism involves 'wetting'—changing water droplets on lens surface into a flat, transparent sheet. Used in clinical settings (OR) by surgeons/staff during minimally invasive procedures. Unlike predicate, requires no lens heating. Provides anti-fogging for >90 minutes. Improves visualization by preventing condensation-related fogging; enhances surgical site clarity; reduces need for repeated lens cleaning/reapplication during procedures.
Clinical Evidence
Bench testing evaluated anti-fogging performance across various lens diameters (1-10mm) using steam-exposure models. Animal study (n=2 male Yucatan swine) compared VitreOx against a control and the predicate (K062779) during 90-minute gastrointestinal procedures. Results: VitreOx showed no fogging throughout the 90-minute duration, whereas the control fogged at 9 minutes and the predicate showed distortion at 31 minutes. No adverse reactions or complications observed in swine during or 1-year post-procedure.
Technological Characteristics
Transparent, water-soluble, aqueous solution containing Hydroxylpropylmethylcellulose (HPMC) and sterile water. Delivered via syringe, dropper bottle, or sterile applicator. Sterilized via autoclave or irradiation. Functions via surface wetting mechanism. Biocompatible (external communicating, limited duration).
Indications for Use
Indicated for use as a temporary anti-fog coating on laparoscopic accessory devices to facilitate intraoperative defogging of the laparoscope lens and maintain visualization of the surgical site during endoscopic and laparoscopic procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features an eagle-like symbol with three stylized human profiles forming its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
SiO2 Nanotech, LLC Ms. Clarizza Watson Chief Executive Officer 275 N. Gateway Drive, Suite 126 Phoenix, Arizona 85034
Re: K163257
Trade/Device Name: VitreOx Anti-fogging Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCT, GCJ Dated: May 11, 2017 Received: May 15, 2017
Dear Ms. Watson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely,
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Jennifer R. Stevenson -
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163257
Device Name VitreOx®
Indications for Use (Describe)
VitreOx® is a temporary anti-fog coating and therein defogs devices such as a laparoscopic accessory intended to facilitate intraoperative defogging of the laparoscope lens thereby maintaining visualization of the surgical site.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### FDA 510(k) Summary
| 1. Applicant & Co-Specification<br>Developer | SiO2 Nanotech, LLC<br>275 N Gateway Drive, Suite 126<br>Phoenix, AZ 85034 USA<br>Tel #: (602)-565-3447 |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Submitter<br>Address | SiO2 Nanotech, LLC<br>275 N Gateway Drive, Suite 126<br>Phoenix, AZ 85034 USA<br>Tel #: (602)-565-3447 |
| Trade/Proprietary Name | VitreOx® |
| Common Names | Endoscope Anti-fogging Device |
| Establishment Number | 404323 |
| Proposed Classification | OCT Anti-fogging solutions and<br>accessories, endoscopy<br>Class II |
### Prepared: October 3, 2016, Revised May 22, 2017
#### Legally marketed devices to which we are claiming equivalence
The different components of the VitreOx® device are functionally equivalent to the following supplies and devices: K062779 Defogging Endoscope Lens Protector.
#### Substantial Equivalence:
The different components of the VitreOx® device are functionally equivalent to the following supplies and devices:
## Predicate devices using sterile anti-fog solutions in laparoscopic and endoscopic procedures: K062779 Defogging Endoscope Lens Protector.
When compared to the primary predicate D.E.L.P. (K062779), it is substantially equivalent, and does not pose any new risk or safety concerns.
#### Device Description:
The device, VitreOx®, is a biocompatible, aqueous solution, when used as indicated. In addition, the solution is alcohol-free. It functions in the same manner as the predicate device. The device is intended to be used prior and during minimal invasive surgeries, to prevent fogging of the lens of devices (laparoscope) that provide visualization during the procedure. The device accessory VitreOx®, can be removed from the lens with typical surgical cleaning methods such as scrubbing with soap and water, autoclaving, etc.
It is a sterile, single-use, disposable, long-lasting (beyond average surgery time of 90 minutes) anti-fog solution that can be applied on endoscope and laparoscope lenses, as well as endoscope sheaths. It is effective instantaneously, wet; but is even more lasting if allowed to dry, or applied prior to surgery to lenses.
{4}------------------------------------------------
#### Intended Use:
VitreOx® is a single use laparoscopic accessory device, to be used prior and during endoscopic and laparoscopic procedures, to facilitate defogging the laparoscopic lens thereby maintaining visualization of the surgical site.
#### Indications for Use:
VitreOx® is a temporary anti-fog coating and therein defogs devices such as a laparoscopic accessory intended to facilitate intraoperative defogging of the laparoscope lens thereby maintaining visualization of the surgical site.
#### Summary of Technological Characteristics
The VitreOx® device accessory, is an anti-fog solution comprised of components that provide similar physical characteristics as the listed predicate devices. The VitreOx is a transparent, water-soluble solution that is packaged in a syringe or bottle for easy use. In addition to preparing the solution in cleanroom conditions, the product can be autoclaved or irradiated to ensure sterility. During in vitro and in vivo studies, the VitreOx® was sterilized via autoclave for sterility.
Similarly to the predicate device D.E.L.P. that uses ShurClens to clean and coat the lens, the VitreOx® solution utilizes a solution on the lens of the endoscope or laparoscope to combat fogging. However, unlike the primary predicate (K062779), it does not utilize the additional component of a heat source, to heat the lens.
The mechanism of defogging or anti-fogging action of the VitreOx® device, works by physically changing the water droplets that form on the lens surface from round droplets, to a flat transparent sheet of water, a phenomenon known as "wetting".
#### Summary of Performance Testing
The VitreOx® device accessory underwent successful bench and animal studies, to determine effectiveness and safety of the product. The performance was compared to the predicate device, Defogging Endoscope Lens Protector (K062779).
Bench Studies: The bench studies involved testing the VitreOx 100 in simulated closed-body cavity surgical conditions.
The VitreOx® application on the surgical lens, was tested using three different methods of delivery, to adapt to the various diameters of endoscope lenses including, but not limited to 1, 1.2, 1.9, 2.7, 3.5, 4, 10 mm:
- 1) Syringe.
- 2) Bottle with dropper applicator, and
- 3) A customized, patented, sterile, single-use applicator.
Steam from 2 Mega Ohm deionized water was heated in a Pyrex container, using a VWR hotplate. Different steam temperatures were tested, from 98° Fahrenheit (32° Celsius), to 122° Fahrenheit (50° Celsius). The steam was directed to a handheld endoscope (Clarus 30000-10), using a stainless steel vessel. Additionally, an electronic endoscope (generic electronic endoscope 14.55 mm, wide angle) was also tested during the bench studies, similarly, steam was directed to the electronic endoscope.
{5}------------------------------------------------
Animal Studies: The VitreOx® device was applied using a bottle with dropper applicator, on a Karl Storz endoscope. The endoscope was used in an exploratory gastrointestinal procedure on 2 male Yucatan swine from the same litter, for 90 minutes. Performance was compared between a control (no anti-fog), and the primary predicate device (New Wave D.E.L.P.).
Surgical conditions were carefully monitored and maintained by the help of veterinarians, who also administered the proper anesthesia for the swine. The surgical room was maintained between 68°F and 73°F. CO2 was used to inflate the swine's abdominal cavity, and the following vitals on the swine were monitored throughout the procedure: Pulse, oxygen level, respiratory rate, and body temperature.
Fogging is noted in the experiment: When there is a change of the light reflection on tissue, the light reflection will not be as crisp. In addition, veins on the tissue as also observed, because it will look less defined when fogging is present. Fogging was observed on the control scope within 9 minutes. Fogging was first observed on the endoscope treated with New Wave D.E.L.P. within 8 minutes, but distortion due to fogging was more evident after 31 minutes. Fogging was not observed on the endoscope with VitreOx®, throughout the 90-minute procedure.
After the exploratory gastrointestinal procedures, the 2 male Yucatan swine were kept at All Pets Animal Hospital in Peoria, Arizona for 2 weeks, for observation. The 2 male Yucatan swine did not suffer from effects of the procedure, and were discharged as "healthy", with no adverse reactions or complications from the procedures. Additionally, there were no signs of a reaction from the VitreOx® device on the swine's skin that came into contact with it. After the swine were discharged, they were monitored for a year to observe any adverse reactions, which there were not.
{6}------------------------------------------------
# Comparison Table
| Characteristic<br>and Material | VitreOx® | D.E.L.P<br>(K)062779 |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| Intended Use | VitreOx™ is intended to be<br>used prior to endoscopic and<br>laparoscopic procedures, to<br>eliminate fogging due to<br>condensation for scopes<br>ranging from 5 to 10 mm. In<br>addition, it can also be used<br>for microscopic lenses,<br>goggles, surgical loupes, and<br>other devices that may fog. | Commonly known as DHELP in<br>the market, is intended to be used<br>prior to and during endoscopic<br>and laparoscopic procedures to<br>prevent fogging of the scope lens<br>for 5mm and 10 mm scopes. |
| Solution | Sterile Water,<br>Hydroxylpropylmethylcellulo<br>se (HPMC) | FDA Approved ShurClens, a<br>wound cleaning surfactant |
| Sponge | Yes | Yes |
| Mechanism of<br>Action | Applying the solution to the<br>lens of the endoscope with a<br>sponge (if needed). | Heating and dipping distal end of<br>endoscope into solution |
| Size of<br>endoscopes used<br>for or<br>accommodated<br>by device | 5 mm, 10 mm, and all other<br>sizes. | 5mm and 10 mm |
| Sterility | Sterile | Sterile |
| Reusable/Dispos<br>able | Disposable | Disposable |
| Meets<br>Biocompatibility<br>needs | Yes | Yes |
| Level of<br>Biocompatibility | External communicating<br>device (Tissue/bone/dentin<br>communicating) | External communicating device<br>(Tissue/bone/ Dentin<br>communicating) |
| Duration of<br>Biocompatibility | Limited | Limited |
{7}------------------------------------------------
## Assessment of Performance:
The VitreOx® device testing was conducted with two live, anesthetized Yucatan swine of the same sex (males) and litter. Two 4 mm Karl Storz endoscopy systems was used, one straight endoscope, the other 45° angle. A gastrointestinal exploratory procedure was conducted for 90 minutes with the VitreOx TM device. The performance of the VitreOx® device was compared to a control (endoscope with no treatment, and the same gastrointestinal procedure conducted for 90 minutes, using the D.E.L.P. device (primary predicate).
# Conclusion:
Our evaluation concluded that VitreOx® raises no new issues of safety and effectiveness. There was no fogging on the scope that VitreOx® was applied through the entire duration of the 90-minute procedure. Unlike the D.E.L.P. device, the VitreOx® device did not require heating, and reapplication during procedure.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.