REPROCESSED ENDOSCOPIC TROCAR

K012578 · Sterilmed, Inc. · NLM · Nov 7, 2001 · Gastroenterology, Urology

Device Facts

Record IDK012578
Device NameREPROCESSED ENDOSCOPIC TROCAR
ApplicantSterilmed, Inc.
Product CodeNLM · Gastroenterology, Urology
Decision DateNov 7, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

The reprocessed endoscopic trocars are intended to provide a pathway for entry of minimally invasive instruments to a body organ or cavity during general, abdominal, thoracic, gynecological or other minimally invasive surgical procedures.

Device Story

Reprocessed endoscopic trocars provide a pathway for minimally invasive surgical instruments to access body organs or cavities. Devices consist of a trocar and cannula assembly with varying lengths, diameters, and blunt or bladed obturator tips. Used in general, abdominal, thoracic, and gynecological surgeries for procedures including observation, dissection, cutting, repair, and tissue/organ manipulation. Reprocessed by SterilMed, Inc. to restore functionality of single-use devices. Healthcare providers use the device to facilitate instrument access during surgery. Benefit includes enabling minimally invasive surgical access.

Clinical Evidence

Bench testing only. Representative samples underwent functional testing and process validation for cleaning, sterilization, and packaging. Manufacturing process includes 100% visual and functional testing.

Technological Characteristics

Endoscopic trocar and cannula assembly; varying lengths and diameters; blunt or bladed obturator tips. Materials and construction methods are equivalent to original manufacturer specifications. Sterilization and cleaning processes validated.

Indications for Use

Indicated for patients undergoing general, abdominal, thoracic, gynecological, or other minimally invasive surgical procedures requiring a pathway for instrument entry into a body organ or cavity.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure. Public Health Service 1 2004 NOV Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Dr. Bruce Lester Vice President, Research and Development SterilMed, Inc. 11400 73rd Ave. North Minneapolis, Minnesota 55369 Re: K012578 - Supplemental Validation Submission Trade/Device Name: Reprocessed Laparoscope General and Plastic Surgery (See enclosed list) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: NLM Dated: August 7, 2001 Received: August 9, 2001 Dear Dr. Lester: The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on November 7, 2001. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register. {1}------------------------------------------------ ## Page 2 - Dr. Bruce Lester Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours. Misiane C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Reprocessed Laparoscope General and Plastic Surgery (Trocar) Models found to be Substantially Equivalent: - 1. Ethicon, 355RT - Ethicon, 35LRT 2. - Ethicon, 35LRL 3. - 4. Ethicon, 511RT - Ethicon, 511RL 5. - Ethicon, 512RT 6. - 7. Ethicon, 512RL {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): K012578 Device Name: Reprocessed Endoscopic Trocars Indications For Use: The reprocessed endoscopic trocars are intended to provide a pathway for entry of minimally invasive instruments to a body organ or cavity during general, abdominal, thoracic, gynecological or other minimally invasive surgical procedures. Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Miriam C. Provost Division Sign Off Division of General, Restorative, and Neurological Devices Page 1 of ____________________________________________________________________________________________________________________________________________________________________ **510(k) Number** k012578 {4}------------------------------------------------ NOV 0 7 2001 # SECTION 2. SUMMARY AND CERTIFICATION ### 510(k) Summary A. SterilMed, Inc. - - August 7, 2001 GCJ , · Contact Person: Submitter: Patrick Fleischhacker SterilMed, Inc. 11400 73rd Avenue North Minneapolis, MN 55369 Ph: 763-488-3400 Fax: 763-488-3350 Date Prepared: Trade Name: Classification Name: and Number: Trocar and Cannula Class II, 21 CFR 876.1500 Product Code: Predicate Device(s): Device Description: Intended Use: The reprocessed endoscopic trocar is substantially equivalent to the Endopath EP Disposable Surgical Trocar (K922608), manufactured by Ethicon; AutoSuture Surgiport® Endoscopic Trocar (K925860), manufactured by US Surgical; and the counterpart devices from the original manufacturer. Reprocessed endoscopic trocars are devices that provide a pathway for entry of minimally invasive instruments to a body organ or cavity during general, abdominal, thoracic, gynecologic or other minimally invasive surgical procedures such as observation, dissecting, cutting, repairing, and removal or manipulation of internal tissues and/or organs. Reprocessed endoscopic trocars are of varying lengths and diameters, and may have either a blunt or bladed obturator tip. This device is designed to provide a pathway for entry of minimally invasive instruments to a body organ or cavity during general, abdominal, thoracic, gynecologic or other minimally invasive surgical procedures. {5}------------------------------------------------ Functional and Safety Testing: Representative samples of reprocessed endoscopic trocars underwent bench testing to demonstrate appropriate functional characteristics. Process validation testing was done to validate the cleaning and sterilization procedures as well as the device's packaging. In addition, the manufacturing process includes visual and functional testing of all products produced. Conclusion: The reprocessed endoscopictrocars are substantially equivalent to the Endopath EP Disposable surgical Trocar (K922608), manufactured by Ethicon; AutoSuture Surgiport® Endoscopic Trocar (K925860), manufactured by US Surgical; and the counterpart devices from the original manufacturer. This conclusion is based upon the devices' similarities in functional design, materials, indications for use and methods of construction.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...