4K Endoscopic Camera System (SV-800, SV-800I, SV-800R, SV-800N)

K250551 · Shenzhen Sophway Technology Co., Ltd. · GCJ · Oct 2, 2025 · Gastroenterology, Urology

Device Facts

Record IDK250551
Device Name4K Endoscopic Camera System (SV-800, SV-800I, SV-800R, SV-800N)
ApplicantShenzhen Sophway Technology Co., Ltd.
Product CodeGCJ · Gastroenterology, Urology
Decision DateOct 2, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

It is used to connect with an optical endoscope during the endoscopic diagnosis, treatment, and observation. And to capture, process and transmit images in the human body cavity under the field of view observed by the endoscope to the monitor.

Device Story

4K Endoscopic Camera System (models SV-800, SV-800I, SV-800R, SV-800N) comprises a camera control unit (CCU) and 4K camera head. Device connects to optical endoscopes to capture, process, and transmit internal body cavity images to a monitor for physician observation during diagnosis or surgery. System supports white light imaging (WLI) and automatic white balance. Features include real-time image recording to external USB or internal hard disk. Operated by clinicians in clinical settings. Output allows real-time visualization of surgical sites, aiding clinical decision-making during endoscopic procedures.

Clinical Evidence

Bench testing only. Performance testing verified device functions against design specifications. Reprocessing (cleaning, disinfection, sterilization) validated. Software and cybersecurity validation performed per FDA guidance. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18 confirmed.

Technological Characteristics

System includes CCU and 4K camera head with CMOS image sensor. Resolution: 3840x2160/1920x1080. Interfaces: 12G SDI, 3G SDI, HDMI. Power: 100-240V, 50/60Hz. Sterilization: ASP Sterrad 100NX plasma. Standards: IEC 60601-1:2020, IEC 60601-1-2:2021, IEC 60601-2-18:2009. Software-defined functionality differentiates the four models.

Indications for Use

Indicated for use with an optical endoscope for visualization during endoscopic diagnosis, treatment, and observation in human body cavities.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 2, 2025 Shenzhen Sophway Technology Co., Ltd. % Eva Li Consultant Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave. Shanghai, 200122 China Re: K250551 Trade/Device Name: 4K Endoscopic Camera System (SV-800, SV-800I, SV-800R, SV-800N) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, EOB, EOQ, FGB, HET, NWB Dated: September 2, 2025 Received: September 2, 2025 Dear Eva Li: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250551 - Eva Li Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250551 - Eva Li Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2025.10.02 09:43:26 -04'00' Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250551 Device Name 4K Endoscopic Camera System (SV-800, SV-800I, SV-800R, SV-800N) Indications for Use (Describe) It is used to connect with an optical endoscope during the endoscopic diagnosis, treatment, and observation. And to capture, process and transmit images in the human body cavity under the field of view observed by the endoscope to the monitor. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K250551 SHENZHEN SOPHWAY TECHNOLOGY CO., LTD. Rev01 # 510(K) Summary ## 1. General Information Applicant: Shenzhen Sophway Technology Co., Ltd. Address: 7th Floor, Building 9#, Zhongyuntai Technology Industrial Park, Songbai Rd., Shiyan Avenue, Bao'an, Shenzhen, 518108 Guangdong, P.R. China Contact: Shuting Chen Tel: +86 755 36368352 Date of preparation: September 30, 2025 Device Identification: Common Name: 4K Endoscopic Camera System Trade Name: Gynecologic laparoscope and accessories Device Classification: Class II Regulation: 21 CFR 884.1720 Product Code: HET, EOB, EOQ, FGB, GCJ, NWB ## 2. Predicate device K151011 VISERA 4K UHD SYSTEM Olympus Medical Systems Corp. ## 3. Indication for Use It is used to connect with an optical endoscope during the endoscopic diagnosis, treatment, and observation. And to capture, process and transmit images in the human body cavity under the field of view observed by the endoscope to the monitor. ## 4. Device Description The 4K Endoscopic Camera System consists of a camera control unit and a 4K camera head. The device is intended to be used to connect with an optical endoscope during the endoscopic diagnosis and/or treatment/surgery, and to capture, process and transmit images in the human body cavity under the field of view observed by the endoscope to the monitor. There are 4 models of the device. The hardware configuration of the device of 4 models are same, and difference is different functions which opened up through software. {5} K250551 SHENZHEN SOPHWAY TECHNOLOGY CO., LTD. Rev01 5. Technology comparison with the predicate device | | Subject device | Predicate device | Comparison | | --- | --- | --- | --- | | Device name | 4K Endoscopic Camera System | VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB | NA | | Device code | HET, EOB, EOQ, FGB, GCJ, NWB | HET, EOB, EOQ, FGB, GCJ, NWB | Same | | Indication for use | It is used to connect with an optical endoscope during the endoscopic diagnosis, treatment, and observation. And to capture, process and transmit images in the human body cavity under the field of view observed by the endoscope to the monitor. | VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400 The camera control unit has been designed to be used with Olympus endoscopes, camera heads, light source, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EB The camera head has been designed to be used with Olympus endoscopes, camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. | Similar1 | | electrical safety | Comply to IEC 60601-1:2020, IEC 60601-2-18: 2009 | Comply to IEC 60601-1: 2005+A1, IEC 60601-2-18: 2009 | same | | EMC | Comply to IEC IEC60601-1-2:2021 | Comply to IEC IEC60601-1-2:2007 | same | | CCU | | | | | Dimension | 364(L)*325(W)*137(H)mm | 390(W) × 160(H) × 506(D) mm | similar | | Resolution | 4K (3840×2160) (1920×1080) | 4K (4096pxl×2160pxl) | similar | | Signal output (interface type) | 12G SDI × 1, 3G SDI × 1, HDMI × 2 | SDI (3G/HD) | Different1 | | Video output format | 3840×2160, 1920×1080 | 4096×2160, 3840×2160, or 1920×1080 | similar | | Image recording | The data is stored in real time to the external access USB interface device or internal hard disk by photographing or recording functions. | No | Different2 | {6} K250551 SHENZHEN SOPHWAY TECHNOLOGY CO., LTD. Rev01 | Operation mode | Continuous operation | Continuous operation | same | | --- | --- | --- | --- | | White balance | Automatic white balance | Automatic white balance | same | | Observation light imaging | White light imaging(WLI) | White light imaging(WLI) | same | | details about the display | Touchscreen | Touchscreen | same | | input/output (signals, power) | 70VA | 350VA | Different3 | | Operating temperature | 10~30°C | 10~35°C | Different4 | | Operating humidity | 30% ~ 85%RH | 30 ~ 85% RH | Same | | Operating atmosphere | 700~1060 (hpa) | 700 ~ 1060 hPa | Same | | Power Supply | 100-240V 50/60Hz | 100-240V 50/60Hz | same | | Weight | 4.8Kg | 13.5Kg | Similar | | Camera head | | | | | image resolution | 1/3" Three Chip CMOS: 1920×1080/CMOS | Undisclosed | N/A | | number of pixels | 1/3" Three Chip CMOS: 1920×1080/CMOS | Undisclosed | N/A | | Size | W51 mm × H48 mm × L101 mm | W43.6 mm × H49.5 mm × L122.5 mm | similar | | Weight | 570 kg | 280 g | similar | | Operating temperature | 10~30°C | 10~35°C | similar | | Operating humidity | 30% ~ 85%RH | 30 ~ 85% RH | Same | | Operating atmosphere | 700~1060 hpa | 700 ~ 1060 hPa | Same | | size of pixels | 2.9um | Undisclosed | N/A | | size of active area | 1/3" | Undisclosed | N/A | | focal length | 14mm | 23.5mm | similar | | focus control details | Automatic; Manual | Automatic; Manual | same | | image sensor | CMOS image sensor | CMOS image sensor | same | | Single use or reuse | reuse | reuse | same | | Reprocessing | End user sterilized ASP Sterrad 100NX plasma sterilizer | End user sterilized EOG/STERRAD | similar | 3 / 4 {7} K250551 SHENZHEN SOPHWAY TECHNOLOGY CO., LTD. Rev01 ## 6. Non-Clinical performance tests: The differences of technological characteristics between the predicate device and the subject device are confirmed that they are substantially equivalent through the following tests and standards. - A performance testing were conducted to demonstrate that the 4K Endoscopic Camera System performs according to specifications and functions as intended. Tests results obtained verified the safety and effectiveness of the devices in accordance with design specifications and applicable standards. - The instructions for reprocessing of the subject devices include Cleaning, Disinfection, and Sterilization. All cleaning, disinfection, and sterilization methods have been validated. - The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." - The cybersecurity validation activities were performed in accordance with the FDA Guidance, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" - Electromagnetic compatibility, electric safety, and thermal safety had been confirmed. - Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2019. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment. The following voluntary standards to comply with have been applied to the 4K Endoscopic Camera System: IEC 60601-1: 2020, IEC 60601-1-2: 2021, IEC 60601-2-18: 2009, ## 7. Conclusion The subject device is substantially equivalent to its predicate devices. The nonclinical testing demonstrates that the device is as safe, as effective as well as the legally marketed devices under K151011.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...