The EndoFLIP® System is indicated for use in a clinical setting to measure pressure and dimensions in the esophagus, pylorus, and anal sphincters. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders. The EF-325 series of catheters can make pressure and dimensional measurements in the esophagus, pylorus, and anal sphincters; whereas the BF-325 series and EF-825 catheters can make dimensional measurements in the esophagus, pylorus, and anal sphincters. The EF-620 catheter can make dimensional measurements in the esophagus. Other indications for use include: · To estimate the size of a stoma produced by a gastric band (all EndoFLIP® catheters) · For use as an adjunct to a bougie for measuring the size of a gastric sleeve created during bariatric surgery, where it is suitable for diameter measurements for 22 to 60Fr sleeves (EF-620 catheter)
Device Story
EndoFLIP® System measures pressure and dimensions in esophagus, pylorus, and anal sphincters using specialized catheters (EF-325, BF-325, EF-825, EF-620). Catheters are positioned in target anatomy; system performs pre-programmed isovolumetric inflation steps or patient-controlled steps. FLIP Topography module is a software accessory running on an off-the-shelf medical-grade PC; connects via cable to EndoFLIP® unit. Displays real-time data on larger screen; provides historical graphs of readings over last 40 seconds. Used in hospitals and physician offices by clinicians. Output assists in comprehensive evaluation of motility disorders, stoma sizing for gastric bands, and gastric sleeve sizing during bariatric surgery. Benefits include enhanced visualization of motility data to aid clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only; verification activities focused on software data acquisition, display characteristics, data management, and communication.
Technological Characteristics
System consists of catheter-based pressure/dimension measurement hardware and software module. Connectivity via cable to off-the-shelf medical-grade PC. Complies with IEC 60601-1 and IEC 60601-1-2. No changes to materials or fundamental sensing technology.
Indications for Use
Indicated for patients with symptoms consistent with gastrointestinal motility disorders, including those undergoing gastric band surgery and post-operative adjustment. Contraindicated where endoscopy is contraindicated or in patients with actively bleeding esophageal varices.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, abstract design connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 17, 2017
Crospon Ltd. Paul Dryden Consultant 24301 Woodsage Dr. Bonita Springs, Florida 34135
Re: K170833
Trade/Device Name: FLIP Topography Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal Motility Monitoring System Regulatory Class: Class II Product Code: FFX Dated: March 16, 2017 Received: March 20, 2017
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Charles Viviano -S
For
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K170833
Device Name
# EndoFLIP® System with FLIP Topography module
Indications for Use (Describe)
The EndoFLIP® System is indicated for use in a clinical setting to measure pressure and dimensions in the esophagus, pylorus, and anal sphincters. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.
The EF-325 series of catheters can make pressure and dimensional measurements in the esophagus, pylorus, and anal sphincters; whereas the BF-325 series and EF-825 catheters can make dimensional measurements in the esophagus, pylorus, and anal sphincters. The EF-620 catheter can make dimensional measurements in the esophagus.
Other indications for use include:
- · To estimate the size of a stoma produced by a gastric band (all EndoFLIP® catheters)
· For use as an adjunct to a bougie for measuring the size of a gastric sleeve created during bariatric surgery, where it is suitable for diameter measurements for 22 to 60Fr sleeves (EF-620 catheter)
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 5 4/14/17
| Company: | Crospon Ltd.<br>Galway Business Park<br>Dangan, Galway, Ireland<br>Tel: +011 [353] (91) 519882 |
|----------------------------|------------------------------------------------------------------------------------------------|
| Official Contact: | John O'Dea PhD - CEO |
| Proprietary or Trade Name: | EndoFLIP® System with FLIP Topography module |
| Common/Usual Name: | Gastrointestinal motility monitoring system |
| Classification Name: | FFX |
| Device: | FLIP Topography module |
| Modified Device: | K160725 - EndoFLIP® System – Crospon Ltd. |
## Device Description:
The modification to the EndoFLIP® System is the addition of display functionality. This functionality is software and is referred to as the FLIP Topography module which is a supplemental accessory.
The addition of the supplemental accessory FLIP Topography module does not modify the predicate EndoFLIP® System or change the intended use of the device. This modification does not alter the fundamental scientific technology; the EndoFLIP® System algorithm or change the hardware and accessories associated with the EndoFLIP® System.
As there are no changes in hardware or software that will impact performance, there is no need to validate the changes through a clinical investigation.
# Modification:
The modification to the EndoFLIP® System is the addition of display functionality. This functionality is software and is named the FLIP Topography module. FLIP Topography module is software which runs on an off-the-shelf PC that connects to the EndoFLIP® via a cable and displays the data on a larger screen.
FLIP Topography displays the traditional EndoFLIP® data plus FLIP Topography adds historical data graphs (live data for the last 40 seconds) for use by the clinician.
# Indications for Use:
The EndoFLIP® System is indicated for use in a clinical setting to measure pressure and dimensions in the esophagus, pylorus, and anal sphincters. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.
{4}------------------------------------------------
#### 510(k) Summary Page 2 of 5 4/14/17
The EF-325 series of catheters can make pressure and dimensional measurements in the esophagus, pylorus, and anal sphincters; whereas the BF-325 series and EF-825 catheters can make dimensional measurements in the esophagus, pylorus, and anal sphincters. The EF-620 catheter can make dimensional measurements in the esophagus.
Other indications for use include:
- To estimate the size of a stoma produced by a gastric band (all EndoFLIP® catheters)
- . For use as an adjunct to a bougie for measuring the size of a gastric sleeve created during bariatric surgery, where it is suitable for diameter measurements for 22 to 60Fr sleeves (EF-620 catheter)
The indications for use are unchanged.
## Patient Population:
Patients with symptoms consistent with gastrointestinal motility disorders.
#### Environment of Use:
Hospitals, Physician offices.
## Contraindications:
The EndoFLIP® System is contraindicated where endoscopy is contraindicated.
The EndoFLIP® System is contraindicated for use in patients with actively bleeding varices in the esophagus.
#### Predicate Device Comparison
The EndoFLIP® System with FLIP Topography module is compared to the EndoFLIP® System in the device comparison table below.
#### Summary of Modifications:
The modification to the EndoFLIP® System is the addition of display functionality; this functionality is software and is named the FLIP Topography module. FLIP Topography is software which runs on an off-the-shelf PC that connects to the EndoFLIP® via a cable and displays the data on a larger screen.
#### Verification and Validation Activities
Verification activities for the modification were performed and the results demonstrated that the predetermined acceptance criteria were met. These verification activities included:
- . Data acquisition
- . Display characteristics (labels, accuracy, etc.)
- . Data management
- . Communication
{5}------------------------------------------------
# 510(k) Summary Page 3 of 5
| | EndoFLIP® System with FLIP<br>Topography module | Unmodified Predicate<br>EndoFLIP® System<br>K160725 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attributes | | |
| Indications for Use | The EndoFLIP® System is indicated for<br>use in a clinical setting to measure pressure<br>and dimensions in the esophagus, pylorus,<br>and anal sphincters. It is intended to be<br>used as an adjunct to other diagnostic<br>methods as part of a comprehensive<br>evaluation of patients with symptoms<br>consistent with gastrointestinal motility<br>disorders.<br>The EF-325 series of catheters can make<br>pressure and dimensional measurements in<br>the esophagus, pylorus, and anal<br>sphincters; whereas the BF-325 series and<br>EF-825 catheters can make dimensional<br>measurements in the esophagus, pylorus,<br>and anal sphincters. The EF-620 catheter<br>can make dimensional measurements in<br>the esophagus.<br>Other indications for use include:<br>• To estimate the size of a stoma produced<br>by a gastric band (all EndoFLIP®<br>catheters)<br>• For use as an adjunct to a bougie for<br>measuring the size of a gastric sleeve<br>created during bariatric surgery, where it is<br>suitable for diameter measurements for 22<br>to 60Fr sleeves (EF-620 catheter) | The EndoFLIP® System is indicated for<br>use in a clinical setting to measure pressure<br>and dimensions in the esophagus, pylorus,<br>and anal sphincters. It is intended to be<br>used as an adjunct to other diagnostic<br>methods as part of a comprehensive<br>evaluation of patients with symptoms<br>consistent with gastrointestinal motility<br>disorders.<br>The EF-325 series of catheters can make<br>pressure and dimensional measurements in<br>the esophagus, pylorus, and anal<br>sphincters; whereas the BF-325 series and<br>EF-825 catheters can make dimensional<br>measurements in the esophagus, pylorus,<br>and anal sphincters. The EF-620 catheter<br>can make dimensional measurements in<br>the esophagus.<br>Other indications for use include:<br>• To estimate the size of a stoma produced<br>by a gastric band (all EndoFLIP®<br>catheters)<br>• For use as an adjunct to a bougie for<br>measuring the size of a gastric sleeve<br>created during bariatric surgery, where it is<br>suitable for diameter measurements for 22<br>to 60Fr sleeves (EF-620 catheter) |
| Anatomical Sites | Esophagus, stomach, pylorus, and anal<br>sphincters | Esophagus, stomach, pylorus, and anal<br>sphincters |
| Environments of<br>use | Hospitals, Physician offices | Hospitals, Physician offices |
| Patient Population | Patients with symptoms consistent with<br>gastrointestinal disorders<br>Patient undergoing gastric band surgery<br>and post-operative band adjustment | Patients with symptoms consistent with<br>gastrointestinal disorders<br>Patient undergoing gastric band surgery<br>and post-operative band adjustment |
| | 4/14/11/ | |
| | EndoFLIP® System with FLIP | Unmodified Predicate |
| | Topography module | EndoFLIP® System |
| | | K160725 |
| Contraindications | The EndoFLIP® System is contraindicated | The EndoFLIP® System is contraindicated |
| | where endoscopy is contraindicated. | where endoscopy is contraindicated. |
| | | |
| | The EndoFLIP® System is contraindicated | The EndoFLIP® System is contraindicated |
| | for use in patients with actively bleeding | for use in patients with actively bleeding |
| | varices in the esophagus. | varices in the esophagus. |
| Principle of | The catheter is positioned in the desired | The catheter is positioned in the desired |
| Operation | location one volume setting and inflation | location one volume setting and inflation |
| | rate to be made. There are up to ten | rate to be made. There are up to ten |
| | isovolumetric steps which are pre- | isovolumetric steps which are pre- |
| | programmed. These steps are programmed | programmed. These steps are programmed |
| | in terms of the volume to be delivered and | in terms of the volume to be delivered and |
| | the time to pause between each step. | the time to pause between each step. |
| | Alternatively the system can be | Alternatively the system can be |
| | programmed to allow the patient to decide | programmed to allow the patient to decide |
| | when the next step commences. | when the next step commences. |
| FLIP Topography | Software that displays EndoFLIP® data on | EndoFLIP® displays data on its integrated |
| | a larger screen and provides a historical | screen. |
| | graph that shows readings over time. | |
| | | Does not have a historical graph option to |
| | | show readings over time.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.