EG-530N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The EG-530N can be inserted orally or transnasally.
Device Story
The FUJIFILM EG-530N is a gastroscope used for upper digestive tract visualization and treatment. It consists of a control section, insertion tube, and umbilicus. The device utilizes glass fiber bundles for illumination and a distal CCD image sensor to capture video. It connects to a FUJIFILM video processor and light source to display images on a monitor. The insertion tube includes channels for air/water delivery and accessory passage (e.g., forceps). Operated by clinicians in a clinical setting, the device allows for real-time observation and therapeutic intervention. The output is a visual feed on a monitor, assisting the physician in diagnosing and treating conditions of the esophagus, stomach, and duodenal bulb.
Clinical Evidence
Bench testing only. Performance testing included field of view, bending capability, air/water supply rates, suction rate, working length, forceps channel diameter, viewing direction, resolution, and light guide output. Electrical safety was verified per ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60601-2-18. Biocompatibility was confirmed per ISO 10993-5 and ISO 10993-10. Reprocessing validation followed AAMI TIR12, AAMI TIR30, and FDA guidance.
Technological Characteristics
Gastroscope featuring a CCD image sensor and fiber optic illumination. Materials include a new epoxy resin. Connectivity via video connector to external processor/light source. Complies with ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18, ISO 10993-5, and ISO 10993-10. Reprocessing validated per AAMI TIR12/TIR30.
Indications for Use
Indicated for visualization, observation, diagnosis, and endoscopic treatment of the upper digestive tract (esophagus, stomach, duodenal bulb) in patients requiring oral or transnasal endoscopy.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2019
FUJIFILM Corporation % Jeffrey Wan Specialist, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 81 Hartwell Avenue, Suite 300 Lexington, MA 02421
Re: K183686
> Trade/Device Name: FUJIFILM Endoscope Model EG-530N Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDS Dated: December 27, 2018 Received: December 28, 2018
Dear Jeffrey Wan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark J. Antonino -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183686
Device Name
FUJIFILM Endoscope Model EG-530N
#### Indications for Use (Describe)
EG-530N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The EG-530N can be inserted orally or transmasally.
Type of Use (Select one or both, as applicable)
| <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
#### FUJIFILM Endoscope Model EG-530N
Date: December 27, 2018
#### Submitter's Information:
FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928
#### Contact Person:
Jeffrey Wan Specialist, Regulatory Affairs Telephone: (201) 675-8947 E-M ail: jeffrey.wan@fujifilm.com
#### Identification of the Proposed Device:
| Proprietary/Trade Name: | FUJIFILM Endoscope Model EG-530N |
|-------------------------|-------------------------------------------------|
| Common Name: | Gastroscope |
| Device Class: | Class II |
| Review Panel: | Gastroenterology/Urology |
| Classification: | Endoscope and accessories, 21 C.F.R. § 876.1500 |
| Product Code: | FDS |
#### Predicate Device:
Fujinon EG-530N (K063316)
#### Reference Device:
FUJIFILM Video Colonoscope EC-600WL v2 (K160196)
#### Intended Use / Indications for Use:
EG-530N is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenal bulb. The EG-530N can be inserted orally or transnasally.
{4}------------------------------------------------
## Device Description:
FUJIFILM Endoscope Model EG-530N is comprised of three general sections: a control portion, an insertion portion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary Charge-Coupled Device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart.
#### Technological Characteristics:
The proposed device EG-530N differs from the predicate device in the following minor modifications:
- . Expansion of transport and storage conditions to a temperature range of -20°C to 60°C and a humidity range of 10 to 85% RH.
- . Use of a new epoxy resin in the manufacturing process. Both the new and original epoxy resins can be used for repair.
- . Change in design of the video connector.
- . Compatibility with different components and accessories.
#### Performance Data:
Electrical safety of the proposed devices were evaluated using the following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.
The proposed device EG-530N was adopted into biocompatibility testing conducted on the predicate device EG-530N using the following consensus standards: ISO 10993-5:2009, and ISO 10993-10:2010.
The proposed device EG-530N was adopted into the reprocessing validation of the predicate device EG-530N using the following consensus standards: AAMI TIR12:2010, AAM TIR30:2011. Validation of the cleaning, disinfection, and sterilization instructions was performed in accordance with FDA's quidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling," published March 17, 2015.
Fujifilm conducted the following performance testing on the proposed device EG-530N to ensure that the modified device performs equivalently to the predicate device:
- . Field of view
- Bending capability
- . Rate of air supply
- Rate of water supply
- . Suction rate
- Working length
{5}------------------------------------------------
- . Forceps channel diameter
- . Viewing direction
- Resolution ●
- . LG output
In all cases, the devices met the pre-defined acceptance criteria for the test.
## Substantial Equivalence:
The company's EG-530N has the same intended use and indications for use as the previously cleared predicate EG-530N (K063316). In addition, the proposed device has similar technological characteristics and principles of operation as the predicate. The minor differences between the proposed and predicate devices do not raise new or additional questions of safety or effectiveness of the proposed devices. Thus, the proposed device EG-530N is substantially equivalent to the predicate device.
## Conclusions:
The modified EG-530N is substantially equivalent to the predicate EG-530N and conforms to applicable medical device safety and performance standards.