EG-3800T, VIDEO GASTROSCOPE

K962460 · Pentax Precision Instrument Corp. · FDS · Sep 16, 1996 · Gastroenterology, Urology

Device Facts

Record IDK962460
Device NameEG-3800T, VIDEO GASTROSCOPE
ApplicantPentax Precision Instrument Corp.
Product CodeFDS · Gastroenterology, Urology
Decision DateSep 16, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The EG-3800T, Video Gastroscope, is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Upper Gastrointestinal Tract. The Upper Gastrointestinal Tract includes, but is not limited to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, and Small Bowel. The instrument is introduced per orally when indications consistent with the requirement for procedure(s) are observed in Adult and Pediatric patient populations.

Device Story

EG-3800T is a video gastroscope for upper GI tract visualization and therapy. Input: optical images captured via CCD sensor; illumination provided by 300W xenon lamp in external video processor. Operation: light transmitted via fiber bundles; CCD collects image data synchronized with RGB color wheel strobes; video processor compiles full-color frames. Output: real-time video displayed on monitor. Used in clinical settings by physicians. Features flexible insertion tube, angulation controls, air/water delivery, suction, and two working channels for biopsy forceps. Benefits: enables minimally invasive diagnostic and therapeutic access to GI organs.

Clinical Evidence

No clinical data; substantial equivalence based on bench testing and technical comparison.

Technological Characteristics

Flexible video gastroscope with CCD imaging sensor and fiber optic illumination. Connects to external video processor (300W xenon lamp, RGB color wheel). Features two working channels, angulation controls, air/water/suction ports. System requires software-controlled video processor for image compilation and display.

Indications for Use

Indicated for adult and pediatric patients requiring optical visualization and therapeutic access to the upper gastrointestinal tract, including the esophagus, stomach, duodenum, and small bowel.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} AUG-07-96 WED 11:54 PENTAX SESC FAX NO. 305 430 8505 P. 01 DOCUMENT NUMBER: KS034.B7 DATE PREPARED: 08-07-96 PREPARED BY: PAUL SILVA PAGE 1 OF 1 510(K) SUMMARY DEVICE: EG-3800T, VIDEO GASTROSCOPE 1. SUBMITTER INFORMATION: Pentax Precision Instrument Corporation 30 Ramland Road Orangeburg, NY, 10962 TEL: (914)-365-0700 SEP 16 1996 K962460 2. NAME OF DEVICE: Trade Name: EG-3800T, Video Gastroscope Classification Name: Gastroscope, Gastro- Urology (78FDS), 876.1500 3. PREDICATED DEVICE(S) INFORMATION: 1. FG-36TA, Fiber Gastroscope Pentax K833924 2. EC-3800TL, Video Colonoscope Pentax K951579 3. EPM-3300, Video Processor Pentax K934918 4. DEVICE DESCRIPTION: The EG-3800T, Video Gastroscope, must be used with a Video Processor (a software controlled device). The endoscope has a flexible insertion tube, a control body, and umbilicus. The umbilicus provides connection to the video processor. The control body includes controls for up/ down/ left/ right angulation, air/water delivery, suction, two (2) accessory inlet ports, and a suction channel selector. The device contains light carrying bundles, to illuminate the body cavity, and a charge couple device (CCD) to collect image data. The instrument contains two (2) working channels through which biopsy devices, or other devices, may be introduced (the instrument is supplied with two biopsy forceps). The Video Processor contains a 300 watt xenon lamp which provides white light that is filtered, via a Red, Green, Blue color wheel, and is focused at the connected video endoscope lightguide prong. The endoscope light carrying bundles present the color strobes to the body cavity and the CCD collects image data for each strobe of color. The video processor stores the CCD information until all three color strobes are completed and a full color frame is compiled. Image data and other screen display information are formatted and presented to the video outputs of the video processor for display. 5. INTENDED USE: The EG-3800T, Video Gastroscope, is intended to provide optical visualization (via a video monitor) of, and therapeutic access to, the Upper Gastrointestinal Tract. The Upper Gastrointestinal Tract includes, but is not limited to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, and Small Bowel. The instrument is introduced per orally when indications consistent with the requirement for procedure(s) are observed in Adult and Pediatric patient populations. 6. COMPARISON TO PREDICATED DEVICE(S): The submission for substantial equivalence included EG-3800T literature including specifications, the identification of standard set components, and identification of optional accessories, comparison tables were provides to illustrate the comparisons to the predicated devices. The submission for substantial equivalence was not based on an assessment of clinical performance data.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...