K062222 · Unisensor AG · FFX · Mar 15, 2007 · Gastroenterology, Urology
Device Facts
Record ID
K062222
Device Name
UNITIP HIGH RESOLUTION CATHETER
Applicant
Unisensor AG
Product Code
FFX · Gastroenterology, Urology
Decision Date
Mar 15, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1725
Device Class
Class 2
Indications for Use
The catheter probe is used to obtain a high resolution mapping of peristaltic activity or pressure within the organs of the gastrointestinal tract, and to allow storage of the corresponding data. The device is manufactured in the basic model variations listed below: 1. High Resolution Gastrointestinal Catheter with up to 40 pressure sensors: This model variation is intended for simple direct esophageal pressure, either at, above, or below the specific site of interest. Pressure information is provided at multiple sensor sites. Pressures may be measured continuously, at rest, or dynamically. 2. Gastrointestinal catheter, with pH sensors: This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except that this model variation includes up to 7 pH sensors. PH probes are added to allow measurements of pH in the stomach and/or esophagus. 3. Gastrointestinal catheter, with sleeve: This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except one or more of the pressure sensor sites will include a sleeve. The sleeve is added to broaden the pressure sensitive one for the sensor site, and reduce positioning sensitivity. 4. Gastrointestinal catheter, with impedance rings: This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except it includes up to 14 conductive rings for impedance studies. 5. Gastrointestinal catheter, with lumen: This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except it includes a lumen.
Device Story
UniTip High Resolution GI Catheter; ultra-miniature solid-state pressure sensors; pH sensors; impedance rings; lumen; silicon rubber sleeves. Input: physiological pressure, pH, and impedance signals from GI tract. Transformation: multiplexing sensor signals via scanning electronics; transmission to external recording equipment. Output: high-resolution pressure/peristaltic mapping data. Usage: clinical setting; operated by trained personnel. Benefit: enables detailed assessment of GI motility and pressure dynamics; aids clinical decision-making regarding GI function.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and risk management documentation.
Indicated for patients requiring high-resolution mapping of peristaltic activity or pressure within the gastrointestinal tract, including esophageal pressure monitoring, pH measurement, and impedance studies.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
Sierra Scientific Instruments, Inc., Motility Visualization System (K031169)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Number: K0622222 Date: Page 1 of 2
# 510(k) Summary
MAR 1 5 2007
#### Introduction
This summary is intended to comply with requirements of the SMDA and 21CFR §807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.
#### 510(k) Applicant & Holder
Unisensor AG Bahnstrasse 12a Attikon CH-8544 Switzerland
### 510(k) Correspondent
Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors 13605 West 7th Ave., Golden, CO USA Tel: 303-234-9412 / Fax: 303-234-9413
### Date Prepared
July 19, 2006
#### Proprietary Name of Device
UniTip High Resolution GI Catheter
#### Classification Name
Gastrointestinal motility monitoring system
#### 510(k) Classification
Class II FFX
#### Device Description and Intended Use
The Unisensor UniTip Gastrointestinal Pressure Sensor Catheter product line consists of multiple ultra-miniature solid-state pressure sensors, located at the distal end of a gastrointestinal catheter. pH sensors, silicon rubber sleeves, lumen, and conductive impedance rings, may also be included. There are scanning electronics for multiplexing the sensor signals, and an electrical connection located at the proximal end to carry the sensor element signals to an external gastrointestinal motility recording equipment. The UniTip GI catheters are produced in various lengths and catheter sizes.
This modification consists of increasing the maximum number of pressure sensor elements to 40. The closer proximity of the pressure sensor elements to each other allows for high-resolution
{1}------------------------------------------------
510(k) Number: K0622222 Date: Date: Date: Date: Page 2 of 2
mapping of pressures within the organs of the human gastrointestinal tract. This modification also increases the maximum number of pH sensors to 7, and the maximum number of impedence rings to 14.
### Predicate Devices
Unisensor AG, UniTip Pressure Sensor Catheter K003580
Sierra Scientific Instruments, Inc., Motility Visualization System K031169
### Biocompatibility
The requirements of the following standards have been used in part to establish substantial equivalence:
EN 30993-1/ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing"
The company did not conduct, nor depend on, clinical studies in order to establish substantial equivalence.
### Risk Management
This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program.
Unisensor AG believes that the UniTip High Resolution GI Catheter is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate devices.
.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing the wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
AR 1 5 2007
Unisensor AG c/o Mr. Robert N. Clark Medical Device Regulatory Advisors 13605 West 7" Avenue GOLDEN CO 80401
Re: K062222
Trade/Device Name: Unitip High Resolution Catheter Regulation Number: 21 CFR §876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: II Product Code: FFX Dated: February 22, 2007 Received: February 26, 2007
Dear Mr. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indivious for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Coometic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The greeral controls provisions of the Act include requirements for annual registration, listing of devines, good manufacturing practice, labeling, and prohibitions against misbranding and adulter of on.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 899. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a circular logo or seal. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo, and the word "Centennial" is below the letters. The text around the edge of the circle is difficult to read.
*Protecting and Promoting Public Health*
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): _________________
Device Name:
Indications for Use:
The catheter probe is used to obtain a high resolution mapping of peristaltic activity or pressure within the organs of the gastrointestinal tract, and to allow storage of the corresponding data.
The device is manufactured in the basic model variations listed below:
- 1. High Resolution Gastrointestinal Catheter with up to 40 pressure sensors:
This model variation is intended for simple direct esophageal pressure, either at, above, or below the specific site of interest. Pressure information is provided at multiple sensor sites. Pressures may be measured continuously, at rest, or dynamically.
- 2. Gastrointestinal catheter, with pH sensors:
This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except that this model variation includes up to 7 pH sensors. PH probes are added to allow measurements of pH in the stomach and/or esophagus.
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Ingram
Division of Reproductive, Abdominal 510k) Num
l of 2
{5}------------------------------------------------
- 3. Gastrointestinal catheter, with sleeve:
This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except one or more of the pressure sensor sites will include a sleeve. The sleeve is added to broaden the pressure sensitive one for the sensor site, and reduce positioning sensitivity.
- 4. Gastrointestinal catheter, with impedance rings:
This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except it includes up to 14 conductive rings for impedance studies.
- 5. Gastrointestinal catheter, with lumen
This model variation is the same as the High Resolution Gastrointestinal (GI) catheter described above, except it includes a lumen.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.