K013356 · Wilson-Cook Medical, Inc. · FCG · Oct 19, 2001 · Gastroenterology, Urology
Device Facts
Record ID
K013356
Device Name
WILSON-COOK ULTRASOUND BIOPSY NEEDLE
Applicant
Wilson-Cook Medical, Inc.
Product Code
FCG · Gastroenterology, Urology
Decision Date
Oct 19, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
Used for soft tissue biopsy through the accessory channel of an ultrasound endoscope.
Device Story
The Wilson-Cook Ultrasound Biopsy Needle is a sterile, single-use device designed for soft tissue biopsy. It is used in conjunction with an ultrasound endoscope, which provides ultrasound guidance for needle placement. The device features a side-cutting needle configuration and an echogenic tip to enhance visibility under ultrasound. It is operated by a clinician during endoscopic procedures to obtain tissue samples for diagnostic purposes. The device is a modification of previous designs, utilizing a PEEK sheath and a Nitinol (Titanium Nickel) stylet. The device is intended to assist in the diagnosis and staging of lesions by facilitating accurate tissue sampling.
Clinical Evidence
No clinical data. Substantial equivalence is supported by design verification, dimensional/functional testing, and biocompatibility assessment of patient-contacting materials.
Technological Characteristics
Biopsy needle with side-cutting configuration and echogenic tip. Materials: PEEK sheath, Nitinol (Titanium Nickel) stylet. Sterile, single-use, disposable. Operates via manual insertion through the accessory channel of an ultrasound endoscope. No electronic components or software.
Indications for Use
Indicated for soft tissue biopsy performed via the accessory channel of an ultrasound endoscope.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
# OCT 1 9 2001
### RE: SPECIAL 510(K): DEVICE MODIFICATION FOR THE WILSON-COOK ULTR BIOPSY NEEDLE
#### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS l.
# Submitted By:
Wilson-Cook Medical Inc. 4900 Bethania Station Road & 5951 Grassy Creek Boulevard Winston-Salem, NC 27105
### Device Description:
The Wilson-Cook Modified Ultrasound Biopsy Needle is used for soft tissue biopsy through the accessory channel of an ultrasound endoscope. This device is supplied sterile and intended for single use only.
| Trade Name: | Wilson-Cook Ultrasound Biopsy Needle |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Ultrasound Biopsy Needle |
| Classification Name/Code: | Set, Biopsy Needle and Needle, GU, 78 FCG |
| Classification: | FDA has classified similar devices as Class II, as<br>per 21 CFR § 876.1075. This device falls within<br>the purview of the Gastroenterology and Urology<br>Device Panel. |
| Performance Standards: | To the best of our knowledge, performance<br>standards for this device do not exist. |
| Intended Use: | Used for soft tissue biopsy through the accessory<br>channel of an ultrasound endoscope. |
### Predicate Device:
| PREDICATE DEVICE | MANUFACTURER | DOCUMENT CONTROL<br>NUMBER |
|----------------------------------|--------------------------|----------------------------|
| ProAct Biopsy Needle | ProAct Ltd. | K926559/A |
| Wilson-Cook Ultrasound<br>Needle | Wilson-Cook Medical Inc. | K934356 |
## Substantial Equivalence:
The Wilson-Cook Modified Ultrasound Biopsy Needle is substantially equivalent to the referenced predicate devices with respect to design, materials of construction and intended use.
Biopsy Needle in the side-cutting This device is identical to the ProAct configuration of the needle, as well as the echogenic needle tip. In addition, both of these devices operate in the same manner in order to obtain biopsies.
The modified Ultrasound Biopsy Needle and the predicate Wilson-Cook Ultrasound Needle are both used endoscopically, employing ultrasound technology to quide the needle. The patient-contacting materials of these devices differ slightly, with the sheath of the predicate being polytetrafluoroethylene and the sheath of the modified device being PEEK. In addition, the stylet of the predicate is comprised of ANSI Grade 304 Stainless Steel while the modified device stylet is comprised of Titanium Nickel (Nitinol). Applicable biocompatibility data is included.
{1}------------------------------------------------
K06 3356
Page 2 of 2
### RE: SPECIAL 510(K): DEVICE MODIFICATION FOR THE WILSON-COOK ULTR BIOPSY NEEDLE
| DEVICE<br>CHARACTERISTIC | Wilson-Cook Modified<br>Ultrasound Biopsy<br>Needle<br>[Subject of Special<br>510(k)] | ProAct Biopsy<br>Needle<br>(K926559/A) | Predicate<br>Wilson-Cook<br>Ultrasound Needle<br>(K934356) |
|--------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Used for soft tissue<br>biopsy through the<br>accessory channel of an<br>ultrasound endoscope. | Used for<br>percutaneous,<br>intra-operative, or<br>through a<br>laparoscope<br>biopsies. | Used for aspiration<br>biopsy to diagnose<br>and stage GI lesions,<br>providing sampling<br>by employing<br>ultrasound<br>technology to guide<br>the needle. |
| Sterility | Sterile, Disposable | Sterile, Disposable | Sterile, Disposable |
#### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS (continued) l.
# Biocompatibility:
Reasonable assurance of biocompatibility for the patient-contacting materials has been established through a history of use in similar patient-contacting medical devices and as applicable biocompatibility test results.
# Design Control/Risk Analysis/Design Verification:
Design Control, Risk Analysis, Design Verification activities for the subject of this special 510(k) have been conducted in accordance with all applicable internal procedures. The design control process employed is inclusive of the elements as stipulated by 21 CFR Part 820.30, as applicable to the project. The risk analysis performed identified the risks relative to the performance requirements, as specified by our internal procedure for Risk Analysis. The failure mode, effect of failure, severity, potential cause, rate of occurrence, design control element/production controls to eliminate, the potential to detect and our recommended actions were also documented. During Design Verification, dimensional and functional testing to ensure the performance and design integrity of this product line were conducted. A l l results obtained during our Design Verification met our predetermined acceptance criteria for this product line.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol featuring a staff with two snakes coiled around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH &" which is arranged in a semi-circular fashion. The text is in all caps.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 9 2001
Re: K013356
Trade/Device Name: Wilson-Cook Ultrasound Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II
Product Code: 78 FCG Dated: September 28, 2001 Received: October 10, 2001
Dear Ms. Posner:
Ms. Margaret J. Posner
Wilson-Cook Medical
GI Endoscopy
Regulatory Affairs Specialist
4900 Bethania Station Road
WINSTON-SALAM NC 27105
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your booker only of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the encreated of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r read of acrease a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page __ 12__ of ___ 12__
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: ____Wilson-Cook Ultrasound Biopsy Needle
Indications for Use:
Used for soft tissue biopsy through the accessory channel of an ultrasound endoscope.
(Please do not Write Below This Line – Continue on another Page, if Needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div style="display:inline-block; vertical-align:top;">✓</div> | OR | Over-The-Counter |
|----------------------|----------------------------------------------------------------|----|--------------------------|
| (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) |
Nancy C Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devices
| 510(k) Number | K013356 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.