HEMOBLAST™ Bellows

P170012 · Dilon Technologies, Inc. · PMX · Dec 15, 2017 · General, Plastic Surgery

Device Facts

Record IDP170012
Device NameHEMOBLAST™ Bellows
ApplicantDilon Technologies, Inc.
Product CodePMX · General, Plastic Surgery
Decision DateDec 15, 2017
DecisionAPPR
Regulation21 CFR 878.4490
Device ClassClass 3
AttributesTherapeutic

Intended Use

The HEMOBLAST™ Bellows is indicated in surgical procedures as an adjunct to hemostasis when control of minimal, mild, and moderate bleeding by conventional procedures is ineffective or impractical, except in neurosurgical, ophthalmic, and urological procedures.

Device Story

HEMOBLAST™ Bellows is a sterile, single-use applicator containing a dry hemostatic powder composed of porcine collagen, bovine chondroitin sulfate, and human pooled plasma-derived thrombin. Operated manually by a surgeon in an OR setting, the bellows system delivers powder directly to active bleeding sites. The powder absorbs blood, concentrates coagulation factors/platelets, and provides a scaffold for fibrin formation; chondroitin sulfate aids tissue cohesion. The device requires no preparation or mixing. It is used as an adjunct to conventional methods (pressure, sutures, cautery) to achieve hemostasis in minimal to moderate bleeding. The powder is absorbed in vivo over 4 weeks. Clinical benefit includes faster time-to-hemostasis compared to traditional gelatin-thrombin agents.

Clinical Evidence

PMA approval based on a prospective, randomized, controlled, multicenter pivotal study (n=242) comparing HEMOBLAST™ to a gelatin-thrombin (G+T) control. Primary endpoint: non-inferiority for hemostasis within 6 minutes. Results: 93.1% success for HEMOBLAST™ vs 73.5% for G+T (p<0.0001). Secondary endpoints: HEMOBLAST™ showed superiority in hemostasis at 3 and 6 minutes and significantly faster preparation time (0.37 min vs 2.40 min). Safety profile was comparable to control; no unanticipated adverse device effects. Increased porcine collagen antibody titers observed but without clinical sequelae.

Technological Characteristics

Materials: Porcine collagen (with glucose), bovine chondroitin sulfate, human pooled plasma-derived thrombin. Applicator: Polycarbonate bellows and 10 cm nozzle extension. Energy: None (manual). Sterilization: Gamma-irradiation (ISO 11137). Biocompatibility: ISO 10993 compliant. Form factor: Powder-filled bellows applicator. Connectivity: None.

Indications for Use

Indicated for patients 21 years or older undergoing open cardiothoracic, abdominal, or orthopedic lower extremity surgery where conventional hemostasis is ineffective or impractical. Contraindicated in patients with known allergies to porcine, bovine, or human plasma-derived materials; do not inject into vessels or tissue; do not use in absence of active blood flow; do not use for severe/extreme bleeding or skin incision closure.

Regulatory Classification

Identification

An absorbable hemostatic agent or dressing is a device intended to produce hemostasis by accelerating the clotting process of blood. It is absorbable.

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED) I. GENERAL INFORMATION Device Generic Name: Hemostatic agent Device Trade Name: HEMOBLAST™ Bellows Device Procode: PMX Applicant’s Name and Address: Biom’Up SA 8, Allée Irène Joliot-Curie 69800 Saint Priest France Date(s) of Panel Recommendation: None Premarket Approval Application (PMA) Number: P170012 Date of FDA Notice of Approval: December 15, 2017 II. INDICATIONS FOR USEC The HEMOBLAST™ Bellows is indicated in surgical procedures as an adjunct to hemostasis when control of minimal, mild, and moderate bleeding by conventional procedures is ineffective or impractical, except in neurosurgical, ophthalmic, and urological procedures. III. CONTRAINDICATIONS Do not inject HEMOBLAST™ Bellows into a vessel or tissue. There is a risk of allergic-anaphylactoid reaction and/or thromboembolic events, which may be life-threatening. Do not apply HEMOBLAST™ Bellows in the absence of active blood flow, e.g., while the vessel is clamped or bypassed. Extensive intravascular clotting and even death may result. Do not use the HEMOBLAST™ Bellows for treatment of severe or extreme bleeding. Do not administer to patients with known allergies or hypersensitivity to materials of porcine or bovine origin. Do not use in the closure of skin incisions because it may interfere with the healing of the skin edges due to mechanical interposition of the powder. IV. WARNINGS AND PRECAUTIONS The warnings and precautions can be found in the HEMOBLAST™ Bellows labeling. PMA P170012: FDA Summary of Safety and Effectiveness Data Page 1 {1} # V. DEVICE DESCRIPTION The HEMOBLAST™ Bellows consists of the hemostatic powder (HEMOBLAST™ Bellows Hemostatic Powder) supplied in an applicator system incorporating a bellows design. The HEMOBLAST™ Bellows Hemostatic Powder is dry, sterilized, biocompatible, and non-pyrogenic. No intraoperative preparation, mixing, or heating is required. It absorbs in vivo over a 4-week period. HEMOBLAST™ Bellows Hemostatic Powder is composed predominantly of highly purified porcine collagen (with glucose) with smaller amounts of bovine chondroitin sulfate (CS) and human pooled plasma derived thrombin. These components, all of which are supplied in powder form, are shown in Table 1 below. Table 1: HEMOBLAST™ Bellows Hemostatic Powder Composition | HEMOBLAST™ Bellows Hemostatic | Source | | --- | --- | | Collagen Powder (with glucose) | Porcine | | Chondroitin Sulfate | Bovine | | Thrombin | Human pooled plasma | The HEMOBLAST™ Bellows Applicator contains $1.65 \pm 0.05$ g of HEMOBLAST™ Bellows Hemostatic Powder. The product is sterilized using gamma-sterilization and provided in double-packaging. ![img-0.jpeg](img-0.jpeg) Figure 1: HEMOBLAST™ Bellows Applicator The HEMOBLAST™ Bellows 10 cm Nozzle Extension, which is made of polycarbonate tubing, serves to assist in the delivery of the HEMOBLAST™ Bellows Hemostatic Powder, during surgery. The HEMOBLAST™ Bellows 10 cm Nozzle Extension can be used to assist with the application of the powder to active bleeding sites where a slightly longer tip is desired by the surgeon. It has not been tested or designed for laparoscopic PMA P170012: FDA Summary of Safety and Effectiveness Data {2} application or for application at locations where the surgeon cannot clearly visualize the site of active bleeding. ![img-1.jpeg](img-1.jpeg) Figure 2: HEMOBLAST™ Bellows 10 cm Nozzle Extension ## Principles of Operation The HEMOBLAST™ Bellows applies the hemostatic agent to the source of the bleeding via manual squeezing of the bellows. The implant material is applied to cover the entire target bleeding site. The hemostatic agent absorbs excess blood and helps in achieving hemostasis. The hemostatic effect of the HEMOBLAST™ Bellows Hemostatic Powder is due to its collagen composition, which activates the coagulation process by absorbing blood, concentrating coagulation factors and platelets and providing a surface for coagulation to initiate. Thrombin from pooled human plasma, as an ancillary blood derivative, is added to the formulation to aid the effect of the hemostatic agent. The powder thrombin facilitates the conversion of fibrinogen to fibrin. In addition, CS, in powder form, is included in the HEMOBLAST™ Bellows Hemostatic Powder formulation in order to provide cohesion between the hemostatic wound and the surrounding tissue. Once applied to the target bleeding site, the powder must remain in contact with the source of bleeding to ensure hemostasis. A wet laparotomy pad, temporarily applied to the treated bleeding site, keeps the powder in contact with the bleeding area. ## VI. ALTERNATIVE PRACTICES AND PROCEDURES There are several other alternatives to control bleeding. Conventional procedures include the use of direct pressure, sutures, and/or electrocautery. Other commercially available devices are hemostats, sealants, and adhesives. These devices include: hemostats composed of gelatin, bovine collagen, cellulose, polysaccharide spheres, thrombin, gelatin and thrombin, and thrombin and fibrinogen; sealants composed of fibrinogen and thrombin, polyethylene glycol polymers with or without albumin, albumin and aldehyde, and cyanoacrylate; and adhesives composed of cyanoacrylate, albumin and aldehyde, and thrombin and fibrinogen. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets their expectations and lifestyle. PMA P170012: FDA Summary of Safety and Effectiveness Data {3} PMA P170012: FDA Summary of Safety and Effectiveness Data Page 4 # VII. MARKETING HISTORY HEMOBLAST™ Bellows was approved for marketing and sale in the European Union in 2016, and the product is currently marketed in France and Germany. HEMOBLAST™ has never been removed from any market for any reason. The HEMOBLAST™ Bellows has not been marketed in the United States. # VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH Below is a list of the potential adverse effects (e.g., complications) associated with the use of HEMOBLAST™ include: - Adhesion formation; - Allergy or anaphylaxis; - Blockage of cardiopulmonary bypass system and cell saver devices; - Compromised attachment of orthopedic implants; - Creutzfeldt-Jakob disease (CJD) agent; - Increased infection; - Lack of effectiveness; - Nerve compression; - Thrombosis or thromboembolism; - Transmissible Spongiform Encephalopathies (TSE); and - Viral disease transmission. For the specific adverse events that occurred in the clinical studies of HEMOBLAST™, please see Section X below. # IX. SUMMARY OF NONCLINICAL STUDIES ## A. Laboratory Studies Biocompatibility testing of HEMOBLAST™ Bellows has been performed according to GLP regulations and pursuant to ISO 10993 and FDA’s guidance document entitled “Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process” (June 16, 2016). Cytotoxicity, Sensitization, and Intradermal Irritation testing were completed on the HEMOBLAST™ Bellows Applicator and HEMOBLAST™ Bellows 10 cm Nozzle Extension; and Cytotoxicity, Sensitization, Intradermal Irritation, Systemic Toxicity, Genotoxicity, and Implantation testing were completed on the HEMOBLAST™ Bellows Hemostatic Powder. The testing demonstrated that the HEMOBLAST™ Bellows is biocompatible as summarized in the tables below. {4} Table 2: HEMOBLAST™ Bellows Applicator (With Powder) - Biocompatibility | Test | Test Article | Results | | --- | --- | --- | | Qualitative Cytotoxicity Study on Extract [According to ISO 10993-5 (2009) / GLP (21 CFR 58 - 2002)] | Hemostatic powder in a bellows | PASS | | ISO Intracutaneous Study in Rabbits [According to ISO 10993-10 (2010) / GLP (21 CFR 58 - 2002)] | Hemostatic powder in a bellows | PASS | | ISO Guinea Pig Maximization Sensitization [According to ISO 10993-10 (2010) / GLP (21 CFR 58 - 2002)] | Hemostatic powder in a bellows | PASS | | ISO Acute Systemic Toxicity Study in Mice [According to ISO 10993-11 (2006) / GLP (21 CFR 58 - 2002)] | Hemostatic powder in a bellows | PASS | | Mouse Peripheral Blood Micronucleus Study [According to ISO 10993-3 (2003) / GLP (21 CFR 58 - 2002)] | Hemostatic powder in a bellows | PASS | | Genotoxicity: Mouse Lymphoma Assay [According to ISO-10993-3 (2003) / ISO 10993-12 (2012) / GLP (21 CFR 58 - 2002)] | Hemostatic powder in a bellows | PASS | | Genotoxicity: Bacterial Reverse Mutation Study [According to ISO-10993-3 (2003) / GLP (21 CFR 58 - 2002) | Hemostatic powder in a bellows | PASS | | Qualitative Cytotoxicity Study on Extract [According to ISO 10993 Part5 / GLP (21 CFR 58– 2002)] | Hemostatic powder in a bellows | PASS | Table 3: HEMOBLAST™ Bellows Applicator (Without Powder) - Biocompatibility | Test | Test Article | Results | | --- | --- | --- | | ISO MTS Cytotoxicity Test [According to ISO 10993-5 (2009) / GLP (21 CFR 58 - 2002)] | Bellows | PASS | | ISO Intracutaneous in Rabbits [According to ISO 10993-10 (2010) / GLP (21 CFR 58 - 2002)] | Bellows | PASS | | ISO Guinea Pig Maximization Sensitization Test [According ISO 10993-10 (2010) / GLP (21 CFR 58 - 2002)] | Bellows | PASS | | 4-Week Systemic Toxicity Study in Rats Following Subcutaneous Implantation [According to ISO 10993-6 (2007) and -11 (2006) / GLP (21 CFR 58 - 2002)] | Bellows | PASS | Table 4: HEMOBLAST™ Bellows 10 cm Nozzle Extension - Biocompatibility | Test | Test Article | Results | | --- | --- | --- | | ISO MTS Cytotoxicity Test according to ISO 10993-5 [According to ISO 10993-5 (2009) / GLP (21 CFR 58 - 2002)] | Plastic cannula | PASS | | ISO Intracutaneous Study in Rabbits [According to ISO 10993-10 (2010) / GLP (21 CFR 58 - 2002)] | Plastic cannula | PASS | | ISO Guinea Pig Maximization Sensitization Test [According to ISO 10993-10 (2010) / (21 CFR 58 - 2002)] | Plastic cannula | PASS | The additional non-clinical tests were conducted to provide additional device characterization. This testing is summarized in the tables below. PMA P170012: FDA Summary of Safety and Effectiveness Data {5} PMA P170012: FDA Summary of Safety and Effectiveness Data Page 6 # Applicator System Testing Table 5: Applicator System Testing Studies Overview | Test | Test Purpose | Results | | --- | --- | --- | | Bench Testing on the HEMOBLAST™ Bellows Applicator: Coverage Test of Target Disks by the Hemostatic Powder Delivered by the Bellows | Evaluated: -Delivery accuracy and performance. | PASS | | Handling Performance Evaluation of a Bellows (HEMOBLAST™ Bellows) Delivering Powdered Hemostat | Evaluated: Leakage; Delivery. | PASS | | HEMOBLAST™ Bellows Applicator Verification Program | Evaluated: Mechanical performance of the HEMOBLAST Bellows Applicator; The effects of ageing, acute temperature and gamma irradiation on HEMOBLAST™ Bellows Applicator. | PASS | # Nozzle Extension Testing Table 6: Applicator System Testing Studies Overview | Test | Test Purpose | Results | | --- | --- | --- | | Powder Leak Evaluation of HEMOBLAST™ Bellows Without or With Applicator in a Bench Top Model | Evaluated: -Usability, handling and delivery characteristics. | PASS | | Delivery Performance Evaluation of HEMOBLAST™ Bellows Without or With Applicator in a Bench Top Model | Evaluated: -Delivery performance. | PASS | | Coverage of Target Disks by HEMOBLAST Bellows Without or With Cannula in a Bench Top Model | Evaluated: -Delivery performance. | PASS | | HEMOBLAST Bellows Applicator Verification Program | Evaluated: -Mechanical performance characteristics. | PASS | Viral inactivation for the collagen and CS was validated. Stability testing for the product was conducted and the results were acceptable. {6} The device is packaged in a double-tray (i.e., blister) configuration and sterilized using gamma-irradiation per ISO 11137. Validation of the packaging, including sealing of the inner and outer trays, has been performed. Real-time stability and shelf-life of the finished sterilized product is ongoing. The testing completed to date supports the labeled shelf-life. The non-clinical tests of the HEMOBLAST™ Bellows demonstrated that HEMOBLAST™ Bellows is safe for use in surgical procedures as an adjunct to hemostasis when control of minimal, mild, and moderate bleeding by conventional procedures is ineffective or impractical, except in neurosurgical, ophthalmic, and urological procedures. ## B. Animal Studies Nine (9) porcine and three (3) rabbit non-clinical studies of the safety and effectiveness of the HEMOBLAST™ Bellows were performed. This testing is summarized in the table below. Table 7: Non-clinical Animal Studies Overview | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Resorption, local tissue effects, and performance in a rabbit orthopedic model | This study was conducted to assess resorption, local tissue effects, and hemostatic performance of the device in comparison to a control device. Cortico-trabecular defects in the femoral condyle and periosteal defects in the tibia were evaluated. | Rabbit = 7 Time points = day 0, 4 weeks, 6 weeks Cortico-trabecular defect in each medial femoral condyle (2 per rabbit for 6 rabbits and 1 per rabbit for T0 rabbit). Periosteal defect in the diaphysis of each tibia (2 per rabbit for 6 rabbits and 1 per rabbit for T0). Hemoblast Bellows: 14 defects Control Device: 12 defects | 100% of sites tested achieved hemostasis 3 minutes after a single application. **Femoral sites:** Device: Bone/cartilage outgrowth identified at test sites in 2 of 4 animals at 4 weeks and 1 of 2 rabbits at 6 weeks Control: Bone/cartilage outgrowth identified in 3 of 4 animals at 4 weeks and in 1 of 2 rabbits at 6 weeks. **Tibial sites:** Device: Bone/cartilage outgrowth identified at test sites in 2 of 4 rabbits at 4 weeks and 1 of 2 rabbits at 6 weeks Control: Similar swellings were observed in 2 of 4 rabbits at 6 weeks. The tibial sites were not fully healed at 4 weeks. | PMA P170012: FDA Summary of Safety and Effectiveness Data {7} | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Hemoblast Bellow Hemostatic Agent GLP Safety and Efficacy Study in a Swine Vascular Model | The safety and efficacy of Hemoblast Bellows was evaluated in a swine vascular surgery model using procedural data, adverse events, and histopathology in comparison to a control. | Swine = 12 Time points = 72 hours, 30 days Carotid arteriotomy (simple closure or patch closure) Femoral arteriotomy (simple closure or patch closure) Hemoblast Bellows: 18 sites (9 simple closure, 9 patch) Control: 18 sites (9 simple closure, 9 patch) | **Hemostasis:** **Carotid artery:** Hemoblast Bellows: 6 minutes: 50% of sites 10 minutes: 83.3% of sites Control: 6 minutes: 100% of sites 10 minutes: 100% of sites **Femoral artery:** Hemoblast Bellows: 6 minutes: 83.3% of sites 10 minutes: 100% of sites Control: 6 minutes: 100% of sites 10 minutes: 100% of sites **Simple Closure:** Hemoblast Bellows: 6 minutes: 88.9% of sites 10 minutes: 99.9% of sites Control: 6 minutes: 100% of sites 10 minutes: 99.9% of sites **Patch Closure:** Hemoblast Bellows: 6 minutes: 33.3% of sites 10 minutes: 82.1% of sites Control: 6 minutes: 100% of sites 10 minutes: 99.9% of sites All treated vessels were observed angiographically at day 0, 72 hours, and day 30 post-treatment. All vessels were observed to be patent at these time points. No thrombi were seen in any of the systemic tissues evaluated at either time point. There was no histologic evidence of the article at the 30 day time point. | PMA P170012: FDA Summary of Safety and Effectiveness Data Page 8 {8} | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Resorption, Local Tissue Effects and Performance of a Hemostatic Device Following Application on Liver Wounds in Pigs | Resorption, hemostatic performance, and local tissue effects were evaluated in a pig liver model with post-treatment follow-up at 4 weeks. | Swine = 2 Time points: 3, 6, and 10 minutes, 4 weeks Liver: Defects (8 mm in diameter 2 to 3 mm deep) Hemoblast Bellows: 10 defects Sham: 6 defects | Hemostasis Hemoblast Bellows: 3 minutes: 50% (5 of 10) 6 minutes: 90% (9 of 10) 10 minutes: 100% (10 of 10) Sham: 3 minutes: 0% (0 of 0) 6 minutes: 0% (0 of 0) 10 minutes: 0% (0 of 0) Histopathological analysis classified the device as a non-irritant in comparison to a sham procedure site. The device was not macroscopically or microscopically detected at the 4 week time point. | PMA P170012: FDA Summary of Safety and Effectiveness Data Page 9 {9} | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Resorption, Local Tissue Effects, and Performance of a Hemostatic Device Following Application on Bone Defects in Rabbits | Resorption, local tissue effects and hemostatic performance were evaluated in a rabbit orthopedic surgical model. | Rabbits = 10 Time points: day 0 and 4 weeks Cortico-trabecular defect in the medial femoral condyle. Hemoblast Bellows: 8 defects Sham: 7 defects | Hemostasis: 3 minutes: 100% (8 of 8) 6 minutes: 100% (8 of 8) 10 minutes: 100% (8 of 8 )) Sham: 3 minutes: 1 of 6 (14%) 6 minutes: 1 of 6 (14%) 10 minutes: 1 of 6 (14%) Histopathological analysis classified the device as a slight irritant in comparison to a sham procedure site. The device was not macroscopically or microscopically detected at the 4 week time point. 2 of 9 device treated defects were not healed at the 4 week time point. | PMA P170012: FDA Summary of Safety and Effectiveness Data Page 10 {10} | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Hemoblast Bellows Hemostatic Agent Acute GLP Efficacy Study in Swine Model | Hemostatic efficacy was assessed in a swine vascular, orthopedic, and hepatic injury model. | Swine = 5 Time points: 3, 6, and 10 minutes Vascular: Arteriotomies with simple closure (5 to 12 mm in length) Hemoblast Bellows: 13 defects Control Device 1: 13 defects Control Device 2: 13 defects Liver: Defects (8 mm in diameter and 2-7 mm deep) Hemoblast Bellows: 18 defects Control Device 1: 18 defects Control Device 2: 18 defects Orthopedic: Femoral condyle defects (1.5 to 4.2 cm²) Hemoblast Bellows: 12 defects Control Device 1: 12 defects Control Device 2: 12 defects | Hemostasis: Hemoblast Bellows: 3 minutes: 63.9% of sites 6 minutes: 92% of sites 10 minutes: 100% of sites Control Device 1: 3 minutes: 41.0% of sites 6 minutes: 53.8% of sites 10 minutes: 100% of sites Control Device 2: 3 minutes: 61.4% of sites 6 minutes: 76.3% of sites 10 minutes: 100% of sites | PMA P170012: FDA Summary of Safety and Effectiveness Data Page 11 {11} | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Hemostasis Maintenance at 24 hours and Performance of a Hemostatic Device Following Application on Liver Wounds in Pigs | Acute hemostasis and maintenance of hemostasis at 24 hours was evaluated in a swine hepatic model using two versions of the hemostatic device (Test article 1: Subject device, Test article 2: 7% chondroitin sulfate). | Swine = 3 Time points: 3, 6, and 10 minutes, 24 hours Liver: Defects (8mm in diameter and 2 to 3 mm deep) Test Article 1: 12 defects Test Article 2: 12 defects | **Hemostasis** Test Article 1: 3 minutes: 67% (8 of 12) 6 minutes: 100% (12 of 12) 10 minutes: 100% (12 of 12) 24 hours: 100% (12 of 12) Test Article 2: 3 minutes: 58% (7 of 12) 6 minutes: 83% (10 of 12) 10 minutes: 92% (11 of 12) 24 hours: 92% (11 of 12) | | Hemostasis Maintenance at 24 hours and Performance of a Hemostatic Device Following Application on Bone Defects in Rabbits | Acute hemostasis and maintenance of hemostasis at 24 hours was evaluated in a rabbit orthopedic model using two versions of the hemostatic device (Test article 1: Subject Device, Test article 2: 7% chondroitin sulfate). | Rabbits = 10 Time points: 3, 6, and 10 minutes, 24 hours Femoral condyle defects Test Article 1: 10 defects Test Article 2: 10 defects | **Hemostasis** Test Article 1: 3 minutes: 100% (10 of 10) 6 minutes: 100% (10 of 10) 10 minutes: 100% (10 of 10) 24 hours: 100% (10 of 10) Test Article 2: 3 minutes: 100% (10 of 10) 6 minutes: 100% (10 of 10) 10 minutes: 100% (10 of 10) 24 hours: 100% (10 of 10) | PMA P170012: FDA Summary of Safety and Effectiveness Data Page 12 {12} PMA P170012: FDA Summary of Safety and Effectiveness Data Page 13 | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Performance, resorption, and local tissue effects in a porcine liver model | Hemostatic performance, local tissue effects, and preparation time were evaluated for two methods of hemostatic powder delivery: bellows and prototype | Swine = 1 Time points: 3, 6, and 10 minutes, 24 hours Liver defect (2 cm² and 2 to 3 cm in depth) Bellows: 4 sites Prototype: 4 sites | **Hemostasis:** Hemoblast Bellows: 3 minutes: 0% (0 of 4) 6 minutes: 75% (3 of 4) 10 minutes: 100% (4 of 4) Prototype: 3 minutes: 50% (2 of 4) 6 minutes: 50% (2 of 4) 10 minutes: 100% (4 of 4) No delayed bleeding was observed 24 hours after product application for both delivery devices. | | Study of the Efficacy and Tolerance of a New Hemostatic Device in a Swine Model | Hemostatic performance, adhesion formation, risk of air embolism, and local tissue reaction was assessed in a swine hepatic injury model where the hemostatic powder was applied with a prototype device with a short applicator (10 cm) and a long applicator (30 cm). | Swine = 8 Time points: 3, 6, and 10 minutes Liver defect (12 mm in diameter and 3 to 4 mm deep) Short applicator: 6 sites Long applicator: 6 sites Control: 4 sites | **Hemostasis:** Prototype: 6 minutes: 67% (8 of 12) 10 minutes: 92% (11 of 12) Control: 6 minutes: 100% (4 of 4) 10 minutes: 110% (4 of 4) Tissue attachments/adhesions were present at 100% of treated sites (prototype and control at 1 week and 4 weeks. No significant difference in tenacity was observed. No device was visible at the 4 week endpoint. | {13} | Study Name | Purpose | Study Design | Study Results | | --- | --- | --- | --- | | Acute Hemostatic Performance Study in a Pig Hepatic Bleeding Model | Hemostatic performance was evaluated for two methods of hemostatic powder delivery. | Swine = 1 Time points: 3, 6, and 10 minutes Liver defect (8 mm in diameter and 1 mm deep) 10 sites per article. | Hemostasis: 3 minute: 70% (7 of 10) 6 minute: 90% (9 of 10) 10 minute: 100% (10 of 10) | | Acute Hemostatic Performance in a Pig Bony Tissue Bleeding Model | Hemostatic performance was evaluated in a pig orthopedic bleeding model for two methods of hemostatic powder delivery. | Swine = 2 Time points: 3, 6, and 10 minutes Femoral condyle defect (3 to 5 cm²) 10 sites per article. 2 sites for negative control. | Hemostasis: 3 minutes: 70% (7 of 10) 6 minutes: 100% (10 of 10) 10 minutes: no evaluation performed | | Study of the Efficacy of a Hemostatic Device in an Acute Anticoagulated Swine Model | Hemostatic efficacy of the device when applied with a prototype device was assessed in an open hepatic injury model in pigs undergoing treatment with aspirin. | Swine = 3 Time points: 3, 6, and 10 minutes Liver defect (8 mm in diameter, 3 to 4 mm deep) Prototype device: 48 sites Control: 24 sites Sham: 24 sites | Hemostasis: 3 minutes: 50% (24 of 48) 6 minutes: 85% (41 of 48) 10 minutes: 98% (47 of 48) | X. SUMMARY OF PRIMARY CLINICAL STUDIES The applicant performed clinical studies to establish a reasonable assurance of safety and effectiveness of general surgical procedures with Hemoblast Bellows as an adjunct to hemostasis when control of minimal, mild, and moderate bleeding by conventional procedures is ineffective or impractical, except in neurosurgical, ophthalmic, and urological procedures in the US under IDEs G150037 and G160063. Data from these clinical studies were the basis for the PMA approval decision. A summary of the clinical studies is presented below. PMA P170012: FDA Summary of Safety and Effectiveness Data {14} # Pilot Clinical Study: # A. Study Design This was a prospective, multicenter, single-arm pilot clinical investigation to collect data to support the use of the Surface Bleeding Severity Scale ("SBSS"), and to collect initial safety and efficacy data on HEMOBLAST™ Bellows in clinical use. The Pilot Study enrolled 31 subjects; 27 subjects were included in the safety population and 24 subjects were included in the effectiveness population. Subjects enrolled into the study were required to have a target bleeding site with SBSS score of 1 (minimal bleeding), 2 (mild bleeding), or 3 (moderate bleeding). This 6-week acute study was a single arm study which included patients undergoing only orthopedic and abdominal surgeries with associated bleeding sites; there was no cardiothoracic arm. The bleeding scale elements used in the SBSS are provided in the table below. Table 8: Surface Bleeding Severity Scale | Surface Bleeding severity Scale Score | 0 | 1 | 2 | 3 | 4 | 5 | | --- | --- | --- | --- | --- | --- | --- | | Verbal Descriptor | None | Minimal | Mild | Moderate | Severe; not immediately life-threatening | Extreme; immediately life-threatening | | Visual Descriptor | Dry | Oozing | Pooling | Flowing | Streaming | Gushing | | Expected Intervention(s) | None | Manual pressure, cautery, adjuvant hemostat(s) | Manual pressure, cautery, suture, adjuvant hemostat(s) | Manual pressure, cautery, suture, adjuvant hemostat(s) | Manual pressure, cautery, suture, staples, tissue repair | Manual pressure, cautery, suture, staples, tissue repair | | Maximum Expected ACS-ATLS Shock Risk Class | 1 | 1 | 1 | 2 | 3 | 4 | | ACS-ATLS Shock Risk Class: 1 – involves up to 15% of blood volume; typically no change in vital signs and fluid resuscitation is not usually necessary. Class 2 – involves 15-30% of total blood volume; patient is often tachycardic with a narrowing of the difference between the systolic and diastolic blood pressures; the body attempts to compensate with peripheral vasoconstriction; skin may start to look pale and be cool to the touch; volume resuscitation with crystalloids is all that is typically required; blood transfusion is not typically required. Class 3 – involved loss of 30-40% of circulating blood volume; patient's blood pressure drops; heart rate increases, peripheral hypoperfusion worsens; fluid resuscitation with crystalloid and blood transfusion are usually necessary. Class 4 – involves loss of >40% of circulating blood volume; the limit of the body's compensation is reached and aggressive resuscitation is required to prevent death.References:1. Spotnitz WD, Zielske D, Centis V et al. The SPOT GRADE: a new method for reproducibly quantifying surgical | | | | | | | PMA P170012: FDA Summary of Safety and Effectiveness Data {15} wound bleeding. Spine (Phila Pa 1976). 2017 Oct 10. doi: 10.1097/BRS.0000000000002447. [Epub ahead of print]. 2. Kortbeek JB, Al Turki SA, Ali J et al. Advanced trauma life support. 8th edition, the evidence for change. J Trauma 2008;64:1638-50. # 1. Clinical Inclusion and Exclusion Criteria Enrollment in the Pilot Study was limited to subjects who met the following preoperative inclusion criteria: - Subject is undergoing elective open abdominal or orthopedic lower extremity surgery; - Subject or an authorized legal representative is willing and able to give prior written informed consent for investigation participation; - Subject on antiplatelets, including aspirin, will discontinue medication at least 10 days prior to surgery; and - Subject is 21 years of age or older. In addition, subjects must have met the following intraoperative inclusion criteria: - Subject does not have an active or suspected infection at the surgical site; - Subject in whom the Investigator is able to identify a target bleeding site (TBS) for which any applicable conventional means for achieving hemostasis are ineffective or impractical; and - Subject has a TBS with an SBSS score of 1, 2, or 3. Subjects were not permitted to enroll in the Pilot Study if they met any of the following exclusion criteria: - Subject is undergoing a laparoscopic, thoracoscopic, or robotic surgical procedure; - Subject is undergoing a spinal surgical procedure; - Subject is undergoing a neurologic surgical procedure; - Subject is undergoing an emergency surgical procedure; - Subject is pregnant, planning on becoming pregnant during the follow-up period, or actively breast-feeding; - Subject has a clinically significant coagulation disorder or disease, defined as a platelet count &lt; 100,000 per microliter and/or International Normalized Ratio &gt; 1.5 within 4 weeks of surgery; - Subject had chronic corticosteroid use within 2 weeks prior to surgery; - Subject receiving intravenous heparin or oral Coumadin within 24 hours of surgery; - Subject has an active or suspected infection at the surgical site; - Subject has had or has planned any organ transplantation; PMA P170012: FDA Summary of Safety and Effectiveness Data {16} - Subject has a known sensitivity or allergy to bovine and/or porcine substance(s) or any other component(s) of the hemostatic agent; - Subject has ASA classification of &gt; 4; - Subject has a life expectancy of less than 3 months; - Subject has a known psychiatric disorder, which in the opinion of the Principal Investigator, would preclude the subject from completing this clinical study; - Subject has documented severe congenital or acquired immunodeficiency; - Subject has religious or other objections to porcine or bovine components; - Subject in whom the investigational device will be used at the site of a cemented or uncemented porous coated joint implant; - Subject is currently participating or has participated in another clinical trial within the past 30 days and is receiving/has received an investigational drug, device, or biologic agent; and - Subject is not appropriate for inclusion in the clinical trial, per the medical opinion of the Principal Investigator. 2. Follow-up Schedule All enrolled subjects were evaluated in the immediate postoperative period (within 24 hours after surgery) and at 6 weeks postoperatively (visit 4-8 weeks postoperatively). 3. Clinical Endpoints Primary endpoint: The primary endpoint of this clinical investigation is the mean paired Kappa statistic for the assignment of SBSS scores by two (2) Investigators. Secondary endpoints: Secondary endpoints of this clinical investigation consist of: - Proportion of subjects achieving hemostasis within 6 minutes of HEMOBLAST™ Bellows application; - Proportion of subjects achieving hemostasis within 10 minutes of HEMOBLAST™ Bellows application; - Proportion of subjects achieving hemostasis within 3 minutes of HEMOBLAST™ Bellows application; and - Incidence of adverse events through final follow-up. B. Accountability A total of 31 subjects met preoperative eligibility criteria, including roll-in subjects. Four (4) subjects were excluded from the study intraoperatively due to not meeting intraoperative inclusion criteria, resulting in a total of 27 enrolled subjects, including the roll-in subjects. These 27 subjects are considered the safety analysis population. PMA P170012: FDA Summary of Safety and Effectiveness Data Page 17 {17} There was one roll-in subject at each site and roll-in subjects are not included in the effectiveness analysis. Therefore, 24 subjects enrolled in the study are considered the effectiveness analysis population. The exploratory analysis population includes 25 subjects total (the effectiveness analysis population plus one subject who met preoperative eligibility requirements but failed intra-operative eligibility criteria). Enrollment was fairly balanced across all three (3) sites, with nine (9) subjects being enrolled at Site 2 and 8, and 10 subjects enrolled at Site 6 and Site 8, respectively. Site 2 had three (3) subjects that were excluded intraoperatively, Site 08 had one (1) subject that was excluded intraoperatively, while no subjects were excluded intraoperatively at Site 6. Table 9: Number of Subjects per Site | Population | All | Site 2 | Site 6 | Site 8 | | --- | --- | --- | --- | --- | | Total: Preoperative eligible + roll-in subjects | 31 | 12 | 8 | 11 | | Intraoperative ineligible subjects | 4 | 3 | 0 | 1 | | SAFETY ANALYSIS POPULATION: Preoperative AND Intraoperative eligible (ENROLLED and includes roll-in subjects) | 27 | 9 | 8 | 10 | | Roll-in subjects | 3 | 1 | 1 | 1 | | EFFICACY ANALYSIS POPULATION: Preoperative AND Intraoperative eligible (ENROLLED and EXCLUDES roll-in subjects) | 24 | 8 | 7 | 9 | | EXPLORATORY ANALYSIS POPULATION*: Meet preoperative eligibility criteria but fail intraoperative eligibility criteria due to ineligible SBSS scores PLUS EFFICACY ANALYSIS POPULATION (excludes roll-in subjects) | 25 | 9 | 7 | 8 | *Four subjects were preoperatively eligible but then did not meet intraoperative criteria 02-04, 02-07, 02-08 and 08-03. Only subject 02-07 was excluded intraoperatively due to an SBSS score. Per the Statistical Analysis Plan, the exploratory analysis population should include the subjects intraoperatively ineligible resulting from one or both investigators providing a baseline SBSS score for the TBS that is in the ineligible range (i.e., SBSS 0, 4, or 5). Therefore, subject 02-07 was the only preoperatively eligible subject included in the exploratory analysis population. Table 10: Number of Subjects in Each Analysis Population | Population | All | Abdominal | Orthopedic | | --- | --- | --- | --- | | TOTAL: Preoperative eligible + roll-in subjects | 31 | 12 | 19 | | Intraoperative ineligible subjects | 4 | 3 | 1 | | SAFETY ANALYSIS POPULATION: Preoperative AND Intraoperative eligible (ENROLLED and includes roll-in subjects) | 27 | 9 | 18 | | Roll-in subjects | 3 | 1 | 2 | PMA P170012: FDA Summary of Safety and Effectiveness Data Page 18 {18} PMA P170012: FDA Summary of Safety and Effectiveness Data Page 19 # C. Study Population Demographics and Baseline Parameters The average age for all subjects was 62.8 and gender was roughly evenly distributed with 51.9% of all subjects being male and 48.1% female. In terms of race, the majority of subjects self-identified as Caucasian (77.8%) and the remaining subjects self-identified as African-American (22.2%). Self-identified Hispanic subjects were 22.2% of the entire subject pool. There were not any notable differences between sites in terms of demographic characteristics other than Site 2 enrolling all of the Hispanic subjects in the study while Sites 6 and 8 enrolled all of the African-American subjects in the study. Table 11: Study Population Demographics | Measure | All (n=27) | Abdominal (n=9) | Orthopedic (n=18) | | --- | --- | --- | --- | | Age | 62.8 ± 8.64 | 60.4 ± 10.70 | 64.0 ± 7.48 | | | 63.0 [55.0, 68.0] | 55.0 [53.0, 66.0] | 64.5 [62.0, 68.0] | | Gender | | | | | Male | 14/27 (51.9%) | 6/9 (66.7%) | 8/18 (44.4%) | | Female | 13/27 (48.1%) | 3/9 (33.3%) | 10/18 (55.6%) | | Ethnicity | | | | | Hispanic or Latino | 6/27 (22.2%) | 6/9 (66.7%) | 0/18 (0.0%) | | Not Hispanic or Latino | 21/27 (77.8%) | 3/9 (33.3%) | 18/18 (100.0%) | | Race | | | | | Caucasian | 21/27 (77.8%) | 9/9 (100.0%) | 12/18 (66.7%) | | African American | 6/27 (22.2%) | 0/9 (0.0%) | 6/18 (33.3%) | | American Indian or Alaska Native | 0/27 (0.0%) | 0/9 (0.0%) | 0/18 (0.0%) | | Asian | 0/27 (0.0%) | 0/9 (0.0%) | 0/18 (0.0%) | | Native Hawaiian or other Pacific Islander | 0/27 (0.0%) | 0/9 (0.0%) | 0/18 (0.0%) | | Other | 0/27 (0.0%) | 0/9 (0.0%) | 0/18 (0.0%) | Numbers are mean ± Standard Deviation/ median. Table 12: Study Population Baseline Parameters | Measure | All | Abdominal | Orthopedic | | --- | --- | --- | --- | | Liver cell carcinoma | 1/27 (3.7%) | 1/9 (11.1%) | 0/18 (0.0%) | | Metastatic colon cancer to the liver | 1/27 (3.7%) | 1/9 (11.1%) | 0/18 (0.0%) | | Metastatic gastroesophageal junction cancer | 1/27 (3.7%) | 1/9 (11.1%) | 0/18 (0.0%) | {19} Table 13: Study Population Surgical Baseline Parameters | Measure | All | Site 2 | Site 6 | Site 8 | Abdominal | Orthopedic | | --- | --- | --- | --- | --- | --- | --- | | Surgical Indication | | | | | | | | Abdominal | 9/27 (33.3%) | 9/9 (100.0%) | 0/8 (0.0%) | 0/10 (0.0%) | 9/9 (100.0%) | 0/18 (0.0%) | | Orthopedic | 18/27 (66.7%) | 0/9 (0.0%) | 8/8 (100.0%) | 10/10 (100.0%) | 0/9 (0.0%) | 18/18 (100.0%) | | Location | | | | | | | | Medial Retinaculum | 1/27 (3.7%) | 0/9 (0.0%) | 0/8 (0.0%) | 1/10 (10.0%) | 0/9 (0.0%) | 1/18 (5.6%) | | Quad | 3/27 (11.1%) | 0/9 (0.0%) | 0/8 (0.0%) | 3/10 (30.0%) | 0/9 (0.0%) | 3/18 (16.7%) | | Quad Tendon | 3/27 (11.1%) | 0/9 (0.0%) | 0/8 (0.0%) | 3/10 (30.0%) | 0/9 (0.0%) | 3/18 (16.7%) | | Segment 2 | 1/27 (3.7%) | 1/9 (11.1%) | 0/8 (0.0%) | 0/10 (0.0%) | 1/9 (11.1%) | 0/18 (0.0%) | | Segment 3 | 3/27 (11.1%) | 3/9 (33.3%) | 0/8 (0.0%) | 0/10 (0.0%) | 3/9 (33.3%) | 0/18 (0.0%) | | Segment 4 | 2/27 (7.4%) | 2/9 (22.2%) | 0/8 (0.0%) | 0/10 (0.0%) | 2/9 (22.2%) | 0/18 (0.0%) | | Surgical History | | | | | | | | Surgical History | | | | | | | | Surgical History | | | | | | | Numbers are n/N (percent). PMA P170012: FDA Summary of Safety and Effectiveness Data {20} | Measure | All | Site 2 | Site 6 | Site 8 | Abdominal | Orthopedic | | --- | --- | --- | --- | --- | --- | --- | | Segment 5 | 2/27(7.4%) | 2/9 (22.2%) | 0/8 (0.0%) | 0/10 (0.0%) | 2/9 (22.2%) | 0/18 (0.0%) | | Segments 4 & 6 | 1/27(3.7%) | 1/9 (11.1%) | 0/8 (0.0%) | 0/10 (0.0%) | 1/9 (11.1%) | 0/18 (0.0%) | | Suprapatellar pouch | 10/27(37.0%) | 0/9 (0.0%) | 8/8(100.0%) | 2/10(20.0%) | 0/9 (0.0%) | 10/18 (55.6%) | | Tendon | 1/27(3.7%) | 0/9 (0.0%) | 0/8 (0.0%) | 1/10(10.0%) | 0/9 (0.0%) | 1/18 (5.6%) | | Tissue type | | | | | | | | Bone | 1/27(3.7%) | 0/9 (0.0%) | 1/8(12.5%) | 0/10 (0.0%) | 0/9 (0.0%) | 1/18 (5.6%) | | Liver | 9/27(33.3%) | 9/9(100.0%) | 0/8 (0.0%) | 0/10 (0.0%) | 9/9 (100.0%) | 0/18 (0.0%) | | Muscle And Soft Tissue | 3/27(11.1%) | 0/9 (0.0%) | 3/8(37.5%) | 0/10 (0.0%) | 0/9 (0.0%) | 3/18 (16.7%) | | Quad/Tendon | 1/27(3.7%) | 0/9 (0.0%) | 0/8 (0.0%) | 1/10(10.0%) | 0/9 (0.0%) | 1/18 (5.6%) | | Subcutaneous soft tissue | 3/27(11.1%) | 0/9 (0.0%) | 0/8 (0.0%) | 3/10(30.0%) | 0/9 (0.0%) | 3/18 (16.7%) | | Synovium | 2/27(7.4%) | 0/9 (0.0%) | 2/8(25.0%) | 0/10 (0.0%) | 0/9 (0.0%) | 2/18 (11.1%) | | Synovium And Muscle Tissue | 2/27(7.4%) | 0/9 (0.0%) | 2/8(25.0%) | 0/10 (0.0%) | 0/9 (0.0%) | 2/18 (11.1%) | | Tendon | 5/27(18.5%) | 0/9 (0.0%) | 0/8 (0.0%) | 5/10(50.0%) | 0/9 (0.0%) | 5/18 (27.8%) | | Tendon/Soft Tissue | 1/27(3.7%) | 0/9 (0.0%) | 0/8 (0.0%) | 1/10(10.0%) | 0/9 (0.0%) | 1/18 (5.6%) | | TBS approximate dimensions (cm2) | 21.1 ± 75.88(27)4.0 [4.0,9.0] | 51.0 ± 131.04(9)4.0 [4.0,16.0] | 3.7 ± 1.53(8)4.0 [3.1,4.0] | 8.1 ± 4.07(10)8.3 [4.0,12.0] | 51.0 ± 131.04(9)4.0 [4.0,16.0] | 6.1 ± 3.84(18)4.0 [4.0,9.0] | | Conventional Procedures for Hemostasis | | | | | | | | Pressure | 12/27(44.4%) | 0/9 (0.0%) | 2/8(25.0%) | 10/10(100.0%) | 0/9 (0.0%) | 12/18 (66.7%) | | Suture, ligation | 0/27(0.0%) | 0/9 (0.0%) | 0/8 (0.0%) | 0/10 (0.0%) | 0/9 (0.0%) | 0/18 (0.0%) | | Cautery | 2/27(7.4%) | 0/9 (0.0%) | 1/8(12.5%) | 1/10(10.0%) | 0/9 (0.0%) | 2/18 (11.1%) | PMA P170012: FDA Summary of Safety and Effectiveness Data {21} | Measure | All | Site 2 | Site 6 | Site 8 | Abdominal | Orthopedic | | --- | --- | --- | --- | --- | --- | --- | | Other | 0/27 (0.0%) | 0/9 (0.0%) | 0/8 (0.0%) | 0/10 (0.0%) | 0/9 (0.0%) | 0/18 (0.0%) | | None (impractical) | 14/27 (51.9%) | 9/9 (100.0%) | 5/8 (62.5%) | 0/10 (0.0%) | 9/9 (100.0%) | 5/18 (27.8%) | Numbers are mean ± Standard Deviation/ median. ## D. Safety and Effectiveness Results The mean paired kappa across all sites was 0.9301 indicative of almost perfect agreement. Given the pre-specified criteria to advance to the pivotal clinical investigation was that the mean paired kappa statistic for assignment of SBSS scores was &gt;0.80, this endpoint is deemed successful. These results validated the use of the SBSS for the pivotal study. There were 41 adverse events not believed to be device related. One serious adverse event, portal vein thrombosis, occurred in a patient who underwent a right hepatic lobectomy for metastatic colon cancer with metastectomy of additional metastases in the left hepatic lobe. This required prolonged operative time and Pringle maneuver all of which more likely contributed to the portal vein thrombosis. This event was not likely caused by embolization of Hemoblast applied to the liver parenchymal edge of resection and resolved with anticoagulation. Table 14: Adverse Events | Measure | All | Site 2 | Site 6 | Site 8 | Abdominal | Orthopedic | | --- | --- | --- | --- | --- | --- | --- | | Number of AEs | 41 | 29 | 3 | 9 | 29 | 12 | | Number of subjects with AEs | 15/27 55.6% (37.3%,72.4%) | 8/9 88.9% (56.5%, 98.0%) | 2/8 25.0% (7.1%, 59.1%) | 5/10 50.0% (23.7%, 76.3%) | 8/9 88.9% (56.5%, 98.0%) | 7/18 38.9% (20.3%, 61.4%) | | Number of AEs by severity | | | | | | | | Mild | 16 | 13 | 0 | 3 | 13 | 3 | | Moderate | 22 | 13 | 3 | 6 | 13 | 9 | | Severe | 3 | 3 | 0 | 0 | 3 | 0 | | Number of subjects with AEs by severity* | | | | | | | PMA P170012: FDA Summary of Safety and Effectiveness Data {22} | Measure | All | Site 2 | Site 6 | Site 8 | Abdominal | Orthopedic | | --- | --- | --- | --- | --- | --- | --- | | Mild | 5/2718.5% (8.2%,36.7%) | 4/944.4%(18.9%,73.3%) | 0/80.0% (0.0%,32.4%) | 1/1010.0%(1.8%,40.4%) | 4/944.4% (18.9%,73.3%) | 1/185.6% (1.0%,25.8%) | | Moderate | 8/2729.6% (15.9%,48.5%) | 2/922.2% (6.3%,54.7%) | 2/825.0% (7.1%,59.1%) | 4/1040.0%(16.8%,68.7%) | 2/922.2% (6.3%,54.7%) | 6/1833.3% (16.3%,56.3%) | | Severe | 2/277.4% (2.1%,23.4%) | 2/922.2%(6.3%,54.7%) | 0/80.0% (0.0%,32.4%) | 0/100.0%(0.0%,27.8%) | 2/922.2% (6.3%,54.7%) | 0/180.0% (0.0%,17.6%) | | Number AEs related to device | 0 | 0 | 0 | 0 | 0 | 0 | | Number of subjects with AEs related to device | 0/270.0% (0.0%,12.5%) | 0/90.0%(0.0%,29.9%) | 0/80.0%(0.0%,32.4%) | 0/100.0%(0.0%,27.8%) | 0/90.0%(0.0%,29.9%) | 0/180.0%(0.0%,17.6%) | | Number of SAEs | 8 | 7 | 1 | 0 | 7 | 1 | | Number of subjects with SAEs | 4/2714.8% (5.9%,32.5%) | 3/933.3%(12.1%,64.6%) | 1/812.5% (2.2%,47.1%) | 0/100.0%(0.0%,27.8%) | 3/933.3% (12.1%,64.6%) | 1/185.6% (1.0%,25.8%) | | Number of SADEs | 0 | 0 | 0 | 0 | 0 | 0 | | Number of subjects with SADEs | 0/270.0% (0.0%,12.5%) | 0/90.0% (0.0%,29.9%) | 0/80.0% (0.0%,32.4%) | 0/100.0%(0.0%,27.8%) | 0/90.0% (0.0%,29.9%) | 0/180.0% (0.0%,17.6%) | | Number of UADEs | 0 | 0 | 0 | 0 | 0 | 0 | | Number of subjects with UADEs | 0/270.0% (0.0%,12.5%) | 0/90.0% (0.0%,29.9%) | 0/80.0% (0.0%,32.4%) | 0/100.0%(0.0%,27.8%) | 0/90.0% (0.0%,29.9%) | 0/180.0% (0.0%,17.6%) | Numbers are n/N percent (95% CI). Wilson confidence limits (score based) are used in the table. *The most severe AE is counted for each subject. PMA P170012: FDA Summary of Safety and Effectiveness Data {23} Table 15: Thromboembolic Events | | Total Thromboembolic Events | Pulmonary Embolism | Deep Venous Thrombosis | Stroke | Thrombosed AV Fistula | Other(portal vein thrombosis) | | --- | --- | --- | --- | --- | --- | --- | | HEMOBLAST™ US Pilot | 1/27 (3.7%) | 0/27 (0.0%) | 0/27 (0.0%) | 0/27 (0.0%) | 0/27 (0.0%) | 1/27 (3.7%) | Hemoblast induced hemostasis at 3 minutes for half of all subjects in the effectiveness analysis population and 79.2% of this group (19/24) achieved hemostasis at 6 minutes. There were 2/24 subjects that did not achieve hemostasis at 10 minutes. Comparing abdominal surgery subjects to orthopedic surgery subjects at the 6-minute time point, it appears the orthopedic group performed better with 93.8% of subjects achieving hemostasis compared to 50.0% of subjects in the abdominal surgery group. Table 16: Proportion of Subjects Achieving Hemostasis at 3, 6, and 10 Minutes - Efficacy Analysis Population | Time | All | Site 2 | Site 6 | Site 8 | Abdominal | Orthopedic | | --- | --- | --- | --- | --- | --- | --- | | 3 minutes | 12/24 50.0% (31.4%, 68.6%) | 2/8 25.0% (7.1%, 59.1%) | 3/7 42.9% (15.8%, 75.0%) | 7/9 77.8% (45.3%, 93.7%) | 2/8 25.0% (7.1%, 59.1%) | 10/16 62.5% (38.6%, 81.5%) | | 6 minutes | 19/24 79.2% (59.5%, 90.8%) | 4/8 50.0% (21.5%, 78.5%) | 6/7 85.7% (48.7%, 97.4%) | 9/9 100.0% (70.1%, 100.0%) | 4/8 50.0% (21.5%, 78.5%) | 15/16 93.8% (71.7%, 98.9%) | | 10 minutes | 22/24 91.7% (74.2%, 97.7%) | 6/8 75.0% (40.9%, 92.9%) | 7/7 100.0% (64.6%, 100.0%) | 9/9 100.0% (70.1%, 100.0%) | 6/8 75.0% (40.9%, 92.9%) | 16/16 100.0% (80.6%, 100.0%) | Numbers are n/N percent (95% CI). Wilson confidence limits (score based) are used in the table. Cumulative numbers of subjects achieving hemostasis at each time are counted. ## Pediatric Extrapolation In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population. ## E. Financial Disclosure The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pilot clinical study included nine (9) investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data. PMA P170012: FDA Summary of Safety and Effectiveness Data {24} Pivotal Study: A. Study Design This was a prospective, randomized, controlled, multicenter pivotal study to evaluate the safety and effectiveness of the HEMOBLAST™ Bellows (“HEMOBLAST™”) in cardiothoracic, abdominal (both soft tissue and organ space), and orthopedic lower extremity surgeries. Subjects were randomized intraoperatively in a 2:1 ratio to HEMOBLAST™ or the control treatment, an absorbable gelatin sponge, USP and recombinant thrombin (“G+T”). The study also included a lead-in phase, to ensure consistent assessment of bleeding severity using the SBSS scale by study investigators, wherein subjects were not randomized. Lead-in subjects received HEMOBLAST™ and were followed for safety only. The specific objective of the study was to evaluate the safety and effectiveness of HEMOBLAST™ compared to G+T. The primary hypothesis in this study was that HEMOBLAST™ is non-inferior relative to G+T for success at achieving hemostasis within 6 minutes. A Clinical Events Committee (“CEC”) and Independent Data Monitoring Committee (“IDMC”) were utilized to review all serious adverse events and safety and effectiveness data, respectively. 1. Clinical Inclusion and Exclusion Criteria Enrollment in the Pivotal Study was limited to subjects who met the following preoperative inclusion criteria: - Subject is undergoing an elective open cardiothoracic, abdominal, or orthopedic lower extremity surgery; - Subject or an authorized legal representative is willing and able to give prior written informed consents for investigation participation; - Subject undergoing cardiothoracic surgery is not allergic to protamine; and - Subject is 21 years of age or older. In addition, subjects must have met the following intraoperative inclusion criteria: - Subject does not have an active or suspected infection at the surgical site; - Subject undergoing cardiothoracic surgery with anticoagulation must have anticoagulation reversed prior to Target Bleeding Site (TBS) identification and treatment; - Subject in whom the Investigator is able to identify a TBS for which any applicable conventional means for hemostasis are ineffective or impractical; and - Subject has a TBS with an SBSS score of 1, 2, or 3. Subjects were not permitted to enroll in the Pivotal Study if they met any of the following exclusion criteria: PMA P170012: FDA Summary of Safety and Effectiveness Data Page 25 {25} - Subject is undergoing a laparoscopic, thoracoscopic, or robotic surgical procedure; - Subject is undergoing a neurologic surgical procedure; - Subject is undergoing a spinal surgical procedure; - Subject is undergoing an emergency surgical procedure; - Subject is pregnant, planning on becoming pregnant during the follow-up period, or actively breast-feeding; - Subject has a clinically significant coagulation disorder or disease, defined as a platelet count &lt; 100,000 per microliter or International Normalized Ratio &gt; 1.5 within 4 weeks of surgery; - Subject receiving intravenous heparin within 12 hours before surgery or oral Coumadin within 2 days before surgery; - Subject receiving antiplatelet medications within 5 days prior to surgery; - Subject undergoing abdominal or orthopedic lower extremity surgery receiving aspirin within 7 days prior to surgery; - Subject has an active or suspected infection at the surgical site; - Subject has had or has planned to receive any organ transplantation; - Subject has a known sensitivity or allergy to bovine and/or porcine substance(s) or any other component(s) of the hemostatic agent; - Subject has ASA classification of 5; - Subject has a life expectancy of less than 3 months; - Subject has a known psychiatric disorder, which in the opinion of the Principal Investigator, would preclude the subject from completing this clinical study; - Subject has a documented severe congenital or acquired immunodeficiency; - Subject has religious or other objections to porcine, bovine, or human components; - Subject in whom the investigational or control device will be used at the site of a valve replacement or repair; - Subject in whom the investigational or control device will be used at the site of a synthetic graft or patch implant; - Subject is currently participating or has participated in another clinical trial within the past 30 days and is receiving/has received an investigational drug, device, or biologic agent; and - Subject is not appropriate for inclusion in the clinical trial, per the medical opinion of the Principal Investigator. 2. Patient Follow-up Schedule All patients underwent the same intraoperative investigational evaluations. During the surgery, hemostasis was evaluated by the investigator 3, 6, and 10 minutes after application of the investigational or control treatment until hemostasis was achieved. Reapplication of the randomized hemostat was performed at the 3 and 6 minute evaluation time points, as needed. In cases where hemostasis was not achieved by 10 minutes, the Investigator may have used whatever means necessary in order to PMA P170012: FDA Summary of Safety and Effectiveness Data Page 26 {26} control bleeding, except for any hemostatic products containing thrombin or aprotinin. Thrombin should not have been used in subjects randomized to receive HEMOBLAST, but may have been used in subjects randomized to the G+T arm. Investigators were permitted to use any of the remaining randomized hemostat for bleeding sites other than the target bleeding site ("TBS") or any hemostatic product not containing thrombin or aprotinin. Safety assessments occurred one day and 6 weeks postoperatively. Blood draws for antibody evaluation were performed preoperatively (within 4 weeks of surgery) and 6 weeks postoperatively. ## 3. Clinical Endpoints With regards to safety, endpoints were used to characterize the safety profile include the rate of occurrence of all AEs and SAEs, reoperation rate due to bleeding, mean volume of intraoperative transfusions, total operative time, mean duration of hospitalization, intraoperative blood product administration, and post-operative blood product administration: - The total operative time was measured as the time from entry into surgical suite to time of exit; - The duration of hospitalization was measured as the time between admission and discharge; and - The intraoperative and post-operative administration of blood products was tracked and the associated mean volume administered was measured in units. Blood products included red blood cells, platelets, fresh frozen plasma (FFP), and cryoprecipitate. With regards to effectiveness, the primary effectiveness endpoint of this clinical investigation was non-inferiority of HEMOBLAST™ relative to G+T for success at achieving hemostasis within 6 minutes. The secondary effectiveness endpoints of this clinical investigation were: - Superiority of HEMOBLAST™ relative to G+T in mean preparation time from the opening of package to product being ready to use; - Non-inferiority of HEMOBLAST™ relative to G+T for success at achieving hemostasis within 3 minutes; - Superiority of HEMOBLAST™ relative to G+T for success at achieving hemostasis within 6 minutes; and - Superiority of HEMOBLAST™ relative to G+T for success at achieving hemostasis within 3 minutes. The pivotal study was designed to test the non-inferiority hypothesis of HEMOBLAST™ relative to G+T for success at achieving hemostasis within 6 minutes and within a 10% margin. The study pre-specified a single interim PMA P170012: FDA Summary of Safety and Effectiveness Data Page 27 {27} analysis for early stopping due to futility or effectiveness after outcome data had been observed on 240 treated subjects. If the study were to continue to the final analysis with a decision in favor of comparable effectiveness, it was anticipated that effectiveness data would be available on a maximum of 400 treated subjects (approximately 267 patients treated with HEMOBLAST™ under a 2:1 randomization scheme). Effectiveness analyses were conducted on the time-to-hemostasis ("TTH") population, defined as all subjects who were randomized, received study intervention, and had a TTH assessment recorded regardless of whether the measurement was censored (defined as the use of an additional hemostatic product or surgical rescue prior to the end of observation time, or failing to achieve and maintain complete hemostasis prior to the end of observation time). Lead-in subjects were not part of the TTH Population. Safety analyses were conducted on the Full Analysis population, defined as all subjects who were randomized into the study and received study intervention and all lead-in subjects. In the primary analyses, missing TTH values were not imputed. All values right censored prior to 6 minutes were considered treatment failures for the purpose of the primary analysis. With regard to success/failure criteria, individual subject success will be defined as hemostasis of the target bleeding site within 6 minutes of hemostat application. The proportion of subjects, overall, and in each surgical indication, meeting this success criterion will be estimated and a 95% confidence interval for the true proportion of subjects meeting the success criteria will be presented. Overall study success will be defined as non-inferiority of HEMOBLAST™ relative to G+T in the proportion of subjects achieving hemostasis within 6 minutes of hemostat application. ## B. Accountability of PMA Cohort The study was stopped early for effectiveness, per the IDMC recommendation, based on the pre-specified stopping rules. At the time of the interim analysis, a total of 258 subjects were enrolled in the study, including 16 lead-in subjects and 242 randomized subjects. The interim analysis on completion of half the intended total number of patients in the pivotal study demonstrated non-inferiority at the 6 minute hemostasis primary endpoint and superiority at the 10-30% superiority at all the secondary endpoints including hemostasis at 6 minutes, 3 minutes, and 10 minutes, as well as, superiority on time to preparation for use. Furthermore, the pivotal study demonstrated non inferiority of Hemoblast for hemostasis at 10 minutes. Table 17 represents the enrollment for each treatment group and Table 18 represents the enrollment for each surgical arm. Randomization was stratified by surgical arm. PMA P170012: FDA Summary of Safety and Effectiveness Data Page 28 {28} Table 17: Enrollment Details for Each Treatment Group | Population | All | HEMOBLAST™ | G+T | | --- | --- | --- | --- | | TOTAL: Preoperative eligible + lead-in subjects | 260*; | 175 | 83 | | Intraoperative ineligible subjects | 1 | 0 | 0 | | SAFETY ANALYSIS POPULATION: Preoperative AND Intraoperative eligible (ENROLLED and includes lead-in subjects) | 258; | 175 | 83 | | Lead-in subjects | 16 | 16 | 0 | | EFFICACY ANALYSIS POPULATION: Preoperative AND Intraoperative eligible (ENROLLED and EXCLUDES lead-in subjects) | 242; | 159 | 83 | *One subject failed to meet intra-operative eligibility criteria, due to an intraoperative SBSS score of 0, and one subject was withdrawn after randomization/enrollment, but prior to hemostat application. Two (2) subjects were enrolled, that were identified at subsequent monitoring visits, to have not met preoperative eligibility criteria. Because these subjects were enrolled and received HEMOBLAST™ or G+T application, they are considered part of the total, safety analysis and effectiveness analysis populations. Table 18: Enrollment by Surgical Arm | Population | All | Cardiothoracic | Abdominal | Orthopedic | | --- | --- | --- | --- | --- | | TOTAL: Preoperative eligible + lead-in subjects | 260*; | 64 | 98 | 98 | | Intraoperative ineligible subjects | 1 | 0 | 0 | 1 | | SAFETY ANALYSIS POPULATION: Preoperative AND Intraoperative eligible (ENROLLED and includes lead-in subjects) | 258; | 64 | 97 | 97 | | Lead-in subjects | 16 | 6 | 7 | 3 | | EFFICACY ANALYSIS POPULATION: Preoperative AND Intraoperative eligible (ENROLLED and EXCLUDES lead-in subjects) | 242; | 58 | 90 | 94 | * One subject failed to meet intra-operative eligibility criteria, due to an intraoperative SBSS score of 0, and one subject was withdrawn prior to hemostat application due to an intraoperative SBSS score of 0. Two (2) subjects were enrolled, that were identified at subsequent monitoring visits, to have not met preoperative eligibility criteria. Because these subjects were enrolled and received HEMOBLAST™ or G+T application, they are considered part of the total, safety analysis and efficacy analysis populations. # C. Study Population Demographics and Baseline Parameters The demographics of the study population are typical for a hemostatic device study performed in the US. Table 19 presents physical measurement for height, weight, BMI, and blood pressure, which were similar between treatment arms. PMA P170012: FDA Summary of Safety and Effectiveness Data {29} Table 19: Physical Measurements by Treatment Group | Measure | All (n=258) | HEMOBLAST™ (n=175) | G+T (n=83) | | --- | --- | --- | --- | | Height (cm) | 167.4 ± 10.19 | 167.4 ± 10.43 | 167.3 ± 9.72 | | | 167.0 [160.0, 175.0] | 165.0 [160.0, 175.0] | 167.0 [160.0, 175.0] | | Weight (kg) | 83.8 ± 19.121 | 84.6 ± 19.19 | 82.1 ± 19.26 | | | 82.5 [70.5, 97.0] | 83.0 [71.0, 97.5] | 80.9 [69.4, 97.0] | | BMI (kg/m2) | 29.9 ± 6.33 | 30.2 ± 6.14 | 29.3 ± 6.71 | | | 29.1 [25.5, 33.7] | 29.4 [25.6, 34.1] | 27.9 [25.1, 33.1] | | Systolic Blood pressure (mmHg) | 129.0 ± 18.54 | 128.8 ± 18.28 | 129.4 ± 19.18 | | | 128.0 [116.0, 138.0] | 128.0 [116.0, 138.0] | 128.0 [118.0, 141.0] | | Diastolic Blood pressure (mmHg) | 76.6 ± 12.60 | 76.5 ± 12.33 | 76.9 ± 13.24 | | | 78.0 [70.0, 85.0] | 78.0 [69.0, 84.0] | 76.0 [70.0, 85.0] | Numbers are mean $\pm$ SD (N)/ median Table 20 and Table 21 present the baseline SBSS score for the TBS by treatment group and surgical arm, respectively. Table 20: Baseline SBSS Score for Each Treatment Group | Measure | All (N=242) | HEMOBLAST™ (N=159) | G+T (N=83) | | --- | --- | --- | --- | | Investigator SBSS | | | | | 0 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | | 1 | 84 (34.7%) | 59 (37.1%) | 25 (30.1%) | | 2 | 100 (41.3%) | 61 (38.4%) | 39 (47.0%) | | 3 | 58 (24.0%) | 39 (24.5%) | 19 (22.9%) | | 4 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | | 5 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | | SBSS score 2:2 | 158 (65.3%) | 100 (62.9%) | 58 (69.9%) | Numbers are n/N (percent). Table 21: Baseline SBSS Score for Each Surgical Arm | Measure | All (N=242) | Cardiothoracic (N=58) | Abdominal (N=90) | Orthopedic (N=94) | | --- | --- | --- | --- | --- | | Investigator SBSS Score | | | | | | 0 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | | 1 | 84 (34.7%) | 14 (24.1%) | 8 (8.9%) | 62 (66.0%) | | 2 | 100 (41.3%) | 24 (41.4%) | 46 (51.1%) | 30 (31.9%) | | 3 | 58 (24.0%) | 20 (34.5%) | 36 (40.0%) | 2 (2.1%) | | 4 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | PMA P170012: FDA Summary of Safety and Effectiveness Data {30} PMA P170012: FDA Summary of Safety and Effectiveness Data Page 31 | 5 | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | | --- | --- | --- | --- | --- | Numbers are n/N (percent). The orthopedic procedures had a much greater number of patients in the lower SBSS scores. The orthopedic procedures represented a great variety of procedures the largest numbers being in hip and knee replacement and other open surgeries. The abdominal procedure group of patients consisted of 90 patients. The most prevalent abdominal procedures included liver resection surgery for both primary and metastatic cancers with less than 5% of the liver resections performed for biliary cancers and benign hepatic tumors. There were a total of 32 liver resection cases of the 90 patients in the abdominal group. A large portion of the remaining abdominal surgeries included abdominoplasties and abdominal hernia repairs. The cardiothoracic cases comprised the smallest group of patients and consisted of patients requiring application of the subject or control hemostat to the sternal edges, aortotomy sites, saphenous vein graft anastomosis site (proximal or distal not defined), pericardium and myocardial suture lines. Application to synthetic graft bleeding was excluded from this arm of the study. ## D. Safety and Effectiveness Results ### 1. Safety results The analysis of safety was based on the pivotal study cohort of 175 patients available for the 6 week evaluation. The key safety outcomes and adverse effects for this study are presented below. In summary, at the time of the interim analysis, the HEMOBLAST™ group had a lower rate of subjects experiencing adverse events, although this did not reach statistical significance (48.0% vs. 56.6%, p=0.1516). ### Serious adverse events The proportion of patients experiencing a serious adverse event was comparable between treatment groups (10.9% for HEMOBLAST™ vs. 13.3% for G+T, p=0.5415). The two (2) groups were found to have a similar number of subjects experiencing an adverse event related to the device. There were three (3) possible device-related adverse events report in three (3) subjects receiving HEMOBLAST™: - Subject reported symptoms indicating possible systemic inflammatory response syndrome. This event was classified as possibly related to HEMOBLAST™; the event resolved with no sequelae. - Subject experienced a generalized skin reaction with hives 11 days postoperatively; the event resolved with no sequelae. This was deemed as possibly related to HEMOBLAST™ and the CEC adjudicated this event as a Serious Adverse Device Effect (SADE). - Subject was diagnosed with a pulmonary embolus 10 days postoperative; the {31} event resolved with no sequelae. This was deemed as possibly related to HEMOBLAST™, but probably related to the investigational procedure. There were four (4) possible device-related adverse events reported in 4 subjects receiving G+T: - Subject experienced acute blood loss anemia 4 days postoperative and was transfused with packed RBCs; the event resolved with no sequelae. This event was classified as possibly related to G+T. - Subject developed a period of hypotension following surgery; it resolved with no sequelae after an infusion of albumin. This event was classified as possibly related to G+T. - Subject had re-bleeding of the target bleeding site treated with G+T prior to surgical closure. The re-bleeding was controlled using a clip. This was deemed possibly related to the use of G+T. - Subject had a re-bleeding of the target bleeding site treated with G+T prior to surgical closure. The re-bleeding was controlled using cautery. This was deemed possibly related to G+T and definitely related to the investigational procedure. In summary, there was a single event identified as a SADE (subject who experienced skin reaction, described above). PMA P170012: FDA Summary of Safety and Effectiveness Data Page 32 {32} Table 22: Serious Adverse Events | Adverse Event Type* | All | HEMOBLAST™ | G+T | | --- | --- | --- | --- | | Abnormal Bloodwork | 18/258 (7.0%) | 13/175 (7.4%) | 5/83 (6.0%) | | Acute Kidney Injury | 6/258 (2.3%) | 4/175 (2.3%) | 2/83 (2.4%) | | Anemia | 19/258 (7.4%) | 10/175 (5.7%) | 9/83 (10.8%) | | Anxiety | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Arrhythmia | 23/258 (8.9%) | 16/175 (9.1%) | 7/83 (8.3%) | | Atelectasis | 7/258 (2.7%) | 4/175 (2.3%) | 3/83 (3.6%) | | Bile Leak | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Constipation/Ileus | 17/258 (6.6%) | 12/175 (6.9%) | 5/83 (6.0%) | | Dehydration | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Delirium | 2/258 (0.8%) | 1/175 (0.6%) | 1/83 (1.2%) | | Depression | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Dislocation | 2/258 (0.8%) | 1/175 (0.6%) | 1/83 (1.2%) | | Diarrhea | 3/258 (1.2%) | 3/175 (1.7%) | 0/83 (0.0%) | | Dizziness | 3/258 (1.2%) | 2/175 (1.1%) | 1/83 (1.2%) | | Epistaxis | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Fever | 4/258 (1.6%) | 2/175 (1.1%) | 2/83 (2.4%) | | Fluid Overload | 13/258 (5.0%) | 8/175 (4.6%) | 5/83 (6.0%) | | Hypervolemia | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Edema | 6/258 (2.3%) | 3/175 (1.7%) | 3/83 (3.6%) | | Pleural Effusion | 6/258 (2.3%) | 5/175 (2.9%) | 1/83 (1.2%) | | Fracture | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Hypertension | 3/258 (1.2%) | 2/175 (1.1%) | 1/83 (1.2%) | | Hypotension | 3/258 (1.2%) | 2/175 (1.1%) | 1/83 (1.2%) | | Infection (Non wound Related) | 13/258 (5.0%) | 7/175 (4.0%) | 6/83 (7.2%) | | Bacteremia | 3/258 (1.2%) | 2/175 (1.1%) | 1/83 (1.2%) | | Drain | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Phlebitis | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Respiratory | 3/258 (1.2%) | 1/175 (0.6%) | 2/83 (2.4%) | | Urinary Tract | 4/258 (1.6%) | 1/175 (0.6%) | 3/83 (3.6%) | | Yeast | 2/258 (0.8%) | 1/175 (0.6%) | 1/83 (1.2%) | | Other¹ | 4/258 (1.6%) | 3/175 (1.7%) | 1/83 (1.2%) | | Lactic Acidosis | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Lethargy | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Leukocytosis | 3/258 (1.2%) | 3/175 (1.7%) | 0/83 (0.0%) | | Nausea | 18/258 (7.0%) | 15/175 (8.6%) | 3/83 (3.6%) | | Orthostasis | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Pain | 36/258 (14.0%) | 25/175 (14.3%) | 11/83 (13.3%) | | Mouth/Throat | 8/258 (3.1%) | 5/175 (2.9%) | 3/83 (3.6%) | PMA P170012: FDA Summary of Safety and Effectiveness Data Page 33 {33} PMA P170012: FDA Summary of Safety and Effectiveness Data Page 34 | Post Operative | 28/258 (10.9%) | 21/175 (12.0%) | 7/83 (8.4%) | | --- | --- | --- | --- | | Other² | 8/258 (3.1%) | 5/175 (2.9%) | 3/83 (3.6) | | Paresthesia | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Pericarditis | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Pneumothorax | 2/258 (0.8%) | 0/175 (0.0%) | 2/83 (2.4%) | | Pruritis | 4/258 (1.6%) | 3/175 (1.7%) | 1/83 (1.2%) | | Respiratory Insufficiency | 9/258 (3.5%) | 6/175 (3.4%) | 3/83 (3.6%) | | TBS Re-bleed | 5/258 (1.9%) | 1/175 (0.6%) | 4/83 (4.8%) | | Thrombocytopenia | 2/258 (0.8%) | 2/175 (1.1%) | 0/83 (0.0%) | | Thromboembolic Event | 4/258 (1.6%) | 4/175 (2.3%) | 0/83 (0.0%) | | Thrombosed AV Fistula | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Deep Vein Thrombosis | 2/258 (0.8%) | 2/175 (1.1%) | 0/83 (0.0%) | | Other³ | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Urinary Retention/Oliguria | 7/258 (2.7%) | 4/175 (2.3%) | 3/83 (3.6%) | | Vaginal Discharge | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Vomiting | 5/258 (1.9%) | 4/175 (2.3%) | 1/83 (1.2%) | | Wound Related | 16/258 (6.2%) | 12/175 (6.9%) | 4/83 (4.8%) | | Dehiscence | 1/258 (0.4%) | 1/175 (0.6%) | 0/83 (0.0%) | | Hematoma | 3/258 (1.2%) | 2/175 (1.1%) | 1/83 (1.2%) | | Infection | 4/258 (1.6%) | 3/175 (1.7%) | 1/83 (1.2%) | | Non-healing | 2/258 (0.8%) | 1/175 (0.6%) | 1/83 (1.2%) | | Seroma | 6/258 (2.3%) | 6/175 (3.4%) | 0/83 (0.0%) | | Other⁴ | 1/258 (0.4%) | 0/175 (0.0%) | 1/83 (1.2%) | | Other⁵ | 11/258 (4.3%) | 9/175 (5.1%) | 2/83 (2.4%) | ¹ Open text field, like responses were pooled ¹ HEMOBLAST™: Pancolitis, diarrhea positive for clostridium difficile, dental abscess; G+T: tricuspid valve vegetation ² HEMOBLAST™: right leg pain, back pain, right ear pain, left hip pain, right shoulder; G+T: sore shoulder, pain at epidural site and back pain, abdominal pain ³ Intraoperative TEE demonstrated internal jugular vein clot ⁴ Wound irritation from one stitch ⁵ HEMOBLAST™: muscle spasm, thoracic aorta dissection, hematoma, vocal cord paralysis, endoleak, pressure ulcer, decreased right ventricular function during cardiopulmonary bypass removal, arthralgia of right leg, recurrent shingles, gout flare, left thumb tenderness, insomnia, right shoulder acute bursitis, right bronchopleural fistula, blisters around tape covering incision; G+T: right atrium lead dislodgment, left arm weakness, femoral artery perforation Note: Mortality events only occurred in the G+T control group 3/83 patients or 3.6% and serious adverse events tended to be related to re-bleeding episodes. ## Non-serious adverse events Table 23 lists non-serious AEs occurring in 5% or more of all patients, or in one of the treatment arms. Table 23: Number of Subjects Experiencing Each Type of Non-serious Adverse Event | Adverse Event Type* | All | HEMOBLAST | G+T | | --- | --- | --- | --- | | Abnormal Bloodwork | 18/258 (7.0%) | 13/175 (7.4%) | 5/83 (6.0%) | | Anemia | 19/258 (7.4%) | 10/175 (5.7%) | 9/83 (10.8%) | {34} | Arrhythmia | 23/258 (8.9%) | 16/175 (9.1%) | 7/83 (8.3%) | | --- | --- | --- | --- | | Constipation/Ileus | 17/258 (6.6%) | 12/175 (6.9%) | 5/83 (6.0%) | | Fluid Overload | 13/258 (5.0%) | 8/175 (4.6%) | 5/83 (6.0%) | | Infection (Non wound Related) | 13/258 (5.0%) | 7/175 (4.0%) | 6/83 (7.2%) | | Nausea | 18/258 (7.0%) | 15/175 (8.6%) | 3/83 (3.6%) | | Pain | 36/258 (14.0%) | 25/175 (14.3%) | 11/83 (13.3%) | | Wound Related | 16/258 (6.2%) | 12/175 (6.9%) | 4/83 (4.8%) | | Other^{1} | 11/258 (4.3%) | 9/175 (5.1%) | 2/83 (2.4%) | *Open text field, like responses were pooled. $^{1}$ HEMOBLAST™: muscle spasm, thoracic aorta dissection, hematoma, vocal cord paralysis, endoleak, pressure ulcer, decreased right ventricular function during cardiopulmonary bypass removal, arthralgia of right leg, recurrent shingles, gout flare, left thumb tenderness, insomnia, right shoulder acute bursitis, right bronchopleural fistula, blisters around tape covering incision; G+T: right atrium lead dislodgment, left arm weakness, and femoral artery perforation ## Unanticipated Adverse Device Effects There were no Unanticipated Adverse Device Effects. The incidence of adverse events and serious adverse events were statistically the same in both experimental and control groups. The three (3) main categories of potential adverse events identified in the pivotal study included thromboembolic events, wound healing complications primarily in the appearance of sternal dehiscence, and porcine collagen antibody titers. Each of these concerns was further assessed as follows. ## 2. Effectiveness Results The analysis of effectiveness was based on the 242 evaluable patients at the 6-week time point. Key effectiveness outcomes are presented in Tables 24 to 28. ## Primary endpoint The primary efficacy endpoint of non-inferiority of HEMOBLAST™ relative to G+T for success at achieving hemostasis within 6 minutes was met, with 93.1% of the HEMOBLAST™ group achieving hemostasis at 6 minutes versus 73.5% of the G+T group, a difference of 19.5% (9.5% to 29.5%, p&lt;0.0001 for non-inferiority). See Table 24. A 95% repeated confidence interval that accounts for the pre-specified stopping rule is calculated to be (7.1%, 31.9%), ruling out the non-inferiority margin of -10%. PMA P170012: FDA Summary of Safety and Effectiveness Data {35} Table 24: Primary Efficacy Endpoint with Full Efficacy Population | Time | HEMOBLAST | G+T | Difference (95% CI) | Z-statistic | P-value | | --- | --- | --- | --- | --- | --- | | 6 minutes | 148/159 (93.1%) | 61/83 (73.5%) | 19.5% (9.5%, 29.5%) | 5.7926 | <0.0001 | The confidence interval for the estimated difference in the probability of hemostasis at 6 minutes and the P value are computed using the Cochran-Mantel-Haenszel estimator stratified by surgical arms. A summary of the proportion of each treatment group that achieved hemostasis at 3, 6, and 10 minutes is shown in Table 25. The HEMOBLAST™ group showed a higher proportion of patients achieving hemostasis at each time point assessed. Table 25: Proportion of Each Treatment Group Achieving Hemostasis at 3, 6, and 10 Minutes | Time | All | HEMOBLAST | G+T | | --- | --- | --- | --- | | 3 minutes | 151/242 (62.4%) | 113/159 (71.1%) | 38/83 (45.8%) | | 6 minutes | 209/242 (86.4%) | 148/159 (93.1%) | 61/83 (73.5%) | | 10 minutes | 223/238 (93.7%) | 154/158 (97.5%) | 69/80 (86.3%) | Numbers are n/N (percent). The study was stopped early for efficacy, per the IDMC recommendation, based on the pre-specified stopping rules. The proportion of subjects in each treatment group achieving hemostasis at each time point was broken out by surgical arm, see Tables 26, 27, and 28. The HEMOBLAST™ group showed significantly higher rates of hemostasis at 3 and 6 minutes in the abdominal and orthopedic surgical arms. There was no significant difference between groups in the cardiothoracic arm, as shown in the tables below. Table 26: Proportion Achieving Hemostasis for Cardiothoracic Surgical Arm | Time | All | HEMOBLAST™ | G+T | P-value | | --- | --- | --- | --- | --- | | 3 minutes | 31/58 (53.4%) | 20/38 (52.6%) | 11/20 (55.0%) | >0.9999 | | 6 minutes | 46/58 (79.3%) | 31/38 (81.6%) | 15/20 (75.0%) | 0.7343 | | 10 minutes | 51/56 (91.1%) | 34/37 (91.9%) | 17/19 (89.5%) | >0.9999 | Table 27: Proportion Achieving Hemostasis for Abdominal Surgical Arm | Time | All | HEMOBLAST™ | G+T | P-value | | --- | --- | --- | --- | --- | | 3 minutes | 46/90 (51.1%) | 39/59 (66.1%) | 7/31 (22.6%) | 0.0001 | | 6 minutes | 74/90 (82.2%) | 55/59 (93.2%) | 19/31 (61.3%) | 0.0003 | | 10 minutes | 80/88 (90.9%) | 58/59 (98.3%) | 22/29 (75.9%) | 0.0015 | PMA P170012: FDA Summary of Safety and Effectiveness Data {36} Table 28: Proportion Achieving Hemostasis for Orthopedic Lower Extremity Surgical Arm | Time | All | HEMOBLAST™ | G+T | P-value | | --- | --- | --- | --- | --- | | 3 minutes | 74/90 (78.7%) | 54/64 (87.1%) | 20/32 (62.5%) | 0.0082 | | 6 minutes | 89/90 (94.7%) | 62/64 (100.0%) | 27/32 (84.4%) | 0.0037 | | 10 minutes | 92/90 (97.9%) | 62/64 (100.0%) | 30/32 (93.8%) | 0.1135 | ## Secondary endpoints ### Mean preparation time The preparation time for the HEMOBLAST™ group was found to be significantly shorter than the G+T group, with a mean of 0.37 minutes (22 seconds) for the HEMOBLAST™ group and a mean of 2.40 minutes (144 seconds) for the G+T group (-2.03 [-2.10, -1.86], p&lt;0.0001). This shows that HEMOBLAST™ takes 2.03 minutes (122 seconds) less to prepare than the control agent. ### Non-inferiority in achieving hemostasis – 3 min The HEMOBLAST™ group met the secondary endpoint of non-inferiority in success achieving hemostasis at the 3 minute time point, with 71.1% of HEMOBLAST™ subjects achieving hemostasis versus 45.8% in the G+T group; a difference of 27.5% (14.0% to 40.9%, p&lt;0.0001 for non-inferiority). ### Superiority in achieving hemostasis – 6 minutes The secondary endpoint of superiority of HEMOBLAST™ in achieving hemostasis at 6 minutes was met in the overall population. As described in the Primary Endpoint section, the proportion of HEMOBLAST™ subjects achieving hemostasis at 6 minutes was 93.1% versus 73.5% for the G+T group (p=0.0001 for superiority). ### Superiority in achieving hemostasis – 3 minutes The secondary endpoint of superiority of HEMOBLAST™ in achieving hemostasis at 3 minutes was met in the overall population. As described in the Primary Endpoint section, the proportion of HEMOBLAST™ subjects achieving hemostasis at 3 minutes was 71.1% versus 45.8% for the G+T group (p=0.0001 for superiority). In regard to surgical procedure and baseline TBS, the number of subjects in each surgical arm were found to be similar between the two (2) treatment groups, as were the locations for the surgical procedure, the TBS tissue type, and the conventional procedures for hemostasis. The estimated size of the TBS was also found to be similar between treatment groups, with a mean of 5.4 cm² in the HEMOBLAST™ group and 5.8 cm² in the G+T group. ## 3. Pediatric Extrapolation In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population. PMA P170012: FDA Summary of Safety and Effectiveness Data {37} PMA P170012: FDA Summary of Safety and Effectiveness Data Page 38 ## E. Financial Disclosure The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 39 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data. ## XI. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION ### Outside United States (OUS) Study The OUS study was designed to evaluate the efficacy and safety of a HEMOBLAST device prototype in comparison with a control group. The study had three (3) surgical arms cardio-thoracic, abdominal, orthopedic (lower extremity, and spine). This was a prospective comparative randomized, single-blinded, and multicenter, non-inferiority study. The study was prematurely terminated due to changes in device applicator and poolability with the information from the U.S. pivotal study. There were 400 subjects with planned enrollment; however, the actual enrollment consisted of 78 subjects (26 control, 52 HEMOBLAST). All subjects eligible for participation in the study were screened according to pre-defined inclusion/exclusion peri-operative criteria. On the patients assessed prior to study termination, the 6 minutes, hemostatic success was 100% with control devices and 91.5% with Hemoblast; a difference of -8.5% 95% CI: -20.5%, 5.6%. There were 41 minor complications that were not related to Hemoblast. There was one (1) patient with pulmonary emboli in the Hemoblas…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...