P990030 · Angiotech Medical Device Technologies · PMX · Jun 13, 2000 · General, Plastic Surgery
Device Facts
Record ID
P990030
Device Name
VITAGEL SURGICAL HEMOSTAT
Applicant
Angiotech Medical Device Technologies
Product Code
PMX · General, Plastic Surgery
Decision Date
Jun 13, 2000
Decision
APWD
Regulation
21 CFR 878.4490
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
CoStasis® (also known as DynaStat™) Surgical Hemostat (hereafter referred to only as CoStasis in this Summary of Safety and Effectiveness) is indicated in surgical procedures (other than in neurosurgical, ophthalmic, and urological) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.
Device Story
CoStasis is an absorbable hemostatic agent combining bovine collagen, bovine thrombin, and autologous plasma. The system includes a CoStasis syringe (collagen/thrombin/buffer), a transfer syringe for patient plasma, and a delivery system (joiner, spray head/cannula, syringe clip/support). In the OR, surgical assistants assemble the system; the surgeon applies the mixture to bleeding sites. The device acts as a hemostatic sealant. It is used in general, hepatic, cardiac, and orthopedic surgeries to control diffuse bleeding. By accelerating clot formation, it reduces time to hemostasis compared to conventional methods like collagen sponges or electrocautery, potentially reducing blood loss and surgical time.
Clinical Evidence
Clinical evidence includes a randomized, prospective, multicenter trial of 318 patients (general, hepatic, cardiac, orthopedic). Primary endpoint: success rate of complete hemostasis (10 min for non-cardiac, 3 min for cardiac). CoStasis demonstrated superior hemostatic success rates (97% vs 66% non-cardiac; 93% vs 55% cardiac) and significantly faster median time to hemostasis compared to control (collagen sponge or standard of care). Safety evaluated via adverse events and antibody titers (bovine collagen, thrombin, Factor V); no significant safety concerns related to device use were identified.
Technological Characteristics
Components: bovine collagen (20 mg/ml), bovine thrombin (≥300 U/ml), 40 mM CaCl2 buffer, and autologous plasma. Delivery system: radiation-stable plastics meeting USP Class VI requirements. Single-use, sterile, refrigerated storage for CoStasis syringe. Mechanical mixing via spray head or cannula. No electronic components or software.
Indications for Use
Indicated for patients undergoing surgical procedures (excluding neurosurgical, ophthalmic, and urological) requiring adjunct hemostasis when conventional methods are ineffective. Contraindicated in patients with known allergies to bovine-origin materials; do not inject into or allow entry into blood vessels.
Regulatory Classification
Identification
An absorbable hemostatic agent or dressing is a device intended to produce hemostasis by accelerating the clotting process of blood. It is absorbable.
Reference Devices
Thrombin-JMI®
Instat collagen sponge
Tisseel
Submission Summary (Full Text)
{1}
P990030
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION:
**DEVICE GENERIC NAME:** Absorbable Hemostatic Agent
**DEVICE TRADE NAME:** CoStasis® Surgical Hemostat
DynaStat™ Surgical Hemostat
**APPLICANT NAME AND ADDRESS:** Cohesion Technologies, Inc.
2500 Faber Place
Palo Alto, CA 94303 USA
**DATE OF PANEL RECOMMENDATION:** None
**PMA NUMBER:** P990030
**DATE OF GMP INSPECTION:** November 3, 1999
**DATE OF NOTICE OF APPROVAL TO APPLICANT:** JUN 13 2000
## II. INDICATIONS FOR USE:
CoStasis® (also known as DynaStat™) Surgical Hemostat (hereafter referred to only as CoStasis in this Summary of Safety and Effectiveness) is indicated in surgical procedures (other than in neurosurgical, ophthalmic, and urological) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.
## III. CONTRAINDICATIONS:
- Do not inject CoStasis® Surgical Hemostat into blood vessels or allow it to enter blood vessels. Extensive intravascular clotting and even death may result.
- Do not use CoStasis® Surgical Hemostat in patients with known allergies to materials of bovine origin.
See Contraindications in the Package Insert for Thrombin-JMI® enclosed in the CoStasis package.
{2}
2
9
# IV. WARNINGS AND PRECAUTIONS:
The warnings and precautions can be found in the CoStasis® Surgical Hemostat labeling.
# V. DEVICE DESCRIPTION:
## A. COSTASIS SYRINGE:
The CoStasis syringe contains a suspension of 20 mg/ml bovine collagen and at least 300 U/ml bovine thrombin in a 40 mM CaCl₂ buffer that is filled aseptically. After fill, the CoStasis syringe is sealed with a sterile barrier Luer closure and sealed inside a ported pouch. The ported pouch maintains the sterility of the surface of the syringe to allow placement into the sterile operating field. CoStasis is supplied in a fill volume of 5 ml for convenience of use. The CoStasis syringe is single use and is stored at refrigerated temperature.
## B. TRANSFER SYRINGE IN PORTED POUCH:
The transfer syringe is made from radiation stable plastics meeting USP Class VI requirements. The transfer syringe is sealed in a ported pouch with closure and is then radiation sterilized. Patient plasma is aseptically transferred from the CellPaker to the sterile transfer syringe. After plasma transfer, the pouch is opened and the transfer syringe is removed and placed into the sterile operating field. The transfer syringe is single use and holds 5 ml of plasma.
## C. DELIVERY SYSTEM:
The delivery system is provided to combine and mix the contents of the CoStasis syringe with the patient's plasma syringe at the time of administration to the bleeding site. The CoStasis delivery system is comprised of:
- A joiner to connect the CoStasis syringe to the syringe containing the patient's own plasma;
- A spray head incorporating a mixer element;
- A cannula incorporating a mixer element (provided as an alternate delivery component to the spray head);
- A syringe clip to permit the depression of the plunger rods of the two syringes simultaneously; and
- A syringe support to aid the surgeon in holding the assembled device.
{3}
In the event of clogging or accidental contamination during use, an additional mixer/spray head and mixer/cannula are provided. The delivery system parts are single use, sterile, and are made from radiation stable plastics meeting USP Class VI requirements. The parts are sealed in a tray within a tray and are then radiation sterilized. The surgical assistants assemble the delivery system in the operating room. Total time for assembly is less than five minutes. A description of the assembly procedure for the delivery system with the CoStasis syringe and autologous plasma syringe is provided in the CoStasis Directions for Use.
## D. CELLPAKER PLASMA COLLECTION DEVICE:
The CellPaker contains sodium citrate and is used to collect and anticoagulate the patient's blood. After blood collection, the CellPaker is placed into a centrifuge to separate the plasma from cells. The CellPaker is made from radiation stable plastic meeting USP Class VI requirements, and is radiation sterilized prior to aseptic filling with sodium citrate. The CellPaker containing sodium citrate is single use and is stored at room temperature.
The procedure for obtaining autologous plasma is described in the Directions for Use provided with the CellPaker. For convenience and economy of use, the CellPaker is provided in single or multiple unit boxes.
## VI. ALTERNATIVE PRACTICES OR PROCEDURES:
Intra-operative hemostasis may be accomplished by a variety of methods including occlusive procedures (e.g., direct pressure, suture, etc.), electrocautery, lasers, fibrin sealants, and the use of thrombin and absorbable hemostatic agents. The absorbable hemostatic agents consist of porcine gelatin powder and sponges, bovine gelatin matrix, bovine collagen sheets and sponges and oxidized cellulose sponges. Bovine thrombin is also used for hemostasis, but it often used in conjunction with gelatin or collagen hemostatic agents. Thrombin may not be used with oxidized cellulose, since the acidity of the cellulose inactivates the thrombin.
## VII. MARKETING HISTORY:
CoStasis is CE Marked and was introduced into commercial distribution in Europe in January 1999, however, CoStasis has not been marketed in the United States. The CoStasis Surgical Hemostat has not been withdrawn from marketing for any reason relating to the safety and effectiveness of the device.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH:
In a randomized prospective, concurrently controlled clinical trial, 318 patients were treated with CoStasis or the Control (a collagen absorbable hemostat for the non-cardiac procedures, and other conventional methods selected by the surgeon for cardiac
{4}
procedures). The most common adverse events recorded during and after the application of the hemostatic agents were fever, pain, nausea, and atelectasis. Table 1 is a complete list of adverse events reported in greater than 5% of CoStasis treated patients that were observed in the clinical trial. The corresponding adverse events for the Control group are listed for comparison. None of the adverse events listed in the table were determined to be attributed to CoStasis.
Table 1: A Summary of Adverse Events Occurring in Greater Than 1% of the Study Patients.
| Adverse Events Reported in Greater than 5% of Patients in CoStasis Clinical Trial Patients | | |
| --- | --- | --- |
| Adverse Event | CoStasis | Control* (absorbable collagen hemostat) |
| Fever | 51 (30.4%) | 41 (27.3%) |
| Pain | 38 (22.6%) | 33 (22.0%) |
| Nausea | 34 (20.2%) | 29 (19.3%) |
| Atelectasis | 32 (19.0%) | 27 (18.0%) |
| Pleural Effusion | 24 (14.3%) | 28 (18.7%) |
| Anemia | 23 (13.7%) | 27 (18.0%) |
| Peripheral Edema | 19 (11.3%) | 11 (7.3%) |
| Tachycardia | 16 (9.5%) | 20 (13.3%) |
| Constipation | 14 (8.3%) | 9 (6.0%) |
| Rash | 14 (8.3%) | 15 (10.0%) |
| Infection | 13 (7.7%) | 13 (8.7%) |
| Abnormal Healing | 12 (7.1%) | 12 (8.0%) |
| Abdominal Pain | 11 (6.5%) | 15 (10.0%) |
| Edema | 11 (6.5%) | 7 (4.7%) |
| Lung Disorder | 11 (6.5%) | 11 (7.3%) |
| Pericarditis | 10 (6.0%) | 13 (8.7%) |
| Nausea/Vomiting | 10 (6.0%) | 8 (5.3%) |
| Lung Edema | 9 (5.4%) | 9 (6.0%) |
| Pneumonia | 9 (5.4%) | 9 (6.0%) |
| Pruritis | 9 (5.4%) | 12 (8.0%) |
*For the cardiac group, the control was selected by the surgeon, based on his/her conventional methods, including absorbable collagen hemostats, surgical tamponade, electrocautery with or without tamponade, gelatin sponge with thrombin, or no treatment.
Other adverse events observed in less than 5% of the CoStasis treated clinical trial patients were bleeding, hypertension, hypotension, dyspnea, urinary tract infection, back pain, apnea, abscess, allergic reaction, death, chest pain, ecchymosis, hypokalemia, confusion, ileus, pharyngitis, ascites, asthenia, headache, atrial fibrillation, hypervolemia, dizziness, insomnia, paresthesia, increased coughing, oliguria, sepsis, arrhythmia, diarrhea, jaundice, hypomagnesemia, bone pain, anxiety, respiratory disorder, skin ulcer, hematuria, abnormal kidney function, fibrotic surgical wound, necrosis, pain at injection site, angina pectoris, heart arrest, heart failure, ventricular tachycardia, gastritis, GI disorder, melena, vomiting, leukocytosis, leukopenia, acidosis, bilirubinemia, hyperglycemia, hyponatremia, agitation, depression, hallucination, abnormal thinking, pneumothorax, sweating, kidney failure, breast pain, abdominal enlargement, cellulitis, chills, edema at injection site, hernia, anemia, cardiovascular disorder, extrasystoles,
{5}
shock, syncope, superior ventricular tachycardia, anorexia, dyspepsia, GI bleeding, rectal bleeding, abnormal liver function, oral monilia, increased coagulation time, thrombocytopenia, increased creatinine, cyanosis, hypoglycemia, hypophosphatemia, joint disorder, myasthenia, abnormal dreams, stupor, asthma, voice alteration, skin disorder, abnormal urination, vaginitis, asthesia, granuloma, hypothermia, injection site reaction, injury accident, neck rigidity, necrosis at injection site, overdose, sarcoma, seroma, thirst, vascular anomaly, sinus bradycardia, bundle branch block, cardiomegaly, pericardial effusion, ventricular extrasystoles, myocardial infarction, deep thrombophlebosis, thrombosis, thrombophlebosis, peripheral vascular disorder, varicose vein, anastomosis leak, liver carcinoma, colitis, eructation, dysphagia, tongue edema, fecal incontinence, liver failure, stomatitis, tooth disorder, agranulocytosis, hypovolemia, seroma, alkalosis, avitaminosis, dehydration, general edema, gout, hypocalcemia, arthrosis, myalgia, general spasm, aphasia, convulsions, hemiplegia, hypertonia, movement disorder, neuropathy, bronchiectasis, pulmonary embolism, emphysema, hemoptysis, hyperventillation, hypoxemia, hypoxia, infection, lung function decrease, rhinitis, vesiculobullous rash, bone pain, eye pain, breast carcinoma, epididymitis, impotence, urinary incontinence, metrorrhagia, nephrosis, urethral pain, testis disorder, urinary retention.
One adverse event was deemed by the surgeon to be related to the use of CoStasis: blocked inguinal drainage tube reported on day 11 post surgery. No other adverse events were deemed by the surgeon to be related to the use of CoStasis.
Fifteen (15) subjects died during the course of this study. Eleven died before the completion of the 8 week post-op visit and 4 died as of (or after) their final follow-up. None of the deaths were attributable to either CoStasis or the control, rather, it was determined that these patients died as a result of their underlying disease conditions or from unrelated postoperative surgical complications, e.g. respiratory insufficiency, sepsis and multi-organ failure. Nearly half (7 of the 15 deaths) were in the cardiac bypass group, 6 were among the general surgery subjects, and 2 in the hepatic group. Seven deaths occurred in CoStasis treated patients and 8 deaths were in the controls.
## Reported Adverse Events Seen With other Absorbable Hemostatic Agents:
- Hemostatic agents may serve as a nidus for infection and abscess formation and have been reported to potentiate bacterial growth.
- Giant cell granulomas have been observed at implant sites when used in the brain.
- Compression of the brain and spinal cord resulting from the accumulation of sterile fluid has been observed.
- Multiple neurologic events were reported when absorbable hemostatic agents were used in laminectomy operations, including cauda equina syndrome, spinal stenosis,
{6}
meningitis, arachnoiditis, headaches, paresthesias, pain, bladder and bowel dysfunction, and impotence.
- The use of absorbable hemostatic agents during the repair of dural defects associated with laminectomy and craniotomy operations, has been associated with fever, infection, leg paresthesias, neck and back pain, bladder and bowel incontinence, cauda equina syndrome, neurogenic bladder, impotence, and paresis.
- Foreign body reactions, “encapsulation” of fluid, and hematoma have been observed at implant sites.
- Toxic shock syndrome was reported in association with the use of absorbable hemostatic agents in nasal surgery.
- Fever, failure of absorption, and hearing loss have been observed when absorbable hemostatic agents were used during tympanoplasty.
## Adverse Reactions to Thrombin:
See Adverse Reactions in Package Insert for Thrombin-JMI® enclosed in the CoStasis™ Surgical Hemostat package.
## IX. SUMMARY OF PRECLINICAL STUDIES:
There is an extensive history on the safe use of bovine collagen and bovine thrombin in humans. Therefore, standard biocompatibility tests as defined in the ISO 10993 standard were not conducted on CoStasis. Animal studies were done to evaluate the hemostatic performance of CoStasis in a variety of surgical wounds. Key studies and findings are summarized below.
## A. SAFETY STUDIES:
### Evaluation of Composite Tissue Adhesive (CoStasis Surgical Hemostat) as a Hemostat in a Diffuse Organ Bleeding Rabbit Survival Study.
The local biocompatibility of CoStasis was compared with fibrin sealant and Instat collagen sponge. Incisions of controlled dimensions were made in the kidneys of rabbits. The test materials were applied to the defect and animals were evaluated at 7, 14, 30, and 60 days post-operatively. Histologically, all materials demonstrated equivalent inflammation at 7 and 14 days post-operatively. Little to no CoStasis was observed in animals 14 days post-operatively. Instat was observed in animals at 30 days post-operatively, and fibrin sealant was observed in animals at up to 60 days post-operatively.
{7}
7
14
# B. EFFECTIVENESS STUDIES:
Comparison of CoStasis Surgical Hemostat with Instat, Tisseel, and Fibrin Sealant Controls for Hemostatic Qualities in a Diffuse Bleeding Model in Parenchymal Organs of Rabbits.
The performance of CoStasis was compared to some other absorbable hemostatic products. These comparative products were a commercially available absorbable collagen sponge, a commercially available fibrin sealant and a non commercially available fibrin sealant. Incisions of controlled dimensions were made in the kidney and spleen of rabbits. The test materials were applied to the defect and time to hemostasis measured. The mean time to hemostasis was significantly faster using CoStasis as compared to the control materials. In spleen sites, CoStasis had a consistently faster time to hemostasis than all of the controls but only statistically significantly faster than the collagen pad and the non-commercially available fibrin sealant.
Effect of Variation in Platelet and Fibrinogen Levels in Plasma on Hemostatic Performance of CoStasis Surgical Hemostat.
This study evaluated the effect of plasma depleted of platelets and/or fibrinogen on the hemostatic performance of CoStasis. A rabbit model involving incisions of controlled dimensions in kidney and spleen was used. Time to hemostasis was measured. Plasma depleted of platelets or fibrinogen alone did not significantly affect the hemostatic performance of CoStasis. These formulations also had significantly faster times to hemostasis than the control collagen sponge. CoStasis with plasma depleted of platelets and fibrinogen resulted in a significantly slower time to hemostasis than CoStasis with plasma containing normal levels of platelets and fibrinogen.
Hemostatic Performance of CoStasis® Surgical Hemostat in Aspirin and Heparin Anticoagulated Rabbits.
CoStasis was compared to commercially available hemostatic agents (collagen sponge and fibrin sealant) in aspirin and heparin anticoagulated rabbits. A rabbit model involving incisions of controlled dimensions in kidney and spleen was used. Separate groups of rabbits were pretreated with one of two different medications known to impair hemostasis (aspirin and heparin). Impairment of hemostasis was confirmed by laboratory assessment of platelet activity. Mean time to hemostasis and total blood loss was measured. Comparison of CoStasis to an absorbable collagen sponge was done in aspirin treated animals. Comparison of CoStasis, an absorbable collagen sponge, and fibrin sealant was done in heparin treated animals. In all cases, the mean time to hemostasis was faster for CoStasis as compared to the control hemostats in aspirin or heparin treated animals. Blood
{8}
loss was also significantly reduced in all animals treated with CoStasis.
## Evaluation of CoStasis Surgical Hemostat for Hemostatic Qualities in an Ovine Parenchymal Organ Bleeding Model.
The hemostatic performance of CoStasis was compared to fibrin sealant in sheep. In addition, systemic anticoagulation conditions were evaluated using a high dose of heparin. Incisions of controlled dimensions were made in the kidneys, spleen, and liver. The test materials were applied to the defect. Blood loss and time to hemostasis were recorded. In all surgical sites in normal or heparinized animals, CoStasis demonstrated faster mean times to hemostasis than fibrin sealant. Blood loss between the two groups was similar in normal animals. In heparinized animals, treatment with CoStasis resulted in significantly reduced blood loss when compared to treatment with fibrin sealant.
## C. PRECLINICAL CONCLUSIONS:
These preclinical studies showed that CoStasis had a faster time to hemostasis than commercially available control materials (an absorbable collagen sponge and a fibrin sealant) and also a non commercially available fibrin sealant. When the plasma component of CoStasis was depleted of either platelets or fibrinogen, the hemostatic performance of CoStasis remained. In addition, complete hemostasis using CoStasis was achieved in anticoagulated (aspirin or heparin treated) animals.
## X. SUMMARY OF CLINICAL STUDIES:
The objective of the clinical studies was to assess the safety and effectiveness of the CoStasis™ Surgical Hemostat in the treatment of bleeding during surgical procedures. Three Studies were summarized in this PMA, a US feasibility study, an expanded European study and an expanded US study.
### A. US FEASIBILITY STUDY:
A feasibility study was completed in the United States evaluating CoStasis in treating skin graft donor sites. Nine (9) patients were enrolled in this study.
Time to hemostasis using CoStasis was measured from the moment that the fresh wound was completely coated with the device. The clinical protocol used the terms *primary* and *secondary* hemostasis because it was anticipated that sites would pass through an oozing phase (primary hemostasis) before bleeding stopped completely (secondary hemostasis). Because the protocol did not allow reapplication of the device, the recorded times refer to observations made after a single application. In eight of the nine subjects, primary and secondary hemostasis times were identical. Primary hemostasis was achieved in a mean
{9}
time under ten seconds. In one subject, the bleeding was controlled at seven seconds, while a small area of breakthrough bleeding continued to bleed for 142 seconds.
Adverse events were solicited at each visit; a total of 58 events were reported. Investigators assigned causality for all adverse events and no reported event was attributed to the investigational device. All reported events were classified as either related to the surgical procedure or unrelated. Healing of the harvested sites was considered normal with the exception of one patient who removed his own dressing prematurely.
## B. EXPANDED EUROPEAN STUDY:
The European study included treatment of 111 subjects in seven investigational centers in Germany and Austria. The primary effectiveness outcome was a comparison of the rate of failure to achieve hemostasis for CoStasis versus a commercially available fibrin sealant at the anastomotic site during coronary artery bypass grafting. Additional bleeding sites, beyond the primary anastomotic site, were evaluated and included the sternal edge (bone) and the chest wall capillary bed. Standard surgical practice in Europe uses a fibrin sealant only on the anastomotic site. Therefore, other methods and techniques, including cautery, tamponade, and bone wax were used as controls for the sternal edge and chest wall capillary bed.
For the primary performance endpoint of failure to achieve hemostasis on the anastomotic suture line, CoStasis was not different from the fibrin sealant (Fisher’s Exact p = 0.708). In addition to this primary endpoint, analyses of times to hemostasis were conducted. The time to complete hemostasis (median ± 1 standard error) on the anastomotic site was nearly identical for CoStasis (52 ± 16 seconds) and the commercially available fibrin sealant (52 ± 6 seconds), Wilcoxon p-value = 0.68. When comparing median time to complete hemostasis for CoStasis versus the control along the sternal edge, CoStasis was faster (70 ± 14 seconds) than controls (136 ± 28 seconds), Wilcoxon p-value = 0.03. On the capillary beds, the median time to complete hemostasis were similar for CoStasis hemostat (60 ± 11 seconds) and control (120 ± 17 seconds), Wilcoxon p-value = 0.14.
## C. EXPANDED US STUDY:
### Study Design:
The US study was designed to evaluate the safety and effectiveness of CoStasis in hepatic, general, cardiac and orthopedic surgery. As a control, a collagen hemostatic sponge was used in the general, hepatic and orthopedic groups. A
{10}
comparison of CoStasis to standard of care was used in the cardiac group.
## Study Procedure:
The study was performed at 10 centers and randomized 347 subjects. Twenty-nine subjects were removed after randomization but before treatment, 14 CoStasis and 15 controls. A total of 318 patients advanced to treatment and became the intent to treat group. Of the 318 patients treated, 6 were eliminated from effectiveness analyses for various protocol violations leaving 312 evaluable patients. The patients treated in each surgical group are provided in Table 2.
Table 2: Surgical Groups
| Surgical Group | # Treated |
| --- | --- |
| General Surgery | 154 |
| Cardiac Surgery | 74 |
| Hepatic Surgery | 68 |
| Iliac Crest Surgery | 22 |
| Total Patients Evaluated | 318 |
The primary effectiveness outcome was a comparison of the success rate for each treatment group to achieve complete hemostasis (at 10 minutes for all groups except the cardiac group where it was at 3 minutes) for CoStasis versus control for each bleeding site. The statistical analysis used the Fisher's Exact test on the success rates to achieve hemostasis between treatments and controls. As CoStasis was being evaluated as a general hemostatic device, data from all surgical groups were pooled for statistical analysis. In addition to a comparison of the hemostasis success rates, discrete times to achieve hemostasis were compared using a Kaplan-Meier survival plot. Using this plot, the Logrank (Mantel-Cox) and Breslow-Gehan-Wilcoxon p-values were calculated.
This study was also designed to evaluate the safety of CoStasis. The safety evaluation included adverse events reported, clinical laboratory evaluations and an evaluation of antibovine and antihuman antibody titers to thrombin, collagen, and Factor V. Table 3 summarizes the patient accountability data for this clinical study.
{11}
11
Table 3: Patient Accountability Data
| Accountability Category | CoStasis | Control | Total |
| --- | --- | --- | --- |
| Number Randomized | 182 | 165 | 347 |
| Number Treated | 168 | 150 | 318 |
| Number Evaluated for Effectiveness | 166 | 146 | 312 |
| Number Evaluated for Safety | 168 | 150* | 318 |
*One control patient was treated in two surgical groups. This patient was evaluated for safety in each group.
Study Results:
A total of 152 patients were treated in the general surgery group. The general surgery group was analyzed both as a whole group, and also in five subgroups that included, laparotomy (24 patients), retroperitoneal surgery (39 patients), muscle flaps (47 patients), breast/soft tissue (31 patients), and liver transplants (11 patients). The liver transplant patients were included in the general surgical group rather than the hepatic group as the bleeding site treated was the retroperitoneal bed and not the liver itself. A total of 67 patients were treated in the hepatic group. A total of 74 patients were treated in the cardiac group. Sixty-six (66) subjects in the cardiac group were evaluated for control of bleeding at the sternum, fifty-eight (58) for control of bleeding at the anastomotic sites, and thirty-six (36) for control of bleeding in the chest wall capillary bed. A total of 19 patients were treated in the iliac crest surgery group.
Table 4 analyzes the treatment sites achieving complete hemostasis within 10 minutes (intent to treat) for the various surgical groups using the 10-minute time limit for complete hemostasis.
{12}
Table 4: Treatment Sites Achieving Complete Hemostasis Within 10 Minutes Intent-to-Treat Sites (Success/Total)
| Surgical Group | CoStasis | Control |
| --- | --- | --- |
| All Treated Sites (non-cardiac group) | 97% (126/130) | 66% (75/114) |
| General | 97% (77/79) | 67% (50/75) |
| Hepatic | 100% (39/39) | 69% (20/29) |
| Iliac Crest | 83% (10/12) | 50% (5/10) |
The treatment sites achieving complete hemostasis within 3 minutes in the cardiac group is given in Table 5. For the cardiac group each treatment site was timed for 3 minutes as the success time limit rather than the 10-minute limit used for the other surgical groups. For this analysis, intent to treat patients are presented.
Table 5: Treatment Sites Achieving Complete Hemostasis Within 3 Minutes Intent-to-Treat Sites (Success/Total)
| Cardiac Group | CoStasis | Control* |
| --- | --- | --- |
| All Treated Sites | 93% (74/80) | 55% (46/84) |
| Anastomotic | 97% (28/29) | 59% (17/29) |
| Sternal Edge | 83% (25/30) | 44% (15/34) |
| Capillary Bed | 100% (21/21) | 67% (14/21) |
*For the cardiac group, the control was selected by the surgeon, based on his/her conventional methods, including absorbable collagen hemostats, surgical tamponade, electrocautery with or without tamponade, gelatin sponge with thrombin, or no treatment.
The study data for the number of treatment sites achieving complete hemostasis within the success criteria time limit (10 minutes for all surgical types except cardiac surgery and 3 minutes for cardiac surgery) were compared using two statistical tests. CoStasis and control were shown to be equivalent using the Blackwelder and Chang test, using a $\delta$ (clinically significant difference) of 0.25 (p < 0.01). For treatment sites achieving complete hemostasis, the difference between CoStasis and control was also shown to be statistically significant using the Fisher's Exact test (p < 0.05), except for the Iliac Crest Group (p = 0.17).
{13}
Tables 6 shows the cumulative percent of treatment sites with complete hemostasis over the entire 10-minute study period for the combined general, hepatic and iliac crest patients and Table 7 shows the cumulative percent of treatment sites with complete hemostasis for the cardiac patients.
| Table 6: Cumulative Percent of Treatment Sites with Complete Hemostasis Over 10 Minutes Intent-to-Treat Sites for General, Hepatic, and Iliac Crest Groups (Success/Total) | | |
| --- | --- | --- |
| Time Interval | CoStasis | Control |
| 0 – 1 minute | 37% (48/130) | 4% (5/114) |
| 1 – 2 minutes | 60% (78/130) | 17% (19/114) |
| 2 – 3 minutes | 75% (97/130) | 27% (31/114) |
| 3 – 6 minutes | 94% (122/130) | 50% (57/114) |
| 6 – 10 minutes | 97% (126/130) | 66% (75/114) |
| Table 7: Cumulative Percent of Treatment Sites with Complete Hemostasis Over 3 Minutes Intent-to-Treat Sites for Cardiac Group (Success/Total) | | |
| --- | --- | --- |
| Time Interval | CoStasis | Control |
| 0 – 1 minute | 51% (41/80) | 2% (2/84) |
| 1 – 2 minutes | 75% (60/80) | 20% (17/84) |
| 2 – 3 minutes | 93% (74/80) | 55% (46/84) |
For the above analyses, hemostasis time was measured with a stopwatch at all treatment sites. Timing for subjects treated with CoStasis began as the spray made its first contact with the bleeding surfaces. Timing for control subjects began with the first application of the control collagen sponge. Times to complete hemostasis (median ± 1 standard error) for the intent to treat patients were also collected and are presented in Tables 8 and 9. The times to achieve complete hemostasis for all study groups were statistically different using the Breslow-Gehan-Wilcoxon test (p < 0.05).
{14}
| Table 8: Times to Achieve Complete Hemostasis General, Hepatic and Iliac Crest Groups (Median ± 1 standard error) (Intent to Treat Patients) | | |
| --- | --- | --- |
| Surgical Group | CoStasis (seconds) | Control (seconds) |
| All Patients | 90 ± 20 | 347 ± 27 |
| General | 65 ± 13 | 345 ± 63 |
| Hepatic | 150 ± 19 | 360 ± 52 |
| Iliac Crest | 90 ± 35 | 347 ± 343 |
| Table 9: Time to Achieve Complete Hemostasis Cardiac Group (Median ± 1 standard error) (Intent to Treat Patients) | | |
| --- | --- | --- |
| Treatment Site | CoStasis (seconds) | Control (seconds) |
| All Treatment Sites | 55 ± 4 | 180 ± 6 |
| Sternal Edge | 68 ± 33 | > 180 |
| Anastomosis | 62 ± 12 | 180 ± 53 |
| Capillary Bed | 50 ± 5 | 180 ± 19 |
Additional measurements of effectiveness including transfusion requirements and tube drainage were similar between treated and control patients.
## Serology Results:
Patients' sera were evaluated using an indirect enzyme-linked immunosorbent assay (ELISA). Sera samples obtained pre-surgery (Visit 1) and again at approximately eight weeks (Visit 5) following surgery were tested for antibodies to bovine dermal collagen, bovine thrombin, and bovine Factor V/Va.
Of all the patients' sera tested, only 2 of the 172 pretreatment samples (~1%) were positive for any antibodies against bovine collagen, both were in the CoStasis group. At the eight week timepoint (visit 5), 8% of CoStasis treated patients had a positive antibovine collagen antibody titer. Of the control patients, 4% converted
{15}
to a positive titer, all being in the collagen sponge group. None of the serum samples with equivocal or positive titers to bovine collagen showed any reactivity to human collagen type I or III.
Approximately 6% of all pretreatment sera were equivocal or positive for reactivity to bovine thrombin (6/172), equally divided between CoStasis and controls. One of these patients showed equivocal reactivity to human thrombin. At visit 5, positive antibody titers to bovine thrombin were found in 29% of sera from CoStasis treated patients. Twelve percent of the control patients had positive titers. Of the 27 CoStasis treated patients that had positive antibovine thrombin antibody titers, 30% (8/27) showed positive antihuman thrombin antibody titers. In the control group, of the 10 patients that had positive antibovine thrombin antibody titers, 10% (1/10) were positive antihuman thrombin antibody titers.
Three patients had reactivity to bovine Factor V in their pretreatment sera: two with equivocal titers and one positive. All of these patients were in the control group. A single CoStasis patient developed a positive antibovine Factor V antibody titer. This patient was also positive for antibovine thrombin antibodies at visit 5, but not positive for antihuman thrombin or antihuman Factor V antibodies. In the controls, one pretreatment patient with a positive titer remained positive. None of the samples with equivocal or positive titers to bovine Factor V showed any reactivity to human Factor V.
## Clinical Laboratory Evaluations:
Clinical laboratory evaluations were undertaken pre-surgery and again post-surgery. No clinically significant findings were observed in the change-from-baseline laboratory tests for all treated patients.
## XI. CONCLUSIONS DRAWN FROM THE STUDIES:
### A. RISK BENEFIT ANALYSIS:
The absence of any significant adverse event related to product use, and the hemostatic effectiveness of CoStasis, attest to the acceptable risk/benefit ratio of CoStasis in controlling diffuse bleeding.
### B. SAFETY:
The materials used to manufacture CoStasis are well defined with proven acceptable biocompatibility. The number of patients developing antibodies to bovine collagen, thrombin and Factor V were not different from that experienced in other applications of these materials, e.g. topical thrombin used in surgery and injectable collagen products for soft tissue augmentation. The absence of any
{16}
significant adverse event related to use of CoStasis supports the safety of CoStasis for use in controlling diffuse bleeding.
## C. EFFECTIVENESS:
The results of the preclinical and clinical testing demonstrated that there is reasonable assurance of safety and effectiveness for the CoStasis® Surgical Hemostat for the stated indication for use. The sponsor performed a randomized, parallel, controlled, comparative, multicenter clinical trial designed to determine the hemostatic ability of CoStasis in surgical patients. The general, hepatic and orthopedic surgery control groups were treated with a legally marketed absorbable collagen sponge while standard of care was used as the control for the cardiac surgery patients. The preclinical and clinical data support the hemostatic performance and safety of CoStasis when compared to the commercially available control treatments. CoStasis controlled bleeding in less time than the control treatments when time to hemostasis was compared. CoStasis was also observed to have a higher success rate than control when complete hemostasis was measured at 10 minutes for the general surgery, hepatic surgery and iliac crest surgery groups and at 3 minutes for the cardiac surgery group. Additional measurements of effectiveness including transfusion requirements and tube drainage were similar between treated and control patients.
## XII. PANEL RECOMMENDATION:
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the General and Plastic Surgery Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XIII. CDRH DECISION:
A GMP inspection was conducted of the Cohesion Technologies, Inc. facilities in Palo Alto, California on November 3, 1999, and they were found to be in compliance with the Device GMP Regulations.
CDRH issued an approval order on June 13, 2000.
## XIV. APPROVAL SPECIFICATIONS:
Directions for Use: See the labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions and Adverse Events in the labeling.
{17}
17
24
Postapproval Requirements and Restrictions: See approval order.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.