Acclarent Aera Eustachian Tube Balloon Dilation System
Applicant
Acclarent, Inc.
Product Code
PNZ · Ear, Nose, Throat
Decision Date
Sep 16, 2016
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 874.4180
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The system is intended for use in dilating the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction.
Device Story
System includes Eustachian Tube Balloon Catheter (ETBC) and Eustachian Tube Guide Catheter (ETGC); used to dilate cartilaginous portion of Eustachian tube. ETBC features non-compliant nylon balloon (6x16mm) and 'blueberry' tip to prevent advancement into bony isthmus; ETGC provides access with atraumatic angled tip. Operated by ENT physicians in operating room under general anesthesia. Physician uses endoscopic visualization to position device; inflation performed via external inflation device (e.g., Acclarent SE). Dilation relieves ETD symptoms (pressure, pain, clogging). Benefits include improved tympanometry and quality of life (ETDQ-7 scores). Risks include mucosal trauma, bleeding, or carotid artery injury; mitigated by CT screening, specialized tip design, and cadaver-based training.
Clinical Evidence
Prospective, multi-center, randomized controlled trial (2:1 ratio) comparing ETBC dilation + medical management (MM) vs. MM alone in 299 subjects (141 ETBC, 72 MM, 81 lead-in). Primary endpoint: tympanometry normalization at 6 weeks. Results: 51.8% (ETBC) vs 13.9% (MM) (p<0.0001). Secondary endpoint: ETDQ-7 score improvement; 90.6% (ETBC) vs 45.1% (MM) achieved MID. Durability demonstrated at 24 weeks. Safety: 0 serious device/procedure-related AEs; 16 non-serious AEs (mild/moderate) in 13 subjects.
Technological Characteristics
System components: ETBC (nylon balloon, dual-lumen shaft) and ETGC (hypotube, lubricious liner). Materials biocompatible per ISO 10993-1. Sterilization: Ethylene Oxide (EO) per ISO 11135. Max inflation pressure: 12 ATM. Dimensions: 6x16mm balloon. Connectivity: None (mechanical).
Indications for Use
Indicated for dilating the Eustachian tube for treatment of persistent Eustachian tube dysfunction in adults ages 22 and older.
Regulatory Classification
Identification
A Eustachian tube balloon dilation system is a prescription device that includes a flexible catheter attached to an inflatable balloon. The system is intended for use in dilating the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
(i) Mechanical testing, including tensile and flexural testing of catheter joints and materials.
(ii) Durability testing, including fatigue and burst pressure testing of the balloon materials and components.
(iii) Inflation and deflation characterization testing, including time and pressure measurements, and leak testing of the balloon.
(iv) Verification testing of safety features built into the device must be performed, including the characterization of catheter geometries and distal tip insertion limitation mechanisms.
(2) Simulated use testing in a clinically relevant model must demonstrate the reliability of the device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the device.
(5) Performance data must support shelf life by demonstrating continued sterility of the device, package integrity, and device functionality over the identified shelf life.
(6) Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device.
(7) Labeling must include:
(i) Detailed instructions for use.
(ii) A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes.
(iii) A shelf life.
In combination with the general controls of the FD&C Act, the Acclarent Aera™ Eustachian Tube Balloon Dilation System is subject to the following special controls:
- 1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
- a. Mechanical testing, including tensile and flexural testing of catheter joints and materials.
- b. Durability testing, including fatigue and burst pressure testing of the balloon materials and components.
- Inflation and deflation characterization testing, including time and pressure C. measurements, and leak testing of the balloon.
- d. Verification testing of safety features built into the device must be performed. including the characterization of catheter geometries and distal tip insertion limitation mechanisms.
- 2. Simulated use testing in a clinically relevant model must demonstrate the reliability of the device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube.
- 3. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 4. Performance data must demonstrate the sterility of the device.
- 5. Performance data must support shelf life by demonstrating continued sterility of the device, package integrity and device functionality over the identified shelf life.
- 6. Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device.
- 7. Labeling must include:
- a. Detailed instructions for use.
- b. A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes.
- c. A shelf life.
Submission Summary (Full Text)
{0}------------------------------------------------
### DE NOVO CLASSIFICATION REQUEST FOR ACCLARENT AERA™ EUSTACHIAN TUBE BALLOON DILATION SYSTEM
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Eustachian Tube Balloon Dilation System. A Eustachian tube balloon dilation system is a prescription device that includes a flexible catheter attached to an inflatable balloon. The system is intended for use in dilating the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction.
NEW REGULATION NUMBER: 874.4180
CLASSIFICATION: CLASS II
PRODUCT CODE: PNZ
#### BACKGROUND
DEVICE NAME: ACCLARENT AERA™ EUSTACHIAN TUBE BALLOON DILATION SYSTEM
SUBMISSION NUMBER: DEN150056
DATE OF DE NOVO: DECEMBER 4, 2015
CONTACT: ACCLARENT, INC. 33 TECHNOLOGY DRIVE IRVINE, CA 92618
#### INDICATIONS FOR USE
The ACCLARENT AERATM Eustachian Tube Balloon Dilation System is intended to dilate the Eustachian tube for treatment of persistent Eustachian tube dysfunction in adults ages 22 and older.
#### LIMITATIONS
The sale, distribution, and use of the ACCLARENT AERA™ Eustachian Tube Balloon Dilation System are restricted to prescription use in accordance with 21 CFR 801.109.
Limitations on device use are also achieved through the following contraindications. warnings, and precautions included in the Instructions for Use:
{1}------------------------------------------------
# CONTRAINDICATIONS
The ACCLARENT AERA™ Eustachian Tube Balloon Catheter is contraindicated for use in a Eustachian tube with an ipsilateral carotid artery that is dehiscent into the ET lumen or history of ipsilateral patulous Eustachian tube.
## WARNINGS
- Intended for single patient use only. DO NOT REUSE. ●
- . Patients with a history of skull base surgery, prior ear surgery, skull fracture, or anatomic abnormalities may have elevated risk of complications and should be radiographically screened for eligibility before treatment.
- DO NOT use product if the integrity of the sterile packaging has been ● compromised or if the device appears damaged.
- DO NOT use the ACCLARENT AERA™ Eustachian Tube Balloon Catheter if ● the device becomes damaged or touches a non-sterile object outside of the operating field.
- Never advance or retract the device against unknown resistance, as this could . cause tissue trauma or device damage.
- . Advancing the Eustachian Tube Balloon Catheter into the Eustachian tube against resistance may cause injury.
- DO NOT exceed the recommended maximum balloon inflation pressure of 12 ● atmospheres (ATM).
- Use only sterile saline or sterile water for inflation. DO NOT inflate with air. ●
- . The safety of the device as used under local anesthesia has not been evaluated.
## PRECAUTIONS
- DO NOT move the balloon while it is inflated. Ensure balloon is fully deflated ● during insertion and withdrawal.
- Radiographic assessment of the targeted Eustachian tube is recommended prior to ● any procedure involving balloon tuboplasty.
- Certain nasal anatomy such as a deviated nasal septum may preclude access to the . Eustachian tube/s resulting in failure to treat the target anatomy.
- DO NOT inflate the Eustachian Tube Balloon Catheter until it has exited the . ACCLARENT AERA™ Eustachian Tube Guide Catheter.
- DO NOT bend the Guide Catheter shaft. .
## DEVICE DESCRIPTION
The ACCLARENT® AERA™ Eustachian Tube Balloon Dilation System includes the Eustachian Tube Balloon Catheter (ETBC) and the Eustachian Tube Guide Catheter (ETGC). The system is provided with the following additional devices: the Acclarent SE Inflation Device (or Acclarent Balloon Inflation Device) and Relieva Extension Tubing. The Acclarent SE Inflation Device (or Acclarent Balloon Inflation Device) is used to inflate the ETBC. The Acclarent SE Inflation Device and Acclarent Balloon Inflation Device are FDA cleared devices (K150172 and K052198, respectively). The Relieva Extension Tubing may be used to connect the Balloon Catheters and the Inflation Device when additional tubing length is required. The
De Novo Summary (DEN150056)
{2}------------------------------------------------
Relieva Extension Tubing is a Class I exempt device under 21 CFR 874.4420. All devices are provided sterile for single-patient use.
The ETBC is a 6x16mm (inflated diameter x length) flexible Balloon Catheter (Figure 1) with an integrated shaft and a nylon balloon at the distal tip. The non-compliant balloon is designed to dilate the cartilaginous portion of the Eustachian tube (ET). The shaft consists of dual lumen tubing with an actuator component that is designed to assist in rotation and advancement of the ETBC. The balled catheter tip (aka, "blueberry tip") on the ETBC is designed to restrict advancement of the device into the bony portion of the ET, known as the isthmus. There is an endoscopic marker placed at the proximal taper of the balloon to aid in positioning under direct endoscopic visualization.
Image /page/2/Figure/2 description: The image shows a diagram of a medical device, specifically a catheter. The diagram labels several key components of the catheter, including the vent cap, luer, inflation port, actuator, proximal marker, catheter shaft, endoscopic marker, balloon, and catheter tip. The diagram provides a detailed view of the catheter's structure and features.
Figure 1: Illustration of the Eustachian Tube Balloon Catheter
The ETGC (Figure 2) is designed to facilitate ETBC access to the ET. The ETGC incorporates an atraumatic distal tip and distal angled tip profile that facilitates access to the ET. The ETGC contains a lubricious inner liner to allow smooth passage for the ETBC and includes a hypotube for rigidity.
Image /page/2/Figure/5 description: The image shows a medical instrument with several labeled parts. The instrument has a proximal handle/hub, a rigid shaft, a tip angle, and an atraumatic tip. The handle is blue, and the shaft is black.
Figure 2: Diagram of Eustachian Tube Guide Catheter
{3}------------------------------------------------
### SUMMARY OF NONCLINICAL/BENCH STUDIES
The ACCLARENT® AERA™ Eustachian Tube Balloon Dilation System was tested to ensure that it functions according to the design specifications. The device passed the following nonclinical tests as described below.
### BIOCOMPATIBILITY/MATERIALS
The materials used in the ACCLARENT® AERA™ Eustachian Tube Balloon Dilation System are well known and commonly used in the medical device industry, and are appropriate for ethylene oxide (EO) sterilization.
Both the Balloon Catheter and Guide Catheter are considered surface-contacting devices that contact breached or compromised surfaces for limited duration (less than 24 hours) according to ISO 10993-1: 2009/Corr 1:2010. The patient contacting materials including the adhesives were evaluated for cytotoxicity, sensitization, and acute systemic toxicity per ISO 10993-1, and FDA Bluebook Memorandum G95-1 entitled "Required Biocompatibility Training and Testing Profiles for Evaluation of Medical Devices," as listed in Table 1: Biocompatibility TestingTable 1. The results of these tests were found to be acceptable.
| Test Name | Test Method | Results |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Biocompatibility – Balloon Catheter and Guide Catheter | | |
| Cytotoxicity | ISO 10993-5:2009 Biological Evaluation of Medical<br>Devices: Tests for Cytotoxicity: in vitro; MEM<br>Elution method | Acceptable |
| Sensitization | ISO 10993-10:2010: Biological Evaluation of Medical<br>Devices: Tests for Irritation and Delayed Type<br>Hypersensitivity: Guinea Pig Maximization<br>Sensitization method | Acceptable |
| Irritation | ISO 10993-10:2010: Biological Evaluation of Medical<br>Devices: Tests for Irritation and Sensitization: Rabbit<br>Intracutaneous method | Acceptable |
| Acute Systemic<br>Toxicity | ISO 10993-11:2006: Biological Evaluation of Medical<br>Devices: Tests for Systemic Toxicity | Acceptable |
### Table 1: Biocompatibility Testing
### SHELF LIFE/STERILITY
Packaging, shelf life and sterilization testing is summarized in Table 2. The ETBC and the ETGC are packaged together on a b(4) CCI backer card in a b(4) CCI pouch. The pouches are then placed into cartons and sterilized using ethylene oxide (EO). The validation of this sterilization process complies with ISO 11135:2014-"Sterilization of Health Care Products-Ethylene Oxide: Requirements for Development. Validation, and Routine Control of a Sterilization Process for Medical Devices". The sterilization method achieves a sterilization assurance level (SAL) of 106
{4}------------------------------------------------
The proposed shelf life of 2 years has been validated using accelerated aging. The accelerated aging testing was conducted according to ASTM F1980-07; followed by testing on the packaging configuration and functional performance. Seal strength testing was conducted according to ASTM F88-09, and whole package integrity was conducted according to ASTM F2096-11. Real time aging for 2 years has been initiated and will be followed by the same testing on the packaging configuration and functional performance to confirm the proposed shelf life.
Transit testing was performed on the ACCLARENT® AERATM Eustachian Tube Balloon Dilation System after EO sterilization, climate conditioning per ISO 2233:2000 and transit simulations (compression, initial random vibrations, impact, and final random vibration) per ASTM D4169:2009.
| Test Name | Test Method | Acceptance Criteria | Results |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------|
| Sterilization - Balloon Catheter and Guide Catheter | | | |
| Sterilization<br>Validation | ISO 11135:2014:<br>Sterilization of health-<br>care products -<br>Ethylene oxide -<br>Requirements for the<br>development,<br>validation and routine<br>control of a<br>sterilization process for<br>medical devices | Sterility assurance level<br>(SAL) of 10-6 | Passed |
| Ethylene Oxide<br>Residuals | ISO 10993-7:2008:<br>Biological Evaluation<br>of Medical Devices:<br>Ethylene oxide<br>sterilization residuals | Ethylene oxide (EO)b(4)<br>CCLb(4) device and Ethylene<br>chlorohydrin (ECH)b(4)<br>CCLmg/device | Passed |
| Shelf Life and Packaging – Balloon Catheter and Guide Catheter | | | |
| Environmental<br>Conditioning | ISO 2233:2000:<br>Packaging - Complete,<br>filled transport<br>packages and unit<br>loads - Conditioning<br>for testing | N/A | Completed |
| Transportation<br>Simulation | ASTM D4169:2014:<br>Standard Practice for<br>Performance Testing<br>of Shipping Containers<br>and Systems | Testing performed for<br>compression, initial random<br>vibrations, impact, and final<br>random vibration | Completed |
Table 2: Sterilization, Shelf Life and Packaging Testing
{5}------------------------------------------------
| Test Name | Test Method | Acceptance Criteria | Results |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Accelerated Aging | ASTM F1980-07:<br>Standard Guide for<br>Accelerated Aging of<br>Sterile Barrier Systems<br>for Medical Devices | Testing to simulate shelf<br>life of 2 years | Completed |
| Package Integrity<br>(Visual) | ASTM<br>F1886/F1886M:<br>Standard Test Method<br>for Determining<br>Integrity of Seals for<br>Flexible Packaging by<br>Visual Inspection | • No damage to package<br>(pouch or box) that could<br>affect product performance,<br>marketability, or sterility<br>after compression, vibration,<br>and drop challenges.<br>• Package shall have no<br>damage, defects/smudges to<br>the pouch labeling.<br>• The device shall stay in the<br>intended place during<br>compression, vibration, and<br>drop challenges.<br>• Minimum seal width of ¼" | Passed |
| Gross Leak<br>Detection | ASTM F2096-11:<br>Standard Test Method<br>for Detecting Gross<br>Leaks in Packaging by<br>Internal Pressurization<br>(Bubble Test) | No gross leak | Passed |
| Seal Strength | ASTM F88-09:<br>Standard Test Method<br>for Seal Strength of<br>Flexible Barrier<br>Materials | b(4) CCI minimum for<br>individual value and b(4) CCI<br>b(4) CCI minimum for the<br>average value | Passed |
| Test Name | Test Method | Acceptance Criteria | Results |
| Performance Testing - Balloon Catheter | | | |
| Dimensional | Measure the dimensions<br>of the catheter, balloon,<br>location and dimensions<br>of the yellow endoscopic<br>marker, proximal shaft<br>marker, flexible length of<br>the balloon catheter,<br>balloon bond, tip and<br>venting lumen | Per device design<br>specifications | Passed |
| Deflation time | Verify the time to deflate<br>the balloon | Deflation time b(4) CCI | Passed |
| Balloon Burst<br>Pressure | Evaluate that the balloon<br>does not burst when<br>subjected to the maximum<br>rated pressure | Pressure > b(4) CCI | Passed |
| Tensile Strength | Evaluate the tensile<br>strength of all joints | Force at break b(4) CCI | Passed |
| Balloon Catheter<br>Tip Test | Ensure that balloon<br>catheter tip is bonded<br>securely by evaluating its<br>tensile strength | Force at break b(4) CCI | Passed |
| Insertion/Retraction<br>Test | Ensure that balloon<br>catheter can withstand<br>multiple<br>insertion/retraction cycles<br>through the Guide<br>Catheter without<br>structural damage or<br>balloon leakage | After 4<br>insertion/retractions | Passed |
| Fatigue Test | Ensure that the balloon<br>can withstand multiple<br>inflation/deflation cycles<br>at the maximum rated<br>pressure | 8 cycles @ b(4) CCI | Passed |
| Label | Visually verify that<br>correct label is applied to<br>the balloon and balloon<br>inflation port label | Visual confirmation | Passed |
| Performance Testing - Guide Catheter | | | |
| Dimensional | Measure the dimensions<br>of the catheter, distal tip,<br>and transition bond | Per device design<br>specifications | Passed |
| Test Name | Test Method | Acceptance Criteria | Results |
| Fatigue | Ensure that the catheter<br>can withstand multiple<br>insertion/retraction cycles<br>of the balloon catheter<br>through it | After 4<br>insertion/retraction<br>cycles | Passed |
| Materials | Ensure that the materials<br>used to construct distal<br>shaft are clear/ translucent<br>so that balloon catheter tip<br>and yellow endoscopic<br>marker are visible in the<br>guide distal shaft during<br>insertion | Visual confirmation | Passed |
| Torque | Test the torque resistance<br>to ensure no material or<br>joint failure | Torque $\geq$ b(4) CCI | Passed |
| Tensile Strength | Evaluate the tensile<br>strength of all joints | Force at break $\geq$ b(4) CCI | Passed |
| Label | Visually verify that the<br>correct label is applied on<br>the guide handle | Visual confirmation | Passed |
## PERFORMANCE TESTING – BENCH
Functional testing was performed on both the ETBC and the ETGC, as summarized in Table 3Error! Reference source not found.. The testing was performed on devices after EO sterilization, environmental conditioning, transit simulations, and accelerated aging, to simulate worst-case conditions. All testing was performed with a compatible inflation device. The devices passed all functional testing.
{6}------------------------------------------------
Table 3: Performance Testing
{7}------------------------------------------------
### PERFORMANCE TESTING - CADAVER
The ACCLARENT® AERA™ Eustachian Tube Balloon Dilation System was evaluated in cadavers to ensure that it functions in accordance with its design specifications and intended use. This simulated use testing was conducted by ENT physicians using cadaveric specimens.
Each evaluator used the same device to dilate a left and a right ET of the same specimen. Evaluators were asked to assess various attributes such as device placement without excessive advancement to the bony portion of the ET, ability to attach the compatible inflation device or extension tubing to the ETBC, ETBC movement through the ETGC, ability to position and operate the system with one hand, ability to avoid unacceptable trauma to mucosa or anatomical structures, and ability to maintain device functionality throughout treatment. The acceptance criteria for this study were a clinically acceptable rating of "2" or higher on a 1-4 point scale, with 4 defined as "excellent" performance and 1 defined as "clinically unacceptable" performance.
All ratings from each evaluator were above "2". Therefore all performance requirements met the acceptance criteria. The lowest average rating for each requirement was 3.3, while the highest average rating was 3.7 out of a possible rating of 4. The system performed as intended and no safety issues were observed.
{8}------------------------------------------------
## SUMMARY OF CLINICAL INFORMATION
Introduction: Acclarent designed and executed a randomized controlled study evaluating the safety and efficacy of the subject device with the goal of providing sufficient evidence to support a reasonable assurance of safety and efficacy.
Study Objectives: To evaluate efficacy and safety of ET balloon dilation with the Eustachian Tube Balloon Catheter (ETBC) in conjunction with medical management (investigational arm) compared to medical management alone to treat Eustachian tube dysfunction (ETD) in adult patients (control arm).
- . Primary efficacy endpoint evaluation consists of comparison of the proportion of subjects experiencing normalization of tympanometry at 6 weeks post-treatment in the investigational arm versus the proportion of subjects experiencing normalization of tympanometry at 6 weeks post-randomization in the control arm.
- . Secondary efficacy endpoint evaluation employs the Eustachian Tube Dysfunction Questionnaire (ETDO-7) to compare the proportion of subjects achieving at least a minimally important difference (MID) level improvement of 0.5 points at 6 weeks posttreatment in the investigational arm versus the proportion of subjects achieving at least a MID level improvement of 0.5 points at 6 weeks post-randomization in the control arm.
- . Safety evaluation includes occurrence of all anticipated or unanticipated adverse events in both treatment arms.
Study Methodology: The study was a prospective, multi-center, randomized controlled trial to demonstrate superiority of the ET balloon dilation with ETBC in conjunction with medical management (investigational arm) compared to medical management (MM) alone (control arm) to treat ETD. Study population includes subjects age 22 and older with ETD refractory to medical therapy. Each investigator was required to perform 3 successful ETBC procedures in non-randomized Lead-In subjects. MM consists of treatment with triamcinolone acetonide (TA) nasal steroid spray (such as Nasacort® AQ Nasal Spray (NAQ) or equivalent). All enrolled study subjects in both the lead-in and randomized phase are required to adhere to the recommended TA labeling dosage and frequency through the 6 week follow-up visit. All ETBC procedures were performed under general anesthesia in the operating room.
The randomized cohort is conducted in a 2:1 ratio with two subjects randomized to ETBC for every one MM subject. The study is a group sequential design with 3 planned looks (2 interim and 1 final) at 54:27, 108:54, and 162:81 evaluable subjects. The clinical evidence being presented in this summary is supported by the second-look cohort.
The primary endpoint was graded by both the investigator as well as an independent evaluator with appropriate expertise unaffiliated with study patient care. The independent tympanogram evaluator was blinded to subjects' arm assignment and had no knowledge of the outcome of the investigators' tympanogram results.
Each subject was enrolled individually and served as the unit of analysis, regardless of indication for unilateral or bilateral ETD. Subjects had five follow-up visits. Follow-up visits occurred at 2
{9}------------------------------------------------
weeks, 6 weeks, 12 weeks, 24 weeks, and 52 weeks following either randomization (control arm) or the ETBC procedure (investigational arm). MM subjects may cross over to receive an ETBC procedure after the 6 week follow-up visit and are subsequently followed for 12 weeks postprocedure.
Number of Subjects (planned and analyzed): The second interim analysis cohort consisted of 72 MM, 141 ETBC and 81 enrolled Lead-In subjects. Twenty-one (70) MM subjects crossed over to receive an ET dilation procedure and are included in the safety analysis of 299 subjects (80 lead-in + 149 randomized ETBC + 70 MM cross-overs).
Diagnosis and Main Criteria for Inclusion: The study includes adult males or females aged 22 years and older diagnosed with persistent ETD. Persistent ETD is defined by patient-reported symptoms of ETD (including pressure and/or pain and/or clogged and/or muffled sensation in the affected ear(s)) and documented evidence that the subject has experienced persistent ETD symptoms for 12 weeks or greater prior to enrollment. Patients must have failed appropriate medical management consisting of either a minimum of 4 weeks of continuous daily usage of any intranasal steroid spray (INS) or a minimum of one completed course of an oral steroid. The failed medical management may occur any time within 90 days prior to study enrollment. In addition, all patients are required to have a computed tomography (CT) scan taken to confirm absence of internal carotid artery (ICA) dehiscence into the Eustachian tube (ET) lumen bilaterally.
All patients must present with abnormal tympanometry (type B or type C) after failed MM and within 14 days prior to study enrollment. In addition, all included patients must have demonstrated symptomatic dysfunction as documented through a validated quality of life instruments (the ETDQ-7) with a score ≥ 2.1 after failed MM.
Females who are pregnant or lactating were excluded. In addition subjects were excluded if they had (1) anatomy that required an adjunctive surgical procedure on the same day as ETBC surgery to allow balloon catheter access to the ET, or (2) concomitant nasal, sinus or ear procedures planned on the same day as ETBC surgery or any time during study participation. This criterion did not apply to Lead-In subjects. In addition, patients were excluded if they have a history of major surgery of the head or neck within four (4) months prior to randomization, patulous ET in either ear, fluctuating sensorineural hearing loss, active chronic or acute otitis media (AOM), tympanic membrane perforation or presence of a tympanostomy tube in either ear, tympanosclerosis, acute upper respiratory infection, active temporomandibular joint disorder (TMJ), cleft palate or history of cleft palate repair, craniofacial syndrome, cystic fibrosis, ciliary dysmotility syndrome, systemic mucosal diseases or immunodeficiency disorders, intolerance of protocol-defined medication regimen, prior surgical intervention on ET, or absence of dilatory muscular contractions.
## EFFICACY RESULTS:
Technical success was high with 298/299 (99.7%) of all targeted ETs successfully accessed and dilated. The one (1) failure was due to anatomical constraint undetected prior to procedure, resulting from angiofibroma surgery 31 years prior which prevented appropriate balloon access
{10}------------------------------------------------
and positioning. Several events considered as device malfunctions were reported. Two (2) ETBC were reported to have leaked saline upon inflation; of these, one (1) device was disposed following the procedure and was not returned to the company for testing; one (1) device was returned for testing which indicated a functional balloon (i.e., the leak was not confirmed). One (1) ETBC was bent during approach into the ET. No adverse events were associated with the malfunctions.
Primary efficacy results demonstrated superiority of the investigational device plus MM compared to MM alone. Tympanogram normalization was observed in 51.8% (73/141) of investigational subjects compared to 13.9% (10/72) controls (p<0.0001). The Fisher's exact pvalue for the 2nd planned look was smaller than the pre-planned O'Brien-Fleming stopping pvalue at 0.00706. As the primary endpoint did meet the early stopping criteria, enrollment in the study has been ceased. Sensitivity analysis was performed to test the robustness of efficacy results and the impact of missing data in the intent-to-treat (ITT) cohort compared to the primary analysis cohort (PAC). Sensitivity analysis results suggest there is not a large source of uncertainty in the outcomes interpretation since no tipping point was identified.
The data demonstrated a degree of site variation with regards to the treatment effect at the primary endpoint at 6 weeks, which is summarized in Table 4.
| Site | Randomized ETBC<br>(N = 141 Subjects at Baseline) | Medical Management<br>(N = 72 Subjects at Baseline) | Total # of<br>Subjects |
|------|---------------------------------------------------|-----------------------------------------------------|------------------------|
| 1 | 78.9% (15/19) | 18.2% (2/11) | 30 |
| 2 | 33.3% (6/18) | 0.0% (0/9) | 27 |
| 3 | 29.4% (5/17) | 30.0% (3/10) | 27 |
| 4 | 57.1% (8/14) | 0.0% (0/9) | 23 |
| 5 | 91.7% (11/12) | 0.0% (0/6) | 18 |
| 6 | 60.0% (6/10) | 50.0% (3/6) | 16 |
| 7 | 0.0% (0/9) | 25.0% (1/4) | 13 |
| 8 | 77.8% (7/9) | 0.0% (0/4) | 13 |
| 9 | 33.3% (2/6) | 25.0% (1/4) | 10 |
| 10* | 62.5% (5/8) | 0.0% (0/1) | 9 |
| 11 | 50.0% (3/6) | 0.0% (0/2) | 8 |
| 12 | 40.0% (2/5) | 0.0% (0/2) | 7 |
| 13** | 50.0% (2/4) | 0.0% (0/2) | 6 |
| 14 | 25.0% (1/4) | 0.0% (0/2) | 6 |
### Table 4: Primary Endpoint Achievement at 6 weeks
* Data from 5 pooled sites
** Data from 2 pooled sites
The significant treatment by site interaction is believed to be consequence of two sites having a higher percentage of tympanogram normalization for MM subjects than for ETBC subjects (Investigator Sites 3 and 7). However, the results for these sites are inconsistent with the results from all other sites, which show a higher percentage of tympanogram normalization for ETBC
{11}------------------------------------------------
subjects. Possible explanations for the treatment by site interaction could include device and procedural learning curve of the investigators, subject and/or site conformance to the protocol, or statistical variance.
For the secondary efficacy endpoint, a greater proportion of subjects in the investigational arm achieved an MID level change for ETDQ-7 at 6 weeks compared to control. MID level change was observed for 90.6% (126/139) of investigational subjects compared to 45.1% (32/71) controls.
Subjects in the investigational arm (ETBC) also showed a greater improvement in ETDQ-7 score at 6 weeks compared to control (MM). Investigational subjects reported a -2.3 (1.4) mean (SD) improvement in ETDQ-7 score at 6 weeks compared to -0.5 (1.2) for control subjects. Additionally, a greater proportion of subjects in the investigational arm (ETBC) presented without symptomatic dysfunction at 6 weeks compared to control (medical management). Fiftysix percent point one percent (56.1%, 78/139) of investigational subjects reported an ETDO-7 mean item score of less than 2.1 at 6 weeks compared to 8.5% (6/71) for control subjects.
Durability of treatment effect is supported by sustained outcomes for over 170 subjects, where 62.2% (61/98) randomized ETBC and 58.1% (43/74) lead-in subjects demonstrated tympanogram normalization at 24 weeks post-procedure compared to 51.8% (73/141) and 40.0% (30/75) at 6 weeks, respectively. Durability was consistently demonstrated across efficacy outcome measures where improvement in ETDQ-7 scores, proportion of subjects achieving a MID level improvement, and proportion of subjects without symptomatic dysfunction are sustained at 24 weeks post-procedure. Altogether, the efficacy results are supportive of greater improved outcomes for ETBC-treated patients compared to control and a durable treatment effect.
# SAFETY RESULTS
In 444 dilated Eustachian tubes (299 subjects), 0 serious device- or procedure-related adverse events were reported. Sixteen (16) non-serious device or procedure-related adverse events occurred in 13 subjects out of the 299 subjects who underwent Eustachian tube dilation, including 4 events related to the anesthesia component of the procedure. Related events included those identified as "definitely related" and "possibly related" and "possibly related". A conservative approach was taken and events of "unknown relatedness" were also included.
All potentially device- or procedure-related events were not serious, mild or moderate in severity and resolved without sequelae with the most significant treatment interventions consisting of tympanostomy tube placements for 3 subjects or Tylenol for 1 subject.
## LABELING
The labeling includes the Instructions for Use, which includes the device indications for use, a description of the device, contraindications, warnings and precautions, clinical data on the device, and instructions for the safe and effective use of the device. The labeling satisfies the
{12}------------------------------------------------
requirements of 21 CFR 801.109 Prescription devices.
The Instructions for Use addresses the known hazards and risks of the device for the intended use and incorporates safety statements to mitigate these risks, and instructions for how to prevent misuse of the device. The instructions also include information regarding compatibility and selection of appropriate products that can be used with the device, information about the shelf life, and information about the technical parameters such as the maximum allowed inflation pressure, geometry and available balloon sizes. In addition, the instructions provides a summary of the data on safety and effectiveness of the device from the randomized controlled trial comparing balloon dilation plus medical management vs. ongoing medical management alone.
## RISKS TO HEALTH
Table 5 identifies the risks to health that may be associated with use of the Eustachian Tube Balloon Dilation System and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measure |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Introduction of false passages and rupture<br>or damage to carotid artery | Non-clinical performance testing |
| | Simulated use testing |
| | Training |
| | Labeling |
| Injury to mucosal tissue<br>• due to misuse of device on patulous<br>Eustachian tube or following skull<br>base surgery<br>• due to catheter mechanical failure<br>• due to balloon rupture<br>• due to mishandling of device with<br>respect to excessive force and/or<br>incorrect positioning | Non-clinical performance testing |
| | Simulated use testing |
| | Shelf life validation |
| | Training |
| | Labeling |
| Adverse tissue reaction | Biocompatibility evaluation |
| Infection | Sterilization validation |
| | Shelf life validation |
| | Labeling |
Table 5: Identified Risks and Mitigation Measures
## SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the Acclarent Aera™ Eustachian Tube Balloon Dilation System is subject to the following special controls:
- 1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated:
- a. Mechanical testing, including tensile and flexural testing of catheter joints and materials.
{13}------------------------------------------------
- b. Durability testing, including fatigue and burst pressure testing of the balloon materials and components.
- Inflation and deflation characterization testing, including time and pressure C. measurements, and leak testing of the balloon.
- d. Verification testing of safety features built into the device must be performed. including the characterization of catheter geometries and distal tip insertion limitation mechanisms.
- Simulated use testing in a clinically relevant model must demonstrate the reliability of the 2. device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube.
- 3. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 4. Performance data must demonstrate the sterility of the device.
- Performance data must support shelf life by demonstrating continued sterility of the ર. device, package integrity and device functionality over the identified shelf life.
- 6. Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device.
- 7. Labeling must include:
- a. Detailed instructions for use.
- b. A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes.
- c. A shelf life.
## BENEFIT/RISK DETERMINATION
Persistent Eustachian Tube Dysfunction (ETD) can significantly degrade quality of life. Current standard of care such as medication and tympanostomy tubes have not been known to be universally effective or durable. Eustachian tube (ET) dilation using the ACCLARENT AERA™ Eustachian Tube Balloon Dilation System is proposed to be a less invasive technique for treating ETD than treatments such as ablation of mucosa and cartilage. ET dilation may help to relieve symptoms of ETD and sustain benefit of treatment over longer period of time. Patients might perceive these advantages as beneficial and may consider this as safe alternative treatment to other techniques.
The risks of the device are based on nonclinical laboratory and cadaver studies, as well as data collected in a clinical study described above. Probable serious adverse events related to the use of the device or the dilation procedure includes the introduction of false passages and rupture of the internal carotid artery (ICA), which lies immediately adjacent to the ET. Rupture of the
{14}------------------------------------------------
internal carotid artery can lead to excessive bleeding, stroke, and death from fatal bleeding which is near impossible to control in this location. This life-threatening risk is of particular concern in a patient with dehiscence of the bony carotid canal adjacent to the ET. Since the ICA is located close to the ET near the bony isthmus, and because the balloon dilation procedure using the device is visualized only in the nasopharynx and not at the distal end of the ET, the device could damage the bony canal and could lead to catastrophic complications resulting from injury to ICA. In addition, regardless of improvements in visualization with endoscopic approaches, the risk of injury to the ICA and associated severe complications remain, as supported by complications noted in early experience with peritubular ET injections for treatment of patulous ET as well as lateral skull base surgery. Probable device related and/or procedure-related nonserious adverse events include mucosal tears, minor bleeding, conductive deafness, epistaxis, and chronic otitis media.
To mitigate the risk of ICA rupture, the device specifically incorporates an enlarged, soft, distal tip designed to limit dilation to the distal cartilaginous portion of the ET, and physicians received training on the proper usage of the device on lead-in patients for the clinical study, and proper selection of the patients. In addition, CT imaging was performed to ensure dilation was not performed on patients with carotid canal dehiscence within the bony portion of the ET.
In the clinical study, technical success was high with 298/299 (99.7%) of all targeted Eustachian tubes successfully accessed and dilated under general anesthesia. There were several events considered as device malfunctions, including two devices that leaked saline upon inflation, and one device that was bent during approach into the Eustachian Tube.
Among 299 subjects who underwent Eustachian tube dilation (444 dilated ET), there were 0 serious device- or procedure-related adverse events. However, there were 16 non-serious device or procedure-related adverse events that occurred in 13 subjects. There were 2 subjects that had 3 potential device-related adverse events including mucosal tear, worsened ETD, and conductive hearing loss. The most commonly occurring non-serious device- or procedure-related adverse events were epistaxis and ETD. All potentially device- or procedure-related events were mild or moderate in severity and resolved without sequelae.
The lack of serious device- or procedure-related adverse events and the resolution of the low rate of non-serious device- or procedure-related adverse events in the study strongly supports a reasonable assurance of safety for use of the device for ET dilation for adult patients with persistent ETD.
The probable benefits of the device are based on data collected in a clinical study as described above. Persistent Eustachian Tube Dysfunction (ETD) was defined by patient-reported symptoms of ETD (including pressure and/or pain and/or clogged and/or muffled sensation in the affected ear(s)) and documented evidence that the subject has experienced persistent ETD symptoms for 12 weeks or greater. Patients must have failed appropriate medical management consisting of either a minimum of 4 weeks of continuous daily usage of any intranasal steroid spray (INS) or a minimum of one completed course of an oral steroid.
{15}------------------------------------------------
The results of the clinical study show that a greater proportion of subjects in the investigational arm demonstrated tympanogram normalization (51.8%, 73/141) compared to the control arm (13.9%, 10/72). Based on the scores from a widely used validated subjective questionnaire for evaluating and diagnosing ETD, the Eustachian Tube Dysfunction Questionnaire (ETDQ-7), a greater proportion of the subjects in the investigational arm (90.6%. 126/139) compared to the control arm (45.1%, 32/71) also reported enhancement in the quality of life at 6 weeks as measured using the minimally important difference (MID) level improvement of 0.5 points at 6 weeks post-treatment. Subjects in the investigational arm also showed a greater improvement in ETDO-7 score at 6 weeks compared to control. Investigational subjects reported a -2.3 (1.4) mean (standard deviation) improvement in ETDQ-7 score at 6 weeks compared to -0.5 (1.2) for control subjects, and a greater proportion (56.1%, 78/139) of investigational subjects reported an ETDQ-7 mean item score of less than 2.1 at 6 weeks compared to 8.5% (6/71) for control subjects. The analysis of the overall results also showed there was no significant variation in health benefit for different demographic subpopulations; it appeared that the lead in subjects, subjects randomized to the treatment arm and those randomized to the control arm were comparable with respect to demographics and baseline characteristics.
However, the data demonstrated a degree of variation with regards to the treatment effect at the primary endpoint at 6 weeks; the possible explanations for the treatment by site interaction could include device and procedural learning curve of the investigators, subject and/or site conformance to the protocol, or statistical variance. Training on cadavers is therefore necessary to mitigate variations in treatment across multiple users and sites.
Durability of treatment effect is supported by sustained primary effectiveness outcomes for over 170 subjects, where a greater proportion of randomized treated (62.2%. 61/98) and lead-in subjects (58.1%, 43/74) demonstrated tympanogram normalization at 24 weeks post-procedure 51.8% (73/141) and 40.0% (30/75) compared to those at 6 weeks, respectively. Durability was also consistently demonstrated across secondary efficacy outcome measures where improvement in ETDQ-7 scores and proportion of subjects without symptomatic dysfunction are sustained at 24 weeks post-procedure.
Altogether, the efficacy results are supportive of greater improved outcomes for ETBC-treated patients compared to control, as well as a durable treatment effect.
### Patient Perspectives
Patient perspectives considered for the ACCLARENT AERATM Eustachian Tube Balloon Dilation System included:
The use of the improvement in the Eustachian Tube Dysfunction Ouestionnaire (ETDO-7) scores, as part of the secondary efficacy endpoint evaluation. The ETDO-7 is a validated symptom score used in assessing adult patients with Eustachian Tube Dysfunction. The results of the ETDQ-7 score are discussed above.
### Benefit/Risk Conclusion
{16}------------------------------------------------
In conclusion, given the available information above, the data support that for dilating the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction, the probable benefits outweigh the probable risks for the ACCLARENT AERATM Eustachian Tube Balloon Dilation System. The device provides substantial benefits and the risks can be mitigated by the use of general and the identified special controls.
### CONCLUSION
The de novo request for the Acclarent Aera™ Eustachian Tube Balloon Dilation System is granted and the device is classified under the following:
Product Code: PNZ Device Type: Eustachian Tube Balloon Dilation System Class: II Regulation: 21 CFR 874.4180
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.