PNZ · Eustachian Tube Balloon Dilation Device

Ear, Nose, Throat · 21 CFR 874.4180 · Class 2

Overview

Product CodePNZ
Device NameEustachian Tube Balloon Dilation Device
Regulation21 CFR 874.4180
Device ClassClass 2
Review PanelEar, Nose, Throat

Identification

A Eustachian tube balloon dilation system is a prescription device that includes a flexible catheter attached to an inflatable balloon. The system is intended for use in dilating the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated: (i) Mechanical testing, including tensile and flexural testing of catheter joints and materials. (ii) Durability testing, including fatigue and burst pressure testing of the balloon materials and components. (iii) Inflation and deflation characterization testing, including time and pressure measurements, and leak testing of the balloon. (iv) Verification testing of safety features built into the device must be performed, including the characterization of catheter geometries and distal tip insertion limitation mechanisms. (2) Simulated use testing in a clinically relevant model must demonstrate the reliability of the device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube. (3) The patient-contacting components of the device must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the device. (5) Performance data must support shelf life by demonstrating continued sterility of the device, package integrity, and device functionality over the identified shelf life. (6) Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device. (7) Labeling must include: (i) Detailed instructions for use. (ii) A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes. (iii) A shelf life.

In combination with the general controls of the FD&C Act, the Acclarent Aera™ Eustachian Tube Balloon Dilation System is subject to the following special controls: - 1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be evaluated: - a. Mechanical testing, including tensile and flexural testing of catheter joints and materials. - b. Durability testing, including fatigue and burst pressure testing of the balloon materials and components. - Inflation and deflation characterization testing, including time and pressure C. measurements, and leak testing of the balloon. - d. Verification testing of safety features built into the device must be performed. including the characterization of catheter geometries and distal tip insertion limitation mechanisms. - 2. Simulated use testing in a clinically relevant model must demonstrate the reliability of the device to remain mechanically functional throughout the anticipated conditions of use, and validate that the design features limit access to only the cartilaginous portion of the Eustachian tube. - 3. The patient-contacting components of the device must be demonstrated to be biocompatible. - 4. Performance data must demonstrate the sterility of the device. - 5. Performance data must support shelf life by demonstrating continued sterility of the device, package integrity and device functionality over the identified shelf life. - 6. Training must include simulated use on cadavers to ensure users can follow the instructions for use to allow safe use of the device. - 7. Labeling must include: - a. Detailed instructions for use. - b. A detailed summary of the device technical parameters, including maximum allowed inflation pressure, allowable catheter geometries, and available balloon sizes. - c. A shelf life.

Recent Cleared Devices (9 of 9)

RecordDevice NameApplicantDecision DateDecision
K253612Acclarent AERA Eustachian Tube Balloon Dilation SystemAcclarent, Inc.Feb 19, 2026SESE
K230742ACCLARENT AERA Eustachian Tube Dilation SystemAcclarent, Inc.Dec 13, 2023SESE
K223542TubaVent Balloon Dilatation SystemSpiggle & Theis Medizintechnik GmbHAug 3, 2023SESE
K230065VenSure™ Balloon Dilation System, VenSure™ Light Balloon Dilation System, VenSure™ Nav Balloon Dilation System, VenSure™ ET Balloon Dilation SystemFiagon GmbHMay 26, 2023SESE
K220027Audion ET dilation systemEntellus Medical, Inc.Apr 12, 2022SESE
K210841NuVent Eustachian Tube Dilation BalloonMedtronic Xomed, Inc.Aug 16, 2021SESE
K171761ACCLARENT AERA Eustachian Tube Balloon Dilation SystemAcclarent, Inc.Jan 16, 2018SESE
K163509XprESS ENT Dilation SystemEntellus Medical, Inc.Apr 5, 2017SESE
DEN150056Acclarent Aera Eustachian Tube Balloon Dilation SystemAcclarent, Inc.Sep 16, 2016DENG

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...