The Relieva SpinPlus Nav™ Balloon Sinuplasty System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate the sinus ostia and spaces associated with the paranasal sinus cavities for diagnostic and therapeutic procedures; and irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures. For children aged 17 and under, the Relieva SpinPlus Nav™ Balloon Sinuplasty System is intended to provide a means to access the sinus space, within and across nasal and sinus structures; dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures; and irrigate from within the maxillary sinus for therapeutic procedures and to facilitate diagnostic procedures. The Relieva SpinPlus Nav™ Balloon Sinuplasty System may be utilized in conjunction with the ACCLARENT® ENT Navigation System, to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy
Device Story
The Relieva SpinPlus Nav Balloon Sinuplasty System is a manual surgical instrument used by ENT surgeons to access, dilate, and irrigate paranasal sinus ostia. The device integrates a sinus guide catheter, flexible balloon catheter, irrigation port, and an electromagnetically-trackable guidewire into a single handle assembly. The handle features a wire slider for advancing/spinning the guidewire, a balloon slider for positioning the balloon, and a suction port for clearing fluids. When connected to the Acclarent ENT Navigation System, the device provides EM-based real-time tracking of the distal tip relative to pre-loaded CT scans. The surgeon uses the device to navigate to the target ostia, dilate them via saline-inflated balloon, and irrigate the sinus. This system facilitates minimally invasive sinus procedures, potentially reducing surgical trauma and improving access to difficult anatomy compared to traditional methods.
Clinical Evidence
No clinical data was necessary. Substantial equivalence was demonstrated through bench testing, including dimensional specifications, balloon burst pressure, joint separation force, deflation time, cycle fatigue, tensile strength, and irrigation performance. Navigation accuracy was verified using the Acclarent ENT Navigation System. Biocompatibility and sterilization (EtO) were validated per ISO standards.
Technological Characteristics
Single-use, manual surgical instrument. Combines sinus guide catheter, balloon catheter, irrigation, and EM-trackable guidewire. Materials: biocompatible polymers. Sterilization: EtO (AAMI/ANSI/ISO 11135:2014). Connectivity: wired interface to Acclarent ENT Navigation System for EM tracking. Dimensions: 0.035-inch guidewire, 16mm balloon length, 12 ATM max inflation pressure.
Indications for Use
Indicated for patients, including children aged 17 and under, requiring access to nasal and sinus spaces, dilation of sinus ostia/spaces (paranasal for adults, maxillary for children), and irrigation of target sinuses for diagnostic and therapeutic procedures. May be used with ENT navigation systems for anatomical access and placement confirmation.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
Predicate Devices
RELIEVA SPINPLUS® Balloon Sinuplasty System (K143541)
Reference Devices
ACCLARENT® NAVWIRE™ Sinus Navigation System (K161697)
Medtronic NuVent™ EM Sinus Dilation System (K152121)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 13, 2017
Acclarent. Inc. James Patrick Garvey II Associate Director. Regulatory Affairs 33 Technology Drive Irvine, California 92618
Re: K171687
Trade/Device Name: Relieva SpinPlus® Nav Balloon Sinuplasty System Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, and Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: June 6, 2017 Received: June 7, 2017
Dear James Patrick Garvey:
This letter corrects our substantially equivalent letter of September 5, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Eric A. Mann -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171687
Device Name
RELIEVA SPINPLUS NAV Balloon Sinuplasty System
### Indications for Use (Describe)
The Relieva SpinPlus Nav™ Balloon Sinuplasty System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate the sinus ostia and spaces associated with the paranasal sinus cavities for diagnostic and therapeutic procedures; and irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.
For children aged 17 and under, the Relieva SpinPlus Nav™ Balloon Sinuplasty System is intended to provide a means to access the sinus space, within and across nasal and sinus structures; dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures; and irrigate from within the maxillary sinus for therapeutic procedures and to facilitate diagnostic procedures.
The Relieva SpinPlus Nav™ Balloon Sinuplasty System may be utilized in conjunction with the ACCLARENT® ENT Navigation System, to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> |
|------------------------------------------------|-----------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Acclar" is in black, while the "ent" portion is in purple. There is a purple curved line above the "ent" portion of the word.
| 510(K) SUMMARY | |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K171687 |
| [807.92(a)(1)] Submitter Information | |
| Sponsor/Submitter: | Acclarent, Inc.<br>33 Technology Drive<br>Irvine, CA 92618 |
| Contact Person: | Patrick Garvey<br>Associate Director, Regulatory Affairs<br>Email: pgarvey@its.jnj.com<br>Tel: 949-789-8505 Fax: 650-687-5899 |
| Date Summary Prepared: | August 28, 2017 |
| [807.92(a)(2)] Name of Device | |
| Device Trade Name: | RELIEVA SPINPLUS® NAV Balloon Sinuplasty System |
| Common Name: | Sinus Balloon Catheter |
| Device Classification: | Class I |
| Regulation Number: | 21 CFR 874.4420 |
| Classification Name: | Ear, Nose, and Throat Manual Surgical Instrument (21 CFR 874.4420) |
| Product Code: | LRC |
| [807.92(a)(3)] Legally Marketed Devices | |
| Predicate Devices: | RELIEVA SPINPLUS® Balloon Sinuplasty System, (K143541) |
| Reference Devices | ACCLARENT® NAVWIRE™ Sinus Navigation System, (K161697)<br>Medtronic NuVent™ EM Sinus Dilation System, (K152121) |
| [807.92(a)(4)] Device Description | |
| Device Description: | The RELIEVA SPINPLUS® NAV Balloon Sinuplasty System is packaged<br>with a handle with integrated flexible balloon catheter and Sinus Navigation<br>Guidewire, and three RELIEVA® Spin Sinus Guide Catheter Tips. The system<br>features a handle with an integrated flexible balloon catheter. Features of the<br>handle include a sinus guide tip release button, a suction system, a balloon<br>guard, and several gripping features to grip the device. The guide tip release<br>button must be depressed to separate the sinus guide catheter tip from the<br>handle system. The suction system consists of a suction line and a suction<br>port. Suction may be used to clear the field of fluids and/or blood. The suction<br>line is attached to the proximal end of the handle system and may be removed |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Acclar" is in black, while the "ent" is in purple. There is a purple swoosh above the "ent" part of the word.
RELIEVA SPINPLUS® NAV Balloon Sinuplasty System
| | if desired. The suction port may be covered by the user's finger to increase the suction flow rate. A clear balloon guard is connected to the distal end of the handle system and protects the sinus balloon during sinus guide catheter tip exchanges. The Handle features a wire slider, wire spinner, balloon slider, sinus balloon, handle markers, and proximal connections. The wire slider allows the user to advance, retract and spin the sinus navigation guidewire with a single hand while simultaneously supporting the handle. The balloon slider allows the user to advance and retract the sinus balloon catheter to the target sinus ostia and outflow tract, to enable dilation. The integrated connector enables the device to be interfaced to the ACCLARENT ENT™ Navigation System to facilitate EM navigation The RELIEVA SPINPLUS® NAV Balloon Sinuplasty System may be utilized in conjunction with the ACCLARENT ENT™ Navigation System to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy via electromagnetic navigation. |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| [807.92(a)(5)] Intended Use | |
| Indications for Use: | The RELIEVA SPINPLUS® NAV Balloon Sinuplasty System, is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate the sinus ostia and spaces associated with the paranasal sinus cavities for diagnostic and therapeutic procedures; and irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures. |
| | For children aged 17 and under, the RELIEVA SPINPLUS® NAV Balloon Sinuplasty System is intended to: provide a means to access the sinus space, within and across nasal and sinus structures; dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures; and irrigate from within the maxillary sinus for therapeutic procedures and to facilitate diagnostic procedures.” |
| | The RELIEVA SPINPLUS® NAV Balloon Sinuplasty System may be utilized in conjunction with the Acclarent ENT Navigation System, to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy. |
| Difference in<br>Indications from<br>Predicate Device | The difference in indications for use between the subject and predicate devices is supported is presented in Table 1 of this summary. |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Acclarent. The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the "ent" portion in purple. A purple curved line extends over the "ent" portion of the word, adding a visual element to the logo.
RELIEVA SPINPLUS® NAV Balloon Sinuplasty System
## [807.92(a)(6)] Technical Characteristics
| Technological<br>Characteristics: | The RELIEVA SPINPLUS® NAV Balloon Sinuplasty System combines a<br>Sinus Guide Catheter and Handle Assembly (integrated with Sinus Balloon<br>Catheter, Sinus Irrigation and Electromagnetically-trackable Guidewire) into a<br>single device. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | See Table 1 for a comparison of the technological characteristics between the<br>RELIEVA SPINPLUS® NAV Balloon Sinuplasty System and the predicate<br>devices. |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Acclarent. The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the second part in purple. A purple curved line extends over the top of the word, ending in a point to the right.
RELIEVA SPINPLUS® NAV Balloon Sinuplasty System
Table 1: Comparison of Technological Characteristics between the RELIEVA SPINPLUS® NA V Balloon Sinuplasty System and predicate and reference devices.
| Attribute | Primary Predicate Device<br>(RELIEVA SPINPLUS®<br>Balloon Sinuplasty System) | Reference Device<br>(ACCLARENT NAVWIRE™<br>Sinus Navigation Guidewire) | Reference Device<br>(NuVent EM Sinus Dilation<br>System) | Subject Device<br>(RELIEVA SPINPLUS® NAV<br>Balloon Sinuplasty System) |
|------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------|------------------------------------------------------------------------|
| 510(k) number | K143541 | K161697 | K152121 | TBD |
| Manufacturer | Acclarent | Acclarent | Medtronic | Acclarent |
| Trade Name | RELIEVA SPINPLUS® Sinus<br>Dilation System | ACCLARENT NAVWIRE™<br>Sinus Navigation Guidewire | NuVent™ EM Sinus Dilation<br>System | RELIEVA SPINPLUS® NAV<br>Balloon Sinuplasty System |
| Common Name | Sinus Dilation System | Sinus Balloon Catheter | Sinus Balloon Catheter | Sinus Dilation System |
| Class | I | II | I | I |
| Product Code | LRC | PGW | LRC | LRC |
| Classification Section | 21 CFR 874.4420 | 21 CFR 882.4560 | 21 CFR 874.4420 | 21 CFR 874.4420 |
{8}------------------------------------------------
Balloon Sinuplasty System
### Traditional 510(k) Premarket Notification RELIEVA SPINPLUS® NAV
#### The RELIEVA SPINPLUS™ The ACCLARENT® NAVWIRE™ The EM Sinus Dilation System is The RELIEVA SPINPLUS™ Nav Indications for Balloon Sinuplasty System is Sinus Navigation Guidewire is intended for use in conjunction Balloon Sinuplasty System is Use intended to: provide a means to intended for use as a navigable with the Medtronic Computer intended to provide a means to access the sinus space and guidewire to provide access to Assisted Surgery System during access the sinus space, within and sinus procedures when surgical across nasal and sinus structures; illuminate within and nasal and sinus spaces, and transilluminate across nasal and confirmation of placement in the navigation or image-guided surgery dilate the sinus ostia and spaces sinus structures; dilate the sinus accessed anatomy. The Acclarent may be necessary. When used associated with the paranasal sinus ostia and spaces associated with the NavWire is designed for use during concomitantly, these systems may cavities for diagnostic and paranasal sinus cavities for procedures where reference to a be used to locate and move tissue, therapeutic procedures; and irrigate diagnostic and therapeutic rigid anatomical structure in the bone or cartilaginous tissue from within a target sinus for procedures; and irrigate from field of ENT surgery, such as the surrounding the drainage pathways therapeutic procedures and to within a target sinus for therapeutic paranasal sinuses, can be identified of frontal, maxillary, and sphenoid facilitate diagnostic procedures. procedures and to facilitate relative to a CT-based model of the sinuses to facilitate dilation of the diagnostic procedures. anatomy. sinus ostia: or locate and move For children aged 17 and under, the RELIEVA SPINPLUS™ Nav tissue, bone or cartilaginous tissue For children aged 17 and under, the surrounding the drainage pathways Balloon Sinuplasty System is RELIEVA SPINPLUS™ Balloon of frontal, maxillary, and sphenoid intended to provide a means to Sinuplasty System is intended to: sinuses that is scarred, granulated access the sinus space, within and or previously surgically altered to provide a means to access the sinus across nasal and sinus structures: space and illuminate within and facilitate dilation of the sinus ostia. dilate sinus ostia and spaces associated with the maxillary sinus transilluminate across nasal and The Medtronic computer-assisted sinus structures; dilate sinus ostia surgery system and its associated for diagnostic and therapeutic and spaces associated with the applications are intended as an aid procedures; and irrigate from maxillary sinus for diagnostic and for precisely locating anatomical within the maxillary sinus for therapeutic procedures; and irrigate structures in either open or therapeutic procedures and to from within the maxillary sinus for percutaneous procedures. Their use facilitate diagnostic procedures. therapeutic procedures and to is indicated for any medical facilitate diagnostic procedures. condition in which the use of The RELIEVA SPINPLUS™ Nav stereotactic surgery may be Balloon Sinuplasty System may be appropriate, and where reference to utilized in conjunction with the a rigid anatomical structure, such as ACCLARENT® ENT Navigation System, to provide access to nasal the skull, can be identified relative to a CT- or MR-based model, or and sinus spaces, and to confirm digitized landmarks of the placement in the accessed anatomy. The system and its anatomy." associated applications should be used only as an adjunct for surgical guidance. They do not replace the surgeon's knowledge, expertise, or judgment.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Acclar" is in black, while the "ent" is in purple. There is a purple curved line above the "ent" portion of the word.
RELIEVA SPINPLUS® NAV
Balloon Sinuplasty System
| Attribute | Primary Predicate Device<br>(RELIEVA SPINPLUS®<br>Sinus Dilation System) | Reference Device<br>(ACCLARENT NAVWIRE™<br>Sinus Navigation Guidewire) | Reference Device<br>(NuVent EM Sinus Dilation<br>System) | Subject Device<br>(RELIEVA SPINPLUS® NAV<br>Balloon Sinuplasty System) |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization | EtO | EtO | EtO | EtO |
| Packaging | Thermoformed tray in pouch | Thermoformed tray in pouch | Unknown | Thermoformed tray in pouch |
| Single Use | Yes | Yes | Yes | Yes |
| Patient Contact | Direct | Direct | Direct | Direct |
| Labeled as Non-<br>Pyrogenic? | No | No | No | No |
| Technological<br>Characteristics | Combines a Sinus Guide<br>Catheter and Handle Assembly<br>(integrated with Sinus Balloon<br>Catheter, Sinus Irrigation and<br>Sinus Illumination System) into<br>a single device. The device is<br>connected to any standard light<br>source via accessory light cable<br>and adapter. | The ACCLARENT®<br>NAVWIRE™ Sinus Navigation<br>Guidewire utilizes<br>electromagnetic image-guided<br>sinus surgery and is used in<br>conjunction with a compatible<br>surgical navigation system<br>which consists of computer-<br>aided software, CT-imaging,<br>patient tracker, registration<br>probe, and various instruments<br>used in sinus surgery.<br>ACCLARENT® NAVWIRETM<br>Sinus Navigation Guidewire<br>provides real-time tracking at<br>the distal tip of the guidewire in<br>the nasal anatomy. | Combines a Sinus Guide<br>Catheter and Handle Assembly<br>(integrated with Sinus Balloon<br>Catheter and Sinus Navigation<br>System) into a single device.<br>The device is connected to the<br>Medtronic Fusion Navigation<br>System to provide location<br>information via EM navigation<br>relative to a pre-loaded scan. | Combines a Sinus Guide<br>Catheter and Handle Assembly<br>(integrated with Sinus Balloon<br>Catheter, Sinus Irrigation and<br>Sinus Navigation System) into a<br>single device. The device is<br>connected to the Acclarent ENT<br>Navigation System to provide<br>location information via EM<br>navigation relative to a pre-<br>loaded scan. |
| Balloon Length | 16 mm | N/A | 17 mm | 16 mm |
| Guidewire<br>Diameter | 0.035 inches | 0.035 inches | N/A | 0.035 inches |
| Maximum<br>Inflation Pressure | 12 ATM | N/A | Unknown | 12 ATM |
| Flexible Balloon<br>Catheter | Yes | N/A | N/A | Yes |
| Attribute | Primary Predicate Device<br>(RELIEVA SPINPLUS®<br>Sinus Dilation System) | Reference Device<br>(ACCLARENT NAVWIRE™<br>Sinus Navigation Guidewire) | Reference Device<br>(NuVent EM Sinus Dilation<br>System) | Subject Device<br>(RELIEVA SPINPLUS® NAV<br>Balloon Sinuplasty System) |
| Balloon<br>Radiopaque<br>Marker Bands | No | N/A | EM-Navigation | No |
| Suction<br>Incorporated | Yes | No | No | Yes |
| Irrigation<br>Incorporated | Yes | No | No | Yes |
| Balloon Slide<br>Mechanism | Yes | No | No | Yes |
| Indicated for<br>Children | Yes, maxillary | Yes, maxillary1 | No | Yes, maxillary |
| Compatible with<br>ACCLARENT®<br>ENT Navigation<br>System | No | Yes | No<br>(Compatible with<br>Medtronic Fusion® ENT<br>Navigation System) | Yes |
| Principles of<br>Operation | Manually operated device.<br>Balloon inflated with sterile<br>saline or water to mechanically<br>dilate sinus ostia. | Manually operated device.<br>Balloon inflated with sterile<br>saline or water to mechanically<br>dilate sinus ostia. | Manually operated device.<br>Balloon inflated with sterile<br>saline or water to mechanically<br>dilate sinus ostia. | Manually operated device.<br>Balloon inflated with sterile<br>saline or water to mechanically<br>dilate sinus ostia. |
| Flexible | Yes – flexible sinus balloon<br>catheter and guidewire. | Yes – Flexible guidewire | No- fixed, rigid instrument not<br>designed to bend. | Yes – flexible balloon sinus<br>catheter and guidewire. |
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Acclarent" in black font. The last part of the word, "rent" is in purple font. There is a curved line above the "rent" part of the word.
## RELIEVA SPINPLUS® NAV Balloon Sinuplasty System
<sup>1</sup> Based upon use with Relieva UlttirraNav Sinus Balloon Catheter, cleared on K161698
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Acclarent. The text "Acclarent" is in black font. There is a purple arc above the text.
## [807.92(b) (1)] Determination of Substantial Equivalence
| Non-Clinical Performance<br>Data: | The RELIEVA SPINPLUS Nav™ Balloon Sinuplasty System met all<br>performance acceptance criteria including dimensional specifications;<br>balloon burst pressure, joint separation force, deflation time, and<br>balloon cycle fatigue. Additional testing to demonstrated that wire<br>stiffness, tensile strength and irrigation testing were also performed to<br>demonstrate that the device met performance specifications.<br><br>Packaging shelf life was established through accelerated aging via<br>ASTM F1980-07, ASTM F88/F88M-09, and ASTM F2096-04<br>requirements and confirmed to meet a shelf life of three months.<br><br>Biocompatibility testing successfully completed to determine that the<br>RELIEVA SPINPLUS® NAV Balloon Sinuplasty System is<br>biocompatible.<br><br>The sterilization process has been validated per AAMI/ANSI/ISO<br>11135:2014 and demonstrated a sterility assurance level of 10-6. The<br>method used for sterilization validation was the overkill (half-cycle<br>approach) in a fixed chamber. Ethylene oxide residuals were tested and<br>met ISO 10993-7:2008 requirements. The subject device is not tested<br>nor labeled as "non-pyrogenic”.<br><br>Simulated use testing was performed with ENT surgeons performing<br>balloon dilation of the paranasal sinuses utilizing the RELIEVA<br>SPINPLUS® NAV Balloon Sinuplasty System and the ACCLARENT<br>ENT™ Navigation System. The testing demonstrated that the<br>RELIEVA SPINPLUS® NAV Balloon Sinuplasty System and the<br>ACCLARENT ENT™ Navigation System could effectively access the<br>paranasal sinuses. Clinical data was not necessary for the RELIEVA<br>SPINPLUS® NAV Balloon Sinuplasty System.<br><br>Testing was performed to verify the navigation accuracy of the device<br>when used with the Acclarent ENT Navigation System (K161701).<br>Testing included sensitivity and connection verification.<br>The performance data demonstrated that the device performs as<br>intended. |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| [807.92(b) (2)] Determination of Substantial Equivalence | |
| Clinical Performance Data | Clinical data was not necessary for the RELIEVA SPINPLUS Nav™<br>Balloon Sinuplasty System. The performance data demonstrated that<br>the device performs as intended.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.