K152121 · Medtronic Xomed, Inc. · LRC · Dec 1, 2015 · Ear, Nose, Throat
Device Facts
Record ID
K152121
Device Name
NuVent EM Sinus Dilation System
Applicant
Medtronic Xomed, Inc.
Product Code
LRC · Ear, Nose, Throat
Decision Date
Dec 1, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The EM Sinus Dilation System is intended for use in conjunction with the Medtronic Computer-Assisted Surgery System during sinus procedures when surgical navigation or image-guided surgery may be necessary. When used concomitantly, these systems may be used to - locate and move tissue, bone or cartilaginous tissue surrounding the drainage pathways of frontal, maxillary, and sphenoid sinuses to facilitate dilation of the sinus ostia; or - locate and move tissue, bone or cartilaginous tissue surrounding the drainage pathways of frontal, maxillary, and sphenoid sinuses that is scarred, granulated or previously surgically altered to facilitate dilation of the sinus ostia. The Medtronic computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT- or MR-based model, or digitized landmarks of the anatomy. The system and its associated applications should be used only as an adjunct for surgical guidance. They do not replace the surgeon's knowledge, expertise, or judgment.
Device Story
The NuVent EM Sinus Dilation System consists of sterile, single-use sinus seekers (frontal, maxillary, sphenoid) and an inflator. Each seeker contains an electromagnetic (EM) tracker. Used with the Medtronic Computer-Assisted Surgery System, the device utilizes a low-energy magnetic field generated by an emitter to track the seeker's tip. The Fusion software displays the tip's location within patient CT/MR image planes or anatomical renderings. Surgeons use the seeker to navigate to the sinus ostia; once positioned, the integrated balloon is inflated with saline via the inflator to dilate the sinus outflow tract. The system provides real-time surgical guidance, aiding in precise anatomical localization during sinus procedures. It is intended for use by surgeons in clinical settings. The device benefits patients by enabling precise, image-guided dilation of sinus pathways, including in complex cases involving scarred or previously altered anatomy.
Clinical Evidence
Prospective, non-randomized, non-blinded, single-arm, multicenter clinical study. Enrolled subjects with scarred, granulated, or previously surgically altered tissue. Primary endpoint: performance of the device in the intended population. Results: device performed as intended; no adverse events attributed to the device.
Technological Characteristics
Single-use sinus seekers with integrated EM tracking; balloon dilation mechanism; inflator (plunger, barrel, extension tube). Operates with Medtronic Computer-Assisted Surgery System (Fusion software). Electromagnetic sensing principle. Sterile, single-use.
Indications for Use
Indicated for patients undergoing sinus procedures requiring surgical navigation or image-guided surgery for the frontal, maxillary, and sphenoid sinuses, including cases with scarred, granulated, or previously surgically altered tissue, to facilitate dilation of the sinus ostia.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized image of three human profiles facing right, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2015
Medtronic Xomed, Inc. Ms. Gabriela Anchondo Regulatory Affairs 6743 Southpoint Drive North Jacksonville, FL 32216
Re: K152121
Trade/Device Name: Nuvent EM Sinus Dilation System Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: October 30, 2015 Received: November 2, 2015
Dear Ms. Anchondo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K152121
# Device Name:
EM Sinus Dilation System
# Indications for Use:
The EM Sinus Dilation System is intended for use in conjunction with the Medtronic Computer-Assisted Surgery System during sinus procedures when surgical navigation or image-quided surgery may be necessary. When used concomitantly, these systems may be used to
- locate and move tissue, bone or cartilaginous tissue surrounding the drainage pathways ● of frontal, maxillary, and sphenoid sinuses to facilitate dilation of the sinus ostia; or
- locate and move tissue, bone or cartilaginous tissue surrounding the drainage pathways 0 of frontal, maxillary, and sphenoid sinuses that is scarred, granulated or previously surgically altered to facilitate dilation of the sinus ostia.
The Medtronic computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT- or MR-based model, or digitized landmarks of the anatomy.
The system and its associated applications should be used only as an adjunct for surgical quidance. They do not replace the surgeon's knowledge, expertise, or judgment.
× Prescription Use _ (Per 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (Per 21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 5. 510(k) Summary
# [807.92(a)(1)] 510(k) Owner
| Name | Medtronic Xomed, Inc. | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address | 6743 Southpoint Drive North<br>Jacksonville, FL 32216 USA | |
| Phone and Fax Numbers | Phone: (904) 279-7550<br>Fax: (904) 296-2386 | |
| Name of Contact Person | Gabriela Anchondo<br>Regulatory Affairs Manager | |
| Date Summary Prepared | July 24, 2015 | |
| [807.92(a)(2)] Name of Device | | |
| Trade or Proprietary Name | NuVent™ EM Sinus Dilation System | |
| Common or Usual Name | EM Sinus Dilation System | |
| Classification Name | Ear, nose and throat manual surgical instrument<br>(21 CFR 874.4420, Product Code LRC) | |
| [807.92(a)(3)] Legally Marketed Device | | |
| Predicate Device | K132297 EM Sinus Dilation System | |
| [807.92(a)(4)] Description of Device | | |
| Device Description | The NuVent™ EM Sinus Dilation System<br>comprises sterile, single-use instruments that<br>combine electromagnetic (EM) "plug and play"<br>tracking capability with the pathway expansion<br>effects of balloon dilation technology and an<br>inflator. Each of the three types of sinus seekers<br>(frontal, maxillary and sphenoid) has a unique<br>shape and angle that allows for entry into the sinus<br>outflow tract. The inflator consists of a plunger,<br>barrel and extension tube. | |
| | Each sinus seeker is intended for use in<br>conjunction with the Fusion software on a<br>Medtronic computer-assisted surgery system.<br>Inside each sinus seeker is an EM tracker.<br>The emitter on the EM Computer-Assisted Surgery<br>System generates a low-energy magnetic field to<br>locate the tracker mounted inside of the sinus | |
{4}------------------------------------------------
seeker. The software displays the location of the sinus seeker's tip within multiple patient image planes and other anatomic renderings. After confirmation of placement, the sinus seekers balloon can be inflated with saline solution by using the inflator to expand the outflow of the targeted sinus.
#### [807.92(a)(5)] Intended Use
#### Indications for Use Statement
The EM Sinus Dilation System is intended for use in conjunction with the Medtronic Computer-Assisted Surgery System during sinus procedures when surgical navigation or image-guided surgery may be necessary. When used concomitantly, these systems may be used to
- locate and move tissue, bone or . cartilaginous tissue surrounding the drainage pathways of frontal, maxillary, and sphenoid sinuses to facilitate dilation of the sinus ostia: or
- locate and move tissue, bone or . cartilaginous tissue surrounding the drainage pathways of frontal, maxillary, and sphenoid sinuses that is scarred, granulated or previously surgically altered to facilitate dilation of the sinus ostia.
The Medtronic computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CTor MR-based model, or digitized landmarks of the anatomy.
The system and its associated applications should be used only as an adjunct for surgical quidance. They do not replace the surgeon's knowledge, expertise, or judgment.
The difference in indications for use is supported by clinical data and does not affect the safety and effectiveness of the device.
Difference in Indications from Predicate Device
{5}------------------------------------------------
# [807.92(a)(6)] Technological Characteristics
## Comparison
The subject device has the same technological characteristics as the predicate device. The only difference is in the indications for use. Refer to the table that follows for a comparison.
| Attributes | EM Sinus Dilation System<br>(Subject Device) | EM Sinus Dilation System<br>(Predicate Device) | Comparison |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | K152121 | K132297 | -- |
| 510(k)<br>Submitter | Medtronic Xomed | Medtronic Xomed | SAME |
| Class & Reg | Class I -21 CFR 874.4420 | Class I - 21 CFR 874.4420 | SAME |
| Product Code<br>& Class Name | LRC - ENT Manual Surgical<br>Instrument | LRC - ENT Manual Surgical<br>Instrument | SAME |
| Intended Use | The EM Sinus Dilation System is<br>intended for use in conjunction<br>with the Medtronic Computer-<br>Assisted Surgery System during<br>sinus procedures when surgical<br>navigation or image-guided<br>surgery may be necessary.<br>When used concomitantly, these<br>systems may be used to:<br>locate and move tissue,<br>bone or cartilaginous tissue<br>surrounding the drainage<br>pathways of frontal,<br>maxillary, and sphenoid<br>sinuses to facilitate<br>dilation of the sinus ostia; or<br>locate and move tissue,<br>bone or cartilaginous tissue<br>surrounding the drainage<br>pathways of frontal,<br>maxillary, and sphenoid<br>sinuses that is scarred,<br>granulated or previously<br>surgically altered to<br>facilitate dilation of the<br>sinus ostia. The Medtronic computer-<br>assisted surgery system and its<br>associated applications are<br>intended as an aid for precisely<br>locating anatomical structures in<br>either open or percutaneous<br>procedures. Their use is<br>indicated for any medical<br>condition in which the use of<br>stereotactic surgery may be<br>appropriate, and where | The EM Sinus Dilation System<br>is intended for use in sinus<br>procedures when surgical<br>navigation or image-guided<br>surgery may be necessary to<br>locate and move tissue, bone or<br>cartilaginous tissue surrounding<br>the drainage pathways of the<br>frontal, maxillary, and sphenoid<br>sinuses.<br><br>The EM Sinus Dilation system is<br>used in conjunction with the<br>Medtronic computer-assisted<br>surgery system.<br><br>The Medtronic computer-<br>assisted surgery system and its<br>associated applications are<br>intended as an aid for precisely<br>locating anatomical structures in<br>either open or percutaneous<br>procedures. Their use is<br>indicated for any medical<br>condition in which the use of<br>stereotactic surgery may be<br>appropriate, and where<br>reference to a rigid anatomical<br>structure, such as the skull, can<br>be identified relative to a CT- or<br>MR-based model, or digitized<br>landmarks of the anatomy.<br><br>The system and its associated<br>applications should be used<br>only as an adjunct for surgical<br>guidance. They do not replace<br>the surgeon's knowledge,<br>expertise, or judgment. | SAME with exception of:<br><b>"scarred, granulated and<br/>previously surgically<br/>altered"</b> :<br>This use has been<br>validated in a human<br>clinical study.<br><br><b>"facilitate dilation of the<br/>sinus ostia."</b><br>This use was previously<br>cleared via K132297 as<br>supported by cadaver data.<br>The cleared instructions for<br>use included instructions<br>about the dilation of the<br>sinus ostia. However, the<br>indications for use<br>statement did not. The<br>dilation of the sinus ostia is<br>now being added to the<br>indications for use for<br>clarity. |
| Attributes | EM Sinus Dilation System<br>(Subject Device) | EM Sinus Dilation System<br>(Predicate Device) | Comparison |
| | be identified relative to a CT- or<br>MR-based model, or digitized<br>landmarks of the anatomy.<br><br>The system and its associated<br>applications should be used only<br>as an adjunct for surgical<br>guidance. They do not replace<br>the surgeon's knowledge,<br>expertise, or judgment. | | |
| Location<br>Method | Location of tissue accomplished<br>via electromagnetic (EM) "plug<br>and play" tracking capability<br>using image guidance (IG). | Location of tissue accomplished<br>via electromagnetic (EM) "plug<br>and play" tracking capability<br>using image guidance (IG). | SAME |
| Movement<br>Method | Movement of tissue<br>accomplished using a rigid<br>seeker with unique angles to<br>allow entry of frontal, maxillary or<br>sphenoid sinuses. | Movement of tissue<br>accomplished using a rigid<br>seeker with unique angles to<br>allow entry of frontal, maxillary<br>or sphenoid sinuses. | SAME |
| Dilation<br>Method | Balloon is inflated with saline<br>solution by using the inflator to<br>expand outflow track of targeted<br>sinus ostia. | Balloon is inflated with saline<br>solution by using the inflator to<br>expand outflow track of targeted<br>sinus ostia. | SAME |
| Balloon Size | 5mm, 6mm and 7mm | 5mm, 6mm and 7mm | SAME |
| Patient<br>Contacting<br>Materials | SAME | SAME | SAME |
| Tracking<br>Method | Electromagnetic | Electromagnetic | SAME |
| System<br>Accuracy | Clinical environment | Benchtop and simulated<br>environment | No device changes were<br>implemented as a result of<br>the proposed expanded<br>indications; therefore, no<br>additional benchtop or<br>simulated environment<br>testing was required to<br>demonstrate substantial<br>equivalence. The system<br>accuracy for a patient<br>population with scarred,<br>granulated and previously<br>surgically altered tissue<br>was verified with a clinical<br>study. |
| Method of<br>Action | Reusable instrument for dilation<br>of sinus ostia with balloon | Reusable instrument for dilation<br>of sinus ostia with balloon | SAME |
{6}------------------------------------------------
{7}------------------------------------------------
# [807.92(b)(1)] Determination of Substantial Equivalence
#### Non-Clinical Performance Data The subject device did not undergo any design changes as a result of the proposed expanded indications. As a result, no additional non-clinical performance testing was required to demonstrate substantial equivalence.
#### [807.92(b)(2)] Determination of Substantial Equivalence
| Clinical Performance Data | A clinical study was conducted in subjects with<br>scarred, granulated or previously surgically altered<br>tissue to support the proposed expanded indication.<br>The subject device performed as intended and did<br>not raise new safety concerns; there were no<br>adverse events attributed to the subject device. |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| [807.92(b)(3)] Conclusion | |
Conclusions from Non-Clinical and Clinical Tests
The subject and predicate EM Sinus Dilation System devices are identical in design and performance specifications. Bench and animal testing was not required to demonstrate substantial equivalence. The proposed expanded indications were supported with a prospective, non-randomized, non-blinded, single arm, multicenter clinical study. Subjects with scarred, granulated or previously surgically altered tissue were enrolled and treated in at least one frontal, sphenoid, or maxillary sinus with the subject device. The subject device performed as intended and did not raise new safety concerns; there were no adverse events attributed to the subject device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.