DTAD

K121678 · Gyrus Acmi, Inc. · ERL · Aug 30, 2012 · Ear, Nose, Throat

Device Facts

Record IDK121678
Device NameDTAD
ApplicantGyrus Acmi, Inc.
Product CodeERL · Ear, Nose, Throat
Decision DateAug 30, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DTAD is intended for cutting, coagulation, and removal of tissue in general ENT procedures. Specific indications include: Tonsillectomy and Adenoidectomy.

Device Story

Disposable, battery-operated microdebrider for ENT surgery; performs tissue cutting, coagulation, and removal. Input: manual activation via handpiece button for blade oscillation; electrosurgical generator footswitch for monopolar coagulation. Operation: internal motor oscillates inner blade gear at ~3500 rpm; monopolar lead connects to outer blade for coagulation. Features: flexible blade (0°-50° bend), nosecone for 90° tip rotation, suction port for fluid/tissue removal. Used in OR by surgeons. Output: mechanical resection of tissue and monopolar RF coagulation. Benefits: single-use convenience, integrated suction/cutting/coagulation, adjustable tip geometry for tonsillectomy/adenoidectomy access.

Clinical Evidence

No clinical testing conducted. Evidence provided via non-clinical bench testing (electrical safety per IEC 60601-1/2-2, mechanical/functional performance, stability/shelf-life) and preclinical evaluations (ex vivo bovine tissue for thermal impact/coagulation; in vivo porcine models for ergonomics, usability, and tissue removal).

Technological Characteristics

Single-use battery-operated microdebrider. Materials: patient-contacting parts identical to K034004; silicone RF insulator same as K041285. Energy: battery-powered motor (oscillation) and external electrosurgical generator (monopolar RF). Dimensions: flexible blade (0°-50°). Connectivity: monopolar cable to standard OR generator. Sterilization: Ethylene Oxide (EtO) per ISO 11135. Standards: IEC 60601-1, IEC 60601-2-2, ISO 10993-1.

Indications for Use

Indicated for patients undergoing general ENT procedures, specifically tonsillectomy and adenoidectomy.

Regulatory Classification

Identification

An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Gyrus ACMI DTAD Gyrus ACMI, Inc. K121678 Traditional 510(k) Notification June 2012 Page 1 of 6 # 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. DTAD AUG 30 2012 ## General Information Manufacturer: Gyrus ENT, L.L.C. (subsidiary of Gyrus ACMI, Inc.) 2925 Appling Road Bartlett , TN 38113 Phone: 1-800-262-3540 Fax: 1-901-373-0237 Establishment Registration Number: 510(k) Submitter: Establishment Registration Number: Contact Person: Date Prepared: ## Device Description Classification Name: 1037007 Gyrus ACMI, Inc. 136 Turnpike Rd. Southborough, MA 01772-2104 3003790304 Dolan Mills, RAC Specialist, Regulatory Affairs June 6, 2012 Electrosurgical Cutting & Coagulation Device and Accessories, ENT Electric or Pneumatic Surgical Drill Class 2 21 CFR 878.4400 GEI General and Plastic Surgery Panel 21 CFR 874.4250 ERL Ear, Nose, Throat Gyrus ACMI DTAD Project Name: ﻏ {1}------------------------------------------------ Gyrus ACMI DTAD Gyrus ACMI, Inc. 121678 Traditional 510(k) Notification June 2012 Page 2 of 6 Trade Name(s): Generic/Common Name: Disposable Tonsil and Adenoid Device (DTAD) Electrical Surgical Drill / Shaver or Electrosurgical cutting and coagulation device and accessories # Predicate Devices | Gyrus ACMI Inc. Diego® RF Powered Dissector & Drill System: | K034004 | |-------------------------------------------------------------|---------| | Covidien / ValleyLab Suction Coagulator: | K091223 | | Bovie SEER: | K082568 | | Peak TnA Plasmablade: | K083415 | ## Device Description and Technological Characteristics The DTAD (Disposable Tonsil and Adenoid Device) is a disposable microdebrider blade with a built-in motor that will plug into the users electrosurgical generator. The sole energy application mode is Monopolar coagulation. The monopolar use will require a return electrode for the generator. The cutting performance is the same as the current Gyrus ACMI Diego® tonsil and adenoid blades cleared under K034004, with the addition of monopolar coagulation. The monopolar effect is completely dependent upon the generator used and its settings, which is the same as the other monopolar predicates. The device will be a single-use battery operated microdebrider with monopolar capability. A button will activate the blade oscillation and the standard OR electrosurgical unit footswitch will power the monopolar effect. A nosecone will allow the tip to rotate 90° left or right. The blade will be provided at a 40° bend for adenoids with the capability of being reduced to 15° for tonsillectomies. The blade angle is flexible, and the design allows the blade to be bent between 0° and 50°, The microdebrider device will include a monopolar cable that connects to a separate standard operating room electrosurgical generator. A standard suction tube will be attached to the suction port and a clip will attach the tubing to the cable. Two batteries, provided with a three year shelf life from the manufacturer, will be provided in the DTAD cable to power the blade oscillation. The battery leads connect internally to the PCB in the housing which connects to the oscillation switch and motor. When the handpiece button is pressed, power is sent to the motor which oscillates a gear which in turn oscillates the inner blade gear at approximately 3500 rpm. The monopolar lead from the cable is connected directly to the outer blade and is powered by the electrosurgical generator footswitch. For user convenience the package will also include a disposable stylet for declogging if needed. The stylet is the same as the one provided in the current Diego® tubeset {2}------------------------------------------------ Gyrus ACMI DTAD Gyrus ACMI, Inc. K121678 Traditional 510(k) Notification June 2012 Page 3 of 6 package (K034004). The device packaging is similar to the predicate Diego blades and disposables, and the devices are not labeled as non-pyrogenic. # Material The DTAD uses patient-contacting materials that are utilized in the predicate devices. as well as other legally marketed devices manufactured by Gyrus ACMI. The composition of the DTAD is similar to the devices within the previously cleared predicate, Diego® RF Powered Dissector and Drill System, K034004. The handle and RF cable are not patient-contacting materials, and therefore no biocompatibility testing is warranted for those materials. The composition of the patient contacting shaver blade portion is similar to the devices cleared under K034004. The silicone material used as an RF insulator is the same material used in the disposable Gyrus ACMI Plasmacision devices tied to K041285. These devices are used in a similar manner, for the same procedures, manufactured in the same value stream, and are subjected to the same EtO sterilization cycle and process as the DTAD. The DTAD is classified in accordance with ISO 10993-1, as an external communicating, tissue/bone/dentin device for limited exposure (<24hrs.). ISO10993-1 and FDA Blue Book memo #G95-1 guidelines recommend that these direct patient contact parts have supporting data for cytotoxicity, sensitization and irritation. Full biocompatibility testing (Cytotoxicity, sensitization, and irritation) to ISO10993-1 is available for the predicate devices containing the same patient contacting materials in similar quantities. Since Gyrus ACMI does currently utilize the patient contacting materials in other devices with the same intended use no biocompatibility testing is required. ## Intended Use / Indications The DTAD is intended for cutting, coagulation, and removal of tissue in general ENT procedures. Specific indications include: Tonsillectomy and Adenoidectomy. #### Compliance to Standards The following standards were used during the design and validation of the DTAD: IEC 60601-1: 2005 IEC 60601-2-2: 2009 ISO 10993-1: 2009 ISO 10993-7: 2008 12 {3}------------------------------------------------ Gvrus ACMI DTAD Gyrus ACMI, Inc. K121678 Traditional 510(k) Notification June 2012 June 2012 Page 4 of 6 ISO 11135-1: 2007 ISO 11607-1: 2006 ISO 11607-2: 2006 ISO 15223-1: 2008 ASTM F88 - 09 ASTM F1886-98: 2004 The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests were identified and performed as a result of risk analysis assessment. ## Summary of Sterilization and Shelf Life Discussion The DTAD will be shipped sterile, for single use only. Each device is sterilized using Ethylene Oxide, using an over-kill cycle validated in accordance with ISO 11135 to provide a sterility assurance level of 106. The maximum allowable levels of residues of ethylene oxide, and ethylene chlorohydrin are established. The residual levels are less than those specified in ISO 10993-7. The product will be released with a one year shelf life, which will later be extended to 3 years after functional accelerated age testing is complete. The Shelf Life period for the DTAD was determined through an analysis of the shelf-life stability of the materials used in the design and packaging of the DTAD, which are the same as those used in the predicate Diego® devices, as well as an analysis of the packaging materials and processes used with other Gyrus ACMI devices. Accelerated shelf-life functionality studies were conducted to support an initial one year shelf life, with real time testing in process to confirm an initial one year expiration date. Functional shelflife is also dependent upon the useful life of the batteries. Additional functionality studies (accelerated and real time) are in process to support a three-year expiration date in the future. ## Summary of Performance Testing The following non-clinical, preclinical tests and usability studies were conducted: #### Non-Clinical / Preclinical Performance Evidence of safety and effectiveness was obtained from two primary areas: - 1) non-clinical (electrical, mechanical, functional, stability) performance testing - 2) preclinical (bench tissue, animal) evaluations and testing Non-clinical: Basic safety and performance testing was performed in accordance with IEC 60601-1 and IEC 60601-2-2. In addition, verification and comparison bench studies were conducted to evaluate the mechanical and functional performance. Specifically, test results in the following areas were provided: functionality, RF {4}------------------------------------------------ Gyrus ACMI DTAD Gyrus ACMI, Inc. K121678 Traditional 510(k) Notification June 2012 Page 5 of 6 isolation, tip vibration, fluid ingress, battery and blade endurance, cable and connector pull force, monopolar effect, strain relief strength, suction and clog performance, blade flex capability, ability of the cutting edge to resist damage, blade torque, generator compatibility, and operational temperature. Stability: Representative samples were subjected to accelerated and real time ageing to confirm that the device maintains functionality and continues to meet their specifications over time. The results of the accelerated age testing demonstrate that the device will be stable for the stated shelf-life. In addition, real time age testing will confirm the results of the accelerated age testing. Representative samples were also subjected to ship testing. Preclinical: Evidence obtained from preclinical bench tissue (ex vivo) and animal (in vivo) studies demonstrate that the DTAD performs substantially equivalent to the predicate devices in relevant aspects associated with usability, cutting, coagulation, and removal of tissue. Bench tissue - evaluated ex vivo using bovine tissue: - Thermal margin - - -Thermal impact - Visual comparison of coagulation ・ Animal - evaluated in vivo using porcine models: - Cut, coagulation, and suction - - Ergonomics ー - Usability aspects such as device setup, tip rotation, and tip malleability - - -Overall design confidence Testing demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate devices. No clinical testing was conducted. The use of Electrosurgical Cutting and Coagulation Devices / Microdebriders has been documented in published literature and indicates safe and effective use for the target procedures and patient populations. ## Substantial Equivalence to Predicates The DTAD utilizes features incorporated into the following legally marketed predicate devices: The DTAD utilizes high frequency electrosurgical energy to cut and coagulate tissue in a similar manner as the predicate devices. With the exception of monopolar capability and the built-in battery powered motor, the DTAD cuts and removes tissue in the same manner as the predicate Diego® RF {5}------------------------------------------------ Gyrus ACMI DTAD Gyrus ACMI, Inc. K121678 Traditional 510(k) Notification June 2012 Page 6 of 6 Dissector and Drill System. The DTAD blade offers monopolar coagulation at the cutting tip only when connected to a monopolar electrosurgical generator. The DTAD blade consists of a rotating inner member with a serrated tip within a stationary outer tube with a cutting window at the tip. The DTAD inner blade has a double helix laser cut that allows the inner blade to be flexible. The outer is spiral cut under an annealed aluminum tube and shrink tubing that allows the blade to be bent between 0° and 50°. The DTAD inner blade oscillates at a high rate of speed and as tissue is pulled into the cutting window it is resected and removed. The oscillating portion of the DTAD blade is driven by an electric motor inside the handle portion of the device that incorporates a gear transmission to provide the desired range of speeds and torque. The speed is limited by the motor within the handle. The hollow inner blade is attached to a suction supply to facilitate removal of tissue and fluid from the surgical site. The Covidien / Valleylab suction monopolar coagulators are intended for general and ENT surgical procedures where the coagulation of tissue and suction of fluids are desired. The Bovie resection device is intended for cutting and monopolar coagulation of soft tissue using radiofrequency. The Peak Plasmablade TnA Tonsil and Adenoid Tissue Dissection device is intended to be used for the cutting and coagulation of soft tissue during otolaryngology (ENT) surgery including adenoidectomy and tonsillectomy. #### Conclusion: The cutting and coagulation performance of the DTAD was compared against performance requirements and the predicate systems listed above. Testing demonstrated that the performance requirements were met, and that the Gyrus ACMI DTAD exhibited comparable performance characteristics to the predicates. In summary, the Gyrus ACMI DTAD is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Gyrus ACMI, Incorporated % Mr. Dolan Mills, RAC Specialist, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772 AUG 300 2012 Re: K121678 Trade/Device Name: DTAD Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, nose, and throat electric or pneumatic surgical drill Regulatory Class: II Product Code: ERL Dated: August 17, 2012 Received: August 20, 2012 Dear Mr. Mills: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {7}------------------------------------------------ ## Page 2- Mr. Dolan Mills, RAC device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Erind Keith Image /page/7/Picture/7 description: The image shows a stylized letter 'f' on the left, followed by the letters 'N' and 'R' on the right. The 'f' is written in a cursive style with a loop at the top and a curved tail. The letters 'N' and 'R' are in a simple, sans-serif font and are positioned close to the 'f'. Mark Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ Traditional 510(k) Notification June 2012 Page 1 of 1 # Indications for Use 510(k) Number: K121678 Device Name: DTAD # Indications for Use: . : The DTAD is intended for cutting, coagulation, and removal of tissue in general ENT procedures. Specific indications include: Tonsillectomy and Adenoidectomy. Prescription Use: X (Per 21 CFR 801.109) AND/OR Over-the-Counter Use: (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) E.C. Keith (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K121678
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...