Electric Nasal Aspirator (BC026, BC025, BC023)

K250697 · Shenzhen Kingboom Technology Co., Ltd. · BTA · Apr 23, 2025 · General, Plastic Surgery

Device Facts

Record IDK250697
Device NameElectric Nasal Aspirator (BC026, BC025, BC023)
ApplicantShenzhen Kingboom Technology Co., Ltd.
Product CodeBTA · General, Plastic Surgery
Decision DateApr 23, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4780
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.

Device Story

Portable electric nasal aspirator (models BC026, BC025, BC023) for home use by caregivers on children. Device consists of main unit with motor pump and suction funnel nozzle. Motor generates negative pressure (52-60Kpa) to remove nasal secretions. User interface includes buttons and LED display to control vacuum pressure; some models include light function for distraction. Powered by rechargeable 3.7V Li-ion battery. Device provides non-invasive suction to clear nasal passages, potentially improving patient comfort and breathing. Caregiver operates device manually; output is physical removal of mucus.

Clinical Evidence

No clinical test data was used to support the decision of substantial equivalence. Bench testing only.

Technological Characteristics

Portable powered suction pump. Materials: ABS, PC, Silicone. Sensing/Actuation: DC motor pump. Energy: 3.7V 700mAh rechargeable Li-ion battery. Dimensions: ~160x41x39-41mm. Connectivity: None. Sterilization: Not specified. Software: Complies with ANSI AAMI IEC 62304. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133-2, ISO 10993-1.

Indications for Use

Indicated for intermittent removal of nasal secretions and mucus in children aged 2-12 years. Used in home environments.

Regulatory Classification

Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION April 23, 2025 Shenzhen Kingboom Technology Co., Ltd % Reanny Wang General Manager Shenzhen Reanny Medical Devices Management Consulting Co. Room 1509, Jingting Building, Dongzhou Community Guangming Street, Guangming District Shenzhen, Guangdong 518107 China Re: K250697 Trade/Device Name: Electric Nasal Aspirator Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: February 26, 2025 Received: March 7, 2025 Dear Reanny Wang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250697 - Reanny Wang Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K250697 - Reanny Wang Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shuchen Peng -S Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250697 Device Name Electric nasal aspirator Indications for Use (Describe) The Electric Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} Traditional 510(k) Summary 510(k) number: K250697 1. Information of Submitter and Correspondent Submitter's information: Company Name: Shenzhen Kingboom Technology Co., Ltd Street Address: #201, 2nd fl, No.5, Tongxin Road, Pingdong Community, Pingdi Street, Longgang District City: Shenzhen State/Province: Guangdong Country: China Telephone: +86-13590244726 Fax: / Contact Person: Ma Qiang Contact Title: General Manager Contact Email: 156673750@qq.com Date Prepared: Feb. 26, 2025 Submission correspondent's information: Shenzhen Reanny Medical Devices Management Consulting Co., Ltd Address: Room 1509, Jingting Building, Dongzhou Community, Guangming Street, Guangming District, Shenzhen 518107, China Contact Person: Reanny Wang E-mail: reanny@reanny.com Phone: +86(755) 27391220 2. Device Information Trade Name: Electric nasal aspirator {5} Model: BC026, BC025, BC023 Common Name: Powered suction pump Classification Name: Pump, portable, aspiration (powered) Regulation: 21 CFR § 878.4780 Device Class: Class 2 Product Code: BTA ## 3. Identification of Predicate Device(s) | Manufacturer | Shenzhen XinLianFeng Technology Co., LTD | | --- | --- | | Legally Marketed Device | Electric nasal aspirator, model: BC026 | | 510(K) Number | K222547 | ## 4. Description of Device Electric nasal aspirator (model: BC026, BC025 and BC023) consists of main unit, and suction portion working together as one unit. The Electric nasal aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. The motor pump operates on a rechargeable battery. The rechargeable battery can be charged from the external power adapter (not included in this device) through the provided charging line. The user interface consists of buttons and LED display, and the user can control the vacuum pressure through the button. The differences between BC026, BC025, and BC023 are as follows: 1) BC026 has 5 suction levels, while BC023 and BC025 have 3 suction levels. 2) BC026 and BC023 have "Light" function for distracting and pacifying the child. BC025 does not have this function. 3) BC026 and BC023 can pause suction during the suction process. BC025 does not have this feature. 4) BC026 and BC023 have 4 buttons, BC025 has 2 buttons. 5) In addition, BC026, BC025, and BC023 have different Display, Size, Weight, and Appearance. ## 5. Indications for Use The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. {6} 6. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: 6.1 Non-clinical testing A series of safety and performance tests were conducted on the subject device. - Product service life - Software validation - Electromagnetic compatibility and electrical safety - Function test All the test results demonstrate Electric nasal aspirator meets the requirements of its predefined acceptance criteria and intended use, and it is substantially equivalent to the predicate devices. 6.2 Clinical Testing No clinical test data was used to support the decision of substantial equivalence. 7. Performance Summary The devices conform to applicable standards as follow table: | Test Type | Standard Designation Number | FDA Recogniti on Status | Outcome for Device | | --- | --- | --- | --- | | Safety | ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | Yes | Conforms | | EMC | IEC 60601-1-2:2020 IEC TR 60601-4-2: 2016 | Yes | Conforms | | Home healthcare environment | IEC 60601-1-11:2020 | Yes | Conforms | | Performance | Enterprise standard | Yes | Conforms | | Biocompatibility | ISO 10993-1:2018 | Yes | Conforms | | Software | ANSI AAMI IEC 62304:2006/A1:2016 | Yes | Conforms | | Safety of Lithium battery | IEC 62133-2:2017 | Yes | Conforms | {7} Biocompatibility: The patient contact part (Funnel nozzle) utilized in the subject device is the same as the FDA cleared BC026 Electric nasal aspirator in K222547, their relationship to the patient for the subject device are identical in formation, processing, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.). # 8. Discussion of Comparison to Predicate Devices. The Electric nasal aspirator submitted in this 510(k) submission is substantially equivalent in intended use, technological characteristics, materials, and performance to the cleared BC026 Electric nasal aspirator K222547. Differences between the subject and predicate device do not raise new questions of safety and effectiveness. | Device | Subject device | Primary Predicate device | Comparison | | --- | --- | --- | --- | | Manufacturer | Shenzhen Kingboom Technology Co., Ltd | Shenzhen XinLianFeng Technology Co., LTD | / | | 510(K) number | TBD | K222547 | / | | Product name | Electric nasal aspirator, model: BC026, BC025, BC023 | Electric nasal aspirator, model: BC026 | / | | Classification | Class II Device, BTA (21 CFR § 878.4780) | Class II Device, BTA (21 CFR § 878.4780) | Identical | | Classification Panel | General & Plastic Surgery | General & Plastic Surgery | Identical | | Indication for Use | The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. | The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. | Identical | | Patient Population | Age 2-12 years old | Age 2-12 years old | Identical | | Intended Environment | Home use | Home use | Identical | | Device Description | The BC026, BC025, BC023 Electric nasal aspirator is a portable device which is intended for suction of nasal | The BC026 Electric nasal aspirator is a portable device which is intended for suction of nasal | Identical | {8} | Device | Subject device | Primary Predicate device | Comparison | | --- | --- | --- | --- | | | passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. | passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. | | | Vacuum pressure | 52-60Kpa | 52-60Kpa | Identical | | Music function | Yes | Yes | Identical | | Light function | BC026 and BC023: yes BC025: No | Yes | Similar Note 1 | | Noise Level | <80dBA | <80dBA | Identical | | Power consumption | 2.2W | 2.2W | Identical | | Motor Type | 3.7V DC | 3.7V DC | Identical | | Power Source | DC 3.7 V / 700mAh Rechargeable Li-ion battery | DC 3.7 V / 700mAh Rechargeable Li-ion battery | Identical | | Device Dimension | BC026 and BC025: 160 (H) x 41 (L) x 41 (W)mm; BC023: 160 (H) x 41(L) x 39(W)mm | 160 (H) x 41 (L) x 41 (W)mm | Similar Note 2 | | Weight | BC026: 320±5g BC025: 305±5g BC023: 317±5g | 320±5g | | | Tips Dimension (ψ) | OD4.3/ID2.4 | OD4.3/ID2.4 | Identical | | Main Materials | ABS, PC, Silicone | ABS, PC, Silicone | Identical | | Operating condition | 5°C(41°F) to 40°C(104°F); 15% to 93% R.H. | 5°C(41°F) to 40°C(104°F); 15% to 93% R.H. | Identical | | Storage condition | -10°C(-23°F) to 70°C (158°F); 10% to 95% R.H. | -10°C(-23°F) to 70°C (158°F); 10% to 95% R.H. | Identical | | Expected service life | 2 years | 2 years | Identical | | Type BF | Type BF applied part | Type BF applied part | Identical | {9} | Device | Subject device | Primary Predicate device | Comparison | | --- | --- | --- | --- | | applied part | | | | | Safety | IEC 60601-1 IEC 60601-1-11 IEC 62133-2 IEC 62471 | IEC 60601-1 IEC 60601-1-11 IEC 62133-2 IEC 62471 | Identical | | EMC | IEC 60601-1-2 | IEC 60601-1-2 | Identical | | Water-resistance | IP22 | IP22 | Identical | | Biocompatibility | BC026, BC025, BC023 operate in conjunction with silicone nasal aspiration tips, which come into the contact with nasal skin and mucosa for less than 24 hours. | BC026 operates in conjunction with silicone nasal aspiration tips, which come into the contact with nasal skin and mucosa for less than 24 hours. | Identical | | Contacted Parts | Silicone Tip (funnel nozzle) | Silicone Tip (funnel nozzle) | Identical | | Material of contacted parts | Silicone | Silicone | Identical | The discussion of differences exist between the subject and predicate devices is listed as follows: Note 1 Light function is used to distract and pacify the child. This function is optional and is not required for the device to achieve its intended use. Without the light function components, BC025 still complied with IEC 60601-1, IEC 60601-1-11 and IEC 60601-1-2, which demonstrates substantial equivalence. Note 2 Although there are some minor differences in the "Device Dimension" and "Weight" between the predicate device and BC025, BC023, the patient contact part (Funnel nozzle) is identical for all models. Therefore, differences in dimension and weight do not affect the substantial equivalence comparison. 9. Conclusions Based on performance testing, comparison and analysis, the Electric nasal aspirator (model: BC026, BC025, BC023) subject device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...