K103047 · Ceremed , Inc. · KHJ · Jan 5, 2011 · Ear, Nose, Throat
Device Facts
Record ID
K103047
Device Name
CERETENE SOLUBLE IMPLANT MATERIAL
Applicant
Ceremed , Inc.
Product Code
KHJ · Ear, Nose, Throat
Decision Date
Jan 5, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3620
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Ceretene" Soluble Implant Material is indicated for use as a water-soluble implant material and as a water-soluble space occupying material as an adjunct during the natural healing process.
Device Story
Ceretene Soluble Implant Material is a sterile, odorless, opaque, wax-like substance composed of water-soluble alkylene oxide copolymers (AOC). Provided in bars and sheets (0.5 to 5 grams), it is intended for manual manipulation by clinicians to soften and increase stickiness before application. The material acts as a space-occupying implant to support the natural healing process. It is supplied sterile via irradiation and is intended for single use.
Clinical Evidence
Bench testing only.
Technological Characteristics
Material: Water-soluble alkylene oxide copolymers (AOC). Form: Wax-like bars and sheets (0.5-5g). Sterilization: Irradiation. No additives or colorants.
Indications for Use
Indicated for use as a water-soluble implant and space-occupying material as an adjunct during the natural healing process.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
K062280 — OSTENE, AOC, OSTEOTENE, CERETENE, CEREPOR, APTENE, APATENE AND ACTIPASTE · Ceremed , Inc. · Dec 11, 2006
K122561 — ADAPTAIN FASTWRAP, ENVELOCK, GRAFTAIN · Ceremed , Inc. · Nov 27, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
K103047 lofz
Ceremed, Inc.
Page 6 - 1
Traditional 510(k) - Ceretene" Soluble Implant Material
#### VI - 510 (K) SUMMARY
#### Submitted by:
(f
P
Tadeusz Wellisz, M.D. Ceremed, Inc. 3643 Lenawee Ave. Los Angeles, California 90016 Tel: (310) 815-2125 Fax: (310) 815-2130
Contact Person:
Date Prepared
Common/Usual Name:
Proprietary Name:
Regulation Number:
Regulation Name:
Regulatory Class:
Product Code:
Predicate Device:
Tadeusz Wellisz, M.D.
September 24, 2010
Soluble Synthetic Polymer Implant Material
Ceretene™ Soluble Implant Material
21 CFR 874.3620
Ear, nose and throat synthetic polymer material
II
KHJ
Ceremed, Inc. Ceretene" (K081531) Ceremed, Inc. Ostene®CT (K091636)
#### Description of the device:
Ceretene" Soluble Implant Material is an odorless, opaque wax-like material designed to be utilized directly out of the package. It is best used immediately following removal from the package, and can be softened and increased in stickiness by warming and by additional handling and manipulation, if so desired.
Ceretene" Soluble Implant Material is comprised of a sterile mixture of water-soluble alkylene oxide copolymers (AOC). Ceretene™ Soluble Implant Material contains no other additives or colorants. Ceretene" Soluble Implant Material is formed in bars and sheets of various weights ranging from 0.5 to 5 grams each.
JAN 5 20111
{1}------------------------------------------------
## Ceremed, Inc.
# Traditional 510(k) - Ceretene™ Soluble Implant Material
Ceretene" Soluble Implant Material is provided sterile by irradiation and must not be resterilized.
## Intended use:
7
Ceretene" Soluble Implant Material is indicated for use as a water-soluble implant material and as a water-soluble space occupying material as an adjunct during the natural healing process.
### Substantial equivalence:
The non-clinical evaluations used to determine substantial equivalence included indications, intended use, design, materials, sterilization, and performance. The comparison demonstrates that the device in this submission has the same fundamental scientific technology as the legally marketed predicate Ceretene" (K081531) and is substantially equivalent in indications, intended use, and performance; and is identical in design, materials, and sterilization to the predicate Ostene®CT (K091636).
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
111 5 2011
Ceremed, Inc. % Tadeusz Wellisz, M.D. 3643 Lenawee Avenue Los Angeles, California 90016
Re: K103047
Trade/Device Name: Ceretene™ Soluble Implant Material Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, nose, and throat synthetic polymer material Regulatory Class: Il Product Code: KHJ Dated: September 24, 2010 Received: October 15, 2010
Dear Dr. Wellisz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
. .
{3}------------------------------------------------
Page 2 - Tadeusz Wellisz, M.D.
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
lf you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), please go to http://www.fda.gov/Aboutf-DA/CentersOffices/CDRH/CDRHOffices/ucm1 1 5800.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21)FOR Preci. 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll me (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
A. B. R.
for.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Ceremed, Inc.
# Traditional 510(k) - Ceretene™ Soluble Implant Material
## Y. INDICATIONS FOR USE:
510 (k) Number (if known):
Device Name: Ceretene™ Soluble Implant Material,
Indications For Use:
Ceretene" Soluble Implant Material is indicated for use as a water-soluble implant material and as a water-soluble space occupying material as an adjunct during the natural healing process.
| Prescription Use | X |
|----------------------|---|
| (Per 21 CFR 801.109) | |
OR
Over-The-Counter Use (Optional Format 1-2-96)
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED.)
CONCURRANCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
Daniel Krave for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103047
Division Sign-Off
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.