OSTENE, AOC, OSTEOTENE, CERETENE, CEREPOR, APTENE, APATENE AND ACTIPASTE
K062280 · Ceremed , Inc. · KHJ · Dec 11, 2006 · Ear, Nose, Throat
Device Facts
Record ID
K062280
Device Name
OSTENE, AOC, OSTEOTENE, CERETENE, CEREPOR, APTENE, APATENE AND ACTIPASTE
Applicant
Ceremed , Inc.
Product Code
KHJ · Ear, Nose, Throat
Decision Date
Dec 11, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ostene® is intended for use as a water-soluble space occupying material as an adjunct during the natural healing process and it provides for control of bleeding from bone surfaces by acting as a mechanical barrier.
Device Story
Ostene is a sterile, water-soluble, wax-like material composed of alkylene oxide copolymers (ethylene oxide and propylene oxide). Used as a mechanical barrier to control bleeding from bone surfaces and as a space-occupying material during healing. Supplied in sterile packaging; intended for immediate use post-removal. Can be softened or made stickier via manual manipulation or warming. Applied by clinicians during surgical procedures. Provides hemostasis through physical occlusion of bone surfaces. Biocompatible per ISO-10993 standards.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics, intended use, and biocompatibility testing in accordance with ISO-10993 standards.
Technological Characteristics
Water-soluble synthetic polymer material; alkylene oxide copolymer blend (ethylene oxide and propylene oxide). Sterile, wax-like consistency. Sterilized via irradiation; non-resterilizable. Biocompatibility compliant with ISO-10993.
Indications for Use
Indicated for use as a water-soluble implant material to control bleeding from bone surfaces and as a space-occupying material adjunct during natural healing processes.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
Predicate Devices
Ceremed, Inc. AOC™ Porous Polyethylene Surgical Implants (K043133)
Medtronic Xomed, Inc. MeroGel™ Injectable (K002972)
Submission Summary (Full Text)
{0}------------------------------------------------
K062280 Page 11
## Ceremed, Inc. 510 (k) Premarket Notification - Ostene Implant Material
# VII - 510 (K) SUMMARY OF SAFETY AND EFFECTIVENESS:
### Submitted by:
Tadeusz Wellisz, M.D. Ceremed, Inc. 3643 Lenawee Ave. Los Angeles, California 90016 Tel: (310) 815-2125 Fax: (310) 815-2130
# DEC 1 1 2006
| Contact Person: | Tadeusz Wellisz, M.D. |
|----------------------|-------------------------------------------------------------------------------|
| Date Prepared | November 14, 2006 |
| Common/Usual Name: | Soluble Polymer Surgical Implant Material |
| Proprietary Names: | Ostene®, AOC™, Osteotene™, Ceretene™, Cerepor™, Aptene™, Apatene™, Actipaste™ |
| Classification Name: | Synthetic Polymer Material, (per 21 CFR 874.3620) |
### Product Code:
KHJ, MTT
### Predicate Devices
- 1 . Ceremed, Inc. AOC™ Porous Polyethylene Surgical Implants K043133
- 2. Ceremed, Inc. AOC™ Bone Wax K041363
- 3. Ceremed, Inc. AOC" Bone Wax (Ostene®) K052528
- 4. Genzyme Corp. Sepragel™ ENT K043035
- 5. Medtronic Xomed, Inc. MeroGel™ Injectable K002972
Page 7
{1}------------------------------------------------
K062280 Page 2/2
Page 8
## Ceremed. Inc. 510 (k) Premarket Notification - Ostene Implant Material
### Description of the device:
Ostene® is a water-soluble odorless, opaque wax-like material consisting of a sterile mixture of alkylene oxide copolymers, derived from ethylene oxide and propylene oxide. It is best used immediately following removal from the package, and can be softened and increased in stickiness by warming and by additional handling and manipulation, if so desired. Ostene® is provided sterile by irradiation and must not be resterilized.
### Intended use:
Ostene® is intended for use as a water-soluble space occupying material as an adjunct during the natural healing process and it provides for control of bleeding from bone surfaces by acting as a mechanical barrier.
### Substantial equivalence:
The Ostene® in this application has the same intended use fundamental scientific technology as the legally marketed AOC™ Bone Wax (Ostene®) and the coating of AOC™ Porous Polyethylene Surgical Implants. The classification of the Ostene® as water-soluble synthetic polymer material for use as a space occupying material is supported by its comparison to the legally marketed device Sepragel (K043035) and MeroGel (K002972).
The Ostene® in this application as a bone hemostatic agent is identical to, and has the same intended use and indications for use as the predicate device AOC™ Bone Wax (Ostene®). The biocompatibility of the alkylene oxide copolymer blend is in accordance with the standards set forth in ISO-10993 Biological Testing of Medical and Dental Materials and Devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a symbol and text. The symbol is a stylized representation of an eagle or bird, with three curved lines forming its body and wings. The text is arranged in a circular fashion around the symbol, spelling out "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ceremed, Inc. DEC 1 1 2006 % Tad Wellisz, M.D. President 3643 Lenawee Avenue Los Angeles, California 90016 Re: K062280 R002280
Trade/Device Name: Ostene", AOC"", Osteotene™, Ceretene™, Aptene™",
Actipaste", Apatene™, Actipaste™ Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, nose, and throat synthetic polymer material Regulatory Class: II Product Code: KHJ, MTJ Dated: November 14, 2006 Received: November 16, 2006
Dear Dr. Wellisz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Tad Wellisz, M.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Yol N
Mark N. Melkerson
DEP Director
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Ceremed, Inc. 510 (k) Premarket Notification – Ostene Implant Material
## XII. INDICATIONS FOR USE:
510 (k) Number (if known): __________________________________________________________________________________________________________________________________________________
Device Name: Ostene®, AOC™, Osteotene™, Ceretene™, Cerepor™, Aptene™, Apatene™, Actipaste™
Indications For Use:
OSTENE® is indicated for use as a water-soluble implant material for use in the control of bleeding from bone surfaces and as a water-soluble space occupying material as an adjunct during the natural healing process.
Prescription Use
(Per 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office Of Device Evaluation (ODE)
**Sign Off**
(Division Sign-Off) Division of General, Restorative, Page 1 of and Neurological Devices
**510(k) Number** L062280
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.