Post-market clinical data was used to confirm the safety and efficacy of the Lyric2 by assessing device wear time, patient comfort, usability, and clinical performance (gain, speech understanding, sound quality) in a real-world setting.
Post-market clinical testing; Usability; Patient comfort; Device wear time
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Post-market clinical testing; Post-market clinical assessment; Follow-up/Duration: Up to 4 months (device wear duration)
Hearing-impaired persons
Cleared Lyric device
Device sizing, gain, speech understanding, sound quality, length of device wear, patient ratings of comfort, post-removal health of the ear, and usability
Indications for Use
The Lyric 2 hearing aid is a disposable, extended-wear air conduction hearing aid, designed to be used and worn by hearing-impaired persons. Hearing assistance is achieved through amplification of sound pressure waves, which are transmitted to the external ear canal via air conduction. The hearing aid is placed in the ear canal by an appropriately trained ENT Physician, Audiologist or Hearing Aid Dispenser and can remain in the ear canal for up to 4 months or until the battery is depleted. Upon device removal the hearing aid is discarded.
Device Story
Disposable, extended-wear, deep-canal air conduction hearing aid; receives sound waves via analog microphone; processes signals using programmable ASIC with WDRC (Wide Dynamic Range Compression) circuit; transmits amplified sound to ear canal. Placed by ENT physician, audiologist, or dispenser; remains in canal up to 4 months. Includes magnetic sensor for remote control of volume/settings. User/professional removes device via removal loop; device discarded after use. Benefits patient through continuous, extended-wear hearing assistance. Performance adjusted via proprietary fitting software.
Clinical Evidence
Bench testing per ANSI S3.22:2009; biocompatibility, electrical safety, and EMC testing performed. Post-market clinical testing confirmed comparable performance to cleared device, including assessments of device sizing, gain, speech understanding, sound quality, wear duration, patient comfort, and post-removal ear health.
Technological Characteristics
Disposable, deep-canal, air conduction hearing aid. Features: analog microphone, programmable ASIC for analog signal processing, custom built-in battery, magnetic sensor for remote control, cerumen protection, and removal loop. Digitally programmable WDRC circuit. Connectivity via proprietary fitting system/software. Form factor designed for deep-canal placement.
Indications for Use
Indicated for hearing-impaired persons requiring air conduction hearing assistance. Device is placed in the ear canal by trained professionals and worn for up to 4 months or until battery depletion.
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.
{0}------------------------------------------------
KI30790
## 510(k) SUMMARY
#### Phonak's Lyric2
# JUN 2 7 2013
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Phonak LLC 4520 Weaver Pkwy Warrenville, IL 60555
Phone: 630-821-5058 Facsimile: 630-393-7400
Contact Person: Laura Eilman
Date Prepared: May 29, 2013
# Name of Device and Name/Address of Sponsor:
Lyric2
Phonak LLC 4520 Weaver Pkwy Warrenville, IL 60555
# Common or Usual Name: Lyric or Lyric2
Classification Name: Hearing Aid, air conduction
Predicate Devices
| | MANUFACTURER'S NAME | DEVICE'S TRADE<br>NAME | 510(K) NUMBER |
|---|-----------------------|------------------------|---------------|
| 1 | Insound Medical, Inc. | Insound XT Series | K021867 |
| 2 | Insound Medical, Inc. | Lyric Hearing Aid | K081136 |
# Intended Use / Indications for Use
The Lyric 2 hearing aid is a disposable, extended-wear air conduction hearing aid, designed to be used and worn by hearing-impaired persons. Hearing assistance is achieved through amplification of sound pressure waves, which are transmitted to the external ear canal via air conduction. The hearing aid is placed in the ear canal by an appropriately trained ENT Physician, Audiologist or Hearing Aid Dispenser and can remain in the ear canal for up to 4 months or until the battery is depleted. Upon device removal the hearing aid is discarded.
{1}------------------------------------------------
### Technological Characteristics
Lyric2 is a disposable, programmable, deep-canal, extended-wear, air conduction hearing device, It uses an analog, digitally programmable WDRC ("Wide Dynamic Range Compression") circuit with very low power consumption. Hearing assistance is achieved through amplification of sound pressure waves, which are transmitted to the external ear canal via air conduction. The amplification characteristics are contained in digitally programmable memory, and adjustment of device parameters is achieved through the proprietary Fitting System and Software.
The Lyric2 consists of:
- A programmable application specific integrated circuit ("ASIC") for analog signal processing. . An analog microphone receives sound waves that will be amplified and transmitted to the speaker.
- A custom made built-in battery to power the amplification circuit. .
- A magnetic sensor to program device parameters and to remotely control user settings � (On/Off/Sleep/Volume).
- Mechanical components to protect the device from cerumen. .
- A removal loop to allow the user and/or professional to remove the device. .
- . Seals that allow the device to fit comfortably in the ear canal
Accessories available for Lyric2:
- User Remote Control
- . Device sizer to determine best device size
- Length sizer to determine insertion depth .
- . An insertion tool for insertion and removal of device and device sizers
- A dedicated software and programming interface to program the device with customer · ● specific settings
### Performance Data
The performance characteristics of the Lyric Hearing Aid have been evaluated in accordance with ANSI S3.22:2009, "Specification of Hearing Aid Characteristics." The device met all applicable specifications developed by the company in accordance with design input specifications. Additional performance and functional testing, including biocompatibility testing, electrical safety, and electromagnetic compatibility testing, has confirmed the equivalent performance of the modified Lyric compared to the cleared Lyric. Post-market clinical testing also confirmed the comparable performance of the Lyric2 to the cleared device. These data included an assessment for appropriate device sizing, as well as clinical assessments of gain, speech understanding, and sound quality of the study device, length of device wear, patient ratings of comfort, assessment of the proportion of the otherwise eligible population who meet the sizing parameters of the study device, post-removal health of the ear, and usability evaluations by patients and audiologist. The clinical assessments performed demonstrated the comparable safety and efficacy of the Lyric2.
# Substantial Equivalence
The Lyric2 has the same general intended use and same indications, similar technological characteristics, and same principles of operation as the cleared Lyric. The minor differences in the
{2}------------------------------------------------
Lyric2's technological characteristics do not raise new questions of safety or effectiveness. Bench and clinical testing demonstrates that the Lyric2 is as safe and effective as the cleared Lyric. Thus, the Lyric2 is substantially equivalent to its predicate device.
and the control control of the control of the control of
1999 - 1999 - 1999
.
and the comments of the comments of
.
.
:
-
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
June 27, 2013
Phonak LLC % John. J. Smith, MD, JD Regulatory Counsel Hogan Lovells US LLP 555 Thirteenth Street, NW Washington, DC 20004
Rc: K130790
Trade/Device Name: Lyric 2 Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing aid, air-conduction Regulatory Class: Class I Product Code: ESD Dated: May 29, 2013 Received: May 29, 2013
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r heast be actised and i Drivination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
Food and Drug Administration 10903 New Hampshire Avenue Document Control Cemer - WO66-G609 Silver Spring, MD 20993-0002
{4}------------------------------------------------
Page 2 - Dr. John J. Smith
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Eric AMann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
K130790 510(k) Number (if known):
Device Name: Lyric2
Indications for Use:
The hearing aid is a disposable, extended-wear air conduction hearing aid, designed to be used and wom by hearing-impaired persons. Hearing assistance is achieved through amplification of sound pressure waves, which are transmitted to the external ear canal via air conduction. The hearing aid is placed in the ear canal by an appropriately trained ENT Physician, Audiologist or Hearing Aid Dispenser and can remain in the ear canal for up to 4 months or until the battery is depleted. Upon device removal the hearing aid is discarded.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sageev George -S 2013.06.21 14:12:21 -04'00'
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.