SONIC INNOVATIONS COMPLETELY-IN-THE-CANAL DSP HEARING AID WITH SOFT SHELL
K984518 · Sonic Innovations, Inc. · ESD · Mar 12, 1999 · Ear, Nose, Throat
Device Facts
Record ID
K984518
Device Name
SONIC INNOVATIONS COMPLETELY-IN-THE-CANAL DSP HEARING AID WITH SOFT SHELL
Applicant
Sonic Innovations, Inc.
Product Code
ESD · Ear, Nose, Throat
Decision Date
Mar 12, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3300
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
This air conduction hearing instrument is intended to amplify sound pressure waves and transmit the signal to the external ear through the medium of air to compensate for hearing losses of mild to moderate degree.
Device Story
Completely-in-the-canal (CIC) hearing aid; utilizes digital signal processing (DSP) to amplify sound pressure waves. Input: ambient sound; Output: amplified acoustic signal. Features full dynamic range multiband compression (multiplicative DSP) allowing independent gain/compression adjustment across frequency bands (250-8000 Hz). Device is programmable via PC-based software (HI-Pro/Noah). Used by patients with mild-to-moderate hearing loss; fitted by hearing aid dispensers. Soft shell component is biocompatible, removable, and replaceable. Benefits patient by compensating for hearing impairment through customized amplification.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and compliance with ANSI S3.22 1987 standards.
Technological Characteristics
CIC hearing aid; biocompatible soft shell; zinc-air size 10 battery; digital signal processing; multiband compression; programmable via PC interface; ANSI S3.22 1987 compliant.
Indications for Use
Indicated for individuals with slight, mild, or moderate hearing impairment, including high frequency precipitously sloping, gradually sloping, reverse slope, or flat configurations, and those with low tolerance for loud sounds.
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.
{0}------------------------------------------------
510(k) Premarket Notification
Sonic Innovations CIC DSP Hearing Aid
K984518
SONIC innovations
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
for the
SONIC INNOVATIONS® COMPLETELY-IN-THE-CANAL (CIC) DSP HEARING AID
## Prepared: December 10, 1998
Submitted by: Jeannette Seloover Johnson, Ph.D. Vice President, Research and Planning SONIC innovations Inc. 5330 South 900 East, Suite 240 Salt Lake City, Utah 84117-7261 Telephone 801.288.0993 Fax 801.288.0998
| Classification Name: | Hearing Aid - Air Conduction |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Hearing Aid |
| Proprietary Name: | SONIC innovations® Completely-in-the-Canal DSP<br>Hearing Aid with Soft Shell |
| Device Type: | Completely-in-the-Canal (CIC) Hearing Aid |
| Classification: | Hearing Aid, Air Conduction Panel 77, Procode<br>ESD, 874.3300, Class I |
| Intended Use: | This air conduction hearing instrument is intended to<br>amplify sound pressure waves and transmit the<br>signal to the external ear through the medium of air<br>to compensate for hearing losses of mild to<br>moderate degree. |
| Indications for<br>Use | Indications for use include persons with mild to<br>moderate hearing impairment depending upon<br>specific characteristics of the hearing loss and the<br>patient's environmental situations. |
| Substantially | The SONIC innovations Completely-in-the-Canal<br>DSP w/ |
| Equivalent to: | Soft Shell is substantially equivalent to the SONIC<br>innovations® CIC DSP hearing aid [510(k) not<br>required], the Decibel Instruments Articular<br>(inTune™) Hearing Aids [510(k) No. K964603], and<br>the Hearing Components Ad-Hear Comply<br>Disposable Earmolds [510(k)915794], i.e., it has the<br>SAME intended use and does not raise different<br>questions regarding safety and effectiveness. |
| Materials: | Assembled from standard hearing aid components<br>mounted with custom microchips. The hearing aid<br>soft shell is manufactured from biocompatible<br>materials that have been used in other medical<br>devices. |
| Technical<br>Characteristics: | Technical specifications are in accordance with<br>ANSI S3.22 1987. |
| Power source: | Standard Zinc-Air hearing aid battery, size 10. |
| Features: | The SONIC innovations Completely-in-the-Canal<br>DSP Hearing Aid with Soft Shell is a programmable<br>hearing aid with fully digital signal processing and<br>full dynamic range multiband compression<br>(multiplicative DSPTM) that allows independent<br>adjustment of gain and compression characteristics<br>within each frequency band from 250 to 8000 Hz.<br>The SONIC innovations hearing aid provides<br>database storage of patient information and hearing<br>aid fitting parameters. |
| Assembly: | The Soft Shell is removable and/or replaceable, and<br>can be cleared or replaced as needed. The shells<br>are offered in a few basic sizes and selected by the<br>hearing aid dispenser.<br><br>Except for the Soft Shell, the device is manufactured<br>and delivered completely assembled to the hearing<br>aid dispenser using materials and techniques widely<br>used by other manufacturers of hearing devices.<br><br>Two versions of the Soft Shell are available, both<br>made of biocompatible materials. The two versions<br>of the shell are available in multiple sizes, the<br>specific one being selected by the hearing aid<br>dispenser as suitable for an individual patient.<br><br>The Soft Shell may be cleaned and/or replaced as<br>needed by either the hearing aid dispenser or the<br>individual patient. |
| User Controls: | None |
| Programmability: | All parameters are digitally programmed via<br>proprietary software and the SONIC innovations<br>Hearing Aid Fitting and Programming System or HI-<br>Pro/Noah with PC or HI-Pro with PC. |
J. S. Johnson CONFIDENTIAL Page 102 of 103 Sonic Innovations, Inc. 5330 South 900 East, Suite 240 · Salt Lake City, Utah · 84117-7261 tel 801.288.0993 · fax 801.288.0998
{1}------------------------------------------------
510(k) Premarket Notification Sonic Innovations CIC DSP Hearing Aid
## A USER'S INSTRUCTION GUIDE AND ANSI TECHNICAL DATA ARE SUPPLIED WITH EACH HEARING AID.
J. S. Johnson Sonic Innovations, Inc. CONFIDENTIAL
Page 103 of 103
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a bold, sans-serif font. The words are arranged on a single line and are left-aligned. The background of the image is plain and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 2 1999
Jeannette Seloover Johnson, Ph.D. Vice President, Research & Planning Sonic Innovations, Inc. 5330 South 900 East, Suite 240 Salt Lake City, Utah 84117-7261
Re:
K984518 SONIC Innovations® Completely-in-the Canal (CIC) DSP Hearing Aid with Soft Shell Dated: December 10, 1998 Received: December 21, 1998 Regulatory class: I 21 CFR 874.3300/Procode: 77 ESD
Dear Dr. Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notfication"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CART Daniel C. Schultz, M.D.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Image /page/2/Picture/15 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
{3}------------------------------------------------
1 4 45 1 3
| 510(k) Number (if known): | K984518 |
|---------------------------|---------|
|---------------------------|---------|
Sonic Innovations Completely-In-The-Canal
| Device Name: | DSP Hearing Aid with Softshell |
|--------------|--------------------------------|
|--------------|--------------------------------|
Indications For Use:
Page 1 of 1
A. General Indicacions:
The indication for use of the air conduction hearing aids in this file indication is to amplify sound for individuals with impaired hearing. The devices are indicated for individuals with losses in the folloving category(ies). (Check appropriate space(s)): I
| Severity: | |
|---------------------------------------------|--|
| ✓ 1. Slight | |
| ✓ 2. Mild | |
| ✓ 3. Moderate | |
| 4. Severe | |
| 5. Profound | |
| Configuration: | |
| ✓ 1. High Frequency - Precipitously Sloping | |
| ✓ 2. Gradually Sloping | |
| ✓ 3. Reverse Slope | |
| ✓ 4. Flat | |
| 5. Other | |
| Other | |
| ✓ 1. Low tolerance | |
| 2. | |
| 3. | |
B. Specific Indications (Only if appropriate.):
(Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.)
1.
2.
3.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological 510(k) Number Rescricted device (per 21 CFR 801,420 & 21 CFR 801.421)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.