EVIS EUS Ultrasound Bronchofibervideoscope Olympus BF-UC190F
K183525 · Olympus Medical Systems Corp. · PSV · Sep 6, 2019 · Radiology
Device Facts
Record ID
K183525
Device Name
EVIS EUS Ultrasound Bronchofibervideoscope Olympus BF-UC190F
Applicant
Olympus Medical Systems Corp.
Product Code
PSV · Radiology
Decision Date
Sep 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This instrument has been designed to be used with diagnostic ultrasound system, video system center, light source, documentation equipment, display monitor, EndoTherapy accessories such as an aspiration biopsy needle. This instrument is designed for endoscopic real-time ultrasound imaging, for performing endoscopic ultrasound guided needle aspiration within the airways, tracheobronchial tree, and esophagus.
Device Story
The BF-UC190F is a flexible video endoscope used for real-time ultrasound imaging and ultrasound-guided needle aspiration in the airways, tracheobronchial tree, and esophagus. It integrates with a diagnostic ultrasound system, video center, light source, and display monitor. The device features a control section with an angulation lever for tip steering, an insertion section containing a CCD unit, fiber bundle, instrument channel for therapeutic tools, and an ultrasound transducer. The transducer emits and receives ultrasound waves for imaging. The device is operated by a physician in a clinical setting. The physician views the ultrasound and video output on a monitor to guide needle aspiration or observe tissue. The device aids in diagnosis and therapeutic intervention by providing visual and ultrasound-based guidance for biopsy procedures, potentially improving diagnostic accuracy and patient outcomes.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through thermal safety, mechanical durability, biocompatibility (ISO 10993-1), electrical safety (ANSI/AAMI ES60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), and human factors validation (ISO 14971).
Technological Characteristics
Flexible video endoscope with electrical curved linear array ultrasound transducer (7-10 MHz). Features include CCD imaging, fiber bundle, and instrument channel (2.2mm). Connectivity via integrated electrical/ultrasound connectors. Sterilization via Ethylene oxide or H2O2 (V-PRO maX). Complies with IEC 60601-1, IEC 60601-2-18, and IEC 60601-2-37.
Indications for Use
Indicated for endoscopic real-time ultrasound imaging and ultrasound-guided needle aspiration within the airways, tracheobronchial tree, and esophagus in patients requiring diagnostic or therapeutic endoscopic procedures.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 6, 2019
Olympus Medical Systems Corp. % Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, Pennsylvania 18034-0610
Re: K183525
Trade/Device Name: EVIS EUS Ultrasound Bronchofibervideoscope Olympus BF-UC190F Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: PSV, ITX Dated: August 8. 2019 Received: August 9, 2019
Dear Sheri Musgnung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183525
Device Name
EVIS EUS ULTRASOUND BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-UC190F
#### Indications for Use (Describe)
This instrument has been designed to be used with diagnostic ultrasound system, video system center, light source, documentation equipment, display monitor, EndoTherapy accessories such as an aspiration biopsy needle. This instrument is designed for endoscopic real-time ultrasound imaging, for performing endoscopic ultrasound guided needle aspiration within the airways, tracheobronchial tree, and esophagus.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. There is a registered trademark symbol to the right of the word. There is a yellow line underneath the word.
September 3, 2019
# 510(k) Summary
# 1 GENERAL INFORMATION
- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-machi, Hachioji-shi, Tokyo, Japan 192-8507
- Contact Person: Sheri Musgnung Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung@olympus.com
# 2 DEVICE IDENTIFICATION
| ■ Device Name | EVIS EUS ULTRASOUND<br>BRONCHOFIBERVIDEOSCOPE OLYMPUS BF<br>UC190F |
|------------------------|--------------------------------------------------------------------|
| ■ Common Name | Ultrasound Bronchoscope |
| ■ Regulation Number | 892.1550 |
| ■ Regulation Name | Ultrasonic pulsed doppler imaging system. |
| ■ Regulatory Class | II |
| ■ Product Code | PSV, ITX |
| ■ Classification Panel | Ear, Nose, and Throat |
#### 3 PREDICATE DEVICE
| Device name | 510(k) Submitter | 510(k) No. |
|-----------------------------------|-------------------------------|------------|
| XBF-UC180F-DT8<br>(AKA BF-UC180F) | OLYMPUS MEDICAL SYSTEMS CORP. | K070983 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly blurred, giving them a soft, three-dimensional effect. A registered trademark symbol is located in the upper right corner of the word.
The predicate device. XBF-UC180F-DT8. has been name changed after 510(k) clearance. It is marketed as BF-UC180F.
# 4 DEVICE DESCRIPTIONN
# EVIS EUS BRONCHOVIDEOFIBERSCOPE OLYMPUS BF-UC190F
## 1) General Description of the subject device
The BF-UC190F has been designed to be used with the video system center, light source, documentation equipment, display monitor, and Endo Therapy accessories such as an aspiration biopsy needle. The BF-UC190F is intended for endoscopic real-time ultrasound imaging for performing endoscopic ultrasound guided needle aspiration within the airways, tracheobronchial tree, and esophagus.
The BF-UC190F is a flexible video endoscope and a modification of the XBF-UC180F-DT8 which was previously cleared under K070983.
### 2) Principle of Operation
The BF-UC190F consists of three parts: the control section, the insertion section, and the connector section.
#### The control section
The UP/DOWN angulation control lever on the control section is connected to the tip of the bending section by a series of wires. By operating the UP/DOWN angulation control lever, the bending section at the distal end bends vertically to guide the distal end for insertion and observation within the airways, tracheobronchial tree, and esophagus.
The endoscope contains a cylinder to attach a suction valve for suction. Depressing the suction valve will allow the doctor to use the endoscope to suction any fluids which are obscuring a good view of the tissue. Therapeutic instruments can be passed through the instrument channel for performing endoscopic biopsy and other therapies.
#### The insertion section
The insertion section has main parts including the image guide the light guide, the instrument channel, the balloon channel and the ultrasound transducer.
The light guides bring light from the light source through the endoscope.
The instrument channel outlet (also the suction opening) can be pushed in and out for withdrawing and inserting the therapeutic tools.
The balloon channel feed water to the balloon.
The ultrasound transducer emits and receives the ultrasound wave.
#### The connector section
The connector section connects the endoscope with the light source (CLV-190).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly blurred, giving the image a soft focus. The word appears to be the logo for the company Olympus.
This connector contains the ultrasound cable connector to attach the ultrasound cable for connecting the ultrasound system.
The BF-UC190F is equipped with the CCD unit in the control section, as well as the fiber bundle in the insertion tube. Images are transmitted through the fiber bundle, and the CCD unit in the control section changes them to video signals. This feature is identical to the predicate models (XBF-UC180F-DT8; K070983).
Ultrasound Operation
Operating controls with ultrasound systems
The primary interface between the device and the user is the console's keyboard. The software controlling the keyboard has been designed as the user cannot set the machine to over-ride any of the internally set controls (such as maximum acoustic power).
In any mode, the keyboard menu will only accept valid inputs. Keys and/or commands are not available in a particular mode.
A trackball or a trackpad are provided for calipers, other graphic and imaging adjustments/ controls.
### US CONNECTOR CAP MAJ-2295
This is US connector cap. This is attached to the ultrasound cable connector on the endoscope to protect the ultrasound cable connector and the endoscope from water penetration during reprocessing.
#### 5 INDICATIONS FOR USE
This instrument has been designed to be used with diagnostic ultrasound system, video system center, light source, documentation equipment, display monitor, Endo Therapy accessories such as an aspiration biopsy needle.
This instrument is designed for endoscopic real-time ultrasound imaging, for performing endoscopic ultrasound guided needle aspiration within the airways, tracheobronchial tree, and esophagus.
# 6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE
The BF-UC190F has the same technological characteristics and design as the predicate device except for the following new features:
- Outer diameter of the insertion portion a.
- b. Outer Diameter of Distal End
- Direction of view. C.
- d. Angulations
- Electrical connector (transmit CCD image signal) e.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are slightly blurred, giving the impression of depth. A small, circular symbol is located to the upper right of the word.
- f. More sterilization methods become applicable for endoscope.
- g. Irrigation port angle
- h. Outlet shape of the balloon channel
- i. The compatible cleaning brush is changed from reusable BW-7B and MH-507 to single-use BW-400B and BW-411B
- j. Clarification of cleaning procedure and the brush to be used for Balloon groove
- k. The shape of transducer
Validation from non-clinical testing demonstrated as defined in the performance data below show that these technological features do not raise any new issues of safety or effectiveness of the subject device.
All other technological characteristics of both the subject and predicate devices are identical.
| Item | Subject Device:<br>BF-UC190F | Predicate Device:<br>XBF-UC180F-DT8, K070983 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General information | | |
| Indication for Use | This instrument has been designed to<br>be used with diagnostic ultrasound<br>system, video system center, light<br>source, documentation equipment,<br>display monitor, Endo Therapy<br>accessories such as aspiration biopsy<br>needle.<br>This instrument is designed for<br>endoscopic real-time ultrasound<br>imaging, for performing endoscopic<br>ultrasound guided needle aspiration<br>within the airways, tracheobronchial<br>tree, and esophagus. | This instrument has been designed to<br>be used with a diagnostic ultrasound<br>system (ALOKA CO., LTD), video<br>system center, light source,<br>documentation equipment, video<br>monitor, endo-therapy accessories and<br>other ancillary equipment.<br>This instrument is designed for<br>endoscopic real-time ultrasonic image,<br>ultrasound guided needle aspiration<br>and other endoscopic procedures<br>within the airways, tracheobronchial<br>tree, esophagus and surrounding<br>organs. |
| Clinical Application | Used with EU-ME1:<br>- Transesophageal (non-cardiac)<br>- Intraluminal ultrasound for upper<br>airways and tracheobronchial tree<br>Used with EU-ME2 / EU-ME2<br>PREMIR PLUS:<br>- Transesophageal (non-cardiac)<br>- Intraluminal ultrasound for upper<br>airways and tracheobronchial tree | Used with SSD-Alpha5/10:<br>- Transesophageal (non-cardiac)<br>- Airways and tracheobronchial tree |
| Regulation name | Ultrasonic endoscope, its accessories<br>and ancillary equipment | Ultrasonic Endoscope, its accessories<br>and ancillary equipment |
| Regulation number | 892.1550 | 892.1550 |
| Product code | PSV (ultrasound bronchoscope)<br>ITX (transducer, ultrasonic, diagnostic) | PSV (ultrasound bronchoscope)<br>IYN (system, imaging, pulsed doppler,<br>ultrasonic)<br>ITX (transducer, ultrasonic, diagnostic) |
| Ultrasound specifications | | |
Table 1.1.5-2 Comparison with Predicate Device
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and appears to be a custom typeface. A thin, horizontal, yellow line is located beneath the word. The background is white.
# ULTRASOUND BRONCHOFIBERVIDEOSCOPE BF-UC190F
| Nominal<br>Ultrasonic<br>Frequencies Of<br>The Transducer<br>Assembly | 7-10MHz | 7-10MHz |
|-----------------------------------------------------------------------|---------|---------|
|-----------------------------------------------------------------------|---------|---------|
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are slightly blurred. There is a thin, curved, yellow line underneath the word.
| Item | Subject Device:<br>BF-UC190F | Predicate Device:<br>XBF-UC180F-DT8, K070983 |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mode | Used with EU-ME1:<br>B, Color doppler | Used with SSD-Alpha5/10:<br>B, M, PWD, Color doppler, Amplitude<br>Doppler, Combined |
| | Used with EU-ME2 / EU-ME2<br>PREMIR PLUS: | |
| | B, PWD, Color Doppler, Combined,<br>Harmonic imaging | |
| Applicable Ultrasound<br>system | EU-ME1<br>EU-ME2<br>EU-ME2 Premier Plus | ALOKA SSD-Alpha5 Diagnostic<br>ultrasound system<br>ALOKA SSD-Alpha10 Diagnostic<br>ultrasound system |
| Scanning Direction | Parallel to the Axis of<br>the Insertion Tube | Parallel to the Axis of<br>the Insertion Tube |
| Scanning Field of View | Used with EU-ME1: 60°<br>Used with EU-ME2 / EU-ME2<br>PREMIR PLUS: 65° | 60° |
| Scanning Method | Electrical curved linear array<br>Scanning | Electrical curved linear array Scanning |
| Contact Method | Balloon Method<br>Direct contact method | Balloon Method<br>Direct contact method |
| Endoscope specifications | | |
| Field of View | 80 ° | 80 ° |
| Depth of Field | 2 - 50mm | 2 - 50mm |
| Direction of View | 20° (forward-oblique) | 35° (forward-oblique) |
| Outer Diameter of<br>Distal End | φ6.6mm | φ6.9mm |
| Outer Diameter of<br>Insertion Tube | φ6.3mm | φ6.2mm |
| Angulation<br>UP/DOWN | 160°/70° | 120°/90° |
| Working Length | 600mm | 600mm |
| Instrument Channel | φ2.2mm | φ2.2mm |
| Combination use with<br>Electrosurgical<br>instruments | Not applicable | Not applicable |
| NBI observation | Not available | Not available |
| Scope connector | Electrical connector is integrated to<br>scope connector. Ultrasound connecter<br>is separated. Only when combined with<br>the water-resistant cap, ultrasound<br>connector is water resistant. | Electrical connector (transmit CCD<br>image signal) and ultrasound connecter is<br>separated. When combined with the<br>water-resistant cap, both connecters are<br>water resistant. |
| Protect water penetration<br>during reprocessing. | US connector cap (MAJ-2295) is<br>attached to the ultrasound cable<br>connector on the endoscope to protect<br>the ultrasound cable connector and the<br>endoscope from water penetration<br>during reprocessing. | Water resistant cap (MH-553) is attached<br>to the ultrasound cable connector and the<br>videoscope cable connector on the<br>endoscope to protect the both connectors<br>and the endoscope from water<br>penetration during reprocessing. |
| Sterilization methods for<br>reprocessing | -Ethylene oxide gas<br>-H2O2 (V-PRO maX) | -Ethylene oxide gas |
| Item | Subject Device:<br>BF-UC190F | Predicate Device:<br>XBF-UC180F-DT8, K070983 |
| Compatible Olympus<br>reprocessor for cleaning<br>and disinfection | Not available | Not available |
| High frequency<br>accessories | Not available | Not available |
| RFID tag for<br>communication with<br>endoscope reprocessor | Available | Available |
| Individual scope<br>information<br>(Scope ID) | Available | Available |
| Electric safety | Comply to IEC 60601-1,<br>IEC 60601-2-18 | Comply to IEC 60601-1,<br>IEC 60601-2-18 |
| EMC | Comply to IEC 60601-1 | Comply to IEC 60601-1 |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. There is a registered trademark symbol to the right of the word. The background is white.
# 7 PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### 1) Ultrasound output display
The Output Display for the BF-UC190F were conducted in accordance with the FDA Guidance Document, "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." The subject device conform to the IEC 60601-2-37 Edition 2.0 2007.
#### 2) Reprocessing validation testing
Reprocessing instruction and reprocessing method validation testing for the BF-UC190F were conducted and documentation was provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling."
#### 3) Biocompatibility testing
Biocompatibility testing for the BF-UC190F were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
#### 4) Software verification and validation testing
Software verification and validation testing for the BF-UC190F were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. The word is slightly blurred, suggesting that the image may have been taken out of focus or that the original image was of low resolution. There is a registered trademark symbol to the right of the word.
#### 5) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the BF-UC190F. The system complies with the ANSI/AAMI ES60601-1:2005/A2:2012/(R)2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2007 standards for EMC.
#### 6) Performance testing - Bench
Bench testing for the BF-UC190F as listed below was conducted to ensure that the subject device performs as intended and meet design specifications. Device performance was assessed the design requirements, and included process verification, design verification, and design validation.
- Thermal safety test
- Mechanical durability test
#### 7) Risk analysis
Risk analysis for the BF-UC190F was conducted in accordance with established inhouse acceptance criteria based on ISO 14971:2007 and the human factors validation was conducted in accordance with the FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices." The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
# 8 CONCLUSIONS
Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the BF-UC190F raise no new issue of safety and effectiveness and are substantially equivalent to the predicate devices in terms of safety, effectiveness and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.