The Klinly Ultrasonic Tongue Cleaner is indicated to reduce plaque, food debris, and bacteria from the surface of the tongue, and also to help fight bad breath and promote oral hygiene.
Device Story
Battery-operated (2 AAA) tongue cleaner; used by patients at home to supplement oral hygiene. Device features molded plastic scraping head; operates via manual scraping action combined with 233 Hz tactile vibration and 1.6 MHz ultrasound pressure waves generated by internal piezoelectric transducer. Patient applies pressure to handle while scraping tongue surface. Output is mechanical removal of debris; ultrasound/vibration intended to assist cleaning. Healthcare providers do not operate device; output is self-observed by patient via reduction in debris/breath malodor. Benefits include improved oral hygiene and reduced bacterial load.
Clinical Evidence
Prospective, non-randomized, blinded, single-center study (N=19, ages 23-49). Evaluated breath malodor, tongue plaque, and bacterial counts over 30 days. Results: 37% reduction in breath malodor (p=0.0002), 41% reduction in tongue plaque (p<0.0001), and 75% reduction in tongue bacterial count (p=0.0401). No adverse events reported.
Technological Characteristics
Molded ABS plastic scraping head; piezoelectric transducer for 1.6 MHz ultrasound; 233 Hz tactile vibration motor. Powered by 2x AAA (3.0 VDC) alkaline batteries. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Software level of concern: Moderate.
Indications for Use
Indicated for adults and adolescents 12 years and older to reduce plaque, food debris, and bacteria from the tongue surface, fight bad breath, and promote oral hygiene.
Regulatory Classification
Identification
A powered toothbrush is an AC-powered or battery-powered device that consists of a handle containing a motor that provides mechanical movement to a brush intended to be applied to the teeth. The device is intended to remove adherent plaque and food debris from the teeth to reduce tooth decay.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
September 6, 2019
KEA Products Ltd. % Robert Bock Official Correspondent Robert T. Bock Consultancy LLC 66 Drovers Lane Brewster, New York 10509
Re: K181986
Trade/Device Name: Klinly Ultrasonic Tongue Cleaner Regulation Number: 21 CFR 872.6865 Regulation Name: Powered Toothbrush Regulatory Class: Class I Product Code: QIA Dated: August 7, 2019 Received: August 8, 2019
Dear Robert Bock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael Adjodha Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181986
Device Name
Klinly Ultrasonic Tongue Cleaner
#### Indications for Use (Describe)
The Klinly Ultrasonic Tongue Cleaner is indicated to reduce plaque, food debris, and bacteria from the surface of the tongue, and also to help fight bad breath and promote oral hygiene.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K181986
#### 510(k) SUMMARY
| Applicant: | KEA Products Limited<br>10C, Meyer Ind. Bldg.<br>2 Chong Yip Street<br>Kwun Tong, Kowloon, Hong Kong SAR |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Application Correspondent: | Robert T. Bock<br>Official Correspondent<br>KEA Products Limited<br>66 Drovers Lane<br>Brewster, NY 10509<br>Telephone: 845.729.3176<br>Email: smilex@comcast.net |
| Date of summary prepared: | September 6, 2019 |
| Trade Name: | Klinly Ultrasonic Tongue Cleaner |
| Regulation Description: | Powered toothbrush |
| Regulation Number: | 21 CFR 872.6865 |
| Product code: | QIA |
#### Primary Predicate:
- . Cybersonic (K980075) American Dentronics, Incorporated.
#### Reference Devices:
- . Unik Tongue Cleaner (K983683) Unik Products, Inc.
- . Sonex Ultrasonic Tooth Brush (K913724) Sonex International Corporation.
- . Sonicare Advance Toothbrush, Model 4900 (K040416) Philips Oral Healthcare, Inc.
#### DEVICE DESCRIPTION:
The Klinly Ultrasonic Tongue Cleaner is a battery-operated device intended to be used to supplement daily oral hygiene care. It operates on two (2) AAA Alkaline batteries. All user-contacting components of the Klinly Ultrasonic Tongue Cleaner are constructed of injection-molded plastics which have demonstrated substantially equivalent biocompatibility according to ISO 10993-1.
The Klinly Ultrasound Tongue Cleaner device is comprised of three principles of operation to supplement daily oral hygiene by tongue cleaning. The device has a tongue scraping head, which:
- (1) Operates manually as a standard manual tonque scraper
- (2) Produces a low-frequency tactile vibration at 233 Hz, and
- (3) Emits ultrasound pressure waves at 1.6 MHz frequency.
#### INDICATIONS FOR USE:
The Kinly Ultrasonic Tongue Cleaner is indicated to reduce plaque, food debris, and bacteria from the surface of the tongue, and also to help fight bad breath and promote oral hygiene.
{4}------------------------------------------------
# COMPARISON OF INDICATIONS FOR USE:
| SUBJECT DEVICE | PRIMARY PREDICATE | REFERENCE DEVICE |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Klinly Ultrasonic Tongue Cleaner | Cybersonic (K980075) | Unik Tongue Cleaner (K983683) |
| Indications for use | Indications for use | Indications for Use |
| The Kinly Ultrasonic Tongue<br>Cleaner is indicated to o reduce<br>plaque, food debris, and bacteria<br>from the surface of the tongue,<br>and to help fight bad breath and<br>promote oral hygiene. | Use to remove surface tongue<br>debris, residue, plaque or filmy<br>build-up that results from eating,<br>drinking, smoking or other types of<br>consumption which can lead to<br>bad breath. Also may be used to<br>remove salivary tongue plaques<br>and residue that naturally<br>accumulate during sleep or<br>between cleanings. | To remove bacteria from and<br>prevent plaque build-up on the<br>tongue to help fight bad breath<br>and promote oral hygiene. |
The subject device Indications for Use includes an indication to reduce bacteria from the surface of the tongue, which is not found in the Primary Predicate Indications for Use. Additional differences in wording do not impact the substantial equivalence.
The Reference Device Unik Tongue Cleaner (K983683), is identified for this difference, as it is a manual tongue scraper indicated to remove bacteria (from the surface of the tongue).
Further reference devices used for technological comparisons do not include any component-specific language that would raise any concern related to the substantial equivalence of the subject device.
| Technological<br>Characteristics | SUBJECT DEVICE | PRIMARY PREDICATE |
|----------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Scraping edge | Klinly Ultrasonic Tongue Cleaner<br>Molded plastic (ABS) curved<br>scraping edge, rounded | Cybersonic (K980075)<br>Molded plastic flat scraping edge |
| Scraping action | Manually operated scraping action<br>(patient applies pressure as they<br>are scraping to remove debris) | Same as Subject Device |
| Target Population | Adults and adolescents 12 years<br>and older | Same as Subject Device |
| Anatomical sites | Surface of the tongue | Same as Subject Device |
| Intended use | reducing tongue debris | Same as Subject Device |
| Tactile vibration | Low-frequency (233 Hz) | None |
| Ultrasound pressure<br>waves | 1.6 Mz from piezoelectric<br>transducer in the device | None |
| Energy source | two AAA (3.0 VDC) alkaline<br>batteries | None |
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS :
The subject device is similar to the primary predicate devices are intended as a manual tongue scraper. Both devices are made from molded plastic and intended to be used on the surface of the tongue, operated manually by the patient applies pressure to the device handle while scraping to remove debris). The subject device is different as it is powered by two AA batteries, which power and control the device, piezoelectric transducer for the low-frequency tactile vibrations, and electronics which generate the 1.6 MHz ultrasound pressure waves.
The reference device Sonicare Advanced Toothbrush (K040416) is different than the subject device as it is a powered toothbrush; however, it is also powered and generates low-frequency tactile vibrations (261 Hz), similar to the subject device.
{5}------------------------------------------------
The reference device Sonex Ultrasonic Toothbrush (K913724) is different than the subject device as it is a powered toothbrush; however, it is also similarly powered and generates the same ultrasound pressure waves as the subject device (at 1.6 Mhz).
Any small differences in the technological characteristics between the subject, primary predicate, and reference devices have been addressed through non-clinical performance testing and do not impact the substantial equivalence of the subject device.
# NON-CLINICAL TESTING:
The Klinly Ultrasonic Tongue Cleaner has conducted the following testing to demonstrates substantial equivalence to the predicates according to the following:
Biocompatibility: Assessment per FDA Guidance - Biological Evaluation of medical devices – ISO 10993-1 including:
- . Cytotoxicity Test according to ISO 10993-5: 2009, Biological evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity
- . Sensitization Test according to ISO 10993-10:2010 Part 10: Test for irritation and skin sensitization.
- . Irritation Test according to ISO 10993-10:2010 Part 10: Test for irritation and skin sensitization.
Electrical Safety: Requirements for medical electrical equipment used in the home - IEC 60601-1
EMC: Requirements for Electromagnetic Compatibility of Medical Equipment – IEC 60601-1-2
Software: Moderate Level of Concern - Documents and testing provided according to FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Performance Bench Testing: Bench testing to validate the technical specifications and functionality of the device, including the vibratory and ultrasonic frequencies as well as duty cycles.
Shelf Life: Durability testing to demonstrate that the subject device maintains acceptable performance throughout the use life proposed in the labeling.
#### CLINICAL DATA:
The following data was submitted to demonstrate substantial equivalence in labeling:
A prospective 30 days, non-randomized, blinded, single-center clinical study was performed. The study had nineteen (N=19) male and female patients, ranging in age from 23 to 49 years, who all completed the study using the Klinly Ultrasonic Tongue Cleaner.
The study evaluated the changes in Breath Malodor, Tongue Plaque, and Tongue Bacteria Counts between the baseline and the 30 days endpoint of the study, demonstrating:
- . Breath Malodor reduction of 37% at p=0.0002
- . Tonque Plaque reduction of 41% at p<0.0001
- Tonque Bacterial Count reduction of 75% at p=0.0401
There were no protocol deviations and no adverse events.
#### CONCLUSION:
Based on the the information presented above, the Klinly Ultrasonic Tongue Cleaner demonstrates substantial equivalence to the Primary Predicate and Reference Devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.