BruxZir® Radiant

K240574 · Prismatik Dentalcraft, Inc. · EIH · Mar 20, 2024 · Dental

Device Facts

Record IDK240574
Device NameBruxZir® Radiant
ApplicantPrismatik Dentalcraft, Inc.
Product CodeEIH · Dental
Decision DateMar 20, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.6660
Device ClassClass 2
AttributesTherapeutic

Indications for Use

BruxZir® Radiant is intended to be used in fabrication of dental restorations for the purpose of restoring chewing function. The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior locations. This device is contraindicated for dental restorations greater than 3-units in length.

Device Story

BruxZir® Radiant consists of yttria-stabilized zirconia milling blanks used by dental technicians to fabricate custom all-ceramic dental restorations. The device is processed via CAD/CAM systems for design and milling, followed by high-temperature sintering. Final restorations are used to restore chewing function in anterior and posterior regions. The device is available in multiple shades (Brite, White, 100, 200, 300, 400) and can be adjusted with staining/glazing techniques. The output is a finished dental restoration that provides esthetic and functional replacement for missing or damaged tooth structure. The device is intended for professional use in dental laboratories.

Clinical Evidence

No clinical data included. Bench testing only. Verification activities confirmed compliance with ISO 6872:2015/Amd 1:2018 for flexural strength (>500 MPa), CTE (10-11 x 10^-6/°C), chemical solubility (<100 µg/cm²), and radioactivity (<1.0 Bq/g). Biocompatibility verified per ISO 10993-1:2018 and ISO 10993-5:2009. Translucency and color consistency (ΔE < 2.00) also met predetermined acceptance criteria.

Technological Characteristics

Yttria-stabilized zirconia milling blanks. Manufacturing: Colloidal process. Energy: High-temperature sintering. Dimensions: Various sizes/thicknesses. Connectivity: None (standalone material). Sterilization: Non-sterile. Standards: ISO 6872:2015/Amd 1:2018 (Type II, Class 4), ISO 10993-1:2018, ISO 10993-5:2009.

Indications for Use

Indicated for use by dental technicians to construct custom all-ceramic restorations for anterior and posterior locations. Contraindicated for restorations >3 units in length.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 20, 2024 Prismatik Dentalcraft, Inc. Su Hyun Park Regulatory Affairs Manager 2144 Michelson Drive Irvine, California 92612 Re: K240574 Trade/Device Name: BruxZir® Radiant Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: Class II Product Code: EIH Dated: February 19, 2024 Received: March 1, 2024 Dear Su Hyun Park: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Michael E. Adjodha -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Submission Number (if known) K240574 Device Name BruxZir® Radiant Indications for Use (Describe) The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location. Contraindication: This device is contraindicated for dental restorations greater than 3-units in length. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. On the left side of the logo is a triangle with different colors. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." The text is in a sans-serif font and is blue. #### 510(k) Summary Date Prepared: February 29, 2024 ## CONTACT DETAILS (21 CFR 807.92(a)(1)) Applicant Name: Prismatik Dentalcraft, Inc. Applicant Address: 2144 Michelson Drive, Irvine, CA 92612, USA Applicant Contact Telephone: 949-863-5479 Applicant Contact: So Hyun Park, Regulatory Affairs Manager, MS Applicant Contact Email: so.park(@glidewelldental.com ## DEVICE NAME (21 CFR 807.92(a)(2)) Device Trade Name: BruxZir® Radiant Common Name: Zirconia milling block or dental CAD/CAM block Classification Name: Porcelain powder for clinical use Regulation Number: 872.6660 Regulatory Class: Class II Product Code: EIH #### LEGALLY MARKETED PREDICATE DEVICE (21 CFR 807.92(a)(3)) Predicate 510(k) #: K150872 Predicate Trade Name: BruxZir® Anterior Product Code: EIH #### DEVICE DESCRIPTION SUMMARY (21 CFR 807.92(a)(4)) BruxZir® Radiant are zirconia milling blanks used to produce highly esthetic zirconia dental restorations. The high esthetics and high strength of BruxZir® Radiant restorations make them viable for use in all regions of the mouth. The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. BruxZir® Radiant Blanks are available in 6 shades, Brite, White, 100, 200, 300, 400, that can be easily adjusted to match the final shade. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into sections of different colors, including pink, blue, and yellow. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line. ## INTENDED USE/INDICATIONS FOR USE (21 CFR 807.92(a)(5)) BruxZir® Radiant is intended to be used in fabrication of dental restorations for the purpose of restoring chewing function. The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior locations. This device is contraindicated for dental restorations greater than 3-units in length. #### INDICATIONS FOR USE COMPARISON (21 CFR 807.92(a)(5)) The subject device, BruxZir® Radiant, has the same indications for use as the predicate device, BruxZir® Anterior (K150872). Both devices are indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior locations. Both the subject device, BruxZir® Radiant, and predicate device, BruxZir® Anterior (K150872), are also contraindicated for dental restorations greater than 3-units in length. #### TECHNOLOGICAL COMPARISON (21 CFR 807.92(a)(6)) The subject device, BruxZir® Radiant, is substantially equivalent in technical characteristics to the predicate device, BruxZir® Anterior (K150872) in terms of overall design principles, material, and performance. The subject device, BruxZir® Radiant, is similar in terms of overall design to the predicate device, BruxZir® Anterior (K150872). The subject device, BruxZir® Radiant, is only offered in one style (Zeno) milling blank whereas the predicate device, BruxZir® Anterior (K150872), is offered in various designs. There is no change in packaging for a milling blank. The fundamental principle of operation of the subject device, BruxZir® Radiant, and the predicate device, BruxZir® Anterior (K150872), is the same. The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. Although no color needs to be added additionally, pre-coloring before sintering and/or basic staining and glazing techniques may be used after sintering to achieve the desired shade. Also, the change in the manufacturing process from the press process to the colloidal process does not affect the mechanical properties of the final finished device. The subject device, BruxZir® Radiant, is similar in terms of material composition to the predicate device, BruxZir® Anterior (K150872). The main ceramic component is composed of yttria-stabilized zirconia with varying trace amount of colorants. As the generic type of material has not changed, it does not alter the performance specifications of the subject device. The slight differences in overall chemical composition do not affect the biocompatibility and performance characteristics. The technological comparison table below outlines and provides the similarities between the subject device, BruxZir® Radiant, and the predicate device, BruxZir® Anterior (K150872). Both the subject device and the predicate device have similar physical/mechanical and biocompatibility properties that met the requirements of ISO 6872:2015/Amd 1:2018 (Type {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line. II, Class 4) and ISO 10993. Any differences between the subject device and the predicate device do not raise any new concerns of safety and effectiveness. | Attributes | | Subject Device<br>(TBD) | Predicate Device<br>(K150872) | Comparison | |----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | Device Name<br>Manufacturer | | BruxZir® Radiant<br>Prismatik Dentalcraft, Inc. | BruxZir® Anterior<br>Prismatik Dentalcraft, Inc. | N/A<br>Same | | Product Code | | EIH | EIH | Same | | Prescription Device | | Yes | Yes | Same | | Indications for Use | | The device is indicated<br>for use by dental<br>technicians in the<br>construction of custom<br>made all ceramic<br>restorations for anterior<br>and posterior location. | The device is indicated<br>for use by dental<br>technicians in the<br>construction of custom<br>made all ceramic<br>restorations for anterior<br>and posterior location. | Same | | Contraindications | | This device is<br>contraindicated for<br>dental restorations<br>greater than 3-units in<br>length. | This device is<br>contraindicated for<br>dental restorations<br>greater than 3-units in<br>length. | Same | | Design<br>Characteristic | Chemical<br>Composition | The main ceramic<br>component is<br>composed of yttria-<br>stabilized zirconia. The<br>different shades have<br>varying trace colorants<br>as required to match<br>the desired shades. | The main ceramic<br>component is<br>composed of yttria-<br>stabilized zirconia.<br>The different shades<br>have varying trace<br>colorants as required<br>to match the desired<br>shades. | Similar; Different<br>yttria proportion of the<br>main ceramic<br>components and<br>different oxide<br>colorants for the<br>subject device. | | | Design | Various sizes and<br>thicknesses | Various sizes and<br>thicknesses | Similar; Different<br>sizes and thicknesses<br>are offered for the<br>subject device. | | | Principle of<br>Operation | The manufactured<br>dental restorations are<br>made utilizing a<br>CAD/CAM system for<br>design and<br>manufacture. The<br>designed and<br>manufactured dental | The manufactured<br>dental restorations are<br>made utilizing a<br>CAD/CAM system for<br>design and<br>manufacture. The<br>designed and<br>manufactured dental | Same | | Attributes | Subject Device<br>(TBD) | Predicate Device<br>(K150872) | Comparison | | | Shades | sintered at a high<br>temperature. The final<br>target shades can be<br>easily adjusted to<br>match the final shade.<br>Brite, White, 100, 200,<br>300, 400 | sintered at a high<br>temperature. The final<br>target shades can be<br>easily adjusted to<br>match the final shade.<br>White, 150, 250, 350,<br>450, 550 | Different; different<br>shades are offered for<br>the subject device. | | | Final Vita<br>Shades | A1, A2, A3, A3.5, A4,<br>B1, B2, B3, B4, C1,<br>C2, C3, C4, D2, D3,<br>D4 | A1, A2, A3, A3.5, A4,<br>B1, B2, B3, B4, C1,<br>C2, C3, C4, D2, D3,<br>D4 | Same | | | Flexural Strength | >500 MPa | >500 MPa | Same | | | Coefficient of<br>Thermal<br>Expansion<br>(CTE/25-500°C) | 10-11x10-6/°C | 10-11x10-6/°C | Same | | | Chemical<br>Solubility | <100 µg/cm² | <100 µg/cm² | Same | | | Radioactivity | The activity<br>concentration of<br>Uranium-238 is no<br>more than 1.0 Bq/g. | The activity<br>concentration of<br>Uranium-238 is no<br>more than 1.0 Bq/g. | Same | | | Biocompatibility | Biocompatible and<br>non-toxic | Biocompatible and<br>non-toxic | Same | | | Sterility | Non-sterile | Non-sterile | Same | | | Manufacturing<br>Process | Colloidal Process | Biaxial Pressing<br>Process | Different<br>manufacturing process | | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line. ## NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS 21 CFR 807.92(b)) Design verification according to BruxZir® Radiant Design Verification Protocol was conducted by Prismatik Dentalcraft, Inc. The tests conducted are listed below: - . Flexural Strength according to ISO 6872:2015/Amd 1:2018 - Translucency - Visual Shade Evaluation ● - Color Consistency - Coefficient of Thermal Expansion according to ISO 6872:2015/Amd 1:2018 ● - Chemical Solubility according to ISO 6872:2015/Amd 1:2018 - Radioactivity according to ISO 6872:2015/Amd 1:2018 ● - . Biocompatibility according to ISO 10993-1:2018 and ISO 10993-5:2009 {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle with a color gradient on the left and the company name on the right. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line. No clinical data is included in this submission. The test samples were made using the standard manufacturing process and QC process for BruxZir® Radiant to ensure that the test samples meet the following criteria for the finished device: - Flexural Strength must be greater than 500 MPa after sintering, as measured per ISO . 6872:2015/Amd 1:2018, Type II Class 4. - . Translucency of each formulation must be higher than the values shown in the table below at 700nm. | Shade | BruxZir® Radiant % Translucency at 700nm | |-------|------------------------------------------| | White | >59 | | Brite | >65 | | 100 | >57 | | 200 | >57 | | 300 | >57 | | 400 | >57 | - Grade of "Pass" for Visual Shade Match for all shades. ● - Color consistency ΔE calculation needs to be less than 2.00 between samples measured. - The Coefficient of Thermal Expansion must be between 10-11 x 10-9/℃ and the standard deviation of CTE must be no greater than 0.5 x 10-9℃ per ISO 6872:2015/Amd 1:2018. - The solubility must be lower than 100 µg/cm² per ISO 6872:2015/Amd 1:2018. ● - The activity concentration of Uranium-238 must not exceed 1.0 Bq'g-1 per ISO ● 6872:2015/Amd 1:2018. - BruxZir® Radiant must be biocompatible, safe, and non-toxic. The results of the verification activities confirmed that the design of the subject device met all the predetermined criteria according to the design specifications. The results of the testing were used to address questions related to substantial equivalence based on differences in technological features between the subject device. BruxZir® Radiant, and the predicate device, BruxZir® Anterior (K150872), by verifying that the modified device met all the design specification requirements, and the design changes were implemented successfully without causing any unexpected issues.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...