Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank
K221242 · Hunan Vsmile Biotechnology Co., Ltd. · EIH · Jul 29, 2022 · Dental
Device Facts
| Record ID | K221242 |
| Device Name | Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank |
| Applicant | Hunan Vsmile Biotechnology Co., Ltd. |
| Product Code | EIH · Dental |
| Decision Date | Jul 29, 2022 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.6660 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.
Device Story
Device consists of zirconia powder blanks (regular or pre-shaded with inorganic pigments) used as raw material for dental restorations. Dental professionals or laboratories mill blanks into final shapes (crowns, bridges, veneers, inlay/onlay) using CAD/CAM or manual milling machines. Blanks are sintered at temperatures >1450°C to achieve final ceramic properties. Output is a custom-fitted prosthetic dental device. Benefits include restoration of tooth function and aesthetics. Device is non-sterile and intended for single use.
Clinical Evidence
No clinical data. Bench testing only. Performance verified per ISO 6872:2015 (flexural strength, chemical solubility, density, fracture toughness). Biocompatibility verified per ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, and muscle implantation) with all tests passing.
Technological Characteristics
Materials: Yttria-stabilized tetragonal zirconia (Y-TZP) conforming to ISO 13356; inorganic pigments (Fe2O3, Er2O3, MnO). Form factor: Blocks and rods. Processing: CAD/CAM or manual milling followed by sintering >1450°C. Non-sterile. Performance conforms to ISO 6872.
Indications for Use
Indicated for dental restorations (crowns, bridges, veneers, inlay/onlay) fabricated by dental professionals or laboratories using CAD/CAM or manual milling machines.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Predicate Devices
- Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank (K141724)
Related Devices
- K221051 — Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank · Changsha Honchon Technology Co., Ltd. · Jul 8, 2022
- K141724 — UPCERA DENTAL ZIRCONIA BLANK & DENTAL ZIRCONIA PRE-SHADED BLANK · Liaoning Upcera Co., Ltd. · Oct 22, 2014
- K230003 — Topzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank · Topzir Biotech Co., Ltd. · Mar 31, 2023
- K240672 — Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank ( ZD-HT-PLUS, ZD-ST, ZD-SHT, ZD-HT-C, ZD-ST-C, ZD-SHT-C, ZD-UT-C, ZD-ST-ML, ZD-SHT-ML, ZD-UT-ML, ZD-3D-Pro-ML, ZD-4D-Pro-ML) · Zhejiang Zahndent Biotechnology Co., Ltd. · Jun 7, 2024
- K212259 — Zirdent Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank · Zirdent New Material Co., Ltd. · Nov 12, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2022
Hunan Vsmile Biotechnology Co., Ltd % Shanfeng Jiang Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 304, Building A, No. 62 Nanyun 2nd Road, Science Town Huangpu District, Guangzhou City, Guangdong 510663 China
Re: K221242
Trade/Device Name: Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: May 2, 2022 Received: May 2, 2022
Dear Shanfeng Jiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K221242
#### Device Name
Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank
Indications for Use (Describe)
Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------|------------------------------------------------------------|
|-------------------------------------------------------------|------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
## Section 5 - 510(k) Summary
#### K221242
Date of Summary Preparation: April 20, 2022 Date of Modification: July 19, 2022
### 1. Submitter's Identifications
Submitter's Name: Hunan Vsmile Biotechnology Co.,Ltd 501-5,501-6, Lujing High-tech Juxing Road No.8. Lugu Address: Zone,Changsha,Hunan,China Zip Code: 410000 Contact Person: Xiaomi Tang Contact Title: General Manager Contact E-mail Address: 2839189612@qq.com Tel: +86-731-85682958
## 2. Correspondent's Identifications
Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 304, Building A, No. 62 Nanyun 2nd Road, Science Town, Huangpu District, Guangzhou City, Guangdong, 510663, China ZIP Code: 510663 Contact Person: Shanfeng Jiang Contact Title: Regulation Control Manager Contact E-mail Address: jiang13620586569@126.com Tel: +86-20-82329549 Fax: +86-20-82329549
### 3. Name of the Device
Device Classification Name: Powder, Porcelain Regulation Description: Porcelain powder for clinical use Trade Name: Vsmilehappyzir Dental Zirconia Blank &Dental Zirconia Pre-Shaded Blank Model: HT-plus, ST, ST-C, ST-ML, SHT, SHT-C, SHT-ML, 3D-pro-ML, 3D-Prism-ML, UT, UT-C, UT-ML Regulation Medical Specialty: Dental Review Panel: Dental Product Code: EIH Regulation Number: 21 CFR 872.6660 Device Classification: Class II
### 4. The Predicate Devices
Primary Predicate device: K141724 Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank
{4}------------------------------------------------
Liaoning Upcera Company Limited
# 5. Device Description
Vsmilehappyzir Dental Zirconia Blank are derived from zirconia powder that has been processed into their final shapes. These blanks are then being further fabricated into all-ceramic restorations such as crowns, bridges, veneers, inlay/onlay. The zirconia powder is composed of ZrO2+ Y2O3+ HfO2+ Al2O3 with its composition conforms to ISO 13356, Implants for surgery - Ceramic materials based on yttriastabilized tetragonal zirconia (Y-TZP). The performance of the dental blanks conforms to ISO 6872, Dentistry - Ceramic materials.
Vsmilehappyzir Dental Zirconia Pre-Shaded Blank are derived from the same Zirconia powder as the regular Dental Zirconia Blank with the addition of very small amount of inorganic pigments before the composite material is processed into their final net shapes. These blanks are then being further fabricated into all-ceramic restorations such as crowns, bridges, veneers, inlay/onlay. The purpose of the inorganic pigments is to generate the color on the prosthetic dental devices, after sintering at dental labs, that matches natural color of patient's teeth. The performance of the dental blanks conforms to ISO 6872, Dentistry -- Ceramic materials.
# 6. Intended Use of Device
Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.
# 7. Summary of Substantial Equivalence
| | Table 1 Comparison to Predicate Device | | |
|--|----------------------------------------|--|--|
|--|----------------------------------------|--|--|
| | Proposed Device | Predicate device | Comparison |
|------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------------|
| 510k Number | K221242 | K141724 | ------ |
| Product Code | EIH | EIH | Same |
| Proprietary Name | Vsmilehappyzir<br>Dental Zirconia Blank & Dental<br>Zirconia Pre-Shaded Blank | Upcera Dental Zirconia Blank &<br>Dental Zirconia Pre-Shaded Blank | ------ |
| Model: | HT-plus, ST, ST-C,<br>ST-ML, SHT,<br>SHT-C, SHT-ML,<br>3D-pro-ML,<br>3D-Prism-ML, UT,<br>UT-C, UT-ML | ------ | ------ |
{5}------------------------------------------------
| Manufacturer | Hunan Vsmile Biotechnology Co.,Ltd | Liaoning Upcera Company Limited | ----- |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Indications for Use | Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia<br>Pre-Shaded Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals. | Upcera Dental Zirconia Blank & Dental Zirconia<br>Pre-Shaded Blank are used for dental restorations using different CAD/CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals. | Same |
| Basic design | Blocks, and rods | Blocks, disc, and rod | Same |
| Materials | Regular:<br>Zirconia $(ZrO2+ Y2O3+ HfO2+ Al2O3≥99.0%)$<br>Pre-shaded:<br>Zirconia $(ZrO2+ Y2O3+ HfO2+ Al2O3≥99.0%)$<br>Inorganic pigments $(Fe2O3, Er2O3 and MnO <2.0%)$ | Regular:<br>Zirconia $(ZrO2+ Y2O3+ HfO2+ Al2O3≥99.0%)$<br>Pre-Shaded:<br>Zirconia $(ZrO2+ Y2O3+ HfO2+ Al2O3≥98.0%)$<br>Inorganic pigments $(Fe2O3, Pr2O3, and Er2O3 <2.0%)$ | Similar¹ |
| Processing | Sintering at temperature: > 1450°C | Sintering at temperature: > 1500°C | Similar² |
| Dimension | Various | Various | Same |
| Single Use | Yes | Yes | Same |
| Color | None, and Pre-shaded (for pre-shaded series) | None, and Pre-shaded (for pre-shaded series) | Same |
| Sterile | Non-sterile | Non-sterile | Same |
{6}------------------------------------------------
The proposed device is essentially identical to the predicate devices in terms of indication for use, design between our device and the predicate devices. The minor differences are that as below:
Note 1: Zirconia (ZrO2+ Y2O3+ HfO2+ Al2O3) of proposed device which is not less than 99.0% is higher than that of the predicate device. The three inorganic pigments of the proposed device contain Fe2O3, Er2O3 and MnO, while those of the predicate device are Fe2O3, Pr2O3, and Er2O3. These are minor differences and the biocompatibility testing of the overall proposed device passed.
Note 2: Processing (Sintering at temperature) of the proposed device is slightly lower 5°C than that of the predicate device. This is minor difference and the performance testing of the proposed device passed.
# 8. Summary of Non-Clinical Testing
Bench testing was performed per ISO 6872:2015 and internal procedures to ensure that the Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank met its specifications. All tests were verified to meet acceptance criteria. Test results on radioactivity, flexural strength, chemical solubility, linear thermal expansion coefficient, freedom from extraneous materials, uniformity, shrinkage factor, pre-sintered density, sintered density, and fracture toughness of the subject device are very similar to predicate device. Biocompatibility testing was performed to verify the equivalent safety of the materials that are used.
According to ISO 10993-1:2018 and ISO ISO7405:2018, we evaluated and conducted the compatibility test for the proposed device. The following table shows the biocompatibility testing results.
| Item | Proposed device | Result |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Cytotoxicity<br>(ISO 10993-5:2009) | Under the conditions of this study, the test article<br>was non cytotoxic for 2 h and mildly cytotoxic for<br>24 h in the filter diffusion method.<br>Under the conditions of this study, the test article<br>was accepted in the agar diffusion method.<br>Under the conditions of this study, the test article<br>has no potential toxicity to L-929 cells. | Pass |
| Oral Mucosa Irritation<br>(ISO 10993-10:2010) | The test article showed no evidence of causing<br>oral mucosa irritation in the Syrian hamsters. | Pass |
| Delayed Dermal<br>Contact Sensitization<br>(ISO 10993-10:2010) | The test article showed no evidence of causing<br>delayed dermal contact sensitization in the guinea<br>pig. | Pass |
| Subacute Toxicity<br>(ISO 10993-11:2017) | The test result showed that the test article did not<br>induce subacute systemic toxicity in rats under<br>this condition . | Pass |
Table 2 Biocompatibility testing
{7}------------------------------------------------
Hunan Vsmile Biotechnology Co.,Ltd
| Subchronic Toxicity<br>(ISO 10993-11:2017) | There is no obvious histopathological difference in<br>test group and control group. The structure of each<br>argan in test group is normal, no abnormal<br>histopathological changes in the table above was<br>found. | Pass |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Acute Systemic<br>Toxicity<br>(ISO 10993-11:2017) | The test article showed no evidence of causing<br>acute system toxicity in the ICR mice. | Pass |
| In Vitro Mammalian<br>Cell Gene Mutation<br>(ISO 10993-3:2014) | Under the conditions of this study, the test article<br>is considered non-mutagenic. | Pass |
| Micronucleus<br>(ISO 10993-3:2014) | The test result showed that the extract of the test<br>article did not induce the potential genotoxicity<br>toxicity in ICR mice under this condition. | Pass |
| Bacterial Reverse<br>Mutation<br>(ISO 10993-3:2014) | Under the conditions of this study, the number of<br>reverting colonies in the test article group is not<br>equal to or greater that of the spontaneous control,<br>so the test article have no potential mutagenesis. | Pass |
| Muscle Implant<br>(ISO 10993-6:2016) | The test result showed that the test article did not<br>induce local effects after implantation of<br>biomaterials in rabbits under this condition. | Pass |
Note: Testing were Performed on pre-shaded zirconia blank (UT-C) to cover the regular and pre-shaded zirconia blanks.
# 9. Clinical Test Conclusion
No clinical study is included in this submission.
# 10. Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that subject device Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank performs as well as or better than the legally marketed predicate device K141724 Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank. Vsmilehappyzir Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank is substantial equivalent to the legally marketed predicate device K141724 Upcera Dental Zirconia Blank & Dental Zirconia Pre-Shaded Blank.