K234014 · Prismatik Dentalcraft, Inc. · EIH · Feb 27, 2024 · Dental
Device Facts
Record ID
K234014
Device Name
Obsidian® NOW
Applicant
Prismatik Dentalcraft, Inc.
Product Code
EIH · Dental
Decision Date
Feb 27, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Obsidian® NOW is used to fabricate ceramic dental prosthesis in the nature of crowns for posterior and anterior applications using CAD/CAM technology. The block is processed through dental laboratories or by dental professionals.
Device Story
Obsidian® NOW is a fully crystallized lithium disilicate glass-ceramic milling block; used to fabricate monolithic dental restorations (full contour crowns) via CAD/CAM technology. Device processed by dental laboratories or dental professionals; no additional crystallization required prior to placement. Input is anatomical rendering of patient teeth via CAD/CAM equipment; output is final dental restoration. Benefits include restoration of chewing function. Device used in dental clinic or laboratory settings; healthcare providers use output to restore patient dentition.
Clinical Evidence
Bench testing only. Performance verified per ISO 6872:2015/Amd 1:2018, including flexural strength (≥300 MPa), chemical solubility (<100 µg/cm²), radioactivity, and glass transition temperature. Biocompatibility assessed per ISO 10993-1. Packaging validated per ASTM D4169-22. No clinical data included.
Technological Characteristics
Lithium aluminosilicate glass matrix with lithium disilicate crystals, colorants, and opacifiers. Form factor: milling block. Mechanical properties: Type II, Class 3 ceramic per ISO 6872:2015/Amd 1:2018. Non-sterile. No CTE testing required as staining/glazing not needed.
Indications for Use
Indicated for fabrication of ceramic dental crowns for posterior and anterior applications in patients requiring dental restorations.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
K100781 — OBSIDIAN CERAMIC BLOCKS · Prismatik Dentalcraft, Inc. · Jun 21, 2010
K080479 — DENTAL CERAMIC · Dmg USA, Inc. · May 21, 2008
K211916 — IPS e.max One · Ivoclar Vivadent, AG · Aug 20, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 27, 2024
Prismatik Dentalcraft. Inc. Nina Chiang Regulatory Affairs Specialist 2144 Michelson Dr. Irvine, California 92612
Re: K234014
Trade/Device Name: Obsidian® NOW Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: December 19, 2023 Received: December 19, 2023
Dear Nina Chiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
Device Name
Obsidian® NOW
Indications for Use (Describe)
Obsidian® NOW is used to fabricate ceramic dental prosthesis in the nature of crowns for posterior and anterior applications using CAD/CAM technology. The block is processed through dental laboratories or by dental professionals.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a color gradient that starts with blue at the bottom and transitions to pink at the top. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC.", in blue sans-serif font.
# 510(k) Summary
# K234014
Date Prepared: February 20, 2024
#### I. Contact Details
Applicant Name: Prismatik Dentalcraft, Inc.
Applicant Address: 2144 Michelson Drive, Irvine, CA 92612 United States
Applicant Contact Telephone: 949-222-3516
Applicant Contact: Ms. Nina Chiang, Regulatory Affairs Specialist
Applicant Contact Email: nina.chiang@glidewelldental.com
Secondary Contact: So Hyun Park, Regulatory Affairs Manager
Secondary Contact Email: so.park@glidewelldental.com
Secondary Contact Phone: 949-863-5479
#### II. Device Name
Device Trade Name: Obsidian® NOW
Common Name: Glass Ceramic Milling Block or Dental CAD/CAM Block
Classification Name: Porcelain powder for clinical use
Regulation Number: 872.6660
Product Code: EIH
#### III. Legally Marketed Predicate Device
Predicate 510(k) #: K141788
Predicate Trade Name: Obsidian Milling Blocks
Product Code: EIH
#### IV. Device Description
Obsidian® NOW is a fully crystallized lithium disilicate glass-ceramic block that can be milled to produce dental restorations using CAD/CAM technology. The block is intended to be used for fabricating monolithic restorations such as full contour crowns without requiring any additional crystallization prior to placement. The milling block is offered in the commonly used VITA® Classical and Chromascop® Bleach shades.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left side of the logo is a triangular shape with a color gradient. The colors in the triangle range from pink to blue to green to yellow. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." The text is in a sans-serif font and is blue.
#### V. Intended Use/Indications for Use
Obsidian® NOW is used to fabricate ceramic dental prosthesis in the nature of crowns for posterior and anterior applications using CAD/CAM technology. The block is processed through dental laboratories or by dental professionals.
#### VI. Technological Comparison
The subject device, Obsidian® NOW, is substantially equivalent to the predicate device, Obsidian Milling Block (K141788) in intended use, material, design principles, and technological characteristics.
The subject device, Obsidian® NOW, has the same intended use as the predicate device, Obsidian Milling Block (K141788), as the material is used in fabrication of dental restorations such as crowns for the purpose of restoring function. The subject device, Obsidian® NOW, has the same indication for use except for the device trade name and the addition of a chairside indication for use where the blocks are processed through dental laboratories or by dental professionals.
The subject device, Obsidian® NOW, is substantially equivalent to the predicate device, Obsidian Milling Block (K141788) in technological characteristics. The same mechanical property testing in terms of flexural strength that was conducted on the predicate device, Obsidian Milling Block (K141788), was also conducted on the subject device, Obsidian® NOW, according to ISO 6872:2015/Amd 1:2018. The flexural strength of the subject device, Obsidian® NOW, met performance criteria in ISO 6872 for Type II, Class 3.
The subject device, Obsidian® NOW, and the predicate device, Obsidian Milling Block (K141788) are similar in material composition. Both devices utilize glass components as the base material, colorants, and opacifiers that make up the lithium aluminosilicate glass matrix. Despite the differences in the colorants and opacifiers to achieve the desired shades, the slight differences in chemical formulation do not affect the safety and effectiveness of the device as verified by the biocompatibility and performance testing.
The subject device, Obsidian® NOW, is substantially equivalent to the predicate device, Obsidian Milling Block (K141788) in terms of design. Both devices are offered in a block form in different shades and are used to make final dental restorations based on the anatomical rendering of patient's teeth using CAD/CAM equipment.
| Attributes | Subject Device<br>(K234014) | Predicate Device<br>(K141788) | Comparison | |
|-----------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Obsidian® NOW | Obsidian Milling<br>Block | N/A | |
| Manufacturer | Prismatik Dentalcraft,<br>Inc. | Prismatik Dentalcraft,<br>Inc. | Same | |
| | Attributes | Subject Device<br>(K234014) | Predicate Device<br>(K141788) | Comparison |
| | Product Code | EIH | EIH | Same |
| | Prescription Device | Yes | Yes | Same |
| | Intended Use | Obsidian® NOW is<br>intended to be used in<br>fabrication of dental<br>restorations for the<br>purpose of restoring<br>chewing function. | Obsidian Milling<br>Block is intended to be<br>used in fabrication of<br>dental restorations for<br>the purpose of<br>restoring chewing<br>function. | Same; except for<br>the device trade<br>name. |
| | Indications for Use | Obsidian® NOW is<br>used to fabricate<br>ceramic dental<br>prosthesis in the<br>nature of crowns for<br>posterior and anterior<br>applications using<br>CAD/CAM<br>technology. The block<br>is processed through<br>dental laboratories or<br>by dental<br>professionals. | The Obsidian Milling<br>Blocks are used to<br>fabricate ceramic<br>dental prostheses in<br>the nature of crowns<br>and bridges for<br>posterior and anterior<br>applications using<br>CAD/CAM methods. | Similar; except for<br>the device trade<br>name and the<br>subject device has<br>the addition of a<br>chairside<br>indication for use. |
| | Prescription Device | Yes | Yes | Same |
| Design<br>Characteristics | Material<br>Composition | The device is<br>composed of lithium<br>aluminosilicate glass<br>matrix embedded with<br>lithium disilicate<br>crystals, colorants, and<br>opacifiers to achieve<br>the desired shade. | The device is<br>composed of lithium<br>aluminosilicate glass<br>matrix embedded with<br>lithium silicate<br>crystals, colorants, and<br>opacifiers to achieve<br>the desired shade. | Similar; the<br>subject device is<br>made of lithium<br>disilicate glass,<br>fewer colorants,<br>and opacifiers as<br>opposed to the<br>predicate device. |
| | Design | Fully crystallized<br>lithium disilicate glass<br>ceramic available in<br>milling block form for<br>single-unit<br>restorations. | Partially crystallized<br>lithium silicate glass<br>ceramic available in<br>milling block form for<br>single-unit<br>restorations. | Similar; the<br>subject device<br>does not require<br>additional<br>crystallization<br>prior to placement. |
| | Shades | A1, A2, A3, A3.5, B1 | A1, A2, A3, A3.5, B1,<br>B2, B3, C1, C2, C3,<br>D2, D3, BL1, and BL4 | Similar; no bleach<br>shades and fewer<br>VITA shades are<br>offered for the<br>subject device. |
| Attributes | Subject Device<br>(K234014) | Predicate Device<br>(K141788) | Comparison | |
| Flexural Strength | ≥300 MPa<br>Type II Class 3 per ISO<br>6872:2015/Amd<br>1:2018 | >300 MPa<br>Type II Class 3 per ISO<br>6872:2015/Amd<br>1:2018 | Same | |
| Chemical<br>Solubility | <100 µg/cm² | <100 µg/cm² | Same | |
| Biocompatibility | Biocompatible per ISO<br>10993-1 | Biocompatible per ISO<br>10993-1 | Same | |
| Radioactivity | The activity<br>concentration of<br>Uranium-238 is no<br>more than 1.0 Bq/g-1. | The activity<br>concentration of<br>Uranium-238 is no<br>more than 1.0 Bq/g-1. | Same | |
| Freedom from<br>Extraneous<br>Materials | Shall be free from<br>extraneous materials<br>when assessed by<br>visual inspection | Shall be free from<br>extraneous materials<br>when assessed by<br>visual inspection | Same | |
| Glass Transition<br>Temperature | Shall not deviate by<br>more than ±20°C | Shall not deviate by<br>more than ±20°C | Same | |
| Sterility | Non-sterile | Non-sterile | Same | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo features a colorful triangular prism on the left side. To the right of the prism, the text "PRISMATIK DENTALCRAFT, INC." is displayed in a clear, sans-serif font.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions VII.
Non-clinical data submitted to demonstrate substantial equivalence include:
- Flexural Strength according to ISO 6872:2015/Amd 1:2018 .
- Visual Shade Evaluation
- Color Consistency ●
- Cementation Testing ●
- Printing & Gluing ●
- Uniformity of Color according to ISO 6872:2015/Amd 1:2018
- Chemical Solubility according to ISO 6872:2015/Amd 1:2018 ●
- Radioactivity according to ISO 6872:2015/Amd 1:2018 ●
- Freedom from Extraneous Materials according to ISO 6872:2015/Amd 1:2018
- Glass Transition Temperature according to ISO 6872:2015/Amd 1:2018 ●
- Biocompatibility assessment per the FDA Guidance "Use of International Standard . ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."
- Packaging Validation according to ASTM D4169-22 ●
No clinical data is included in this submission.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a gradient of colors, ranging from blue to pink to yellow. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC." in blue font.
# Flexural Strength
Flexural strength was tested on all shades offered for Obsidian® NOW. The average flexural strength met the acceptance criteria of 300 MPa, which is the value to be achieved for Type II, Class 3 ceramic product according to ISO 6872:2015/Amd 1:2018. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device. Obsidian® NOW, and predicate device, Obsidian Milling Blocks (K141788).
# Visual Shade Evaluation
All shades offered for Obsidian® NOW were evaluated by qualified evaluators using the sample dental restorations milled from the final crystallized products against corresponding VITA® Classical shade guide. Evaluations concluded that Obsidian® NOW meets visual shade match requirements and works as intended. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device, Obsidian® NOW, and predicate device, Obsidian Milling Blocks (K141788).
# Color Consistency
Color consistency within a batch was tested on all shades offered for Obsidian® NOW. A ΔE of less than 2.0 between two coordinates in the CEILAB color space is considered indicative that different samples within a batch being perceptibly identical to one another. Obsidian® NOW meets the color consistency requirement since the ΔE calculation between the samples measured was less than 2.00. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device. Obsidian® NOW, and predicate device. Obsidian Milling Blocks (K141788).
### Uniformity of Color
Uniformity of color was tested on all shades offered for Obsidian® NOW to meet the requirement of ISO 6872:2015/Amd 1:2018. Obsidian® NOW meets the uniformity of color requirement since all samples milled into restorations from a representative final finished block were uniform in color when assessed by visual inspection. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device, Obsidian® NOW, and predicate device, Obsidian Milling Blocks (K141788).
### Chemical Solubility
The worst cases were tested for solubility and the measured value was <100 µg/cm², meeting the ISO 6872:2015/Amd 1:2018 requirement. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device, Obsidian® NOW, and predicate device, Obsidian Milling Blocks (K141788).
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a color gradient from blue to pink to yellow. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC." in blue text.
# Radioactivity
Obsidian® NOW was tested on the worst cases to meet the radioactivity requirement of ISO 6872:2015/Amd 1:2018. Obsidian® NOW meets the radioactivity requirement since the sample measured below the Uranium-238 activity threshold of 1.0 Bq-g-1 per ISO 6872:2015/Amd 1:2018. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device, Obsidian® NOW, and predicate device, Obsidian Milling Blocks (K141788).
# Freedom from Extraneous Materials
All shades offered for Obsidian® NOW were tested to meet the freedom from extraneous materials requirement of ISO 6872:2015/Amd 1:2018. Obsidian® NOW meets the freedom from extraneous materials requirement since the samples were all free from extraneous materials when assessed by visual inspection. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device, Obsidian® NOW, and predicate device, Obsidian Milling Blocks (K141788).
# Glass Transition Temperature
All shades offered for Obsidian® NOW were tested to meet the glass transition temperature requirement of ISO 6872:2015/Amd 1:2018. Obsidian® NOW meets the glass transition temperature requirement since the samples did not deviate by more than ±20℃ of the average To obtained across all lots. The results of the testing were used to address questions related to substantial equivalence based on differences in technical specifications between the subject device. Obsidian® NOW, and predicate device. Obsidian Milling Blocks (K141788).
### Coefficient of Thermal Expansion (CTE)
Coefficient of thermal expansion was not tested on the subject device, Obsidian® NOW, as it is not applicable. Coefficient of Thermal Expansion (CTE) is an important property if staining and glazing were required, for which the subject device, Obsidian® NOW, is not. This difference in technical specification does not raise any issues related to substantial equivalence between the subject device, Obsidian® NOW, and predicate device, Obsidian Milling Block (K141788).
### Biocompatibility
Biological evaluation within a risk management process was performed in accordance with ISO 10993-1:2018 and ISO 14971:2019. Based on the cytotoxicity testing results from the subject device, Obsidian® NOW, it was determined that there is no biocompatibility concern regarding dental glass material, colorants, or other elements. The results of the testing were used to address questions related to substantial equivalence based on
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.
differences in chemical composition between the subject device, Obsidian® NOW, and predicate device, Obsidian Milling Blocks (K141788).
### Packaging Validation
Packaging validation was conducted to ensure that the packaging configurations for Obsidian® NOW are suitable to withstand the distribution environment. Per ASTM D4169-22, Obsidian® NOW was tested to check resistance against manual handling, vehicle stacking, loose load vibration, low pressure (high altitude) hazard, vehicle vibration, and concentrated impact. After the test, the shipping containers were visually inspected for any damage. It was determined that the respective packaging for Obsidian® NOW is suitable for use. The results of the testing were used to address questions related to substantial equivalence based on differences in product packaging between the subject device, Obsidian® NOW, and the predicate device, Obsidian Milling Block (K141788).
#### VIII. Conclusion
Based on the technological characteristics and non-clinical test data included in this submission, the subject device, Obsidian® NOW, has been shown to be substantially equivalent to the predicate device. Obsidian Milling Block (K141788).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.