The Midmark Elevance Instrument Delivery System is intended for use by professional dental practitioners in providing treatment to dental patients in a dental operatory. The system is designed to deliver air, water, vacuum and lowvoltage electricity to hand-held dental instruments.
Device Story
Elevance is a dental delivery unit providing air, water, vacuum, and low-voltage electricity to hand-held dental instruments. System components include a doctor's unit, assistant's unit, cuspidor, mount arms, and a junction box with power supply and regulators. Operated by dental practitioners in a clinical operatory. Device facilitates connection of various accessories (pneumatic handpieces, electric motors, scalers, cameras, curing lights, syringes, vacuum instruments). Features include centrally located controls, a touch pad for settings, and proportional solenoid valves for air/water adjustment. Enhances clinical workflow by centralizing instrument power and utility delivery.
Clinical Evidence
Bench testing only. Testing included qualification run evaluation, EN 60601-1-2 (EMC), EN 61000-3-2, IEC 60601-1 (electrical safety), and ISO 7494-1/2 (dental unit performance).
Technological Characteristics
Dental delivery unit; pneumatic and electrical power distribution. Features proportional solenoid valves for air/water control and touch pad interface. Standards: IEC 60601-1, EN 60601-1-2, EN 61000-3-2, ISO 7494-1, ISO 7494-2. Connectivity: Integrated junction box for utility supply.
Indications for Use
Indicated for use by professional dental practitioners to provide treatment to dental patients in a dental operatory setting. No specific age or gender restrictions; no contraindications listed.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
Predicate Devices
Midmark Procenter Instrument Delivery System (K003090)
Related Devices
K231297 — Firstar Dental Unit · Firstar Dental Company · Jan 27, 2025
K173608 — Forest Dental Unit · Forest Dental Products, Inc. · Aug 17, 2018
K020833 — BOYD DELIVERY UNIT · Boyd Industries, Inc. · Aug 8, 2002
K251626 — Midmark Dental Delivery System · Midmark Corporation · Nov 3, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
K120239
510(k) Special Midmark Corporation Elevance
| 6. 510(k) Summary or 510(k) Statement | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 510(k) SUMMARY |
| | AUG 2 4 2012 |
| 510(k) Owner | Midmark Corporation<br>60 Vista Drive<br>Versailles, Ohio 45380-0286<br>TEL: 937-526-8249 |
| Contact person | Robyn Scopis<br>Regulatory Consultant to Midmark Corp.<br>Regulatory Specialists, Inc.<br>3722 Ave. Sausalito<br>Irvine, CA 92606<br>TEL: 949.422.3853<br>FAX: 949.552.2821<br>EMAIL: robyn@regulatoryspecialists.com |
| Date summary was prepared | August 22, 2012 |
| Name of device | Elevance |
| Common Name | Dental Delivery Unit |
| Classification Name | Unit, Operative Dental |
| Regulation | 872.6640 |
| Product Code | EIA |
| Unmodified Device | Midmark Procenter Instrument Delivery System<br>K003090 |
| Description | |
### Description
The Elevance Delivery Unit includes components to deliver air, water, electrical power, and vacuum to dental handpieces and accessories. The controls are contained in a Doctor's Unit, an Assistant's Unit, and a Cuspidor. Additional parts include mount arms, console mount housing, and a junction box that houses a power supply and air/water regulators. Handpiece accessories or instruments can be added to the Unit, Midmark does not manufacture these accessories but, do provide means to connect them into the Unit. These include high and low speed pneumatic handpiece tubing, electric handpiece motors, scalers, intraoral cameras, curing lights, air/water syringe, and SE and HVE vacuum instruments.
{1}------------------------------------------------
K120239
510(k) Special Midmark Corporation Elevance
## Intended Use
The Midmark Elevance Instrument Delivery System is intended for use by professional dental practitioners in providing treatment to dental patients in a dental operatory. The system is designed to deliver air, water, vacuum and lowvoltage electricity to hand-held dental instruments.
Technological Characteristics
The Unmodified Device and the Elevance were compared in the following areas and found to have similar technological characteristics and to be · equivalent:
・ Indications for Use
> Function of Handpiece Accessories Performance of Handpiece Accessories
The Unmodified Device and the Elevance were compared in the following areas and found to have minor different technological characteristics. The following differences have been determined to not have any impact on the safety or efficacy of the Elevance:
Centrally located controls for instruments Touch Pad on the Control Settings Air and Water Adjustment made via proportional solenoid valves
The following non-clinical performance tests were conducted:
Qualification Run evaluation to verify design EN 60601-1-2:2007 Part 1-2 EN 61000-3-2:2006 +A1:2009 +A2:2009 Part 3-2 EN 60601-1-2:2007 Part 1-2 IEC 60601-1 Part 1 ISO 7494-1:2004 ISO 7494-2:2003
Conclusions from non-clinical performance data
After performing non-clinical performance studies, the data shows that the Elevance demonstrates substantial equivalence to the predicate as a Dental Delivery Unit.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Midmark Corporation C/O Ms. Robyn Scopis Regulatory Consultant Regulatory Specialists, Incorporated 1801 Edgecliff Drive Fullerton, California 92831
Re: K120239
Trade/Device Name: Elevance Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: I Product Code: EIA Dated: August 15, 2012 Received: August 21, 2012
Dear Ms. Scopis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
AUG 2 4 2012
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Runner
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Special Midmark Corporation Elevance
K12.02.32
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ TBD
Device Name: Elevance
Indications for Use:
The Elevance is intended for use by professional dental practitioners in providing treatment to dental patients in a dental operatory. The system is designed to deliver air, water, vacuum and low-voltage electricity to hand-held dental instruments.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runne
Division Sign-Off) . livision of Anesthesiology, General Hospital nfection Control, Dental Devices
Page _ 1 _ of _ 1
10(k) Number: K120039
1 of 1 | Regulatory Specialists, Inc.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.