K020833 · Boyd Industries, Inc. · EIA · Aug 8, 2002 · Dental
Device Facts
Record ID
K020833
Device Name
BOYD DELIVERY UNIT
Applicant
Boyd Industries, Inc.
Product Code
EIA · Dental
Decision Date
Aug 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6640
Device Class
Class 1
Indications for Use
The Boyd Delivery Unit is intended to supply utilities to and serve as a base for dental tools and accessories. The use of this device does not differ significantly from the predicate dental operative units.
Device Story
Boyd Delivery Unit serves as a dental operatory interface device; connects hand instruments to supply utilities including air, water, suction, drain, and electricity. Functions as a system management device; allows operator to control air pressure and water flow to handpieces and syringes via a single control input. Features include stackable design, assistant's instrumentation delivery system, water heater, and cuspidor. Used in dental clinics by dentists or operators to manage hand instruments. Output consists of regulated utility supply to dental tools; enables clinical procedures by providing necessary power and fluid management for dental instrumentation.
Clinical Evidence
Bench testing only.
Technological Characteristics
Dental operative unit; cabinet-mounted. Provides high-speed/low-speed handpiece hookups, 3-way syringe, saliva ejector, and high-volume evacuation. Features master on/off valve, handpiece toggle switches, foot control, and air flush system. Water system is self-contained. Mechanical/pneumatic/hydraulic operation.
Indications for Use
Indicated for use by dental professionals for the effective treatment of patients.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
{0}------------------------------------------------
## 510(k) SUMMARY
- 1) Boyd Industries, Inc. 12900 44th Street North Clearwater, FL 33762 US Phone: 727-561-9292 Fax: 727-561-9393 Contact: Bruce Livingston Establishment Registration Number: 1062917 (reference next page) Date of Submission: March 8, 2002
- 2) Trade Name: Boyd Delivery Units Common Name: Delivery System Classification Name: Dental Operative Unit and Accessories
- 3) Equivalent Device:
The Boyd Delivery Unit is substantial to similar Dental Operative Unit and Accessories previously cleared with the FDA for marketing in the United States, using the same basic technology, design process and attention to safety. Copies of product literature are included in Section 1 and Comparison Data is included in section 2 of this submittal.
- 4) Technical Description:
The Boyd Delivery Unit is a Dental Operative Unit that provides the dentist with primary requirements for a dental operatory. The delivery system is designed to be used as an interface device to connect the dental operatory hand instruments to the appropriate supply utility such as air, water suction, drain and electricity. It functions as a system management device that provides a method of operating various hand instruments from a single control input device. The unit is furnished with controls that allow the dentist or operator to set the air pressure and water flow to the hand pieces and the syringe in the delivery head. The unit is of a stackable design providing a base for dental tools and accessories. Assistant's instrumentation delivery system, water heater, and the cuspidor are all features that this unit can offer.
- 5) Intended use. The Boyd Delivery Unit is intended to supply utilities to and serve as a base for dental tools and accessories. The use of this device does not differ significantly from the predicate dental operative units.
## 6) Comparison of device to marketed devices
Reference chart "Comparison of device to marketed devices" on the following page.
{1}------------------------------------------------
isting from FDA Websit Ticial Correspondent Name: MR. BRUCE LIVING:
Toial Correspondent Phone: Number: 727-561-929
Torial Correspondent Phone: Number: 727-561-9299
000-44TH STREET, INC.
EARWATER, F Setablishment Names
2900 44TA STREET NYON
2017 ATTA STREET NYON
CLEAR ANTAL STREET NYONE Nomber: 10629 17
Stablishment Operationa: Commont Manufactu
Sate of Rugustionar: Cont Owner/Operator:
BOYD INDUSTRUES, INC.
I2900 44TH STRIEST NORTH
CLEARWATERS FLE 33722
CLEARWATERS, FL 33722
Owner/Operator Number: 90303
# omparison of device to marketed device
| Feature | Boyd Industries, Inc. | a-dec (K000966) | DentalEZ (K931204) | Marus Dental |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Mounting unit | Cabinet | Cabinet | Cabinet | Cabinet |
| Instruments<br>Available | High-speed handpiece hookup, low<br>speed handpiece hookup, 3-way<br>syringe, saliva ejector, high volume<br>evacuation with solids collector | High-speed handpiece hookup, low<br>speed handpiece hookup, 3-way<br>syringe, saliva ejector, high volume<br>evacuation with solids collector | High-speed handpiece hookup, low<br>speed handpiece hookup, 3-way<br>syringe, saliva ejector, high volume<br>evacuation with solids collector | High-speed handpiece hookup, low<br>speed handpiece hookup, 3-way<br>syringe, saliva ejector, high volume. |
| Activation | Master off/on valve, handpiece<br>toggle switch and foot control | handpiece toggle switch and foot<br>control | Master off/on valve, handpiece<br>toggle switch and foot control | Automatic activation for each<br>handpiece |
| Cleaning | Air flush system | Water flush | Water flush | Water flush |
| Water System | Self contained | Self contained | City hook up or self contained | City hook up |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 0 8 2002
Mr. Bruce Livingston Boyd Industries, Incorporated 12900 44th Street North Clearwater, Florida 33762
Re: K020833
Trade/Device Name: Boyd Delivery Unit Regulation Number: 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: I Product Code: EIA Dated: June 14, 2002 Received: June 27, 2002
Dear Mr. Livingston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Livingston
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Ulatsowski
A. Ulatowski Timot Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 33
Pege. ........................................................................................................................................................................
510(k) Number (if known)
Device Name:
Indications For Use:
.
Boyd delivery systems are intended for use by Dental Professionals for effective treatment of patients.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurence of CDRH, Office of Device Evaluation (ODF)
Prescription Usey (Per 21 CFR 801,109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Swar Rumpie
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital,
Division Oractrol Devices Division of Anesthesionsy, Dental Devices
510(k) Number. k020833
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.