K052370 · Dentsply International · DYW · Nov 1, 2005 · Dental
Device Facts
Record ID
K052370
Device Name
CP BRACKET SYSTEM
Applicant
Dentsply International
Product Code
DYW · Dental
Decision Date
Nov 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CP Bracket System is indicated for orthodontic movement of teeth.
Device Story
CP Bracket System consists of ceramic orthodontic brackets for orthodontic tooth movement. Single-use devices; applied by orthodontists/dentists during clinical orthodontic treatment. Brackets facilitate tooth alignment; performance relies on proper adhesive application per manufacturer instructions. Biocompatibility testing supports safety; prior use of components in legally marketed devices confirms effectiveness.
Clinical Evidence
No clinical trials performed. Evidence consists of biocompatibility testing and post-market performance data from 20,000 units sold in Japan since 2003, with a low complaint rate (four complaints) attributed to user error regarding adhesive application.
Indicated for orthodontic movement of teeth in patients requiring orthodontic treatment.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
# NOV - 1 2005 510(k) SUMMARY
Ko 5 2370
DENTSPLY International World Headquarters Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405-0872
CONTACT: Helen Lewis August 25, 2005 DATE PREPARED: TRADE OR PROPRIETARY NAME: CP Bracket System Orthodontic bracket (872.5470) CLASSIFICATION NAME: PREDICATE DEVICES: Orthodontic Ceramic Brackets (K042178)
DEVICE DESCRIPTION: The CP Brackets are ceramic orthodontic brackets will be used for a complete orthodontic treatment. The brackets are single-use devices.
INTENDED USE: The CP Brackets are indicated for orthodontic movement of teeth.
TECHNOLOGICAL CHARACTERISTICS: All of the components found in the CP Brackets have been used in DENTSPLY's legally marketed devices.
Biocompatibility testing was completed and the results support the device. A table of the results is included in this submission.
Approximately 20,000 units of this bracket have been sold in Japan since 2003. Only four complaints were received which, upon investigation, were caused by the user failing to follow the adhesive manufacturer's instructions.
We believe that the prior use of the components of the CP Brackets in legally marketed devices, the performance data provided, and the biocompatibility data provided support the safety and effectiveness of the CP Bracket System for the intended uses.
{1}------------------------------------------------
# Predicate Device Information:
Predicate Device 510(k) FDA Substantial Equivalence Letter and Indications for Use Page
Predicate Device Tradename
Mystique® Brackets
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the top half of the circle. The bottom half of the circle contains a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Helen Lewis Director of Corporate Compliance & Regulatory Affairs Dentsply International Susquehanna Commerce Center 221 West Philadelphia Street York, Pennsylvania 17404-0872
Re: K052370
Trade/Device Name: CP Bracket System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: DYW Dated: August 25, 2005 Received: August 30, 2005
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
NOV - 1 2005
{3}------------------------------------------------
#### Page 2 - Ms. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industrv/support/index.html.
Sincerely yours.
Syutie y. Michai Oms.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
(As Required by 21 CFR 807.87(e)
10523 510(K) Number (if known)
CP Bracket System Device Name:
Indications for Use:
The CP Bracket System is indicated for orthodontic movement of teeth.
Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
an... I
### (PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Rumm
(Division Sign-Off)
Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: k052374
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.