K222847 · Orthometric - Industria E Comercio DE Produtos Medicos · NJM · May 4, 2023 · Dental
Device Facts
Record ID
K222847
Device Name
Ceramic Brackets Orthometric
Applicant
Orthometric - Industria E Comercio DE Produtos Medicos
Product Code
NJM · Dental
Decision Date
May 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ceramic Brackets Orthometric are intended for use in orthodontic treatment. The brackets are affixed to teeth so that pressure can be exerted on the teeth.
Device Story
Ceramic Brackets Orthometric are orthodontic appliances made of polycrystalline aluminum oxide; designed for attachment to teeth to exert pressure for alignment correction. Brackets feature rounded contours, low-profile twin design, and integrated color-dot identification systems. Available in various prescriptions, with or without hooks. Used by orthodontists in clinical settings; affixed to teeth to facilitate tooth movement. Device is non-sterile, single-use, and removed upon treatment completion. Benefits include efficient movement mechanics and reduced friction during orthodontic therapy.
Clinical Evidence
No clinical data were included in this submission. Bench testing only, including adhesion strength, friction resistance, and dimensional analysis, performed in accordance with ISO 27020:2010 and ISO 10993 series.
Technological Characteristics
Material: Polycrystalline Aluminum Oxide. Design: Low-profile, twin bracket with rounded contours. Slot size: 0.022 inch. Standards: ISO 27020:2010 (orthodontic brackets), ISO 10993-1:2020, ISO 10993-5:2009, ISO 10993-10:2010, ISO 10993-23:2021 (biocompatibility). Delivery: Non-sterile, single-use.
Indications for Use
Indicated for patients in need of teeth alignment correction requiring orthodontic treatment.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 4, 2023
Orthometric - Industria e Comercio de Produtos Medicos % Tatiana Botura Regulatory Affairs Specialist PR Servicos Regulatorios Administrativos Ltda Rua Alice Aem Saadi, 855/ 2402 Ribeirao Preto, SP 14096570 Brazil
Re: K222847
Trade/Device Name: Ceramic Brackets Orthometric Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: February 2, 2023 Received: February 3, 2023
Dear Tatiana Botura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K222847
Device Name Ceramic Brackets Orthometric
Indications for Use (Describe)
Ceramic Brackets Orthometric are intended for use in orthodontic treatment. The brackets are affixed to teeth so that pressure can be exerted on the teeth.
| Type of Use (Select one or both, as applicable) | <table><tr><td><div> <span> <span style="padding-right: 5px;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </div></td><td><div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <span style="padding-right: 5px;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="padding-right: 5px;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | | |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510 (K) Summary
## K222847
## ADMINISTRATIVE INFORMATION
| Sponsor | Orthometric – Indústria e Comércio de produtos<br>médicos e odontológicos LTDA<br>Rua Irineu Martins, 280 – Distrito de Lacio –<br>Marília, SP – Brazil |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person and Preparer | Tatiana Jabor Botura and Bruno Milhoci<br>Regulatory Affairs Specialist<br>Passarini Regulatory Affairs<br>E-mail: tatiana@passarini.com.br<br>Telephone + 55 (16) 3421 8488 |
| Date Prepared | 04/13/2023 |
## DEVICE NAME AND CLASSIFICATION
| Trade Name | Ceramic Brackets Orthometric |
|------------------------------|------------------------------|
| Common Name | Orthodontic Ceramic Brackets |
| Regulation Number | 21 CFR 872.5470 |
| Regulatory Class | II |
| Product Code | NJM |
| Classification Panel | Dental |
| PREDICATE DEVICE INFORMATION | |
| Predicate Device | K102803 – ClarityTM Advanced Ceramic Brackets |
|------------------|-----------------------------------------------|
| | 3M Unitek Corp (Primary predicate) |
{5}------------------------------------------------
#### Indications for use:
Ceramic Brackets Orthometric are intended for use in orthodontic treatment. The brackets are affixed to teeth so that pressure can be exerted on the teeth.
#### Subject Device Description
The Orthometric's Ceramic Brackets are composed of different designs of brackets that are made of Aluminum Oxide. The ceramic brackets are ICERAM, ICERAM-P, and ICE CLEAR.
#### Substantial Equivalence
As is shown in the equivalence comparison between the subject device and predicate devices, they are equivalent in the product code, regulation number, and common name. The subject and predicate device have the same indications for use. All of them are indicated for the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. The subject device and the predicate K102803 have a color identification system indicating along the axis line a color ink dot. As well as the predicate, the subject device is delivered non-sterile to the end user and is intended to be used once by a single patient.
Like its predicate, the subject devices are manufactured from Aluminum Oxide (ceramic) material, which has a well-documented history of biocompatibility within the oral environment, meeting de requirement of ISO 10993. The subject device is equivalent to the predicate in Polycrystalline material composition.
The subject device and its predicate device incorporate equivalent torques, angulations, and slot sizes. Furthermore, the subject device and the K102803 have bases designed to provide maximum adhesion to the tooth while allowing easy and complete removal when necessary. Ceramic Brackets Orthometric and Clarity Advanced (K102803) have a ceramic bracket body with rounded contours and a low profile, twin design, providing more efficient movement mechanics with reduced friction.
As well as the Clarity Advanced, the subject device has different prescriptions, with and without hooks.
#### 510(k) Summary
{6}------------------------------------------------
Performance tests were carried out to prove the technical equivalence between the Orthometric's Ceramic Brackets and the predicate Clarity Advanced Ceramic Brackets (K102803). The data demonstrates the similarity of the dimensional characteristics when compared. The results of adhesion strength tests demonstrate that both products had similar behavior, showing equivalence, both in the measurement of the external force necessary for detachment and in the shear stress. The Orthometric's Ceramic Brackets demonstrated similarity in the friction resistance test.
| | Subject Device | Primary Predicate device | |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Contents | Ceramic Brackets<br>Orthometric<br>Orthometric Indústria e<br>Comércio de Produtos<br>Médicos e<br>Odontológicos LTDA | Clarity™ Advanced<br>Ceramic Brackets –<br>K102803<br>3M Unitek Corp | Equivalent<br>Discussion |
| Product code | NJM | NJM | Equivalent |
| Common Name | Orthodontic Ceramic<br>Brackets | Orthodontic Ceramic<br>Brackets | Equivalent |
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | Equivalent |
| Indications for<br>use | Ceramic Brackets<br>Orthometric are<br>intended for use in<br>orthodontic treatment.<br>The brackets are affixed<br>to teeth so that pressure<br>can be exerted on the<br>teeth. | Clarity™ Advanced<br>Cerâmicas Brackets are<br>intended for use in<br>orthodontic treatment.<br>The brackets are affixed to<br>teeth so that pressure can<br>be exerted on the teeth. | Equivalent |
| Target<br>Population | Patients in need of teeth<br>alignment correction | Patients in need of teeth<br>alignment correction | Equivalent |
| Single-use | YES | YES | Equivalent |
| Non-sterile | YES | YES | Equivalent |
| Material | Aluminum Oxide | Aluminum Oxide | Equivalent |
| Biocompatibility | Meets the applicable<br>requirement of ISO<br>10993 | Meets the applicable<br>requirement of ISO 10993 | Equivalent |
| Maxillary<br>Angulation | 0 to 11 | 0 to 11 | Equivalent |
| Maxillary<br>Torque (°) | -22 to 18 | -22 to 22 | Equivalent |
| Slot Size | 0.022'' | 0.022'' | Equivalent |
| Bracket Design | Different prescriptions,<br>with and without hooks; | Different prescriptions,<br>with and without hooks; | Equivalent |
| Indication<br>System | Colored-dot | Colored-dot | Equivalent |
Table 5.1: Substantial equivalence comparison
{7}------------------------------------------------
### Non-Clinical Performance Data:
Ceramic Brackets Orthometric was tested following these standards:
- ISO 27020:2010 Dentistry – Brackets and tubes for use in Orthodontics;
- ISO 10993-1:2020 Biological evaluation of medical devices Evaluation and testing within a risk management process;
- . ISO 10993-5:2009- Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity;
- . ISO 10993-10:2010 -Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization;
- . ISO 10993-23:2021 – Biological evaluation of medical devices – Part 23: Tests for irritation;
#### Clinical performance Data:
No clinical data were included in this submission.
#### Conclusion:
The documentation submitted in this premarket notification demonstrates that the Ceramic Brackets have comparable features and performance and, therefore, are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.