Respire Blue Series-EF

K152292 · Respire Medical Holding · LRK · Dec 7, 2015 · Dental

Device Facts

Record IDK152292
Device NameRespire Blue Series-EF
ApplicantRespire Medical Holding
Product CodeLRK · Dental
Decision DateDec 7, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Respire Blue Scries - EF is indicated to treat mild to moderate OSA (Obstructive Sleep Apnea)

Device Story

Respire Blue Series-EF is a custom-fitted intraoral mandibular advancement device; used for treatment of mild to moderate OSA. Device consists of upper and lower trays (acrylic and Wironit chrome alloy) connected by adjustable hinge and telescopic hardware; retained via ball clasps. Patient wears device during sleep; mechanism involves mandibular advancement to increase pharyngeal opening and improve airway exchange. Expansion screws in acrylic side plates allow for titration. Device is prescribed by clinicians; patient-operated for insertion/removal. Output is physical airway support; clinical decision-making based on patient OSA diagnosis and comfort/durability needs. Benefits include reduced snoring and improved breathing during sleep.

Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993 confirmed no skin irritation or sensitization. MEM Elution testing for cytotoxicity passed with reactivity of 0. Material integrity testing compared Wironit/Acrylic to Acrylic-only predicate; both exceeded force requirements for biting (average failure at 275.6 lbf).

Technological Characteristics

Intraoral mandibular advancement device. Materials: Acrylic and Wironit (chrome alloy). Components: Upper/lower plates, telescopic hardware, ball clasps, expansion screws. Dimensions: 6mm mandibular advancement range. Non-powered, mechanical device. Custom-fitted.

Indications for Use

Indicated for adult patients with mild to moderate Obstructive Sleep Apnea (OSA).

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are depicted in a single, continuous line. The logo is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 7, 2015 Respire Medical Holding c/o Mr. Stephen Inglese Founder and CEO Quality Solutions and Support, LLC PO Box 8271 Holland, Michigan 49422 Re: K152292 Trade/Device Name: Respire Blue Series-EF Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: September 7, 2015 Received: September 8, 2015 Dear Mr. Inglese: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang - for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use Statement 4.0 DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K152292 Device Name Respire Blue Series - EF Indications for Use (Describe) The Respire Blue Scries - EF is indicated to treat mild to moderate OSA (Obstructive Sleep Apnea) Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### * DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ # K152292 510(k) Summary Device Common Name: D Device, Anti-Snoring Device Proprietary Name: Respire Blue Series - EF - Submitter: Respire Medical, LLC 18 Bridge St Ste 4J Brooklyn. NY 10021 Phone: 718-643-7326 - Contact: Stephen Inglese Consultant Quality Solutions and Support, LLC Phone: 561-251-0876 Email: swi@qss-llc.com Date Prepared: December 8, 2015 Classification Regulation: 21 CFR §872.5570, Class II - Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea. | Panel: | Dental | |---------------------|---------------------------------------------------------------------------------------------------------| | Product Code: | LRK | | Predicate Device: | K111207 – Submitter's own previously cleared<br>device | | Indication for Use: | The Respire Blue Series – EF is indicated<br>to treat mild to moderate OSA (Obstructive Sleep<br>Apnea) | ### Device Description: The Respire Blue Series – EF (Endurance Frameworks) – is available with a hard device fitting surface. The hard surface consists of Acrylic (side plates / trays) and chrome - Wironit material (upper / Palatal and lower / Lingual plates / trays). Refer to Figure 1 Representative Drawing. The device is retained with ball and clasps which are used to hold elastic bands which guide the jaw to the desired position. In addition there are expansion screws inside the Acrylic side plates / trays that allows for device expansion based on patient needs. The upper and lower components are connected by an adjustable hinge, thus the patient can open and close the mouth while wearing the appliances {4}------------------------------------------------ # Figure 1 - Respire Blue Series - EF - Front View Image /page/4/Picture/2 description: The image shows a dental device, likely a mouthguard or retainer, with a clear central section and colored (red and blue) side components. The device appears to be designed to fit over both the upper and lower teeth simultaneously. The clear part is likely made of a transparent plastic material, while the colored parts are made of a more rigid material. The device is likely used to protect the teeth from grinding or clenching during sleep. # Performance Data: The subject of this 510(k) is a modification to the material used for the manufacturing of the top and bottom plates / trays of the device. The material "Wironit" is a widely used dental material, demonstrated via biocompatibility and cytotoxicity testing. The following nonclinical tests were performed: - Biocompatibility testing was accomplished according to the standards of ● ISO 10993. The results of testing determined that the material didn't cause skin irritation or allergenic sensitization. - Minimal Essential Medium (MEM) Elution testing was designed to ● determine the cytotoxicity of extractable substances. The testing passed with a reactivity of 0 (zero) or none. - Material Integrity testing demonstrates the strength of the Wironit material . as compared to the Acrylic material. The predicate device consisted of only Acrylic. The Wironit / Acrylic and Acrylic only (predicate device) were tested together to determine substantial equivalence as it relates to specimen elongation (stretching) and load bearing force to break. In both tests the points of failure were identified at the same location; yet the amount of energy required to cause failure was less for Wironit /Acrylic. However, the amount of energy shown in the results, still far exceeds the amount of energy required by biting to create such a failure which occurred at an average of 275.6 Ibf using five (5) sample devices. Thus, the proposed Wironit/Acrylic parts are sufficient to survive a load exposure exerted by a human oral cavity. In addition the intended use of this device is indicated to treat mild to moderate OSA (Obstruction Sleep Apnea) and not applying the necessary force for biting. # Substantial Equivalence: The modification of the added material to the originally cleared device is demonstrated in Chart 1. The device function remains the same, the option for the Wironit material for the upper and lower plates / trays provides the patient with a more comfortable fit and durability. Therefore the modified device is substantially equivalent to the previously cleared Respire Blue Series. {5}------------------------------------------------ # Respire Blue Series – EF Traditional 510(k) Respire Medical # Chart 1 | Substantial<br>Equivalence Topic | Respire Blue Series | Respire Blue Series -EF | |-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) | K111207 | K152292 | | Company Name | Respire Medical | Respire Medical | | Regulation Description | Intraoral devices for<br>snoring and intraoral<br>devices for snoring and<br>obstructive sleep apnea<br>(OSA) | Intraoral devices for<br>snoring and intraoral<br>devices for snoring and<br>obstructive sleep apnea<br>(OSA) | | Device Name | Device, Anti Snoring | Device, Anti Snoring | | Product Code | LRK | LRK | | Classification | Class II | Class II | | Intended Use | The Respire Blue Series<br>is indicated to treat mild to<br>moderate OSA. | The Respire Blue Series -<br>EF is indicated to treat mild<br>to moderate OSA. | | Single or Multiple Use | Multiple Use | Multiple Use | | Target Population | Adult Patients | Adult Patients | | Prescription or OTC Use | Prescription | Prescription | | Device Components | Orthodontic Acrylic plates<br>/ trays, Telescopic<br>Hardware and Ball Clasp | Orthodontic Acrylic plates<br>/ trays, and Wironit plates<br>/ trays, Telescopic<br>Hardware and Ball Clasp | | Appliance Design | Customized device Rigid<br>plates / trays two pieces<br>Upper/Lower and sides<br>Acrylic. | Customized device Rigid<br>plates and trays / two pieces /<br>Upper and Lower<br>/ Acrylic and Wironit and<br>sides Acrylic | | Device Functionality | Allows to increase<br>pharyngeal opening, and to<br>improve the ability to exchange<br>air during sleep. Upper and<br>lower plates / trays unhook for<br>easy removal from mouth.<br>Works by mandibular<br>advancement. Adjustable<br>using titration keys. | Allows to increase<br>pharyngeal opening, and to<br>improve the ability to<br>exchange air during sleep<br>Upper and lower plates / trays<br>unhook for easy removal from<br>mouth Works by mandibular<br>advancement Adjustable<br>using titration keys | | Mandibular Advancement<br>Range | 6mm | 6mm | | Raw Material: Side /<br>Upper and Lower Trays | Acrylic (side and upper<br>and lower plates / trays) | Acrylic (side plates / trays)<br>and Wironit (upper and<br>lower / plates / trays) | | Raw Material: Metal<br>Components | Stainless Steel | Stainless Steel | | Colorants | Blue | Blue | Note: Bold "Substantial Equivalence Topic" – Difference between the cleared device and the Subject Device
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...