CLEARPATH ALIGNER

K123514 · Qserve America, Inc. · NXC · Nov 1, 2013 · Dental

Device Facts

Record IDK123514
Device NameCLEARPATH ALIGNER
ApplicantQserve America, Inc.
Product CodeNXC · Dental
Decision DateNov 1, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ClearPath Aligner is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e. all second molars). The ClearPath Aligner is intended for minor anterior tooth movement by way of continuous gentle force.

Device Story

ClearPath Aligner consists of series of custom-fabricated, clear, thermoformed plastic appliances; used for minor anterior tooth movement. Clinician takes dental impressions; manufacturer creates custom aligners based on molds. Patient wears aligners for 14-day intervals, removing only for meals. Mechanism of action involves application of continuous gentle force to teeth to achieve progressive realignment. Device intended for use in dental clinic setting under supervision of dentist or orthodontist. Benefits include aesthetic, removable correction of malocclusion.

Clinical Evidence

No clinical data; no bench testing performed. Safety and effectiveness supported by scientific literature and similarity in design, materials, and use to predicate devices.

Technological Characteristics

Customized thermoformed plastic aligners; non-sterile; patient-specific geometry based on dental impressions. No specific material standards or software/electronic components described.

Indications for Use

Indicated for correction of dental malocclusion in patients with permanent dentition (all second molars) requiring minor anterior tooth movement.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ClearPath Aligner Premarket Notification: Traditional 510(k) ## 510(k) Summary | Submitter Name: | ClearPath Orthodontics | | |---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Submitter Address: | 54E/1, D/1, Gulberg III, Lahore, Pakistan, 54000 | | | Phone Number: | 00 92 42 35752012 | | | Contact Person: | Dr. Waqas Wahab | | | Date Prepared: | 09 November 2012; Revised 24 October 2013 | | | Device Trade Name: | ClearPath Aligner | | | Common Name | Clear Braces | | | Classification Name, Number & Product Code: | Orthodontic Plastic Bracket 21 CFR 872.5470 NXC | NOV 0 1 2013 | | Predicate Devices: | K073121 - NuBrace Invisible Removable Orthodontics (IRO), NuBrace, Inc. K981095 - Align System, Align Technology Inc. | | | Statement of Intended Use | The ClearPath Aligner is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e. all second molars). The ClearPath Aligner is intended for minor anterior tooth movement by way of continuous gentle force. | | | Device Description: | ClearPath Orthodontics uses clear, thin, thermoformed plastic material for the manufacture of its ClearPath Aligner. The aligners are provided non-sterile and are customized for each patient according to the molds provided by the treating dentist or orthodontist. | | | | Depending on the patient need and the treatment plan, a series of aligners may be used. The duration of use for each aligner is 14 days and it is to be worn except during meals. | | | Mechanism of Action | Each ClearPath Aligner exerts gentle force to achieve progressive realignment of the teeth until the final correction has been attained. | | | Summary of Technological Characteristics | ClearPath Orthodontics manufactures the customized aligners based on standard molds sent to the company by the prescribing dentist/orthodontist. The molds are made after the clinician has assessed the patient's teeth, designed a treatment plan, and taken the impressions. | | | | The plastic used for fabrication of the aligners is a commonly used thermoformed plastic, used in many dental appliances. It has a long history of safe and biocompatible use for this purpose. | | {1}------------------------------------------------ | Performance<br>Testing | No bench testing was warranted for this product because the<br>scientific literature and similarity in design, materials and use to<br>the predicate devices assure that the preformed aligners are safe<br>and effective for the specified intended use. | |---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison to the<br>Predicate Devices | The ClearPath Aligner was compared to the two predicate devices<br>with respect to indications for use, design, material composition,<br>and the mechanism of action (gentle continuous pressure on<br>teeth). | | Substantial<br>Equivalence<br>Conclusion | Based upon the same intended use compared to the predicates,<br>the similar base materials used for fabricating the aligners, the<br>same basic design, and the same mechanism of action, it can be<br>concluded that the ClearPath Aligner is substantially equivalent to<br>the predicate devices. | | Verifications<br>regarding this<br>510(k) Summary | The summary includes only information that is also covered in the<br>body of the 510(k). The summary does not contain any puffery or<br>unsubstantiated labeling claims. The summary does not contain<br>any raw data, i.e., contains only summary data. The summary<br>does not contain any trade secret or confidential commercial<br>information. The summary does not contain any patient<br>identification information. | {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ## November 1, 2013 ClearPath Orthodontics C/O Ms. Patsy J. Trisler, J.D., RAC Vice President, Regulatory and Clinical Affairs Oserve America, Incorporated 1 54 Main Street, Suite 2 CHARLESTOWN NH 03603 Re: K123514 Trade/Device Name: ClearPath Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NXC Dated: October 24, 2013 Received: October 30, 2013 Dear Ms. Trisler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mary[S]Bünner -S Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Form 510(k) Number (if known): K123514 Device Name: ClearPath Aligner Indications For Use: The ClearPath Aligner is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e. all second molars). The ClearPath Aligner is intended for minor anterior tooth movement by way of continuous gentle force. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Andrew I. Steen -S 2013.10.31 16:13:12 124 Section 4.0 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...