K180952 · Mem Dental Technology Co., Ltd. · NJM · Jun 7, 2019 · Dental
Device Facts
Record ID
K180952
Device Name
Ceramic Bracket
Applicant
Mem Dental Technology Co., Ltd.
Product Code
NJM · Dental
Decision Date
Jun 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ceramic Bracket is indicated for orthodontic movement of natural teeth.
Device Story
Ceramic brackets are orthodontic appliances bonded to natural teeth by dental professionals; they connect with orthodontic wires to exert pressure and move teeth to preferred positions. The device features a molded polycrystalline alumina ceramic body with an archwire slot, tie-wings for ligation, and hooks for additional tooth movement. It is used in a clinical dental setting. The device is non-self-ligating or self-ligating and is intended for single use. By providing a stable attachment point for orthodontic wires, the device facilitates controlled tooth movement, improving dental alignment for the patient.
Clinical Evidence
No clinical data or animal studies were required. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993 series) and mechanical performance testing (ISO 27020, ISO 6872, ISO 29022, ISO 6474-2).
Technological Characteristics
Materials: Polycrystalline (translucent) alumina. Design: Molded ceramic body with archwire slot, tie-wings, and hooks. Mechanical lock base. Non-sterile. Single-use. Standards: ISO 27020, ISO 6872:2015, ISO 29022:2013, ISO 6474-2:2012.
Indications for Use
Indicated for orthodontic movement of natural teeth in patients requiring orthodontic treatment.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2019
MEM Dental Technology Co., Ltd Anita Chen Official Correspondent 2F., No.22, Ln.31, Sec.1, Huandong Rd., Xinshi Dist. Tainan City, 74146 Tw
Re: K180952
Trade/Device Name: Ceramic Bracket Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: April 20, 2019 Received: May 6, 2019
Dear Anita Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas Nandkumar, Ph.D. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180952
Device Name Ceramic Bracket
Indications for Use (Describe)
Ceramic Bracket is indicated for orthodontic movement of natural teeth.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) summary information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92
| | The assigned 510(k) Number: | K180952 |
|----|-----------------------------|--------------------------------------------|
| 1. | Submitter | |
| | Mailing Address | MEM Dental Technology Co., Ltd |
| | | 2F., No.22, Ln.31, Sec.1, Huandong Rd., |
| | | Xinshi Dist., Tainan City 74146, Taiwan |
| | | Phone: +886 6 5053939 |
| | | Establishment Registration No.: 3009011037 |
| | Contact Person | Anita Chen |
| | Phone: | +886(0) 939-855-759 |
| | Fax: | +886 6 5051212 |
| | E-mail: | m9104303@gmail.com |
| | Date Prepared | 2019.06.06 |
- Device Name 2
| Proprietary Name: | Ceramic Bracket |
|----------------------|----------------------------------------------------------|
| Common or Usual Name | Bracket, Ceramic, Orthodontic |
| Classification Name | Bracket Orthodontic Plastic Bracket (21 CFR<br>872.5470) |
| Product Code | NJM |
| Device | Ceramic Bracket |
| CFR Classification | CFR Part 872.5470 |
| Device Class | II |
| Classification Panel | Dental |
| Primary Predicate | |
| 510(k) number | K160615 |
3
{4}------------------------------------------------
| Device name | Orthodontic Ceramic Brackets 1.1 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | TOMY, INC<br>6 tenko blg. 3-16-7<br>midoricho, fuchu city, JP 183-0006 |
| 4 Device Description | Ceramic bracket are designed to move teeth to<br>improve their alignment. Ceramic bracket are bonded<br>to natural teeth by dental professionals to connect<br>with orthodontic wires to cause tooth movement to a<br>more preferred position.<br><br>Ceramic bracket are intended for use in affixed to a<br>tooth so that pressure can be exerted on the teeth. Use<br>bonding supplies to bond the bracket on to a tooth.<br>Use bonding supplies to bond the bracket on to a<br>tooth. Ceramic bracket are bonded to natural teeth<br>by dental professionals to connect with orthodontic<br>wires to cause tooth movement to a more preferred<br>position. |
| 5. Indication for Use: | Ceramic bracket is indicated for orthodontic<br>movement of natural teeth. |
| 6. Technological | A comparison of the device features, intended use, |
|----------------------------|----------------------------------------------------|
| Characteristics and | and other information demonstrates that the |
| Substantial Equivalence | Ceramic bracket is substantially equivalent to the |
| Comparison with Predicate: | predicate device as summarized in Table 1. |
Table 1
{5}------------------------------------------------
| Manufacturer | MEM Dental Technology Co., Ltd. | Tomy, Inc. |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Primary Predicate Device |
| Device name | Ceramic bracket | Orthodontic Ceramic Brackets 1.1 |
| 510(k) Number | K180952 | K160615 |
| Indication for Use | Ceramic bracket is indicated for<br>orthodontic movement of<br>natural teeth. | Orthodontic Ceramic Brackets<br>1.1 are indicated for orthodontic<br>movement of natural teeth. |
| Classification name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket |
| Rag Number | 872.5470 | 872.5470 |
| Product Code | NJM | NJM |
| Class | II | II |
| Medical Specialty | Dental | Dental |
| Ligation | Non-Self-Ligating | Non-Self-Ligating |
| | Self-Ligating | Self-Ligating |
| Base | Mechanical Lock Base | Mechanical Lock Base |
| Sterility | Non-Sterile | Non-Sterile |
| Materials | Polycrystalline (translucent) alumina | Polycrystalline (translucent) alumina |
| Utility | Single-use only | Single-use only |
| Design | Archwire slot, tiewings for ligature and<br>identification marks for placement<br>Hooks for ligation, for additional tooth<br>movement<br>Molded ceramic body with rounded<br>corners and e Slot to hold orthodontic<br>wires edges | Archwire slot, tiewings for ligature and<br>identification marks for placement<br>Hooks for ligation, for additional tooth<br>movement<br>Molded ceramic body with rounded<br>corners and e Slot to hold orthodontic<br>wires edges |
{6}------------------------------------------------
| Manufacturer | MEM Dental Technology Co., Ltd. | Tomy, Inc. |
|-------------------|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Primary Predicate Device |
| Device name | Ceramic bracket | Orthodontic Ceramic Brackets 1.1 |
| 510(k) Number | K180952 | K160615 |
| Picture of Device | Image: Non-Self-Ligating<br>Image: Self-Ligating | Image: Non-Self-Ligating Mandibular Bicuspid<br>Image: Non-Self-Ligating for all other Teeth<br>Image: Self-Ligating Mandibular Bicuspid<br>Image: Self-Ligating for all other Teeth |
## 7. Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device:
· ISO 27020 First edition 2010-12-15 Dentistry - Brackets and Tubes for use in Orthodontics
Ceramic Bracket.
All of the components found in the Ceramic Bracket have been used in legally marketed devices, such as the primary predicate Orthodontic Ceramic Brackets 1.1.
Biocompatibility Testing
{7}------------------------------------------------
The biocompatibility evaluation and testing of the Product name was conducted in accordance with the following standards and guidance, as recognized by the FDA:
- . FDA Draft Guidance - Use of International Standard ISO- 10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing", dated 04-23-2013
- . ISO 10993-3,Biological evaluation of medical device-Part 3: Test for genotoxicity, carcinogenicity and reproductive toxicity.
- . ISO 10993-5, Biological evaluation of medical device-Part 5: Tests for in vitro cytotoxicity.
- . ISO 10993-10. Biological evaluation of medical device-Part 10: Tests for irritation and skin sensitization.
- ISO 10993-11, Biological evaluation of medical device-Part 11: Tests for systemic toxicity.
- . ISO 10993-12, Biological evaluation of medical device-Part 12: Sample preparation and reference materials.
### Sterilization Status
The Ceramic Bracket are provided non-sterile and are not processed prior to use.
#### Mechanical testing
Ceramic bracket's mechanical function including Shear Bond Strength, Torque Strength test and structure integrity were tested. The results were compared to recognized standards (ISO 6872:2015, ISO 29022:2013, ISO 27020:2010, ISO 6474-2:2012) and clinical literature data to demonstrate substantial equivalence.
No animal studies or clinical testing have been required for these devices.
#### 8. Conclusion
Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the Ceramic bracket is substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.