Anti Snore Mouthpiece

K250028 · Dongguan Yiyao Science & Technology Development Co., Ltd. · LRK · Oct 3, 2025 · Dental

Device Facts

Record IDK250028
Device NameAnti Snore Mouthpiece
ApplicantDongguan Yiyao Science & Technology Development Co., Ltd.
Product CodeLRK · Dental
Decision DateOct 3, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Anti snore mouthpiece is intended as an aid in the reduction of snoring for adults at least 18 years old.

Device Story

Anti Snore Mouthpiece (models TD-02, TD-03) is a mandibular repositioning device (MRD) used at home during sleep. Device consists of maxillary and mandibular trays constructed from ethylene-vinyl acetate copolymer (EVA) and polycarbonate (PC). User performs a 'boil and bite' thermal molding process to custom-fit the trays to their dental arches. Once fitted, the device advances the mandible forward relative to the maxilla, increasing pharyngeal space; this reduces air turbulence, a primary cause of snoring. The device features adjustable jaw advancement increments up to 6mm for user comfort. It is a single-patient, reusable, non-sterile device requiring daily cleaning. The device provides a mechanical barrier and positioning aid to improve airway patency.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including hardness, water absorption, melting point, tensile strength, elongation, forming time, and movement measurements. Biocompatibility testing conducted per ISO 10993-1, including genotoxicity (ISO 10993-3), in vitro toxicity (ISO 10993-5), irritation (ISO 10993-23), systemic toxicity (ISO 10993-11), and skin sensitization (ISO 10993-10).

Technological Characteristics

Intraoral mandibular repositioning device. Materials: Ethylene-vinyl acetate copolymer (EVA) and Polycarbonate (PC). Design: Two-piece (maxillary/mandibular) 'boil and bite' thermal-setting resin trays. Adjustable jaw advancement up to 6mm. Non-sterile. Biocompatibility compliant with ISO 10993-1.

Indications for Use

Indicated for adults at least 18 years old as an aid in the reduction of snoring.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 3, 2025 Dongguan Yiyao Science & Technology Development Co., Ltd. % Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, CHINA Re: K250028 Trade/Device Name: Anti Snore Mouthpiece Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: September 3, 2025 Received: September 3, 2025 Dear Ivy Wang : We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K250028 - Ivy Wang Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250028 - Ivy Wang Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250028 Device Name Anti Snore Mouthpiece (TD-02, TD-03) Indications for Use (Describe) Anti snore mouthpiece is intended as an aid in the reduction of snoring for adults at least 18 years old. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} 510(K) Summary K250028 Prepared date: 2025-09-25 A. Applicant: Dongguan Yiyao Science & Technology Development Co., Ltd. Address: Room501, NO.11 Shennan yiheng street, Shenxiang, Qishi Town, Dongguan City, Guangdong Province, China Contact Person: Xianzhao Tang Tel: +86 13713323614 Email: 494846003@qq.com Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com B. Device: Trade Name: Anti snore mouthpiece Model: TD-02, TD-03 Common Name: Anti-snoring Device Regulatory Information Classification Name: Device, Anti-Snoring Regulation Number: 21 CFR 872.5570 Product Code: LRK Device Class: 2 Review Panel: Dental C. Predicate device: 510(k) Number: K220688 Trade Name: SilentZPro 2.0 Manufacturer: Shinrin-Yoku Traders LLC D. Indications for use of the device: {5} Anti snore mouthpiece is intended as an aid in the reduction of snoring for adults at least 18 years old. # E. Device Description: Anti snore mouthpiece is an intraoral device composed of a maxillary and mandibular tray assembled to position the mandible forward relative to the maxilla to increase users' pharyngeal space and improve the ability to exchange air and decreases air turbulence, a causative factor in snoring. # F. Comparison with predicate device Table 1 General Comparison | Device | Proposed Device | Predicate Device | Result | | --- | --- | --- | --- | | 510K number | K250028 | K220688 | - | | Model name | Anti Snore Mouthpiece | SilentZPro 2.0 | - | | Classification | 2 | 2 | Same | | Regulation | 21 CFR 872.5570 | 21 CFR 872.5570 | Same | | Product code | LRK | LRK | Same | | Indications for use | Anti snore mouthpiece is intended as an aid in the reduction of snoring for adults at least 18 years old. | SilentZPro 2.0 is intended as an aid in the reduction of snoring for adults at least 18 years old. | Same | | OTC use | Yes | Yes | Same | | Target | Adults 18 and older | Adults 18 and older | Same | | Environment | During sleep, at home | Durin sleep, at home | Same | | Mechanism of action | Mandibular repositioning device (MDR) that advances the lower jaw to increase pharyngeal space and alleviate snoring | Mandibular repositioning device (MDR) that advances the lower jaw to increase pharyngeal space and alleviate | Same | | Technological characteristics | Flexible, moldable guard used as a barrier between teeth. | Flexible, moldable guard used as a barrier between teeth. | Same | | Placement of device | In the mouth, on the lower and upper jaws | In the mouth, on the lower and upper jaws | Same | | Preparation / Set-up | Squeeze mouthpiece in a ‘C’ position and adjust. Mold tray Mold the trays to the user’s mouth. | Squeeze mouthpiece in a ‘C’ position and adjust. Mold tray Mold the trays to the user’s mouth. | Same | {6} | Design | Custom-fitted intraoral device using a “boil and bite” approach and thermal setting (heat sensitive) resins. Molded to the entire upper and lower arch of teeth. Consists of an upper and lower tray. Outer shell provides with structural support and inner shell is lined with softer material that is heat sensitive and thus | Custom-fitted intraoral device using a “boil and bite” approach and thermal setting (heat sensitive) resins. Molded to the entire upper and lower arch of teeth. Consists of an upper and lower tray. Outer shell provides with structural support and inner shell is lined with softer material | Same | | --- | --- | --- | --- | | Adjustments | Adjustable jaw advancement position. Ability to reset the adjustment. The upper and lower trays are adjustable increments up to 6mm for use comfort. | Adjustable jaw advancement position. Adjustably positions the mandible forward in 3 positions, 4mm apart anteriorly, while maintaining a 9mm inferior placement for user comfort. | Similar The adjustable position numbers and increment length are different with predicated device, which do not raise any new safety and effectiveness concern. | | Single Patient/ Reusable | Single user, reusable. Clean/rinse daily with toothbrush and toothpaste or with effervescent oral device cleaning tablets. Deep clean once every three days. Non-sterile. | Single user, reusable. Non-sterile. Requires daily cleaning, and reusable with a life use of no more than 30 days. | Different Both are required daily clean. This difference do not raise new safety or effectiveness concern. | | Material | Ethylene-vinyl acetate copolymer (EVA) Polycarbonate (PC) | Polycarbonates (PC); Ethylene-vinyl acetate copolymer resin (EVA) | Same | | Biocompatibility | Assessment conducted for testing for prolonged mucosal surface contact in accordance with ISO 10993-1. | Assessment conducted for testing for prolonged mucosal surface contact in accordance with ISO 10993-1. | Same | G. Non-Clinical Test Conclusion Bench testing has demonstrated that the device is in compliance with pertinent standards and {7} specifications, the expectations of the dental community and the product labeling. Performance testing was performed to the proposed device including hardness, water absorption, melting point, tensile strength & elongation testing, forming time and movement measurement testing to demonstrate its effectiveness, which supports its substantial equivalence to the predicate device. Additionally, as per “Intraoral Devices for Snoring and/or Obstructive Sleep Apnea- Class II Special Controls Guidance Document for Industry and FDA”, tests for genotoxicity (ISO 10993-3), tests for in vitro Toxicity (ISO 10993-5), tests for irritation (ISO 10993-23), tests for systemic toxicity (ISO 10993-11) and tests for skin sensitization (ISO 10993-10) were conducted on the anti snore mouthpiece. The results of the testing met the requirements of the study protocols and the material is considered non-toxic, non-sensitizing and is not an intracutaneous irritant. The results of the studies further support a determination of substantial equivalence to the predicate. ## H. Clinical Test Conclusion No clinical study is included in this submission. ## I. Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K220688. 4
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...