SilentZPro 2.0

K220688 · Shinrin-Yoku Traders, LLC · LRK · Dec 2, 2022 · Dental

Device Facts

Record IDK220688
Device NameSilentZPro 2.0
ApplicantShinrin-Yoku Traders, LLC
Product CodeLRK · Dental
Decision DateDec 2, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

SilentZPro 2.0 is intended as an aid in the reduction of snoring for adults at least 18 years old.

Device Story

SilentZPro 2.0 is an intraoral mandibular repositioning device (MRD) used at home by adults to reduce snoring. The device consists of maxillary and mandibular trays that are custom-fitted by the user via a 'boil and bite' thermal setting process. By advancing the mandible forward relative to the maxilla, the device increases pharyngeal space, improves air exchange, and decreases air turbulence. The device offers three adjustable advancement positions (4mm increments) while maintaining a 9mm inferior placement. It is a single-user, reusable device intended for a maximum 30-day lifespan. The user cleans the device daily. It provides a non-invasive mechanical intervention for snoring management.

Clinical Evidence

No clinical performance testing was conducted. Evidence is based on non-clinical bench testing, including biocompatibility (cytotoxicity per ISO 10993-5, sensitization and irritation per ISO 10993-10) and risk analysis.

Technological Characteristics

Intraoral mandibular repositioning device; boil-and-bite custom-fit design; thermal setting resins; adjustable jaw advancement (3 positions, 4mm increments); non-sterile; reusable (30-day life); biocompatibility per ISO 10993; risk management per ISO 14971.

Indications for Use

Indicated for adults 18+ years old as an aid in the reduction of snoring.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 2, 2022 Shinrin-Yoku Traders LLC % Ryan Bruss Regulatory Consultant RGLM Consulting LLC 6915 225th St SW Mountlake Terrace, Washington 98043 Re: K220688 Trade/Device Name: SilentZPro 2.0 Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: November 4, 2022 Received: November 4, 2022 Dear Ryan Bruss: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Bobak Shirmohammadi -S For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220688 Device Name SilentZPro 2.0 Indications for Use (Describe) SilentZPro 2.0 is intended as an aid in the reduction of snoring for adults at least 18 years old. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a stylized illustration of three evergreen trees. The trees are arranged in a row, with the middle tree being slightly taller than the other two. Above the trees, there are several lines radiating outwards, resembling sun rays. The entire illustration is in a dark green color, giving it a natural and earthy feel. Shinrin-Yoku Traders LLC 18450 SW Seiffert Rd Sherwood, OR 97140 USA +1-971-712-9175 shinrinyokutradersllc@gmail.com www.itusguardz.com #### 510(k) Summary Device Name: SilentZPro 2.0 ### A. General Information | Sponsor | Shinrin-Yoku Traders LLC<br>18450 SW Seiffert Rd, Sherwood, OR 97140 | |----------------|----------------------------------------------------------------------------------| | Contact Person | Ryan Bruss<br>CEO<br>Tel: 971-712-9175<br>Email: shinrinyokutradersllc@gmail.com | | K Number | K220688 | | Date prepared | November 3, 2022 | #### B. Device | Propriety Name | SilentZPro 2.0 | |---------------------|-----------------------------------------------------------| | Common Name | Antisnoring device | | Classification Name | Intraoral devices for snoring and obstructive sleep apnea | | Product Code | LRK | | Class | II | | Regulation Number | 21 CFR 872.5570 | #### C. Predicate Device | Name | PureSleep® | |-------------------|------------------------------------------| | Owner | Sleep Science Partners, Inc. | | K number | K190058 | | Reference Device: | SnoreRx® (K170825, Apnea Sciences, Inc.) | #### D. Description of the Device SilentZPro 2.0 is an intraoral device composed of a maxillary and mandibular tray assembled to position the mandible forward relative to the maxilla to increase users' pharyngeal space and improve the ability to exchange air and decreases air turbulence, a causative factor in snoring. 510(k) Summary {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a stylized illustration of three evergreen trees. The trees are arranged in a row, with the middle tree being slightly taller than the other two. Above the trees, there are several short, dashed lines radiating outwards in a semi-circular pattern, resembling sun rays. The entire illustration is in a dark green color. 18450 SW Seiffert Rd Sherwood, OR 97140 USA +1-971-712-9175 shinrinyokutradersllc@gmail.com www.itusguardz.com The device is intended for Over-the-Counter (OTC) use. #### E. Intended Use Statement SilentZPro 2.0 is intended as an aid in the reduction of snoring for adults at least 18 years old. #### F. Comparison of Technological Characteristics The table below compares the similarities and differences between the SilentZPro 2.0 and the predicate device/Reference device. | Attribute | Proposed device<br>(K220688) | Predicate device<br>(K190058) | Reference device<br>(K170825) | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Proprietary<br>Name | SilentZPro 2.0 | PureSleep® | SnoreRx® | | Manufacturer | Shinrin-Yoku Traders LLC | Sleep Science Partners, Inc. | Apnea Sciences, Inc. | | 510k Number | K220688 | K190058 | K170825 | | Regulation | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | | Product Code | LRK | LRK | LRK | | Class Name | Same | Intraoral Device for Snoring and<br>Obstructive Sleep Apnea | Same | | Rx or OTC | OTC | OTC | OTC | | Intended Use | SilentZ Pro 2.0 is intended as an aid<br>in the reduction of snoring for adults<br>at least 18 years old. | PureSleep® (OTC) is indicated for<br>use for adults 18 years and above as<br>an aid in the reduction of snoring<br>during hours of sleep | The SnoreRx is intended for use on<br>adult patients 18 years of age or<br>older as an aid for the reduction of<br>snoring. | | Population | Same | Adult | Same | | Environment | Same | During sleep, at home | Same | | Mechanism of<br>action | Same as the Predicate Device | Mandibular repositioning device<br>(MRD) that advances the lower jaw<br>to increase pharyngeal space and<br>alleviate snoring | Mandibular repositioning device that<br>advances the lower jaw to increase<br>pharyngeal space. | | Attribute | Proposed device<br>(K220688) | Predicate device (K190058) | Reference device<br>(K170825) | | Placement of<br>device | In the mouth, on the lower and upper<br>jaws | In the mouth, on the lower and upper<br>jaws | In the mouth, on the lower and upper<br>jaws | | Preparation /<br>Set-up | Same as reference device. (Squeeze<br>mouthpiece in a "c" position and<br>adjust. Mold the trays to the user's<br>mouth.) | Connect upper and lower parts per<br>user comfort. Mold the trays to the<br>user's mouth. | Squeeze mouthpiece in a 'C'<br>position and adjust. Mold tray Mold<br>the trays to the user's mouth. | | Design | Custom-fitted intraoral device using<br>a "boil and bite" approach and<br>thermal setting (heat sensitive)<br>resins. Molded to the entire upper<br>and lower arch of teeth.<br>Consists of an upper and lower tray.<br>Outer shell provides with structural<br>support and inner shell is lined with<br>softer material that is heat sensitive<br>and thus allows for custom fitting. | Custom impression to each<br>individual's mouth using a "boil<br>and bite" approach and thermal<br>setting (heat sensitive) resins.<br>Molded to the entire upper and<br>lower arch of teeth.<br>Consists of an upper and lower<br>tray. Outer shell provides with<br>structural support and inner shell is<br>lined with softer material that is<br>heat sensitive and thus allows for<br>custom fitting. | Custom impression to each<br>individual's mouth using a "boil<br>and bite" approach and thermal<br>setting (heat sensitive) resins.<br>Molded to the entire upper and<br>lower arch of teeth.<br>Consists of an upper and lower<br>tray. Outer shell provides with<br>structural support and inner shell is<br>lined with softer material that is<br>heat sensitive and thus allows for<br>custom fitting. | | Adjustments | Adjustable jaw advancement<br>position. Adjustably positions the<br>mandible forward in 3 positions,<br>4mm apart anteriorly, while<br>maintaining a 9mm inferior<br>placement for user comfort | custom fitting.<br>Adjustable jaw advancement<br>position. Adjustably positions the<br>mandible forward in three positions,<br>4mm apart anteriorly, while<br>maintaining a 9mm inferior<br>placement for user comfort | custom fitting.<br>Adjustable jaw advancement<br>position. Ability to reset the<br>adjustment. The upper and lower<br>trays are adjustable in 1mm<br>increments up to 6mm. | | Single Patient/<br>Reusable | Single user, reusable. Non-sterile.<br>Requires daily cleaning, and<br>reusable with a life use of no more<br>than 30 days. | Single user, reusable.<br>Clean/rinse daily with toothbrush<br>and toothpaste or with effervescent<br>oral device cleaning tablets. Deep<br>clean once per week. Non-sterile. | Clean/rinse daily. Deep clean every<br>two weeks. Non-sterile. | | Materials | Same as the Reference Device | Polypropylene homopolymer;<br>ethylene vinyl acetate copolymer<br>(EVA) | Polycarbonates (PC);<br>Ethylene-vinyl acetate copolymer<br>resin (EVA) | | Attribute | Proposed device (K220688) | Predicate device (K190058) | Reference device (K170825) | | Biocompatibility | Same | Assessment conducted for testing for<br>prolonged mucosal surface contact in<br>accordance with ISO 10993-1. | Same | {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a stylized illustration of three evergreen trees. The trees are arranged in a row, with the middle tree being slightly taller than the other two. Above the trees, there are several short, dashed lines radiating outwards in a semi-circular pattern, resembling sun rays. The entire image is rendered in a single color, a shade of green. ## Shinrin-Yoku Traders LLC 18450 SW Seiffert Rd Sherwood, OR 97140 USA +1-971-712-9175 shinrinyokutradersllc@gmail.com www.itusguardz.com {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a stylized logo with three evergreen trees in the center. Above the trees are radiating lines, resembling the sun's rays. A horizontal line runs along the base of the trees, grounding the design. The entire logo is rendered in a single color, a shade of green. 18450 SW Seiffert Rd Sherwood, OR 97140 USA +1-971-712-9175 shinrinyokutradersllc@gmail.com www.itusguardz.com The SilentZPro 2.0 has the same general intended use as the predicate device, and the similar technological characteristics as the predicate device. The biocompatibility performance of the SilentZPro 2.0 met the standard of ISO 10993 biocompatibility testing. The materials of SilentZPro 2.0 are the same as the reference device. The differences between the SilentZPro 2.0 and the predicate device do not raise different questions of safety and effectiveness. ### G. Performance Testing ### Non-Clinical Testing Testing results have been included include cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation (ISO 10993-10). The results of the testing demonstrated that the subject device is biocompatible and meets the applicable requirements of ISO 10993. Device materials were tested for various physical properties. Shinrin-Yoku Traders LLC conducted a risk analysis on SilentZPro 2.0 in accordance with ISO 14971:2007 and the FDA guidance document "Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea". All identified risks have been addressed through device design, biocompatibility tests or through labeling provided to the consumer. {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows a stylized graphic of three evergreen trees. The trees are arranged in a row, with the middle tree being slightly taller than the other two. Above the trees, there are several short, dashed lines radiating outwards in a semi-circular pattern, resembling sun rays. The entire graphic is in a dark green color. 18450 SW Seiffert Rd Sherwood, OR 97140 USA +1-971-712-9175 shinrinyokutradersllc@gmail.com www.itusguardz.com #### Clinical Performance Testing No clinical performance testing was conducted. #### Risk Analysis Shinrin-Yoku Traders LLC conducted a risk analysis on SilentZPro 2.0 in accordance with ISO 14971:2007 and the FDA guidance document "Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea" (November 12, 2002). All identified risks have been addressed through device design, biocompatibility tests or through labeling provided to the consumer. #### H. Conclusions Based on the similarities in intended use and indications for use, together with technologic similarities and results of performance testing, we believe that SilentZPro 2.0 is substantially equivalent to the predicate device, PureSleep® .
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...