K222872 · Vivos Therapeutics, Inc. · LRK · Dec 30, 2022 · Dental
Device Facts
Record ID
K222872
Device Name
DNA Appliance
Applicant
Vivos Therapeutics, Inc.
Product Code
LRK · Dental
Decision Date
Dec 30, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222872 · Dec 30, 2022
DNA Appliance
Vivos Therapeutics, Inc.
Peer-reviewed clinical literature; Clinical data (RWD)
The submission utilizes clinical data and peer-reviewed literature to demonstrate that the DNA appliance reduces the incidence of snoring and mild to moderate obstructive sleep apnea, supporting the substantial equivalence to the predicate device.
Adult patients with snoring and/or mild to moderate obstructive sleep apnea
Not applicable for this study
Reduction in snoring and incidence of apnea-hypopnea episodes
Indications for Use
The DNA Appliance is intended to reduce snoring and/or mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older.
Device Story
The DNA Appliance is a customized intraoral device designed to reduce snoring and mild to moderate obstructive sleep apnea (OSA) in adults. It consists of an upper tray, lower tray, or both, fabricated from standard orthodontic acrylics and wires. The device is fitted by a dentist in a clinical setting for patient use at home during sleep. It functions by applying gentle pressure to tissues at the back of the throat and aims to expand the nasal airway through jaw expansion and mid-facial redevelopment, thereby increasing airway patency. The device features adjustment mechanisms allowing for six degrees of freedom, including antero-posterior, transverse, and vertical dimension of occlusion adjustments. An optional extender on the back of the device further prevents airway collapse. By maintaining an open airway, the device reduces the incidence of apnea-hypopnea episodes, potentially benefiting patients by mitigating health risks associated with OSA, such as hypertension and heart disease.
Clinical Evidence
No new clinical testing was performed. The submission relies on peer-reviewed literature and real-world data (RWD) demonstrating that the DNA appliance reduces snoring and mild to moderate OSA. The device's efficacy is supported by comparisons to oral appliance therapy standards and the predicate device's clinical history.
Technological Characteristics
Custom-manufactured intraoral device; materials include standard orthodontic acrylics and stainless steel orthodontic wires; features orthodontic adjustment screws for six degrees of freedom (AP, TV, VDO); non-sterile; patient-specific design; no electronic components or software.
Indications for Use
Indicated for adult patients 18 years of age or older with snoring and/or mild to moderate obstructive sleep apnea.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
December 30, 2022
Vivos Therapeutics, Inc. % Colette Cozean Regulatory Consultant Colette Cozean, PhD 21581 Midcrest Dr. Lake Forest, California 92630
Re: K222872
Trade/Device Name: DNA Appliance Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: Class II Product Code: LRK, LOZ Dated: September 15, 2022 Received: September 22, 2022
Dear Colette Cozean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222872
Device Name
DNA Appliance
Indications for Use (Describe)
The DNA Appliance is intended to reduce snoring and/or mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5. 510(k) Summary- K222872
| Applicant: | Vivos Therapeutics<br>300 S. 5th Street<br>Murray, KY, 42017, USA<br>(270)-226-9237 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Colette Cozean, PhD<br>21581 Midcrest Drive<br>Lake Forest, CA 92630<br>(949) 855-2885<br>colettecozean@gmail.com |
| Date Prepared | September 15, 2022 |
| Proprietary Name | DNA Appliance |
| Common Name | Dental Device – Anti Snoring /Obstructive Sleep Apnea Device |
| Classification Name | Anti Snoring/ Obstructive Sleep Apnea Device |
| Proposed Product Code | LRK, LQZ |
| Primary Predicate Device | mRNA (K130067) |
| Reference Predicate Devices | mmRNA (K210203) |
## Device Description
The DNA appliance is an intraoral device to reduce snoring, and/or mild to moderate sleep apnea in adults 18 years of age and older. It consists of either an upper tray, or both, and is designed to open the airway during sleep. The device is customized to each patient, and features an adjustment mechanism to allow it to be further customized to each patient.
The device is identical to the already-marketed mRNA appliance except the top and bottom trays do not interlock to advance the jaw.
#### Scientific Principles
During sleep, the muscles in the tongue and back of the throat relax, which can cause them to sag and narrow the airway. Airflow through a narrow airway is the cause of snoring. When this narrowing of the airway is severe, it results in Obstructive Sleep Apnea (OSA), where the airway closes. This can happen up to hundreds of times during the night, lasting for a minute or longer.
{4}------------------------------------------------
With these closures, the brain detects the lack of oxygen and disturbs sleep to draw breath. In many cases, the individual isn't completely aware of the stoppages, which don't fully awaken the sleeper. Sleep apnea has been linked to major medical conditions, including hypertension, headaches, heart disease, diabetes, depression, and more.
#### Device Function
The DNA appliance is a customized oral device featuring a lower tray, upper tray, or both, depending on patient need. These trays put gentle pressure on the tissue at the back of the throat to prevent the airway from collapsing during sleep.
Studies have shown that customized oral devices that function by increasing the patency of the airway show comparable efficacy to continuous positive arway pressure (CPAP) devices, considered the gold standard of treatment for OSA (Oral appliance therapy in Olstructive Skep Apnea syndrome - A clinial study on therapentic outcomes Hoekema A PhD thesis, University Medical Centre Groningen Department of Oral and Maxillofacial Surgery. pp 110, 2007). On the basis of these studies, use of oral devices has been recommended by the American Academy of Sleep Medicine for patients with mild or moderate OSA, or for those with severe OSA who are unable to tolerate the CPAP device.
The DNA Appliance aims to expand the nasal airway through jaw expansion and mid-facial redevelopment. In doing so, an oral device may be able to permanently improve the oropharyngeal airway. Studies have shown that the DNA appliance can increase nasal cavity volume and reduce the incidence of apnea-hypopnea episodes.
The DNA appliance is customized on models of the patient's teeth, using standard orthodontic acrylics and standard orthodontic wires for clasps and retention. The DNA appliance allows for six degrees of freedom in customization, including antero-posterior (AP) adjustment, transverse (TV) adjustment, as well as permitting adjustments of the vertical dimension of occlusion (VDO).
The addition of an optional extender on the back of the device further prevents the patient's airway from collapsing during sleep.
## Intended Use
No change is requested in the indication for use as compared to the predicate "indication" device (mRNA), which remains: "To reduce snoring and/or mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older." In the predicate device submission for use just states adults, but then specifies the target population as 18 years of age and older as stated below.
Target Population: Patients 18 years of age and older with snoring, and /or mild to moderate obstructive sleep apnea.
Environment of Use: Fitting of the DNA appliance in the dental office for patient use at home.
Comparison to Predicate Devices: The DNA appliance with the extender is similar in both mechanism of action and design to the predicate mRNA device, except that it does not advance the mandible. The device allows the airway to remain open during sleep. The device is made of the same materials (standard dental acrylic, stainless steel orthodontic wires, and orthodontic adjustment screws).
The cleaning instructions, instructions for use, and labeling are identical to those currently used for the
{5}------------------------------------------------
already-marketed predicate device. These have been revised slightly since the original application for the predicate device, mRNA about 10 years ago. The precautions, warnings, risk analysis and other critical statements have not been substantially changed.
Shelf Life: The device is provided non-sterile. Shelf life will be identical to the predicate device. No shelf life is required as the device is custom-manufactured and immediately fitted to the patient by the dentist.
Non-clinical Testing: A risk analysis was performed, which considered soreness, obstruction of breathing, tooth movement, and breakage. The product was compared to predicate devices in each area to show the risks were equivalent to the predicate devices. No biocompatibility testing was done as all the components are the same as the predicate device.
Clinical Testing: This submittal relies on peer-reviewed literature and clinical data (RWD) demonstrating that DNA can reduce the incidence of snoring and/or mild to moderate sleep apnea, and raises no new questions of safety and efficacy as compared to mRNA, the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.