Double Tube Herbst Appliance

K243752 · The Tmj Clinic PC · LRK · Sep 17, 2025 · Dental

Device Facts

Record IDK243752
Device NameDouble Tube Herbst Appliance
ApplicantThe Tmj Clinic PC
Product CodeLRK · Dental
Decision DateSep 17, 2025
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Intended Use

to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults

Device Story

Double Tube Herbst Appliance is an intraoral device for mandibular advancement; used during sleep to reduce snoring and mild-to-moderate obstructive sleep apnea. Device consists of upper and lower acrylic base appliances connected bilaterally by a telescopic apparatus. Telescopic assembly includes a double tube and rod mechanism; allows for 15 mm of adjustment in 1/2 mm increments. Prescribing dentist adjusts the device; patient performs daily insertion and removal. Mandibular advancement maintains airway patency. Full arch coverage distributes forces to prevent tooth movement. Lateral range of motion is unrestricted to protect TMJs. Device is fabricated from sheet acrylic, cold cure acrylic resin, and stainless steel.

Clinical Evidence

Bench testing only. Testing assessed component force withstand capabilities compared to the primary predicate. No clinical data provided.

Technological Characteristics

Materials: sheet acrylic, cold cure acrylic resin, stainless steel. Design: bilateral telescopic apparatus with double tube assembly and rod. Adjustment: 15 mm range in 1/2 mm increments. Connectivity: None. Sterilization: Not specified (custom-fabricated dental appliance).

Indications for Use

Indicated for adults with mild to moderate obstructive sleep apnea and nighttime snoring.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 17, 2025 The Tmj Clinic PC John Summer Dentist, President 833 SW 11th Ave. # 810 Portland, Oregon 97205 Re: K243752 Trade/Device Name: Double Tube Herbst Appliance Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK, LQZ Dated: December 5, 2024 Received: December 5, 2024 Dear John Summer: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243752 - John Summer Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243752 - John Summer Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K243752 Device Name double tube herbst appliance Indications for Use (Describe) to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} | 510(k) #: K243752 | 510(k) Summary | Prepared on: 2025-09-08 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | the tmj clinic PC | | | Applicant Address | 833 SW 11th Ave. # 810 portland or 97205 United States | | | Applicant Contact Telephone | 503-329-1810 | | | Applicant Contact | Dr. john summer | | | Applicant Contact Email | john.summer03@gmail.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | double tube herbst appliance | | | Common Name | double tube herbst appliance | | | Classification Name | class 2 special controls | | | Regulation Number | 21 CFR 872.5570 | | | Product Code(s) | Irk, 872.5570 | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K130130 | AT G /M-O SA APPLANCE | Irk | | K222127 | soft palate elevator, reference device | LQZ | | K143244 | Panthera Anti-Snoring Device, reference device | LRK | | Device Description Summary | | 21 CFR 807.92(a)(4) | | SUBMITTER AND CONTACT: John Summer, 833 SW 11th Ave # 810, Portland OR 97205 Phone: 503-241-7353 fax: 503-525-2966 Email:john.summer03@gmail.com DEVICE NAME: DOUBLE TUBE HERBST APPLIANCE REGULATION NUMBER: 21 CFR 872.5570 CLASSIFICATION: CLASS 2, SPECIAL CONTROLS GUIDANCE DOCUMENT: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea SCOPE - Intraoral devices for snoring and/or obstructive sleep apnea. PRODUCT CODES - LRK and LQZ | | | | MATERIALS/BIOCOMPATIBILITY - Materials include sheet acrylic, cold cure acrylic resin, and stainless steel, - the exact same materials as are in the predicates, in about the same proportions. The biocompatibility of these materials in the intended use environment poses no unacceptable risks. NARRATIVE DESCRIPTION - The Double Tube Herbst appliance consists of upper and lower base appliances connected on each side by a telescopic apparatus that advances the mandible. Each telescopic apparatus is loosely but securely attached to wire loops formed at the ends of retentive anchors embedded in the acrylic on the outer (buccal) surfaces of the upper and lower base appliances. A telescopic apparatus consists of a double tube telescopic assembly that is attached to the back end of the upper wire loop; and a rod that is attached to the front end of the lower wire loop. The double tube telescopic assembly consists of a short internally threaded tube | | | {5} welded in parallel onto the back of a long straight tube in which the rod slides. The short internally threaded tube is adjustably attached to the back end of the upper wire loop by engaging the externally threaded lower end of a connector, the upper end of which ends in a helix that is twisted around the upper wire loop. The rod that slides in and out of the long tube is loosely but securely attached to the wire loop on the front end of the lower retentive anchor. The lengths of the telescopic assemblies can be adjusted in 1/2 mm increments along a range of 15 mm by disengaging the telescopic components, rotating the double tube assembly around the connecting arm, and sliding the rod back into the long tube to lock in the adjustment. INTENDED USE – The Double Tube Herbst appliance is worn while sleeping to keep the mandible in a forward position prescribed by the treating dentist to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults. The customized appliance is inserted and removed by the patients every night, and it is adjusted by the prescribing dentist. RISKS – The risk of causing damage to TMJs and teeth is mitigated by the unrestricted lateral range of motion and the easy adjustability of the lengths of the telescopic assemblies without tools coupled with the patient instructions to return to the treating dentist to have it adjusted backward ("walkback") as soon as symptoms arise. The risk of loosening or flaring of the lower teeth or general tooth movement is mitigated in Herbst appliances by full arch coverage which distributes the forces generated by mandibular advancement all around the upper and lower dental arches and thereby prevents excessive force from being delivered to one or more teeth. The risk of dislodging a part that could cause choking or aspiration is mitigated by the retentive design of the anchors, the minimal number of parts, and their secure connections. PERFORMANCE DATA – Bench testing was performed to assess the forces that the components are able to withstand and to compare them to the forces that the primary predicate are able to withstand. PRIMARY PREDICATE DEVICE– is the ATG/SM-OSA APPLIANCE, K130130, cleared in August 2013 and currently widely available. It has the same basic technological characteristics, indications for use, intended use, and mode of action as the subject device. It differs from the subject device in the methodology for attaching the telescopic components to the base appliance and the methodology for adjusting the lengths of the telescopic assemblies. REFERENCE DEVICES – Panthera Anti-Snoring Device, K 143244 for recommended maximum advancement. Soft palate elevator K222127 for identical chemical composition for biocompatibility. COMPARATIVE SAFETY – Compared to the predicate, the subject device poses the same risk of causing TMJ or tooth damage due to the unrestricted lateral range of mandibular movement, the same risk of damage to mucous membranes lining the cheeks and lips due to a lower profile, the same risk of a part breaking free and causing choking or aspiration due to the number of parts and their secure connections, the same risk of toxic leaching due to the same materials and use conditions, and about the same risks of misuse and abuse. COMPARATIVE EFFECTIVENESS – The subject device has the same effectiveness as the predicate, because it has the same mode of action. RISK/BENEFIT ANALYSIS – The potential benefits are significant, because obstructive sleep apnea causes so many serious health problems. The risks generally involve the dentition and its supporting tissues, which must withstand the forces generated by the treatment on a long-term basis. CONCLUSION – The Double Tube Herbst Appliance is similar in intended use and technological characteristics as the predicate device, -ATG/SM-OSA Appliance (K130130). Based on the performance testing, the Double Tube Herbst Appliance is shown to be substantially equivalent. | Intended Use/Indications for Use | 21 CFR 807.92(a)(5) | | --- | --- | | to reduce nighttime snoring and mild to moderate obstructive sleep apnea in adults | | | Indications for Use Comparison | 21 CFR 807.92(a)(5) | | same indications for use | | {6} Technological Comparison 21 CFR 807.92(a)(6) The subject and predicate devices have the same intended use, indications for use, mode of action, basic technological characteristics, performance records, and user interface. They differ in the methodology for attaching the telescopic components to the base appliances and the methodology for adjusting the effective lengths of the telescopic assemblies. These differences from the predicate are designed to increase both safety and effectiveness in the subject device. They are discussed in detail in the attachment below. Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) physical properties testing was performed by Engineering Materials Lab Not Applicable
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...