ProSomnus EVO Sleep and Snore Device, ProSomnus EVO Sleep and Snore Device with Patient Monitoring
K202529 · Prosomnus Sleep Technologies, Inc. · LRK · Nov 20, 2020 · Dental
Device Facts
Record ID
K202529
Device Name
ProSomnus EVO Sleep and Snore Device, ProSomnus EVO Sleep and Snore Device with Patient Monitoring
Applicant
Prosomnus Sleep Technologies, Inc.
Product Code
LRK · Dental
Decision Date
Nov 20, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProSomnus® EVO Sleep and Snore Device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. Optionally, if the DentiTrac micro-recorder is completely embedded in the ProSomnus® EVO Sleep and Snore Device. the micro-recorder is intended to measure patient compliance therapy in combination with the DentiTrac System.
Device Story
Oral appliance for snoring and mild-to-moderate OSA; repositions mandible forward to increase pharyngeal airway space. CAD/CAM designed; digitally milled medical-grade polymer arches; twin-mated posts; no mechanical adjustment components (pistons, screws, straps). Patient-specific; prescribed by physician; used at home or sleep labs. Optional embedded DentiTrac micro-recorder logs wear time for compliance monitoring. Provider uses device to manage airway; patient wears device during sleep. Benefits include reduced snoring and improved sleep apnea symptoms.
Clinical Evidence
Bench testing only. Evaluated anterior-posterior shear strength, stroke, directional torque, lateral shear strength, and durability of hooks. All tests passed against acceptance criteria and compared favorably to predicate device performance.
Technological Characteristics
Medical grade polymer construction; CAD/CAM design; digitally milled; twin-mated posts; non-sterile. Sensing: optional embedded DentiTrac micro-recorder for compliance monitoring. Standards: ISO 10993-1, -2, -5, -10, -11, -12; ISO 7405; ISO 14971.
Indications for Use
Indicated for adult patients with mild to moderate obstructive sleep apnea (OSA) and snoring. Used to reduce snoring and OSA by repositioning the mandible.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
ProSomnus® MicrO2 Obstructive Sleep Apnea Device with Micro-Recorder (K161624)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. To the right of the blue box is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2020
ProSomnus Sleep Technologies, Inc. Divya Mavalli Ouality and Regulatory Manager 5860 West Las Positas Blvd. Pleasanton, California 94588
Re: K202529
Trade/Device Name: ProSomnus EVO Sleep and Snore Device, ProSomnus EVO Sleep and Snore Device with Patient Monitoring Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK, LQZ, PLC Dated: October 23, 2020 Received: October 26, 2020
Dear Divya Mavalli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K202529
#### Device Name
ProSomnus® EVO Sleep and Snore Device ProSomnus® EVO Sleep and Snore Device with Patient Monitoring
#### Indications for Use (Describe)
The ProSomnus® EVO Sleep and Snore Device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Optionally, if the DentiTrac micro-recorder is completely embedded in the ProSomnus® EVO Sleep and Snore Device. the micro-recorder is intended to measure patient compliance therapy in combination with the DentiTrac System.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo features a blue circular design on the left, followed by the company name "ProSomnus" in a larger, bolder blue font. Below "ProSomnus" is the text "SLEEP TECHNOLOGIES" in a smaller, lighter blue font. The logo is clean and professional, suggesting a focus on sleep-related technology.
# K202529 510(K) SUMMARY
## 1. Manufacturer and Submitter
| Company Name: | ProSomnus® Sleep Technology, Inc |
|------------------|-----------------------------------------------------------------------|
| Company Address: | 5860 West Las Positas Blvd.<br>Suite 25<br>Pleasanton, CA 94588 |
| Contact Person: | Divya Mavalli<br>Email: dmavalli@prosomnus.com<br>Phone: 925.307.5337 |
| Date Prepared: | October 23, 2020 |
# 2. Device Name and Classification
3.
| Trade/Proprietary Name: | ProSomnus® EVO Sleep and Snore Device<br>ProSomnus® EVO Sleep and Snore Device with Patient<br>Monitoring |
|-------------------------|---------------------------------------------------------------------------------------------------------------|
| Common Name: | Mandibular Advancement Device |
| Classification Name: | Intraoral devices for snoring and intraoral devices for snoring<br>and obstructive sleep apnea. |
| Regulatory Class: | II |
| Regulatory Number: | 21 CFR 872.5570 |
| Product Code: | LRK (Device, Anti-snoring), LQZ (Device, Jaw Repositioning)<br>PLC (Sleep Appliances with Patient Monitoring) |
| Panel: | Dental |
| Predicate Device | |
| Device Name: | ProSomnus® MicrO2 Obstructive Sleep Apnea Device |
| 510(k) Number: | K133683 |
| Reference Device | |
| Device Name: | ProSomnus® MicrO2 Obstructive Sleep Apnea Device with<br>Micro-Recorder |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left, with the text "ProSomnus" in blue to the right of the circle. Below "ProSomnus" is the text "Sleep Technologies" in a smaller font, also in blue. The logo is clean and modern, with a focus on the company's name and its area of expertise.
510(k) Number: K161624
### 4. Device Description
The ProSomnus® EVO Sleep and Snore Device is an oral device, which improves the flow of air through the patient's pharyngeal space during sleep by repositioning the mandible. The ProSomnus® EVO Sleep and Snore Device and the ProSomnus® EVO Sleep and Snore Device with Patient Monitoring consists of maxillary and mandibular devices that when interfaced together reduce snoring and mild to moderate obstructive sleep apnea by holding the mandible forward during sleep, providing increased pharyngeal space. These separate upper and lower device arches are CAD/CAM designed with twin-mated posts and are digitally milled to be patient-specific according to physician prescription. Customizable ergonomic features are available to improve patient comfort. Prescribed advancements can be achieved by simply removing the current upper or lower device arch and inserting the next upper or lower device arch in the mandibular advancement series. The device does not have any adjustment mechanisms to modify or maintain the mandibular position such as pistons, screws, straps, or repositioning elastics. The device is supplied nonsterile.
A micro-recorder is completely embedded in the ProSomnus® EVO Sleep and Snore Device with Patient Monitoring. The micro-recorder is a compliance sensor, which logs the time the device is worn.
### 5. Intended Use/Indication for Use
The ProSomnus® EVO Sleep and Snore Device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Optionally, if the DentiTrac micro-recorder is completely embedded in the ProSomnus® EVO Sleep and Snore Device, the micro-recorder is intended to measure patient compliance to oral device/appliance therapy in combination with the DentiTrac System.
### 6. Technological Characteristics
Both the predicate and subject devices use upper and lower device arches to reposition the mandible.
The below table provides a comparison of technological characteristics to models of the predicate and reference devices.
| Characteristics | ProSomnus®<br>EVO Sleep and<br>Snore Device<br>(Subject<br>Device) | ProSomnus®<br>EVO Sleep and<br>Snore Device<br>with Patient<br>Monitoring | ProSomnus<br>MicrO2<br>Obstructive<br>Sleep Apnea<br>Device | ProSomnus<br>MicrO2<br>Obstructive<br>Sleep Apnea |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | (Subject Device<br>with Micro-Recorder) | (Predicate<br>Device) | Device with<br>Micro-Recorder<br>(Reference<br>Device) |
| Body Material | Medical Grade<br>Polymer | Medical Grade<br>Polymer | Polymethyl<br>methacrylate<br>(PMMA) | Polymethyl<br>methacrylate<br>(PMMA) |
| Usability | Single-patient,<br>multiple use | Single-patient,<br>multiple use | Single-patient,<br>multiple use | Single-patient,<br>multiple use |
| Splint | Comprised of<br>two or more<br>customizable<br>splints (upper<br>and lower) | Comprised of<br>two or more<br>customizable<br>splints (upper<br>and lower) | Comprised of<br>two or more<br>customizable<br>splints (upper<br>and lower) | Comprised of<br>two or more<br>customizable<br>splints (upper<br>and lower) |
| Technical Method | Advances the<br>patient's<br>mandible in a<br>forward<br>position to<br>increase<br>airway space | Advances the<br>patient's<br>mandible in a<br>forward<br>position to<br>increase airway<br>space and<br>additionally<br>monitors the<br>patient<br>compliance to<br>oral appliance<br>therapy | Advances the<br>patient's<br>mandible in a<br>forward<br>position to<br>increase airway<br>space | Advances the<br>patient's<br>mandible in a<br>forward<br>position to<br>increase airway<br>space and<br>additionally<br>monitors the<br>patient<br>compliance to<br>oral appliance<br>therapy |
| Ergonomic<br>Features - Splint<br>Design | The scalloped<br>and contoured<br>edge allows for<br>patient<br>comfort | The scalloped<br>and contoured<br>edge allows for<br>patient<br>comfort | Straight-line<br>plane at the<br>edge of the<br>splint around<br>the front of the<br>teeth | Straight-line<br>plane at the<br>edge of the<br>splint around<br>the front of the<br>teeth |
| Ergonomic<br>Features - Post<br>Style | The dual radius<br>posts to allow<br>for patient<br>comfort | The dual radius<br>posts to allow<br>for patient<br>comfort | The post angle<br>to the splint is a<br>standard 90<br>degrees | The post angle<br>to the splint is a<br>standard 90<br>degrees |
| Ergonomic<br>Features -<br>Anterior Airway | Anterior<br>airway opening | Anterior airway<br>opening | No anterior<br>airway opening | No anterior<br>airway opening |
| Ergonomic<br>Features -<br>Anterior<br>Discluder | The prescribing<br>provider can<br>request the<br>addition of a<br>vertical<br>element | The prescribing<br>provider can<br>request the<br>addition of a<br>vertical<br>element | No discluder<br>option is<br>provided | No discluder<br>option is<br>provided |
| Ergonomic<br>Features -<br>Lingual Coverage | Prescriber may<br>request<br>posterior or<br>full lingual<br>coverage | Prescriber may<br>request<br>posterior or full<br>lingual<br>coverage | No posterior or<br>full lingual<br>coverage<br>provided | No posterior or<br>full lingual<br>coverage<br>provided |
| Ergonomic<br>Features - Plastic<br>Hooks | Plastic hooks<br>incorporated<br>into the design | Plastic hooks<br>incorporated<br>into the design | Uses stainless<br>steel balls or<br>hooks to secure<br>rubber bands in<br>those<br>prescriptions<br>where they are<br>indicated | Uses stainless<br>steel balls or<br>hooks to secure<br>rubber bands in<br>those<br>prescriptions<br>where they are<br>indicated |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left, followed by the text "ProSomnus" in a larger, bolder blue font. Below "ProSomnus" is the text "SLEEP TECHNOLOGIES" in a smaller, lighter blue font. The logo is clean and modern, suggesting a focus on technology and sleep solutions.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left, followed by the text "ProSomnus" in a larger, bolder blue font. Below "ProSomnus" is the text "SLEEP TECHNOLOGIES" in a smaller, lighter blue font. The overall design is clean and professional, conveying a sense of innovation and expertise in the field of sleep technology.
# 7. Performance Data
Product bench testing was performed to support the above indications for use as well as the claim of Substantial Equivalence.
| Performance<br>Specification | Subject Device<br>ProSomnus®<br>EVO Sleep and<br>Snore Device | Predicate Device<br>ProSomnus<br>MicrO2 OSA<br>Device | Acceptance<br>Criteria | Results<br>(Pass/Fail) |
|----------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------|------------------------|
| Anterior-<br>Posterior (AP)<br>Shear Strength<br>(lbf) | 235.4 | 239.8 | Must be greater<br>than 200 lbf | Pass |
| Stroke (in) | 0.17 | 0.11 | Must be greater<br>than 0.1 | Pass |
| Anterior-<br>Posterior<br>Directional<br>Torque (lbf-in) | $235.4 \text{ lbf } * 0.17 \text{ in } = 40.0 \text{ lbf-in}$ | $239.8 \text{ lbf } * 0.11 \text{ in } = 26.4 \text{ lbf-in}$ | Must be equal<br>or better than<br>predicate<br>device | Pass |
| Lateral<br>Direction Shear<br>Strength (lbf) | 71.5 | 94.2 | Must be greater<br>than 50 lbf | Pass |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left, followed by the company name "ProSomnus" in a larger, bolder blue font. Below "ProSomnus", the words "SLEEP TECHNOLOGIES" are written in a smaller, lighter blue font. The overall design is clean and professional, conveying a sense of innovation and expertise in the field of sleep technology.
| Stroke (in) | 0.17 | 0.11 | Must be greater<br>than 0.1 | Pass |
|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|------|
| Lateral<br>Directional<br>Torque (Ibf-in) | $71.5 \text{ lbf } * 0.17 \text{ in } = 12.2 \text{ lbf-in}$ | $94.2 \text{ lbf } * 0.11 \text{ in } = 10.4 \text{ lbf-in}$ | Must be equal<br>or better than<br>predicate<br>device | Pass |
| Durability when<br>elastic bands<br>pull on hooks at<br>peak force (mm<br>& lbs.) | 135.55 mm<br>stretch length<br>with an average<br>peak force of<br>5.06 lbs. With<br>no visible<br>damage to the<br>device and its<br>metal-free<br>hooks. | 135.55 mm stretch<br>length with an<br>average peak force<br>of 4.74 lbs. With<br>no visible damage<br>to the device and<br>its metal-free<br>hooks. | Must be equal<br>or more durable<br>then predicate<br>device | Pass |
# Non-Clinical Performance Testing
Product biocompatibility testing and risk analysis were performed, and the following standards were utilized:
- > ISO 10993-1: 2018 Biological Evaluation of medical devices Part 1: Evaluation and testing within a risk management process
- A ISO 10993-2 Second edition 2006-07-15 Biological Evaluation of medical devices -Part 2: Animal welfare requirements
- > ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity
- > ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
- > ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
- > ISO 10993-12 Fourth edition 2012-07-01 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
- > ISO 7405: 2018 Dentistry Evaluation of biocompatibility of medical devices used in dentistry
- > ISO 14971:2012 Medical Devices Application of Risk Management to Medical Devices
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left and the company name on the right. The circular design is made up of three curved lines that form a circle. The text "ProSomnus" is in a larger, bolder font, while the text "Sleep Technologies" is in a smaller font below it.
# 8. Substantial Equivalence
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left, followed by the text "ProSomnus" in a larger, bolder blue font. Below "ProSomnus" is the text "SLEEP TECHNOLOGIES" in a smaller, lighter blue font. The logo is clean and modern, suggesting a focus on technology and sleep solutions.
| Substantially<br>Equivalent<br>Connection | ProSomnus®<br>EVO Sleep and<br>Snore Device | ProSomnus®<br>EVO Sleep and<br>Snore Device<br>with Patient<br>Monitoring | ProSomnus<br>Micro2<br>Obstructive<br>Sleep Apnea<br>Device | ProSomnus<br>Micro2<br>Obstructive<br>Sleep Apnea<br>Device with<br>Micro-Recorder |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K202529 | K202529 | K133683 | K161624 |
| Manufacturer | ProSomnus<br>Sleep<br>Technologies | ProSomnus<br>Sleep<br>Technologies | ProSomnus<br>Sleep<br>Technologies | ProSomnus<br>Sleep<br>Technologies |
| Relation | Subject Device | Subject Device<br>with Patient<br>Monitoring | Predicate<br>Device | Predicate<br>Device with<br>Micro-Recorder |
| Device | Device, Anti-<br>Snoring<br>Device, Jaw<br>Repositioning | Sleep<br>Appliances with<br>Patient<br>Monitoring | Device, Anti-<br>Snoring | Sleep<br>Appliances with<br>Patient<br>Monitoring |
| Product Code | LRK<br>LQZ | PLC | LRK | PLC |
| Indications for<br>Use | The device is<br>intended to<br>reduce night<br>time snoring<br>and mild to<br>moderate<br>obstructive<br>sleep apnea<br>(OSA) in adults. | The device is<br>intended to<br>reduce night<br>time snoring<br>and mild to<br>moderate<br>obstructive<br>sleep apnea<br>(OSA) in adults. | The device is<br>intended to<br>reduce night<br>time snoring<br>and mild to<br>moderate<br>obstructive<br>sleep apnea<br>(OSA) in adults. | The device is<br>intended to<br>reduce night<br>time snoring<br>and mild to<br>moderate<br>obstructive<br>sleep apnea<br>(OSA) in adults. |
| Usability | Single patient,<br>multiple use | Single patient,<br>multiple use | Single patient,<br>multiple use | Single patient,<br>multiple use |
| Target<br>Population | Adult patients | Adult patients | Adult patients | Adult patients |
| Where Used | For personal<br>use at home or<br>in sleep<br>laboratories | For personal<br>use at home or<br>in sleep<br>laboratories | For personal<br>use at home or<br>in sleep<br>laboratories | For personal<br>use at home or<br>in sleep<br>laboratories |
| Prescription | Yes | Yes | Yes | Yes |
| Principle of<br>Operation | Mandibular<br>Repositioners | Mandibular<br>Repositioners | Mandibular<br>Repositioners | Mandibular<br>Repositioners |
| Non-Sterile | Yes | Yes | Yes | Yes |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for ProSomnus Sleep Technologies. The logo features a blue circular graphic on the left, followed by the text "ProSomnus" in a larger, bolder blue font. Below "ProSomnus", the words "SLEEP TECHNOLOGIES" are written in a smaller, lighter blue font. The overall design is clean and professional, suggesting a focus on sleep-related technology and solutions.
| Design | Allows space<br>for tongue and<br>mandibular<br>movement of<br>mouth opening<br>and closing<br>during wear | Allows space<br>for tongue and<br>mandibular<br>movement of<br>mouth opening<br>and closing<br>during wear | Allows space<br>for tongue and<br>mandibular<br>movement of<br>mouth opening<br>and closing<br>during wear | Allows space<br>for tongue and<br>mandibular<br>movement of<br>mouth opening<br>and closing<br>during wear |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Adjustment<br>Mechanism | No adjustment<br>mechanisms<br>such as pistons,<br>screws, straps,<br>or repositioning<br>elastics are<br>used to modify<br>or maintain the<br>mandibular<br>position | No adjustment<br>mechanisms<br>such as pistons,<br>screws, straps,<br>or repositioning<br>elastics are<br>used to modify<br>or maintain the<br>mandibular<br>position | No adjustment<br>mechanisms<br>such as pistons,<br>screws, straps,<br>or repositioning<br>elastics are<br>used to modify<br>or maintain the<br>mandibular<br>position | No adjustment<br>mechanisms<br>such as pistons,<br>screws, straps,<br>or repositioning<br>elastics are<br>used to modify<br>or maintain the<br>mandibular<br>position |
# 9. Conclusion
Based on the comparison of technology and the indications for use, we find that ProSomnus® EVO Sleep and Snore Device is substantially equivalent to ProSomnus® MicrO2 Obstructive Sleep Apnea Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.