Post-market commercial sales data (outside the US)
The sponsor utilized real-world clinical experience from international commercial distribution (since October 2014) to support the device's performance, durability, and substantial equivalence to the predicate device.
Commercial experience; Real-time clinical experience; Post-market data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
International commercial experience; Post-market observational experience; Follow-up/Duration: Since October 2014; Study Period: October 2014 to 2016
Adult patients using the device for snoring and obstructive sleep apnea
Not applicable for this study
Device performance, cleaning efficacy, and mechanical durability
Indications for Use
The BluePro® is a mandibular repositioning device intended to reduce or alleviate night-time snoring and mild to moderate obstructive sleep apnea in adult patients 18 years or older.
Device Story
BluePro is a mandibular repositioning device (MRD) for snoring and mild to moderate obstructive sleep apnea. Device consists of upper and lower pre-formed trays; customized via 'boil and bite' method by dentist. Operates by holding mandible in advanced position; integral interlocking mechanism allows for fixed advancement up to 10 mm. Used in home, dental/physician offices, or sleep labs. Patient wears device at night; device allows lateral and vertical jaw movement. Benefits include reduction of snoring and apnea symptoms. Cleaning via water rinse and toothbrush. No electronic or software components.
Clinical Evidence
No clinical testing. Evidence based on non-clinical performance testing, mechanical durability testing (simulated 1-year use via Braem testing), and commercial experience outside the US since 2014.
Technological Characteristics
Mandibular repositioning device; pre-formed thermoplastic trays; 'boil and bite' customization; integral interlocking advancement mechanism; 10 mm maximum adjustment; surface-contacting (mucosal membrane) materials; biocompatibility per ISO 10993 (cytotoxicity, sensitization, irritation).
Indications for Use
Indicated for adult patients 18 years or older with mild to moderate obstructive sleep apnea or snoring. Contraindicated for patients with central sleep apnea, severe respiratory disorders, loose teeth, advanced periodontal disease, loose dental work/dentures, or patients under 18 years of age.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol of medicine, with three figures representing health, hope, and well-being. The seal is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 15, 2016
Bluesom Paul Dryden Consultant BlueSom 24 Rue Leon Gaumont Orvault, 44700 FR
Re: K162192 Trade/Device Name: BluePro® Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: November 14, 2016 Received: November 15, 2016
Dear Mr. Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162192
Device Name
### BluePro®
Indications for Use (Describe)
The BluePro® is a mandibular repositioning device intended to reduce or alleviate night-time snoring and mild to moderate obstructive sleep apnea in adult patients 18 years or older.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
### BLUESOM
24 rue Léon Gaumont 44700 Orvault FRANCE Tel - +33 2 28 25 61 43
| Official Contact: | Erwan Floch - General Manager | |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | BluePro® | |
| Common/Usual Name: | Intra-oral appliance | |
| Classification Name: | LRK - Device, anti-snoring, Intraoral devices for snoring<br>and intraoral devices for snoring and obstructive sleep apnea<br>21 CFR 872.5570, Class 2 | |
| Predicate Device: | K113569 – Apnea Sciences – ApneaRx | |
| | Predicate<br>ApneaRx | Subject Device<br>BluePro® |
| | K113569 | K162192 |
| Indications for Use | The Apnea Sciences Corporation<br>"ApneaRx" is intended for use on<br>adult patients 18 years of age or older<br>as an aid for the reduction of mild to<br>moderate obstructive sleep apnea<br>(OSA), and/or snoring. | The BluePro® is a mandibular<br>repositioning device intended to<br>reduce or alleviate night-time snoring<br>and mild to moderate obstructive<br>sleep apnea in adult patients 18 years<br>or older. |
| Environments of use | Home, dental and Physician offices,<br>Sleep laboratories | Home, dental and Physician offices,<br>Sleep laboratories |
| Patient Population | Adult patients 18 years and older | Adult patients 18 years and older |
| Contraindications | • have central sleep apnea<br>• have severe respiratory disorders<br>• have loose teeth or advanced<br>periodontal disease<br>• are under 18 years of age | • have central sleep apnea<br>• have severe respiratory disorders<br>• have loose teeth or advanced<br>periodontal disease<br>• are under 18 years of age |
| Prescription | Prescription use | Prescription use |
| Single patient, multi-use | Yes | Yes |
| Limitation of duration of use | No limitation | No limitation |
| Principle of operation / means of<br>mandibular advancement | Adjustment of the relative position of<br>the trays by the use of fixed<br>advancement. | Adjustment of the relative position of<br>the trays by the use of fixed<br>advancement. |
| Fixed tray sizes | Yes | Yes |
| Method to advance the mandible | Integral interlocking advancement<br>method | Integral interlocking advancement<br>method |
| Allows lateral and vertical movement | Yes | Yes |
| Employs boil and bite design | Yes | Yes |
| Maximum adjustment by the user | 10 mm | 10 mm |
| Works by holding lower jaw forward | Yes | Yes |
| Cleaned by simple rinsing with water | Yes | Yes |
| Cleaning with toothbrush | Yes | Yes |
| Biocompatibility and Patient contact | Surface Contacting, Mucosal<br>membrane, with duration of use<br>prolonged > 24 hours < 30 days, per<br>FDA Guidance Class II Special<br>Controls Guidance Document:<br>Intraoral Devices for Snoring and/or<br>Obstructive Sleep Apnea; Guidance<br>for Industry and FDA dated<br>11/12/2002 | ISO 10993<br>Cytotoxicity<br>Sensitization<br>Irritation<br>Surface Contacting, Mucosal<br>membrane, with duration of use<br>prolonged > 24 hours < 30 days, per<br>FDA Guidance Intraoral Devices for<br>Snoring and/or Obstructive Sleep<br>Apnea dated 11/12/2002 |
| Age testing | | Yes |
| Real-time clinical experience | | Yes |
### Device Description
The BluePro® oral appliance design concept is based upon the use of a standard set of upper and lower trays, referred also as splints in documentation, that are then customized for fit by a dentist. The design is commonly referred to as a Mandibular Repositioning Device (MRD).
The principle of advancing a lower or upper tray so that it advances the mandible for the treatment of snoring and / or mild to moderate obstructive sleep apnea is well known and there are a number of predicate devices.
### Indications for Use
The BluePro® is a mandibular repositioning device intended to reduce or alleviate night-time snoring and mild to moderate obstructive sleep apnea in adult patients 18 years or older.
### Contraindications
The device is contraindicated for patients who:
- . have central sleep apnea
- . have severe respiratory disorders
- . have loose teeth or advanced periodontal disease
- . have loose dental work, dentures, or other oral conditions which would be adversely affected by wearing dental appliances
- . are under 18 years of age
#### Warnings
Use of this device may cause:
- tooth movement or changes in dental occlusion
- gingival or dental soreness
- . pain or soreness to the temporomandibular joint
- . obstruction of oral breathing
- . excessive salivation
- . excessive dry mouth
- . loosening of dental restorations
- . loosening of teeth
{5}------------------------------------------------
## Environment of Use
Home, Dental and Physician offices, and Sleep laboratories
# Predicate Device Comparison:
Table 1 compares the predicate and the subject device.
{6}------------------------------------------------
### K162192 510(k) Summary 7-Dec-16
The BluePro® is viewed as substantially equivalent to the predicate device because:
Indications - Similar to predicate - ApneaRx - K113569; indicated for treating snoring and mild to moderate obstructive sleep apnea (OSA).
Discussion - The indications for use between the subject device and predicate are similar and therefore they can be found as substantially equivalent.
Technology / Principle of Operation - Similar to predicate - ApneaRx - K113569. Both devices use separate pre-formed trays with a means to advance the mandible / lower jaw. Discussion - Both devices use separate pre-formed trays with a means to advance the mandible / lower jaw are similar and therefore they can be found as substantially equivalent.
Environment of Use - Similar to predicate - ApneaRx - K113569. They are used in Home, Dental offices, Physician offices, Sleep laboratories.
Discussion - Both devices have similar environments of use and therefore they can be found as substantially equivalent.
Patient Population - Similar to predicate - ApneaRx - K113569. 18 years and older Discussion - The patient population is similar and therefore they can be found as substantially equivalent.
### Non-clinical performance testing
#### Commercial experience
The BluePro® has been marketed outside the US since October 2014. The results of this experience support the substantial equivalence of the BluePro® for the intended use, as compared to the predicate device.
The observations support:
- Cleaning via rinsing with water and use of a toothbrush ●
- . Real-time clinical experience from sales outside the US
- . Mechanical and durability after real-time use plus simulated 1 year use via the Braem testing
### Biocompatibility of Materials
All the materials are considered per FDA G95-1 as Surface Contacting, Mucosal membrane, with duration of use prolonged > 24 hours < 30 days, per FDA Guidance Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA dated 11/12/2002
Based upon the Guidance the ISO 10993-1 testing would be:
- Cytotoxicity .
- Sensitization .
- . Irritation
Testing was performed on finished, final devices.
Clinical No clinical testing.
{7}------------------------------------------------
### K162192 510(k) Summary 7-Dec-16
## Discussion of Differences and Conclusion
The subject and predicate devices share equivalent intended use. that of treating snoring and mild to moderate obstructive sleep apnea (OSA). Both the subject and predicate devices share equivalent technology, in that they both use separate pre-formed trays with a means to advance the mandible / lower jaw. Both the subject and predicate devices employ a boil and bite design to provide a custom impression for each patient. Both devices have similar environments of use, patient populations and biocompatibility.
Any differences between the subject and predicate devices are not significant and does not affect the substantial equivalence of the proposed device to the predicate device.
### Substantial Equivalence Conclusion
Based upon the performance testing and comparison to the legally marketed predicate device for indications for use, technology, and performance we believe we have demonstrated that the BluePro® is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.