K143244 · Panthera Dental, Inc. · LRK · Dec 15, 2014 · Dental
Device Facts
Record ID
K143244
Device Name
The Panthera Anti-Snoring Device
Applicant
Panthera Dental, Inc.
Product Code
LRK · Dental
Decision Date
Dec 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Panthera Anti-Snoring Device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Device Story
The Panthera Anti-Snoring Device is a custom-fabricated, removable intraoral mandibular repositioning device. It consists of two splints that fit over the upper and lower teeth, connected by adjustable rods. The device is prescribed by a physician or dentist and fitted in a dental office. By maintaining the lower jaw in a forward position, the device increases pharyngeal space to improve airflow during sleep. The dentist adjusts the mandibular advancement using interlocking rods (15 mm range in 1 mm increments) to achieve optimal therapeutic effect. The device is manufactured using CAD/CAM and selective laser sintering of polyamide type 12 polymer. It is intended for home use by the patient to alleviate snoring and mild to moderate OSA.
Clinical Evidence
No clinical data. Substantial equivalence was established through non-clinical bench testing, including physical property assessment and biocompatibility evaluation based on material similarity to the predicate device.
Technological Characteristics
Custom-fabricated intraoral splints; material: polyamide type 12; manufacturing: CAD/CAM and selective laser sintering; mechanism: mechanical mandibular advancement via interlocking rods; adjustment: 15 mm range in 1 mm increments; vertical opening: up to 4 mm; non-sterile; reusable; metal-free.
Indications for Use
Indicated for adult patients with snoring and mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 15,2014
Panthera Dental, Inc. c/o Mr. David Yungvirt Third Party Review Group, LLC 45 Rockefeller Plaza, Suite 2000 New York, NY 10111
Re: K143244
Trade/Device Name: The Panthera Anti-Snoring Device Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: II Product Code: LRK Dated: December 5, 2014 Received: December 8, 2014
Dear Mr. Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K143244
Device Name
The Panthera Anti-Snoring Device
Indications for Use (Describe)
The Panthera Anti-Snoring Device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position above the words "PANTHERA DENTAL" in purple. The word "PANTHERA" is in a larger font than the word "DENTAL", and the two words are stacked on top of each other.
# 5 .- 510 (k) Summary
[As required by 21 CFR 807.92]
| Date Prepared: | 31 October, 2014 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Panthera Dental Inc.<br>2035 rue du Haut-Bord<br>Québec (QC) G1N 4R7<br>Canada<br>Fax: +1 (418) 431-9942 |
| Official Contact: | Guy Bernard Bantsiele, Ph.D.<br>Regulatory Affairs Manager<br>Guy.bantsiele@bodycad.com<br>Tel: +1 (418) 527-1388<br>Fax: +1 (418) 431-9942 |
| Device Trade Name: | The Panthera Anti-Snoring Device |
| Device Common Name: | Mandibular repositioning device |
| Classification: | 21 CFR 872.5570 (Class II) |
| Product: | LRK |
Predicate: Narval CC (K113201)
Description: The Panthera Anti-Snoring Device is a removable intraoral device used for treating snoring and mild to moderate obstructive sleep apnea. It consists of two custom fabricated splints that fit separately over the upper and lower teeth and engage by means of adjustable rods.
The device functions as a mandibular repositioner, maintaining the lower jaw in a forward position during sleep. This mechanical protrusion acts to increase the patient's pharyngeal space, improving their ability to exchange air during sleep.
The device is a prescription customized for each patient and has an adjustment mechanism enabling the amount of mandibular advancement to be set by the dentist or physician at the time of fitting the device. The device can be adjusted only by the dentist. The maximum protrusion of the device is 15 mm in 1 mm increments.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther, the word "PANTHERA" is written in a bold, sans-serif font, with the word "DENTAL" written in a smaller font below it. A purple line is above the word "DENTAL".
Intended Use: The Panthera Anti-Snoring Device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Technological: The following table displays the differences and similarities between the new Panthera anti-snoring device and one other previously marketed (predicate) device. Equivalence is based on similarities in intended use, materials of construction, design, and operating principles, as summarized in the table below.
| Feature | The Panthera Anti-Snoring | Narval CC (K113201) |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device | |
| Regulation<br>description | Intraoral devices for snoring and<br>obstructive sleep apnea (OSA) | Intraoral devices for snoring and<br>obstructive sleep apnea (OSA) |
| Classification | Class II | Class II |
| Intended Use | To reduce or alleviate snoring and<br>mild to moderate obstructive sleep<br>apnea (OSA) in adults. | To reduce or alleviate snoring and<br>mild to moderate obstructive sleep<br>apnea (OSA) in adults. |
| Materials of<br>construction | Made from polymers (polyamide<br>type 12), Supplied by EOS | Made from polymers (polyamide<br>type 12), Supplied by EOS |
| | Highly resilient and durable<br>biocompatible polymer material and<br>then manufactured through<br>selective laser sintering to allow the<br>device to be strong yet supple as<br>well as flexible yet highly resilient.<br>The device is metal-free, flexible<br>and lightweight | Highly resilient and durable<br>biocompatible polymer material and<br>then manufactured through selective<br>laser sintering to allow the device to<br>be strong yet supple as well as<br>flexible yet highly resilient. The<br>device is metal-free, flexible and<br>lightweight |
| Design | Use the computer-aided design<br>(CAD) and computer-aided<br>manufacturing (CAM) | Use the computer-aided design<br>(CAD) and computer-aided<br>manufacturing (CAM) |
| | Use CAD that enables a high degree<br>of customization according to the<br>physician or dentist prescription to<br>accommodate the complex dental<br>anatomy of individual patients. The<br>CAM and selective laser sintering | Use CAD that enables a high degree<br>of customization according to the<br>physician or dentist prescription to<br>accommodate the complex dental<br>anatomy of individual patients. The<br>CAM and selective laser sintering |
| | | |
| | guarantees precision, accuracy and<br>consistency for each patient.<br><br>Two customized splints that fit<br>separately over the upper and lower<br>teeth inside the mouth. The lower<br>splint contains a triangular<br>protrusion, allowing the splints to<br>engage by means of interlocking<br>rods on the sides. | guarantees precision, accuracy and<br>consistency for each patient.<br><br>Two customized splints that fit<br>separately over the upper and lower<br>teeth inside the mouth. The lower<br>splint contains a triangular<br>protrusion, allowing the splints to<br>engage by means of interlocking<br>rods on the sides. |
| Principle of<br>operation/<br>means of<br>mandibular<br>advancement | Adjustment of the relative position<br>of the splints guides the mandible<br>forward and maintains advancement<br>thus enlarging the airway. The<br>vertical opening of the jaw is not<br>fixed in a single position. | Adjustment of the relative position<br>of the splints guides the mandible<br>forward and maintains advancement<br>thus enlarging the airway. The<br>vertical opening of the jaw is not<br>fixed in a single position |
| Fixed/<br>removable | Removable | Removable |
| Adjustment | Adjusted via the use of interlocking<br>rods placed on the sides of the<br>splints. The shorter the rod, the<br>further the mandible is advanced<br><br>Easy to titrate and highly adjustable<br>with connecting rods that allow for<br>15 mm of protrusive range at 1 mm<br>increments. The dentist can simply<br>select a shorter connecting rod until<br>optimal advancement is achieved.<br><br>Rods are available between 21 mm<br>to 36 mm in 1 mm increments | Adjusted via the use of interlocking<br>rods placed on the sides of the<br>splints. The shorter the rod, the<br>further the mandible is advanced<br><br>Easy to titrate and highly adjustable<br>with connecting rods that allow for<br>15 mm of protrusive range at 1 mm<br>increments. The dentist/patient can<br>simply select a shorter connecting<br>rod until optimal advancement is<br>achieved<br><br>Rods are available between 21 mm<br>to 36 mm in 1 mm increments |
| Maximum<br>protrusion of<br>the device | 15 mm in 1 mm increments | 15 mm in 1 mm increments |
| Supplied<br>sterile/non<br>sterile | Non sterile | Non sterile |
| Target population | Adults patients | Adults patients |
| Single Use/ reusable | Reusable | Reusable |
| Vertical opening | Up to 4 mm | Up to 4 mm |
| Prescription/ OTC | Prescription only | Prescription only |
| Recommended Cleaning and Maintenance | Clean daily in lukewarm water with a soft toothbrush. Rinse, dry and store in case provided. Twice weekly use chlorine-free antibacterial orthodontic cleaning solution. | Clean daily in lukewarm water with a soft brush and a mild soap/baking soda. Rinse, dry, and store in case provided. Twice weekly, use the Sonic Cleaner provided with effervescent antibacterial tablet |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther is the word "PANTHERA" in a purple sans-serif font. The word "DENTAL" is below "PANTHERA" in a smaller font.
## Traditional 510(k) – The Panthera Anti-Snoring Device
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther is the word "PANTHERA" in a stylized font, with the word "DENTAL" in a smaller font underneath.
### Traditional 510(k) – The Panthera Anti-Snoring Device
RISKS: Panthera dental performed no clinical testing. However, an FMEA risk analysis, and evaluation of the materials of construction and design were performed. The function of mandibular advancement devices requires that the prescribing dentist be cognizant of the potential for soreness, soft tissue soreness, and dentition complications (soreness, motion, loosening) by mandibular advancement. Management of these risks is achieved by advising the patient and dentist in the directions for use that early and repeated examination of the fit of the device, and its performance, must be performed in the dental office by the prescribing dentist. The contraindications, warnings, precautions, storage directions, prescription preparation instructions, fitting and adjustment directions are written to avoid potential problems from arising or persisting with the dentition, tissue, or joints, caused by the OSA devices. No new materials are being used in the device; all material (polyamide type12) is already used in the predicate device (Narval CC). No new risks are introduced with the new device that are not present in the predicate device.
Non-Clinical Testing: The non-clinical testing included assessment of the physical properties of the Panthera Ant-Snoring Device and its ability to achieve its intended use. The Panthera Anti-Snoring Device meets the same specifications as set for the predicate device.
A biocompatibility assessment of the device was performed. The purpose of the biocompatibility assessment was to ensure that biocompatibility had been established for the device. The device is biocompatible, based on the similarity of the materials of construction to the predicate device (Narval CC) marketed by ResMed.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo for a dental practice. The logo features a black panther in a crouching position, with its tail curled up. Below the panther, the word "PANTHERA" is written in a purple sans-serif font. Underneath "PANTHERA", the word "DENTAL" is written in a smaller, purple sans-serif font with a line above it.
Clinical Testing: Human clinical studies were not deemed necessary to evaluate the safety or effectiveness of the Panthera Anti-Snoring Device. The Panthera Anti-Snoring Device does not: Use designs dissimilar from the predicate device and other previously cleared devices under a 510(k); The Panthera Anti-Snoring Device does not use new technologies different from legally marketed intramandibular repositioning devices for treating snoring and mild to moderate obstructive sleep apnea; and the Panthera Anti-Snoring Device does not deviate from the indications for use identified in the predicate device: Narval CC.
Substantial Equivalence Conclusion: The new device, the Panthera Anti-Snoring device, is considered to be substantially equivalent to the predicate based on the following: it has essentially the same intended use and is indicated for the same user population; it has equivalent technological characteristics to the predicate; it does not raise new questions of safety and effectiveness; it is at least as safe and effective as the predicative device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.