21 CFR 872.5571 — Closed Loop Auto Titration Device For Oral Appliances

Dental (DE) · Subpart F—Therapeutic Devices · § 872.5571

Identification

An auto-titration device for oral appliances is a prescription home use device that determines a target position to be used for a final oral appliance for the reduction of snoring and mild to moderate obstructive sleep apnea.

Classification Rationale

Class II (special controls). The special controls for this device are:

Product Codes

Product CodeDevice NameClassDevicesAttributes
QCJClosed Loop Auto Titration Device For Oral Appliances21

Special Controls

QCJ — Closed Loop Auto Titration Device For Oral Appliances

In combination with the general controls of the FD&C Act. the auto titration device for oral appliances is subject to the following special controls:

De Novo Order DEN170090

QCJ — Closed Loop Auto Titration Device For Oral Appliances

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate the following: (i) Performance characteristics of the algorithm; and (ii) All adverse events. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, including the following: (i) Validation of the closed loop algorithm; (ii) Mechanical integrity over the expected use life; (iii) Characterization of maximum force, distance, and speed of device movement; and (iv) Movement accuracy of intraoral components. (3) Performance testing must demonstrate the wireless compatibility, electrical safety, and electromagnetic compatibility of the device in its intended use environment. (4) Software verification, validation, and hazard analysis must be performed. (5) The patient-contacting components of the device must be demonstrated to be biocompatible. (6) Performance data must validate the reprocessing instructions for any reusable components. (7) Patient labeling must include: (i) Information on device use, including placement of sensors and mouthpieces; (ii) A description of all alarms; and (iii) Instructions for reprocessing any reusable components. (8) A human factors assessment must evaluate simulated use of the device in a home use setting.

eCFR

QCJ — Closed Loop Auto Titration Device For Oral Appliances

(1) Clinical performance testing must evaluate the following: (i) Performance characteristics of the algorithm; and (ii) All adverse events. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, including the following: (i) Validation of the closed loop algorithm; (ii) Mechanical integrity over the expected use life; (iii) Characterization of maximum force, distance, and speed of device movement; and (iv) Movement accuracy of intraoral components. (3) Performance testing must demonstrate the wireless compatibility, electrical safety, and electromagnetic compatibility of the device in its intended use environment. (4) Software verification, validation, and hazard analysis must be performed. (5) The patient-contacting components of the device must be demonstrated to be biocompatible. (6) Performance data must validate the reprocessing instructions for any reusable components. (7) Patient labeling must include: (i) Information on device use, including placement of sensors and mouthpieces; (ii) A description of all alarms; and (iii) Instructions for reprocessing any reusable components. (8) A human factors assessment must evaluate simulated use of the device in a home use setting.

Ecfr Llm

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...