K132322 · Satelec - Acteon Group · ELC · Mar 11, 2014 · Dental
Device Facts
Record ID
K132322
Device Name
NEWTRON P5 XS B.LED
Applicant
Satelec - Acteon Group
Product Code
ELC · Dental
Decision Date
Mar 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Intended Use
This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis.
Device Story
Dental ultrasonic generator using piezoelectric technology; consists of control unit and NEWTRON SLIM B.LED handpiece. Input: electrical signal from device; Transformation: piezoelectric ceramic transducer converts electrical signal into ultrasonic mechanical vibrations; Output: ultrasonic vibrations transmitted to dental tip/file for plaque/tartar removal. Used in dental clinics by professionals. Features integrated LED ring for illumination. Bench testing confirms irrigation flow and current delivery equivalent to predicates. Benefits: efficient removal of dental deposits during routine and specialized dental procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing comparing irrigation flow and handpiece current values against predicate PMAX NEWTRON XS (K071424).
Technological Characteristics
Piezoelectric ultrasonic generator; handpiece with piezoelectric ceramic transducer. Casings made of self-extinguishing material (UL94V-0). Integrated LED ring for illumination. Powered by electric mains. Sterilization via moist heat (ISO 17665-1/2). Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for dental patients requiring prophylaxis, scaling, periodontal, endodontic, or restorative/prosthetic procedures.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
K151023 — COMPACT PIEZO LED · Mectron S.P.A. · Aug 5, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Additional Information - October 31, 2013
This Summary of Safety And Effectiveness is submitted in accordance with 21 CFR 807.92.c.
#### 01 - Administrative Information
#### 01- a Type of 510(k) submission:
These documents constitute a Traditional 510(k) Submission.
01- b. Submission date: July 19, 2013
#### 01-с. 510(k) Submitter:
SATELEC 17, Avenue Gustave Eiffel BP 30216 33708 Merignac Cedex France Telephone: 011-33-556-340-607 011-33-556-349-292 Fax: E-Mail: philippe.girard@acteongroup.com
Contact Person: 01-d.
> Rick Rosati SATELEC c/o ACTEON, Inc. 124 Gaither Drive, Suite 140 Mt. Laurel, NJ 08054 800-289-6367 Ext. 390 Telephone: Fax: 856-222-4726 E-Mail: rick.rosati@us.acteongroup.com
Establishment Registration Number: 8044015 01- e.
#### 02 - Device Information
| 02-a. Trade Name of Device: | NEWTRON P5 XS B.LED |
|----------------------------------|---------------------|
| 02-b. Common Name of Device: | Ultrasonic Scaler |
| 02-c. Classification Regulation: | 21 CFR 872.4850 |
| 02-d. Medical Device Class: | II |
| 02-e. Panel: | Dental |
| 02-f. Product Code: | ELC |
Pre-Market Notification 510(k) Submission for NEWTRON P5 XS B.LED By SATELEC K132322 - Traditional 510(k) - Confidential Document - Version 02
006 - 1
{1}------------------------------------------------
Additional Information - October 31, 2013
# 03 - Identification of Legally Marketed Predicate(s)
The Substantial Equivalence (SE) of the Satelec New Device is based on the Predicate Devices identified in the Table 01.
| N | Trade Name | Manufacturer | Product Code | 510(k) number | Date Cleared |
|---|---------------------------|--------------|--------------|---------------|-------------------|
| 1 | SUPRASSON P5 NEWTRON | SATELEC | ELC | K050895 | April 20, 2005 |
| 2 | ProUltra Piezo Ultrasonic | SATELEC | ELC | K113430 | February 23, 2012 |
| 3 | PMAX NEWTRON XS | SATELEC | ELC | K071424 | August 24, 2007 |
Table 01 - Identification of Legally Marketed Predicate Devices
#### 04 - Description of the Device
The Satelec New Device is a Dental Ultrasonic Generator. The Satelec New Device uses Piezoelectric Technology. The Satelec New Device uses a Satelec Dental Piezoelectric Handpiece (NEWTRON SLIM B.LED).
The Satelec New Device is designed to be used with Satelec Dental Tips previously cleared with Predicate Devices (K050895, cleared April 20, 2005 and K113430, cleared February 23, 2012).
#### Principles of operation:
An electrical signal emitted by the medical device is supplied to the dental ultrasonic handpiece. This is connected to the medical device via a cord. The handpiece comprises a piezoelectric ceramic transducer, which transforms the electrical signal into ultrasonic vibrations. Mechanical vibrations are transmitted to a tip or a dental file attached to the end of the ultrasonic handpiece.
#### Scientific principles:
The ultrasonic mechanical vibrations transmitted to the dental file attached to the end of the ultrasonic handpiece remove dental plaque and / or dental tartar.
#### 05 — Intended Use
This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis.
#### 06 - Performance Data
It has been determined in the Section 019 "Performance Testing - Bench" that the Satelec New Device and Predicate Device Satelec SUPRASSON PMAX NEWTRON XS (K071424, cleared August 24, 2007) are similar because the measured values of the Irrigation Flow and the values of the Current delivered in the Piezoelectric Handpiece are similar.
Clinical Data is not needed for this 510(k) process.
Pre-Market Notification 510(k) Submission for NEWTRON P5 XS B.LED By SATELEC K132322 - Traditional 510(k) - Confidential Document - Version 02
006 - 2
{2}------------------------------------------------
Additional Information - October 31, 2013
# 07 - Tests and Used Standards
# Electromagnetic Compatibility Test:
The Electromagnetic Compatibility Tests have been performed according to IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3).
# Electrical Safety Tests:
The Electrical Safety Tests have been performed according to IEC60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (General).
# Sterilization Validation:
The Sterilizability Tests have been performed according to ISO 17665-1:2006, Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2 : 2009 of Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1.
# 08 - Technological characteristics of the Device compared to the Predicate Devices
Technological characteristics of the Satelec New Device are the Predicate Device.
# Technological Perspective:
The Satelec New Device and the Predicate Devices use the same technology (Piezoelectricity Technology).
# Material perspective:
The Satelec New Device and the Predicate Devices are very similar because the casings are made in self-extinguishing material (UL94V-0).
# Design perspective:
The Satelec New Device and the Predicate Devices are very similar because they use:
- The same Principle of User's interface. -
- -A Piezoelectric Handpiece.
# Energy source perspective:
The Satelec New Device and the Predicate Device n°3:
- Use the same input energy source (Electric Mains Power Supply). -
- Deliver the same output energy source (ultrasonic micro-vibration). l
- -Deliver the same Handpiece Current Values.
- Deliver a similar irrigation Flow values for considered clinics. -
Pre-Market Notification 510(k) Submission for NEWTRON P5 XS B.LED By SATELEC K132322 - Traditional 510(k) - Confidential Document - Version 02
{3}------------------------------------------------
Additional Information - October 31, 2013
#### Light function perspective:
The Satelec New Device and the Predicate Device nº3:
- Use the same principle of LED ring. -
- Use the same principle of Optical Guide. -
- Use the same quantity of LEDs on the LED ring. -
- Use the same reference of LEDs. -
- Use the same color of light. -
The Light Function does not induce rise temperature on the clinical site.
# 09 - Determination of substantial equivalence
The Satelec New Device has same Indication Statement as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
The Satelec New Device is similar in terms of technological characteristics as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
The Satelec New Device is similar in terms of Performance Data as the Predicate Device Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
The Satelec New Device is similar in terms of Light function as the Predicate Device Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
The identified differences have no impact on the Intended use, Safety and Effectiveness. Effectiveness of the Satelec New Device is the similar as of the Predicate Devices. The Satelec New Device and Predicate Devices are manufactured by SATELEC.
# 10 - Conclusion
Satelec New Device is Substantially Equivalent (SE) to the Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005), Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
End of Section
Pre-Market Notification 510(k) Submission for NEWTRON P5 XS B.LED By SATELEC K132322 - Traditional 510(k) - Confidential Document - Version 02
{4}------------------------------------------------
# Section 006 - 510(k) Summary (SMDA Requirements) Additional Information - October 31, 2013
.
.
This page is intentionally left in blank
Pre-Market Notification 510(k) Submission for NEWTRON P5 XS B.LED By SATELEC K132322 - Traditional 510(k) - Confidential Document - Version 02
006 - 5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and has a simple, clean design. The eagle is a symbol of strength and freedom, and the logo represents the department's mission to protect the health of all Americans.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G605 Silver Spring, MD 20993-0002
March 11, 2014
Satelec-Acteon Group Mr. Rick Rosati Quality Manager 124 Gaither Drive, Suite 140 Mt. Laurel, NJ 08054
Re: K132322
Trade/Device Name: NEWTRON P5 EX B.LED Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: February 10, 2014 Received: February 11, 2014
Dear Mr. Rosati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Rosati
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/6/Picture/7 description: The image shows a signature and some text. The text includes "Sincerely yours," followed by a signature. Below the signature, the text reads "Tejashri Purohit-Sheth, M.D.," "Clinical Deputy Director," and "DAGRID."
Erin J. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Section 005 - Indication for Use Additional Information -- October 31, 2013
#### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K132322
Device Name:
NEWTRON P5 XS B.LED
Indications for Use:
This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosiblesis.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
S. Runner -S -04'00'
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.