NEWTRON P5, NEWTRON P5 B.LED

K132267 · Satelec - Acteon Group · ELC · Mar 10, 2014 · Dental

Device Facts

Record IDK132267
Device NameNEWTRON P5, NEWTRON P5 B.LED
ApplicantSatelec - Acteon Group
Product CodeELC · Dental
Decision DateMar 10, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4850
Device ClassClass 2
AttributesTherapeutic

Intended Use

This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis.

Device Story

Dental ultrasonic generator; uses piezoelectric technology to remove dental plaque and tartar. System consists of main unit, cord, and piezoelectric handpiece (NEWTRON SLIM or NEWTRON SLIM B.LED). Input: electrical signal from device; transformation: piezoelectric ceramic transducer converts electrical signal into ultrasonic mechanical vibrations; output: mechanical vibrations transmitted to dental tip or file. Used in dental clinics by professionals. Provides illumination via LED ring on handpiece. Benefits: efficient removal of dental plaque/tartar during routine dental procedures.

Clinical Evidence

No clinical data provided. Substantial equivalence established via bench testing comparing electrical current delivery and irrigation flow to predicate devices.

Technological Characteristics

Piezoelectric ultrasonic generator; handpiece with piezoelectric ceramic transducer. Casings made of self-extinguishing material (UL94V-0). LED ring illumination. Powered by electric mains. Standards: IEC 60601-1:2005 (electrical safety), IEC 60601-1-2:2007 (EMC), ISO 17665-1:2006/17665-2:2009 (sterilization).

Indications for Use

Indicated for dental prophylaxis, scaling, periodontics, endodontics, and preservation/restoration dentistry including prosthesis. No specific patient population age or gender restrictions provided.

Regulatory Classification

Identification

An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Additional Information - October 30, 2013 This Summary of Safety And Effectiveness is submitted in accordance with 21 CFR 807.92. c. #### 01 - Administrative Information 01- a. Type of 510(k) submission: These documents constitute a Traditional 510(k) Submission. July 16, 2013 01- b. Submission date: 01-c. 510(k) Submitter: > SATELEC 17, Avenue Gustave Eiffel BP 30216 33708 Merignac Cedex France Telephone: 011-33-556-340-607 011-33-556-349-292 Fax: E-Mail: philippe.girard@acteongroup.com 01-d. Contact Person: > Rick Rosati SATELEC c/o ACTEON, Inc. 124 Gaither Drive, Suite 140 Mt. Laurel, NJ 08054 800-289-6367 Ext. 390 Telephone: Fax: 856-222-4726 E-Mail: rick.rosati@us.acteongroup.com · 01- e. Establishment Registration Number 8044015 02 - Device Information 02-a. Trade Name of Device: NEWTRON P5 NEWTRON P5 B.LED 02-b. Common Name of Device: Ultrasonic Scaler 02-c. Classification Regulation: 21 CFR 872.4850 02-d. Medical Device Class: ll 02-e. Panel: Dental ELC 02-f. Product Code: > Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02 > > 006 - 1 {1}------------------------------------------------ Additional Information - October 30, 2013 #### 03 - Identification of Legally Marketed Predicate(s) The Substantial Equivalence (SE) of the Satelec New Device is based on the Predicate Devices identified in the Table 01. | N° | Trade Name | Manufacturer | Product Code | 510(k) number | Date Cleared | |----|---------------------------|--------------|--------------|---------------|-------------------| | 1 | SUPRASSON P5 NEWTRON | SATELEC | ELC | K050895 | April 20, 2005 | | 2 | ProUltra Piezo Ultrasonic | SATELEC | ELC | K113430 | February 23, 2012 | | 3 | PMAX NEWTRON XS | SATELEC | ELC | K071424 | August 24, 2007 | Table 01 - Identification of Legally Marketed Predicate Devices #### 04 - Description of the Device The Satelec New Device is a Dental Ultrasonic Generator. The Satelec New Device uses Piezoelectric Technology. The Satelec New Device uses a Satelec Dental Piezoelectric Handpiece (NEWTRON SLIM or NEWTRON SLIM B.LED). The Satelec New Device is designed to be used with Satelec Dental Tips previously cleared with Predicate Devices (K050895, cleared April 20, 2005 and K113430, cleared February 23, 2012), #### Principles of operation: An electrical signal emitted by the medical device is supplied to the dental ultrasonic Handpiece. This is connected to the medical device via a cord. The Handpiece comprises a piezoelectric ceramic transducer, which transforms the electrical signal into ultrasonic vibrations. Mechanical vibrations are transmitted to a tip or a dental file attached to the end of the ultrasonic Handbiece. #### Scientific principles: The ultrasonic mechanical vibrations transmitted to the tip or to the dental file attached to the end of the ultrasonic Handpiece remove dental plaque and / or dental tartar. #### 05 - Intended Use This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis. #### 06 - Performance Data It has been determined in the Section 019 "Performance Testing - Bench" that the Satelec New Device and Predicate Device Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) are similar because the measured values of the frigation Flow and the values of the Current delivered in the Piezoelectric Handpiece are similar. Clinical Data is not needed for this 510(k) process. Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02 {2}------------------------------------------------ Additional Information - October 30, 2013 #### 07 - Tests and Used Standards #### Electromagnetic Compatibility Test: The Electromagnetic Compatibility Tests have been performed according to IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential Electromagnetic compatibility - Requirements and tests performance - Collateral standard: (Edition 3). #### Electrical Safety Tests: The Electrical Safety Tests have been performed according to IEC60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (General). #### Sterilization Validation: The Sterilizability Tests have been performed according to ISO 17665-1.2006, Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2: 2009 of Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1. #### 08 - Technological characteristics of the Device compared to the Predicate Devices Technological characteristics of the Satelec New Device are the Predicate Devices. #### Technological Perspective: The Satelec New Device and the Predicate Devices use the same technology (Piezoelectricity Technology). #### Material perspective: The Satelec New Device and the Predicate Devices are very similar because the casings are made in self-extinguishing material (UL94V-0). #### Design perspective: The Satelec New Device and the Predicate Devices are very similar because they use: - The same Principle of User's interface. । - A Piezoelectric Handpiece. - #### Energy source perspective: The Satelec New Device and the Predicate Device nº 1: - -Use the same input energy source (Electric Mains Power Supply). - Deliver the same output energy source (ultrasonic micro-vibration). - - Deliver the same Handpiece Current Values. - - Deliver a similar irrigation Flow values for considered clinics. - Pre-Market Notification 510(k) Submission for NEWTRON P5 Bv SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02 {3}------------------------------------------------ Additional Information - October 30, 2013 #### Light function perspective: The Satelec New Device and the Predicate Device nº3: - -Use the same principle of LED ring. - Use the same principle of Optical Guide. l - -Use the same quantity of LEDs on the LED ring. - Use the same reference of LEDs. - - -Use the same color of light. The Light Function does not induce rise temperature on the clinical site. ### 09 - Determination of substantial equivalence The Satelec New Device has same Indication Statement as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007). The Satelec New Device is similar in terms of technological characteristics as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007). The Satelec New Device is similar in terms of Performance Data as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424. cleared August 24, 2007). The Satelec New Device is similar in terms of Light function as the Predicate Device Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007). The Satelec New Device is efficient and safe for the Intended Uses. The identified differences have no impact on the Intended use, Safety and Effectiveness of the Satelec New Device is the similar as of the Predicate Devices. The Satelec New Device and Predicate Devices are manufactured by SATELEC. # 10 - Conclusion Satelec New Device is Substantially Equivalent (SE) to the Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005), Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007). End of Section Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02 006 - 4 {4}------------------------------------------------ # Section 006 - 510(k) Summary (SMDA Requirements) Additional Information - October 30, 2013 This page is intentionally left in blank : Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02 006 - 5 . . {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines representing the agency's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 10, 2014 Mr. Rick Rosati SATELEC C/O ACTEON, Inc. 124 Gaither Drive. Suite 140 Mt. Laurel, NJ 08054 Re: K 132267 Trade/Device Name: Newtron P5, Newtron P5 B.led Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: January 31, 2014 Received: February 3, 2014 Dear Mr. Rosati: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ # Page 2 - Mr. Rosati Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. ErinFDAth-S Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Section 005 - Indication for Use Additional Information - October 30, 2013 #### Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K132267 Device Name: NEWTRON P5 NEWTRON P5 B.LED Indications for Use: This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sean Runner-S 2023.03.07 17:53:36 -05'00' Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%