K132267 · Satelec - Acteon Group · ELC · Mar 10, 2014 · Dental
Device Facts
Record ID
K132267
Device Name
NEWTRON P5, NEWTRON P5 B.LED
Applicant
Satelec - Acteon Group
Product Code
ELC · Dental
Decision Date
Mar 10, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis.
Device Story
Dental ultrasonic generator; uses piezoelectric technology to remove dental plaque and tartar. System consists of main unit, cord, and piezoelectric handpiece (NEWTRON SLIM or NEWTRON SLIM B.LED). Input: electrical signal from device; transformation: piezoelectric ceramic transducer converts electrical signal into ultrasonic mechanical vibrations; output: mechanical vibrations transmitted to dental tip or file. Used in dental clinics by professionals. Provides illumination via LED ring on handpiece. Benefits: efficient removal of dental plaque/tartar during routine dental procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing comparing electrical current delivery and irrigation flow to predicate devices.
Technological Characteristics
Piezoelectric ultrasonic generator; handpiece with piezoelectric ceramic transducer. Casings made of self-extinguishing material (UL94V-0). LED ring illumination. Powered by electric mains. Standards: IEC 60601-1:2005 (electrical safety), IEC 60601-1-2:2007 (EMC), ISO 17665-1:2006/17665-2:2009 (sterilization).
Indications for Use
Indicated for dental prophylaxis, scaling, periodontics, endodontics, and preservation/restoration dentistry including prosthesis. No specific patient population age or gender restrictions provided.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}------------------------------------------------
Additional Information - October 30, 2013
This Summary of Safety And Effectiveness is submitted in accordance with 21 CFR 807.92. c.
#### 01 - Administrative Information
01- a. Type of 510(k) submission:
These documents constitute a Traditional 510(k) Submission.
July 16, 2013 01- b. Submission date:
01-c. 510(k) Submitter:
> SATELEC 17, Avenue Gustave Eiffel BP 30216 33708 Merignac Cedex France Telephone: 011-33-556-340-607 011-33-556-349-292 Fax: E-Mail: philippe.girard@acteongroup.com
01-d. Contact Person:
> Rick Rosati SATELEC c/o ACTEON, Inc. 124 Gaither Drive, Suite 140 Mt. Laurel, NJ 08054 800-289-6367 Ext. 390 Telephone: Fax: 856-222-4726 E-Mail: rick.rosati@us.acteongroup.com
· 01- e. Establishment Registration Number 8044015
02 - Device Information
02-a. Trade Name of Device: NEWTRON P5 NEWTRON P5 B.LED 02-b. Common Name of Device: Ultrasonic Scaler 02-c. Classification Regulation: 21 CFR 872.4850 02-d. Medical Device Class: ll 02-e. Panel: Dental ELC 02-f. Product Code:
> Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02
> > 006 - 1
{1}------------------------------------------------
Additional Information - October 30, 2013
#### 03 - Identification of Legally Marketed Predicate(s)
The Substantial Equivalence (SE) of the Satelec New Device is based on the Predicate Devices identified in the Table 01.
| N° | Trade Name | Manufacturer | Product Code | 510(k) number | Date Cleared |
|----|---------------------------|--------------|--------------|---------------|-------------------|
| 1 | SUPRASSON P5 NEWTRON | SATELEC | ELC | K050895 | April 20, 2005 |
| 2 | ProUltra Piezo Ultrasonic | SATELEC | ELC | K113430 | February 23, 2012 |
| 3 | PMAX NEWTRON XS | SATELEC | ELC | K071424 | August 24, 2007 |
Table 01 - Identification of Legally Marketed Predicate Devices
#### 04 - Description of the Device
The Satelec New Device is a Dental Ultrasonic Generator. The Satelec New Device uses Piezoelectric Technology. The Satelec New Device uses a Satelec Dental Piezoelectric Handpiece (NEWTRON SLIM or NEWTRON SLIM B.LED).
The Satelec New Device is designed to be used with Satelec Dental Tips previously cleared with Predicate Devices (K050895, cleared April 20, 2005 and K113430, cleared February 23, 2012),
#### Principles of operation:
An electrical signal emitted by the medical device is supplied to the dental ultrasonic Handpiece. This is connected to the medical device via a cord. The Handpiece comprises a piezoelectric ceramic transducer, which transforms the electrical signal into ultrasonic vibrations. Mechanical vibrations are transmitted to a tip or a dental file attached to the end of the ultrasonic Handbiece.
#### Scientific principles:
The ultrasonic mechanical vibrations transmitted to the tip or to the dental file attached to the end of the ultrasonic Handpiece remove dental plaque and / or dental tartar.
#### 05 - Intended Use
This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis.
#### 06 - Performance Data
It has been determined in the Section 019 "Performance Testing - Bench" that the Satelec New Device and Predicate Device Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) are similar because the measured values of the frigation Flow and the values of the Current delivered in the Piezoelectric Handpiece are similar.
Clinical Data is not needed for this 510(k) process.
Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02
{2}------------------------------------------------
Additional Information - October 30, 2013
#### 07 - Tests and Used Standards
#### Electromagnetic Compatibility Test:
The Electromagnetic Compatibility Tests have been performed according to IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential Electromagnetic compatibility - Requirements and tests performance - Collateral standard: (Edition 3).
#### Electrical Safety Tests:
The Electrical Safety Tests have been performed according to IEC60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (General).
#### Sterilization Validation:
The Sterilizability Tests have been performed according to ISO 17665-1.2006, Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2: 2009 of Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1.
#### 08 - Technological characteristics of the Device compared to the Predicate Devices
Technological characteristics of the Satelec New Device are the Predicate Devices.
#### Technological Perspective:
The Satelec New Device and the Predicate Devices use the same technology (Piezoelectricity Technology).
#### Material perspective:
The Satelec New Device and the Predicate Devices are very similar because the casings are made in self-extinguishing material (UL94V-0).
#### Design perspective:
The Satelec New Device and the Predicate Devices are very similar because they use:
- The same Principle of User's interface. ।
- A Piezoelectric Handpiece. -
#### Energy source perspective:
The Satelec New Device and the Predicate Device nº 1:
- -Use the same input energy source (Electric Mains Power Supply).
- Deliver the same output energy source (ultrasonic micro-vibration). -
- Deliver the same Handpiece Current Values. -
- Deliver a similar irrigation Flow values for considered clinics. -
Pre-Market Notification 510(k) Submission for NEWTRON P5 Bv SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02
{3}------------------------------------------------
Additional Information - October 30, 2013
#### Light function perspective:
The Satelec New Device and the Predicate Device nº3:
- -Use the same principle of LED ring.
- Use the same principle of Optical Guide. l
- -Use the same quantity of LEDs on the LED ring.
- Use the same reference of LEDs. -
- -Use the same color of light.
The Light Function does not induce rise temperature on the clinical site.
### 09 - Determination of substantial equivalence
The Satelec New Device has same Indication Statement as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
The Satelec New Device is similar in terms of technological characteristics as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
The Satelec New Device is similar in terms of Performance Data as the Predicate Devices Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005) and Satelec PMAX NEWTRON XS (K071424. cleared August 24, 2007).
The Satelec New Device is similar in terms of Light function as the Predicate Device Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
The Satelec New Device is efficient and safe for the Intended Uses. The identified differences have no impact on the Intended use, Safety and Effectiveness of the Satelec New Device is the similar as of the Predicate Devices. The Satelec New Device and Predicate Devices are manufactured by SATELEC.
# 10 - Conclusion
Satelec New Device is Substantially Equivalent (SE) to the Satelec SUPRASSON P5 NEWTRON (K050895, cleared April 20, 2005), Satelec ProUltra Piezo Ultrasonic (K113430, cleared February 23, 2012), and Satelec PMAX NEWTRON XS (K071424, cleared August 24, 2007).
End of Section
Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02
006 - 4
{4}------------------------------------------------
# Section 006 - 510(k) Summary (SMDA Requirements) Additional Information - October 30, 2013
This page is intentionally left in blank
:
Pre-Market Notification 510(k) Submission for NEWTRON P5 By SATELEC K132267 - Traditional 510(k) - Confidential Document - Version 02
006 - 5
. .
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines representing the agency's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2014
Mr. Rick Rosati SATELEC C/O ACTEON, Inc. 124 Gaither Drive. Suite 140 Mt. Laurel, NJ 08054
Re: K 132267
Trade/Device Name: Newtron P5, Newtron P5 B.led Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: January 31, 2014 Received: February 3, 2014
Dear Mr. Rosati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
# Page 2 - Mr. Rosati
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
ErinFDAth-S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Section 005 - Indication for Use Additional Information - October 30, 2013
#### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K132267
Device Name:
NEWTRON P5 NEWTRON P5 B.LED
Indications for Use:
This medical device is intended for prophylaxis, including scaling, for periodontics, for endodontics, and for preservation and restoration dentistry, including prosthesis.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sean Runner-S
2023.03.07
17:53:36 -05'00'
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.