Dental Electrical Motor iRoot Pro

K191276 · Changzhou Bomedent Medical Technology Co.,Ltd · EKX · Jun 30, 2020 · Dental

Device Facts

Record IDK191276
Device NameDental Electrical Motor iRoot Pro
ApplicantChangzhou Bomedent Medical Technology Co.,Ltd
Product CodeEKX · Dental
Decision DateJun 30, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1
AttributesTherapeutic

Intended Use

The Dental Electrical Motor iRoot Pro is a cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length.

Device Story

Cordless endodontic motorized handpiece; integrates apex locator for root canal length measurement. Input: electrical signals via file/lip clip circuit to determine apex position. Operation: electric motor drives rotary file; software controls speed, torque, and rotation modes. Output: LCD display of parameters (speed, torque, apex position); audible warnings. Used in dental clinics by dental professionals. Benefits: enables simultaneous canal enlargement and real-time monitoring of file tip position; improves procedural efficiency and safety. Four operational modes: motor-only; apex locator-only; independent motor/apex locator; and automated motor control (rotate/stop/reverse) based on apex locator feedback.

Clinical Evidence

No clinical data. Bench testing only. Verification included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), dental handpiece performance (ISO 14457, ISO 80601-2-60), biocompatibility (ISO 10993-5, -10), and apex locator accuracy testing.

Technological Characteristics

Cordless, battery-powered (Li-ion 3.7V) handpiece. Materials: 304 stainless steel, silica rubber. Connectivity: USB for charging. Sterilization: Autoclavable contra-angle and lip clip; barrier sleeves for central unit. Software: Moderate level of concern. Sensing: Dual-frequency impedance-based apex location.

Indications for Use

Indicated for endodontic treatment, specifically for root canal preparation (enlarging canals) and root canal length measurement in dental patients.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 30, 2020 ChangZhou BoMedent Medical Technology Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. R912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, 102401 CHINA Re: K191276 Trade/Device Name: Dental Electrical Motor iRoot Pro Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I, reserved Product Code: EKX. LOY Dated: May 29, 2020 Received: June 1, 2020 Dear Ray Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K191276 Device Name Dental Electrical Motor iRoot Pro ### Indications for Use (Describe) The Dental Electrical Motor iRoot Pro is a cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge canals while monitoring the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | | <span> <span style="text-decoration: underline;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</span> | | | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K191276 - 1. Date of Preparation: June 30, 2020 - 2. Sponsor Identification ### ChangZhou BoMedent Medical Technology Co., Ltd. No.9 Changyang Road, West Taihu Science and Technology Industrial Park, ChangZhou, JiangSu 213000, China Contact Person: Lily Zhang Position: Quality Supervisor Tel: +86-0519-88991980 Fax: +86-0519-88991980 Email: qm@bome-dent.com - 3. Designated Submission Correspondent Mr. Ray Wang ## Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401 Tel: +86-18910677558 Fax: +86-10- 56335780 Email: ray.wang@believe-med.com {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: Dental Electrical Motor iRoot Pro Common Name: Dental Handpiece and Accessories Model(s): iRoot Pro Regulatory Information Classification Name: Dental Handpiece and Accessories Classification: 1 Product Code: EKX/LQY Regulation Number: 21 CFR 872.4200 Review Panel: Dental; Indications for Use Statement: The Dental Electrical Motor iRoot Pro is a cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. ### Device Description The Dental Electrical Motor iRoot Pro is a low-speed rotating oral equipment mainly used for root canal preparation and root canal measurement. The product is a portable device powered by built-in lithium batteries and charged by USB interface. LCD displays parameters such as speed, torque, working mode, apex position, etc. Users can also set and modify by keys, and provide design of factory initialization and calibration. The product also provides a complete rotation mode and root canal measurement mode, which can be stored in memory, and provides a key functional mode for users to use quickly. The Dental Electrical Motor iRoot Pro is a cordless endodontic treatment motorized handpiece with root canal measurement capability. It can be used to enlarge canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. The Dental Electrical Motor iRoot Pro is intended to be sterilized prior to use. There are four working modes as: Mode: Only endo motor function enabled; Mode: Endo motor and apex locator work independently, when the file reaching theapical, the apex locator only plays the role of warning and display, will not interfere with the rotation of endo motor, the motor will not stop and reverse when the file reach the apical. Mode: In this mode, the motor will automatically rotate, stop, reverse depends on the length of the root canal measured by the apex locator. {5}------------------------------------------------ The chuck of proposed device needs to be used with the file in accordance with Type 1 in ISO 1797-1, the diameter of chuck interface is @ 1.5. The lubricant to be used with the contra angle is "DO-ALL Dental Handpiece Lubricant" manufactured by ProDrive Systems Inc., which has been cleared as K073353. #### న. Identification of Predicate Device(s) Primary Predicate Device: 510(k) Number: K170275 Product Name: Tri Auto ZX2 Manufacturer: J. Morita USA, Inc. Reference Device: 510(k) Number: K153285 Product Name: EMS-200 Manufacturer: Meta Systems Co., Ltd. - 6. Non-Clinical Test Conclusion Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General requirements for a. basic safety and - essential performance IEC 60601-1-2:2014, Medical Electrical Equipment-Part 1-2: General Requirements For Basic b. Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Tests - C. ISO 14457:2017 Dentistry - Handpieces and motors ISO 80601-2-60:2012 Medical electrical equipment - Part 2-60: Particular requirements for the d. basic safety and essential performance of dental equipment ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro e. cytotoxicity. f. ISO 10993-10:2010 Standard, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity - Cleaning, Intermediate Level Disinfection, and Sterilization of the components of the g. subject device per the FDA Guidance Document Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, AAMI TIR 30, AAMI TIR 12, ISO 17665-1, and ISO 17665-2. In addition, disposable barrier sleeve anti-contamination and ingress verification testing {6}------------------------------------------------ was conducted. - Software documentation for moderate level of concern per the FDA Guidance Document "Guidance h. for the Content of Premarket Submissions for Software Contained in Medical Devices." - Apex Locator Performance Testing to verify the apex locator measuring accuracy performance i. - 7. Clinical Test Conclusion No clinical study is included in this submission. {7}------------------------------------------------ # 8. Substantially Equivalent (SE) Comparison | Table 1 Comparison of Technology Characteristics<br>J | |-------------------------------------------------------| |-------------------------------------------------------| | Item | Proposed Device(s) | Predicate Device(s) | Reference Device | Remark | |-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Device name | Dental Electrical Motor<br>iRoot Pro | Tri Auto ZX2 | EMS-200 | / | | Classification Name | endodontic treatment motorized<br>handpiece/<br>root canal apex locator | endodontic treatment motorized handpiece/<br>root canal apex locator | endodontic treatment motorized handpiece | SAME | | Product Code | EKX/LQY | EKX/LQY | EKX | SAME | | Regulation Number | 872.4200 | 872.4200 | 872.4200 | SAME | | Comparison<br>Statement | The proposed device has same classification information as the predicate device. | | | SAME | | Intended Use | The Dental Electrical Motor iRoot<br>Pro is a cordless endodontic treatment<br>motorized handpiece with root canal<br>measurement capability. It can be<br>used to enlarge canals while<br>monitoring the position of the file tip<br>inside the canal. It can be used as a<br>low-speed motorized handpiece and<br>device for measuring canal length. | The Tri Auto ZX2 device is a cordless<br>endodontic treatment motorized handpiece<br>with root canal measurement capability. It<br>can be used to enlarge canals while<br>monitoring the position of the file tip inside<br>the canal. It can be used as a low speed<br>motorized handpiece and device for<br>measuring canal length. | The EMS-200 is a dental device which<br>combines in a single LCD unit an endo<br>motor which ablates the tooth to expand<br>the root canal, a dental obturator to fill<br>and pressurize various shaped packing<br>elements and an electronic apex locator<br>which assists the operator the location of<br>the front tip in the root canal, for use by<br>trained dental professionals. | SAME | | Usage | Prescription Use | Prescription Use | Prescription Use | SAME | | Comparison<br>Statement | The proposed device has same intended use as the predicate device. | | | SAME | | Technical Specifications | | | | | | Energy used and/or | Li-ion battery (DC 3.7V) | Li-ion battery (DC 3.7V) | DC 12 V | SAME | | delivered | | | | | | Dimension | 280 x 25 x 26mm(central unit include<br>contra angle)<br>123 x 61 x81mm (battery charger) | 30 mm x 30mm x 200 mm<br>Charger: 85 mm x 85mm x 75 mm | 148.1 mm x 187.6 mm x 45 mm | SIMILAR | | Speed | 100-1000 rpm | 100-1000 rpm | 250-800 rpm | SAME | | Gear Ration | 16:1 | Not Available | 16:1 | SAME | | Torque | 0.1-4.0 N.cm | 4 N.cm | 0.6-5.0 N.cm | SAME | | Accuracy of the root<br>apex locator function | -1.5mm to+0.5mm for Apex position | -1.5mm to+0.5mm for Apex position | -0.5mm to +0.5mm | SAME | | Spray Nozzle | Spray nozzle | Spray nozzle | Not Available | SAME | | Sterility | Lip clip, and<br>contra angle are autoclavable.<br><br>The central unit and lip hook cable are<br>covered with an FDA cleared barrier<br>sleeve and intermediate level<br>disinfected. | File holder, contrary electrode, and contra<br>Angle are autoclavable. | Not Available | SAME | | Feature | Canal measurement, alarm, canal<br>length indicator | Canal measurement, alarm, canal length<br>indicator | Canal measurement | SAME | {8}------------------------------------------------ {9}------------------------------------------------ | Operation Mode | <b>R</b> | Mode: Only apex locator<br>measurement function enabled. | EMR mode: This mode is for canal<br>measurement, the motor does not run in this<br>mode. | Not Available | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------| | | <b>M</b> | Mode: Only endo motor function<br>enabled; | CW mode: The motor rotates for forward<br>360, torque reverse and other functions can<br>be used. | SAME | | | <b>MR</b> | Mode: Endo motor and apex<br>locator work independently, when the<br>file reaching the apical, the apex<br>locator only plays the role of warning<br>and display, will not interfere with the | OGP mode: The OGP (Optimum Glide Path)<br>function is used for canal negotiation and<br>making the glide path. | | | | | | OTR mode: The OTR (Optimum Torque<br>Reverse) function is used for canal shaping. | | | | | | CCW Mode: The motor rotates | | | | | | | | | | | | | | | | rotation of endo motor, the motor will<br>not stop and reverse when the file<br>reach the apical.<br><br>Image: MR logo<br>Mode: In this mode, the motor<br>will automatically rotate, stop, reverse<br>depends on the length of the root<br>canal measured by the apex locator. | counterclockwise direction only. | | | | Principle of operation | Electrical motor drives the rotating of<br>file equipped on the rotating to<br>accomplish it's indication for use.<br>And the software control the<br>parameter and method of rotating. | Electrical motor drives the rotating of file<br>equipped on the rotating to accomplish it's<br>indication for use. And the software control<br>the parameter and method of rotating. | The Motor rotates the motor-operated<br>file(Ni-Ti file) by pressing the operation<br>button on the micro-motor handpiece<br>which expands or shapes the root canal by<br>using the rotating power of the electric file<br>(Ni-Ti file).<br>Micro signals consisting of dual<br>frequencies coming from the main unit<br>return to where they are sent after<br>travelling along the electric circuit that is<br>composed of 'main unit - probe cord - file<br>holder-file - patient - lip holder -probe<br>cord'.<br>Pack handpiece provides instantaneous<br>heating and cooling of the heat plugger<br>with precisely controlled temperature and<br>timing.<br>FILL handpiece is designed to inject | SAME | {10}------------------------------------------------ {11}------------------------------------------------ | Patient Contacting<br>Materials | Silica rubber and 304 stainless steel | Not Available | warmed Gutta percha that is specially<br>formulated into the root canal directly.<br>Silica rubber, stainless steel | Difference | |---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|------------| | Comparison<br>Statement: | The proposed device has the same main specifications with the predicate device but only minor difference in dimension, which caused by<br>different appearance design and do not effect the performance.<br>The Patient contacting materials of predicate device could not be identified, so it considered as difference. For this difference, the<br>biocompatibility tests as ISO 10993 series standards have been conducted, the test results shown that the patient contacting materials of<br>proposed could not causes safety concerns. | | | | | Applied Standards: | | | | | | Biocompatibility | ISO10993-5&ISO10993-10&ISO<br>10993-11 | ISO10993-5&ISO10993-10&ISO 10993-11 | ISO10993-10 | SAME | | | Electrical Safety | IEC60601-1 | IEC60601-1 | IEC60601-1 | | EMC | IEC60601-1-2 | IEC60601-1-2 | IEC60601-1-2 | SAME | | Performance | IEC 80601-2-60<br>ISO 14457 | ISO 14457 | ISO 14457 | SAME | | | Comparison<br>Statement | The proposed probe has same applied Standards with the predicate device. | | | {12}------------------------------------------------ 9. Substantially Equivalent (SE) Conclusion The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...