The Endo Motor, model: ZR-Rap, is a cordless endodontic treatment motorized handpiece with root canal measuring capability. It can be used to enlarge canals while monitoring the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length. This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.
Device Story
Cordless, battery-driven endodontic handpiece; integrates motor for canal enlargement and apex locator for length measurement. Input: electrical signals from file tip within root canal. Operation: motor rotates dental file at controlled speeds/torques; apex locator monitors file position via impedance/electrical sensing. Output: visual feedback of file position and motorized canal preparation. Used in dental clinics/hospitals by qualified dental personnel. Healthcare provider uses visual output to guide file depth and canal enlargement, preventing over-instrumentation and ensuring accurate canal preparation. Benefits: improved precision in root canal procedures and reduced risk of apical perforation.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via comparative root canal measurement tests, biocompatibility testing (ISO 10993-5, 10993-10), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and dental handpiece standards (ISO 14457).
Technological Characteristics
Battery-powered (Li-ion 3.7V) cordless handpiece. Materials conform to ISO 10993. Sensing principle: electrical impedance for apex location. Connectivity: none. Sterilization: autoclavable components (contra angle, lip hook, file clip, protective silicon cover). Software: moderate level of concern.
Indications for Use
Indicated for adult patients requiring root canal enlargement and preparation, and root canal length measurement. Contraindications not specified.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
E-Connect S Endo Motor With Built-In Apex Locator (K201993)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 26, 2023
Shenzhen Perfect Medical Instruments Co., Ltd. % Jarvis Wu Consultant Shanghai Sungo Management Consulting Company Limited 14th floor, 1500# Century Ave., Shanghai, Shanghai 200122 CHINA
Re: K222548
Trade/Device Name: Endo Motor Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EKX, LOY Dated: January 30, 2023 Received: January 30, 2023
Dear Jarvis Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222548
Device Name Endo Motor
#### Indications for Use (Describe)
The Endo Motor, model: ZR-Rap, is a cordless endodontic treatment motorized handpiece with root canal measuring capability. It can be used to enlarge canals while monitoring the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length.
This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary K222548
Document Date Prepared:2022/8/12
#### A. Applicant:
Name: Shenzhen Perfect Medical Instruments Co., Ltd.
Address: Room 102, Building A, Ruiji Factory, No. 3 Zaohekeng, Jixia Community, Nanwan Street,
Longgang District.Shenzhen, P.R. China. 518100
Contact Person: Kristy Mo
Tel: +86- 13929326975
Mail: kristy@medtung.com
Submission Correspondent:
Primary contact: Mr. Jarvis Wu
Shanghai SUNGO Management Consulting Co., Ltd.
14th floor, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-58817802
Email: haiyu.wang@sungoglobal.com
Secondary contact: Mr. Raymond Luo
14th floor, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-68828050
Email: fda.sungo@gmail.com
## B. Device:
Proprietary Name: Endo Motor Common Name: Endodontic treatment motorized handpiece with built-in apex locator Model(s): ZR-Rap
#### Regulatory Information
Classification Name: Dental handpiece and accessories Classification: Class I Primary Product code: EKX Secondary Product Code: LQY Regulation Number: 878. 4200
{4}------------------------------------------------
## C. Predicate device:
Manufacturer: Changzhou Sifary Medical Technology Co., Ltd. Device name: E-Connect S Endo Motor With Built-In Apex Locator 510(K) Number: K201993
(Note: Predicate device has NOT been subject to any Medical Device Recalls, including design-related recall.)
## D. Indications use of the device:
The Endo Motor, model: ZR-Rap, is a cordless endodontic treatment motorized handpiece with root canal measuring capability. It can be used to enlarge canals while monitoring the position of the file tip inside the canal. It can be used as a low-speed motorized handpiece and device for measuring canal length.
This device must only be used in hospital environments, clinics or dental offices by qualified dental personnel.
### E. Device Description:
The Endo Motor, model: ZR-Rap, is a battery-driven handpiece with a motor, equipped with a chuck for holding rotary instruments such as a dental file. The Endo Motor can be used for enlargement and preparation of root canals and can also be used as an apex locator.
### F. Comparison with predicate device
Table 1 General Comparison
| Device | Proposed Device | Predicate device | Comparison |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Shenzhen Perfect Medical<br>Instruments Co., Ltd. | Changzhou Sifary<br>Medical Technology Co., Ltd. | - |
| 510K number | K222548 | K201993 | - |
| Device name | Endo Motor | E-connect S Endo Motor | - |
| Classification<br>Regulation | 21CRF 872.4200 | 21CRF 872.4200 | Same |
| Classification | Class I | Class I | Same |
| Product Code | EKX, LQY | EKX, LQY | Same |
| Common name | Endodontic treatment motorized<br>handpiece with built-in apex<br>locator | Endodontic treatment motorized<br>handpiece with built-in apex<br>locator | Same |
| Indications for use | The Endo Motor, model: ZR-Rap,<br>is a cordless endodontic treatment<br>motorized handpiece with root<br>canal measuring capability. It can<br>be used to enlarge canals while<br>monitoring the position of the file<br>tip inside the canal. It can be used<br>as a low-speed motorized<br>handpiece and device for<br>measuring canal length. This<br>device must only be used in<br>hospital environments, clinics or<br>dental offices by qualified dental<br>personnel. | E-connect S is a cordless<br>endodontic treatment motorized<br>handpiece with root canal<br>measuring capability. It can be<br>used to enlarge canals while<br>monitoring the position of the file<br>tip inside the canal. It can be used<br>as a low-speed motorized<br>handpiece and device for<br>measuring canal length. This<br>device must only be used in<br>hospital environments, clinics or<br>dental offices by qualified dental<br>personnel. | Same |
| Patient populations | Adult | Adult | Same |
| Anatomical sites | Root canal, softened<br>dentin | Root canal, softened<br>dentin | Same |
| Where used | Dental clinic, University<br>hospital and the other<br>clinical settings | Dental clinic, University<br>hospital and the other<br>clinical settings | Same |
| Energy used and/or<br>delivered | Li-ion battery (DC 3.7V) | Li-ion battery (DC 3.7V) | Same |
| Exterior Design | 28.2cm × 19.0cm × 6.2cm | 21.5cm × 17.5cm × 9cm | Similar |
| Performance 1 -<br>canal enlargement | 150rpm-650rpm<br>0.6N.cm-4N.cm | 120-1000 rpm<br>0.5N.cm-4N.cm | Different |
| Performance 2 -<br>apex locator | Accuracy of the root apex<br>locator function: -0.5mm | Accuracy of the root apex<br>locator function: -0.5mm | Same |
| | to +0.5mm for Apex position. | to +0.5mm for Apex position. | |
| Materials | Used materials conform | Used materials conform | Same |
| Biocompatibility | to ISO10993. | to ISO10993. | Same |
| Spray Nozzle | Spray nozzle (Oil injector) | Spray nozzle | Same |
| Compatibility<br>with environment<br>and other devices | Conform to IEC60601-1-2 | Conform to IEC60601-1-2 | Same |
| Sterility | Contra Angle, Lip Hook,<br>File clip, Protective silicon cove | Contra Angle, Lip Hook,<br>File clip, Insulating<br>Sleeve autoclavable. | Same |
| Electrical safety | Conform to IEC60601-1 | Conform to IEC60601-1 | Same |
| Mechanical safety | Conform to IEC60601-1 | Conform to IEC60601-1 | Same |
| Thermal safety | Conform to IEC60601-1 | Conform to IEC60601-1 | Same |
| Radiation safety | Conform to IEC60601-1-2 | Conform to IEC60601-1-2 | Same |
{5}------------------------------------------------
Shenzhen Perfect Medical Instruments Co., Ltd. Room 102, Building A, Ruiji Factory, No. 3 Zaohekeng, Jixia Community, Nanwan Street, Longgang District,Shenzhen, P.R. China. 518100
{6}------------------------------------------------
### Different Analysis:
The rotation speed of Endo Motor is within the speed range of the predicate device, and it has passed performance test, no affect on safety or efficacy.
### G. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies with the following standards and FDA Guidance:
- A ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- A IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance
- A IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests
- A IEC 80601-2-60: 2019 Medical Electrical Equipment - Part 2-60: Particular Requirements for Basic
{7}------------------------------------------------
Safety and Essential Performance of Dental Equipment.
- > ISO 14457: 2017 Dentistry - Handpieces and motors
- > Moderate level of software documentation per the FDA Guidance for Software Contained in Medical Devices.
- A Reprocessing validation (i.e., cleaning, disinfection, and sterilization) per the FDA Guidance Document for Reprocessing Medical Devices in Healthcare Setting.
- A Comparative Root Canal Measurement Performance Test to evaluate the root canal length
## H. Clinical Test Conclusion
No clinical study is included in this submission.
### I. Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Endo Motor (Model: ZR-Rap) is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K201993.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.