MEGA ISQ

K132401 · Osstell AB · EKX · Apr 9, 2014 · Dental

Device Facts

Record IDK132401
Device NameMEGA ISQ
ApplicantOsstell AB
Product CodeEKX · Dental
Decision DateApr 9, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.4200
Device ClassClass 1

Intended Use

The MEGA ISQ is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.

Device Story

The MEGA ISQ is a portable, handheld dental instrument used to assess implant stability via Resonance Frequency Analysis (RFA). The system utilizes a Smartpeg (aluminum rod with a magnet) screwed onto a dental implant. A measurement probe on the handheld instrument emits a magnetic pulse to excite the Smartpeg, causing it to vibrate at its resonance frequency. The instrument detects the response signal, calculates the resonance frequency, and displays the result as an Implant Stability Quotient (ISQ) value (0-100). The ISQ value, proportional to implant stiffness, assists clinicians in evaluating stability. The device is used in clinical settings; it does not store data. The MEGA ISQ is a simplified version of the predicate, featuring an LED display and a docking station for USB connectivity, while removing internal RAM and real-time clock functions.

Clinical Evidence

No clinical data was required to support the changes to the Osstell ISQ. Substantial equivalence was established through bench testing, including comparative performance testing against the predicate, material biocompatibility, and conformance to EN IEC 60601-1-2.

Technological Characteristics

Portable handheld instrument; Resonance Frequency Analysis (RFA) principle; LED display (58mm x 37mm); ABS plastic and Dryflex TPE materials; rechargeable power source; USB interface via docking station; sterilization per ANSI/AAMI/ST79:2010/A2:2011; no internal data storage.

Indications for Use

Indicated for measuring dental implant stability in the oral cavity and maxillofacial region.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Applicant/ # Section 6: 510(k) Summary ## APR 0 9 2014 | Manufacturer: | Osstell AB | |-----------------------------------------------|-------------------------------------------| | | Stampgatan 14 | | | 411 01 Göteborg | | | Sweden | | Establishment Registration Number: 3004070020 | | | US Contact: | Cherita James | | | M Squared Associates, Inc. | | | 901 King Street, Suite101 | | | Alexandria, VA 22314 | | | Ph: 703-562-9800 Ext 257 | | | Fax: 703-562-9797 | | | E-mail: CJames@MSquaredAssociates.com | | Date prepared: | 2014-04-07 | | Proprietary Name: | MEGA ISQ | | Common Name: | Dental implant stability analyzer | | Classification Status: | Class I | | Product Codes: | EKX - handpiece, direct drive, ac-powered | | Predicate Device: | Osstell ISQ (K082523) | Karin Breding #### Device Description: The MEGA ISQ is an new generation of the Osstell ISQ, (K082523) the system is designed to measure dental implant stability in the oral cavity and maxillofacial region. Similar to K082523, the MEGA ISQ is a portable, handheld instrument that involves the use of the noninvasive technique, Resonance Frequency Analysis. The system involves the use of a Smartpeg (aluminum rod) attached to the dental implant by means of a screw. The Smartpeg is excited by a magnetic pulse from the measurement probe on the handheld instrument. The resonance frequency, which is the measure of implant stability, is calculated from the response signal. Results are displayed as the Implant Stability Quotient (ISQ). The ISQ is a measurement of the stability of the implant and is derived from the resonance frequency value obtained from the Smartpeg. {1}------------------------------------------------ #### Indication for Use: The MEGA ISO is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region. ### Summary of Technological Characteristics: The modifications to the Osstell ISQ since its previous clearance in K082523 include the following changes: - Replacing the existing display with a LED display . - . Removing the RAM memory - Removing the real time clock . - Move the USB interface to a dock station . - A new material is added to the instrument and docking station to improve grip ability . These minor differences do not affect the safety or performance of the device and do not change the intended use of the MEGA ISQ. These changes were implemented to address the needs of the customer for a basic device which offers the same Resonance Frequency Analysis technology, but does not offer the capability of data storage. The relocation of the USB interface allows a reduce size, as well as allowing for a slightly larger display. #### Summary of Nonclinical Testing: Based on the Risk Analysis, the verification and validation tests that were performed and the acceptance criteria applied included: - . Conformance to EN IEC 60601-1-2 - . Comparative performance testing to predicate - Material biocompatibility . - . Updated steam sterilization parameters in accordance with ``` ANSI/AAMI/ST79:2010/A2:2011 ``` The proposed device passed all applicable test per IEC 60601-1-2, comparative testing of the new LED display, performance of the USB interface, and performance and biocompatibility of the new material demonstrated equivalent performance to the declared predicate. #### Clinical Studies: Clinical data was not required to support the changes to the Osstell ISQ. {2}------------------------------------------------ ## Substantial Equivalence Discussion: The modified MEGA ISQ has the following similarities to the Osstell ISQ previously cleared in K082523: - has the same indicated use, . - . uses the same operating principle, - incorporates the same basic device design and physical properties, . - has no change in autoclave or single use components, or device software . The modified MEGA ISQ has the following differences to the Osstell ISQ previously cleared in K082523: - . Replaces the existing display with a LED display - . Removes the RAM memory - . Removes the real time clock - . Moves the USB interface to a dock station - A new material is added to the instrument and docking station to improve grip ability . | | MEGA ISQ | Predicate Device: Osstell ISQ<br>K082523 | |------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indication for use | The MEGA ISQ is indicated for use in measuring<br>the stability of implants in the oral cavity and<br>maxillofacial region. | No change in Indication for Use | | Operation of<br>System | The MEGA ISQ measures the frequency response<br>from Smartpeg that is directly attached to the<br>implant or abutment. The system includes the<br>following components: Instrument,<br>Smartpeg/Measurement Probe, and PC Data<br>Manager Software.<br><br>The technique involves a smartPeg (10 mm x 3<br>mm) that is attached to the implant or abutment.<br>The SmartPeg is excited over a range of<br>frequencies (1 kHz to 10 kHz) and the resonance<br>frequency is measured with the MEGA ISQ<br>instrument and software. The resonance<br>frequency is determined by the stiffness of the<br>implant system. The MEGA ISQ presents the<br>resonance frequency as an Implant Stability<br>Quotient (ISQ) value (scaled 0-100). The ISQ<br>value is proportional to the stability of the implant. | No change in operation of system | | System<br>Components- | Instrument | Instrument | | | The Instrument is a portable , handheld instrument | The Instrument is a compact unit | | | | Predicate Device: Osstell I:<br>K082523 | | | with a built-in LED display. The unit operates<br>from a rechargeable power source offering over<br>1.5 hours of continuous use between charges. The<br>Measurement Probe, connected to the instrument,<br>is held close to the Smartpeg. The measurement<br>probe sends the excitation signal to the coil in the<br>probe, and also detects the response signal from<br>the second coil in the probe. The microcontroller<br>in the instrument calculates the frequency of the<br>response signal, and presents it on the display as a<br>number, the Implant Stability Index (ISQ).<br>The instrument can be connected to a PC via the<br>USB cable and firmware can be updated.<br>Smartpeg/Measurement Probe | with built-in graphical display.<br>The unit operates from a<br>rechargeable power source<br>offering over 6 hours of<br>continuous use between charges.<br>The Measurement Probe,<br>connected to the instrument, is<br>held close to the Smartpeg. The<br>measurement probe sends the<br>excitation signal to the coil in the<br>probe, and also detects the<br>response signal from the second<br>coil in the probe. The<br>microcontroller in the instrument<br>calculates the frequency of the<br>response signal, and presents it on | | | The stability of the implant is reflected by the<br>resonance frequency of a "Smartpeg" attached to<br>the implant. The Smartpeg is a small aluminum<br>rod, approximately 3 mm in diameter and 10 mm<br>long, with a magnet permanently attached to its<br>top. The Smartpeg is screwed onto the implant.<br>The Smartpeg magnet is excited by a small<br>magnetic pulse generated by a coil in the<br>measurement probe. The Smartpeg vibrates freely<br>at its resonance frequency for some milliseconds.<br>Since the magnet attached to its top is vibrating<br>together with the Smartpeg, the vibration (the<br>"ringing") can be picked up by a second coil in the<br>measurement probe. | the display as a number, the<br>Implant Stability Index (ISQ).<br>Four hundred measurements may<br>be stored in the instrument.<br>The instrument can be connected<br>to a PC via the USB cable and<br>the measurement data can be<br>transferred to the optional ISQ<br>Data Manager Software.<br>Smartpeg/Measurement Probe<br>No change to Smartpeg or<br>Measurement probe | | | PC Data Manager Software<br>Data transfer is not an option for the MEGA ISQ. | PC Data Manager Software<br>The Osstell ISQ Data Manager is<br>a Windows 2000/NT/XP/Vista<br>based software enabling storage,<br>viewing and printing of patient<br>data.<br>The Software is an optional<br>accessory to the Osstell ISQ and<br>is not integral to the functioning | | Instrument and<br>Docking Station<br>materials | ABS Plastic and Dryflex TPE | of the device.<br>ABS Plastic<br>LCD - 64mm x 32mm | | Device Display | LED -58mm x 37mm | | {3}------------------------------------------------ · , : : . . {4}------------------------------------------------ ### Conclusion: The change to display, RAM memory, Clock ,USB interface, and additional device material of the MEGA ISQ do not change the intended use nor do they affect the safety and effectiveness as compared to the Osstell ISQ previously cleared in K082523. Therefore the MEGA ISQ can be found substantially equivalent to the Osstell ISQ cleared in K082523. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The logo is printed in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 9, 2014 Osstell AB C/O Ms. Cherita James Regulatory Consultant M Squared Associates, Incorporated 901 King Street, Suite 101 Alexandria, VA 22314 DEPARTMENT OF HEALTH & HUMAN SERVICES Re: K132401 Trade/Device Name: MEGA ISQ Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: March 14, 2014 Received: March 18, 2014 Dear Ms. James: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Ms. James Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address Sincerely vours. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Image /page/6/Picture/7 description: The image shows a logo or stamp with the text "Tejashri Purohit-Sheth, M.D." and "Clinical Deputy Director". Below that is the word "DAGRID". The image also contains some decorative elements, including a stylized design on the left and a triangular shape on the right. The word "FOR" is at the bottom right. Erin 1. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## Section 5: Indications for Use Statement 510(k) Number (if known): _K132401 MEGA ISQ Device Name: Indications For Use: The MEGA ISQ is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR · Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/7/Picture/11 description: The image shows the text "Sheena A. Green-S" on the first line and "2014.04.07 14:17:46 -04'00" on the second line. The text is written in a clear, sans-serif font. The date and time are separated by colons and periods, and the time zone offset is indicated at the end. Page 1 of __ 1_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...