KLS Martin Individual Patient Solutions

K191028 · KLS-Martin L.P. · JEY · Nov 22, 2019 · Dental

Device Facts

Record IDK191028
Device NameKLS Martin Individual Patient Solutions
ApplicantKLS-Martin L.P.
Product CodeJEY · Dental
Decision DateNov 22, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4760
Device ClassClass 2
AttributesTherapeutic

Intended Use

KLS Martin Individual Patient Solutions implant devices are intended for use in the stabilization, fixation, and reconstruction of the maxillofacial / midface and mandibular skeletal regions.

Device Story

Device consists of patient-specific metallic bone plates and screws for internal fixation in maxillofacial, midface, and mandibular surgery. Input: patient CT scan data. Process: physician collaborates with KLS Martin engineers using COTS software (Materialise Mimics, Geomagic Freeform Plus, IPS CaseDesigner) to virtually plan and design patient-specific implants; design verified by virtual fitting on 3D anatomical models. Output: custom-manufactured titanium plates (subtractive or additive/SLM). Used in clinical settings by surgeons. Benefits: provides patient-specific anatomical fit for reconstruction, stabilization, and aesthetic symmetry.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including mechanical tensile and bending testing per ASTM F382, biocompatibility evaluation per ISO 10993-1, steam sterilization validation per ISO 17665-1, and pyrogenicity testing per AAMI ANSI ST72.

Technological Characteristics

Materials: CP Titanium (ASTM F67) or Ti-6Al-4V (ASTM F136). Manufacturing: Subtractive (milling) or Additive (Selective Laser Melting). Plate thickness: 0.3 mm - 10 mm. Fixation: Titanium screws (1.5 mm - 3.2 mm diameter). Sterilization: Non-sterile, steam. Software: COTS-based planning system. Connectivity: Standalone/networked for image transfer.

Indications for Use

Indicated for stabilization, fixation, and reconstruction of maxillofacial, midface, and mandibular skeletal regions in patients requiring surgical intervention for bone resection, trauma, defects, atrophy, or aesthetic augmentation.

Regulatory Classification

Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. November 22, 2019 KLS-Martin L.P. Katie Rutland Senior Regulatory Affairs Specialist 11201 Saint Johns Industrial Parkway S Jacksonville, Florida 32246 Re: K191028 Trade/Device Name: KLS Martin Individual Patient Solutions Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY Dated: October 24, 2019 Received: October 25, 2019 Dear Katie Rutland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K191028 Device Name KLS Martin Individual Patient Solutions #### Indications for Use (Describe) KLS Martin Individual Patient Solutions implant devices are intended for use in the stabilization, fixation, and reconstruction of the maxillofacial / midface and mandibular skeletal regions. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------| | <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. "L.P." is in black and is smaller than the other words in the logo. 11201 Saint Johns Industrial Pkwy S . Jacksonville, FL 32246 Phone: 904-641-7746 · 800-625-1557 · Fax: 904-641-7378 www.klsmartinnorthamerica.com ### K191028 510(k) Summary 21 CFR 807.92 | Submitter: | KLS-Martin L.P.<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246 | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Katie Rutland<br>Senior Regulatory Affairs Specialist<br>Phone: 800-625-1557<br>Email: katie.rutland(@klsmartinusa.com | | Date Prepared: | November 21, 2019 | | Trade Name: | KLS Martin Individual Patient Solutions | | Common Name: | Plate, Bone | | Classification Name: | Bone plate | | Regulatory Class: | II | | Product Code: | JEY | | Regulation Number: | 21 CFR 872.4760 | | Primary Predicate: | KLS Martin Individual Patient Solutions (K163579) | | Reference Devices: | TruMatch CMF Titanium 3D Printed Implant (K173039)<br>KLS Martin Micro Osteosynthesis System (K944565)<br>KLS Martin IPS Planning System (K182789) | ### Device Description: KLS Martin Individual Patient Solutions is comprised of patient-specific models and metallic bone plates used in conjunction with metallic bone screws for internal fixation of maxillofacial / midface and mandibular bones. The devices are manufactured based on medical imaging (CT scan) of the patient's anatomy with input from the physician during virtual planning and prior to finalization and production of the device. The physician only provides input for model manipulation and interactive feedback by viewing digital models of planned outputs that are modified by trained KLS Martin engineers during the planning session. For each design iteration, verification is performed by virtually fitting the generated implant over a 3D model of the patient's anatomy to ensure its dimensional properties allow an adequate fit. Implants are provided non-sterile, range in thickness from 0.3 mm - 10 mm, and are manufactured using traditional (subtractive) methods from either CP Titanium (ASTM F67) {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black. The logo is simple and modern. or Ti-6Al-4V (ASTM F136) materials or additive methods from Ti-6Al-4V. These patient-specific devices are fixated with previously cleared KLS Martin screws. - Indications for Use: KLS Martin Individual Patient Solutions implant devices are intended for use in the stabilization, fixation, and reconstruction of the maxillofacial / midface and mandibular skeletal regions. ### Technological Characteristics & Substantial Equivalence Discussion The intended use of the subject device. KLS Martin Individual Patient Solutions, is the same as the KLS Martin primary predicate device, KLS Martin Individual Patient Solutions (K163579), differing mainly in the skeletal region for which they are intended and technological specifications. The potential impact on substantial equivalence of each technological difference was addressed through risk analysis and verification and validation testing. #### Similarities to Predicate The subject and primary predicate devices are both intended for reconstructive surgery in the facial skeleton and share the same fundamental principles of operation - patient-specific metallic bone plates used in conjunction with metallic bone screws for facial reconstructive surgery. Both the subject and predicate devices use image data obtained from medical scanners, such as a CT scan. They both use validated commercially off-the-shelf (COTS) software applications to transfer patient imaging from a DICOM format to a .STL format and manipulate the images to produce a final design file. The manufacturing materials used in the subject device are identical to those cleared in KLS Martin Individual Patient Solutions (K163579). The design and dimensions of the plates are based on patient-specific data, using identical methods described in K163579, manufactured from either CP Titanium (ASTM F67) or Ti-6Al-4V (ASTM F136) using traditional (subtractive) or additive manufacturing (Selective Laser Melting (SLM)) methods. The specifications for the subject, primary predicate, and reference devices are similar with respect to plate style, width, length, degree of curvature, fixation hole spacing, and number of fixation holes. Mechanical bench testing performed on both a worst-case midface and orbital plate design, demonstrate that the design expansion to include orbital and maxillofacial / midface implants does not impact substantial equivalence. All subject and predicate devices are provided non-sterile and require the end-user to process the devices using validated cleaning and sterilization methods prior to use as recommended in the device labeling. ### Differences from Predicate The indications for use statement for the subject device of this submission differs from the primary predicate in the addition of the maxillofacial / midface skeletal regions. The skeleton region for which the predicate device was cleared includes only the mandibular region. The subject device expands the indications for use by including the maxillofacial / midface skeletal region and add these devices to the KLS Martin Individual Patient Solutions implant devices. The change in skeletal region presented in the indications for use statement of the subject device compared to the primary predicate device does not impact the substantial {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "martin" is in gray. The letters "L.P." are in black and are slightly smaller than the other letters in the logo. equivalence of the subject device. This difference in skeletal region is mitigated by previously cleared reference devices that are intended for the maxillofacial / midface regions. The subject and primary predicate devices use the same validated commercially off-the-shelf (COTS) software applications to transfer patient imaging from a DICOM format to a .STL format, Materialise Mimics and Geomagic® Freeform Plus™. However, the subject device utilizes an additional COTS software for designing orthognathic implants, IPS CaseDesigner. IPS CaseDesigner was cleared under K161634. This software was also utilized and reviewed in the reference device, K182789, in the creation of orthognathic splints. Previously cleared specifications for the predicate plate thicknesses range from 1.0 mm -3.0 mm and are fixated with previously cleared KLS Martin titanium screws ranging in diameter from 2.0 mm - 3.2 mm in lengths from 5 mm - 22 mm. The subject device plates range in thickness from 0.3 mm - 10 mm and are fixated with previously cleared KLS Martin titanium screws ranging in diameter from 1.5 mm - 3.2 mm in lengths from 3.5 mm - 22 mm. The subject device dimensions differ by offering additional device sizes to accommodate cases in orbital and maxillofacial / midface reconstruction. Subject device implants with thicknesses ranging from 0.6 mm - 10 mm will be used in maxillofacial reconstructive surgeries involving volumetric defects. Responsible clinical conditions for volumetric defects include, but are not limited to, bone resection due to tumors or disease, blunt force trauma, facial defects, bone atrophy, and aesthetic augmentation for facial symmetry. The differences in plate thickness between the subject and predicate devices are addressed with performance data and previously cleared reference devices with plate dimensions that align with the subject device specifications. ### Reference Devices TruMatch CMF Titanium 3D Printed Implant (K173039) has been included as a reference device to address the technological differences between the primary predicate and subject device. The subject device is expanding the primary predicate anatomical regions from the mandible to the orbital and maxillofacial / midface skeletal regions. Both the reference and subject devices offer volumetric implants that augment and contour in the midface skeletal region to allow facial aesthetics and symmetry reconstruction. The reference device is included for the maxillofacial / midface plate design specifications that fall outside of the primary predicate specifications for volumetric implants in the subject device to aid in facial aesthetics and the creations of symmetry in reconstructive surgeries. TruMatch devices are cleared for orbital plate thicknesses ranging from 0.4 mm - 1.5 mm and midface plate thicknesses ranging from 0.8 mm - 10 mm. The KLS-Martin Mini Osteosynthesis System (K944565) has been included as a reference device to address the technological differences between the primary predicate and subject device. The subject device includes implants for the orbital region with a minimum thickness of 0.3 mm, and implants for the midface region with a minimum thickness of 0.6 mm. Because the primary predicate only includes devices for the mandible region with a minimum thickness of 1.0 mm, and the TruMatch CMF Titanium 3D Printed Implant reference device only includes implants with minimum thicknesses of 0.4 mm and 0.8 mm for the orbital and midface regions, respectively, K944565 reference devices have been included to address the subject device minimum plate thicknesses of 0.3 mm -0.6 mm. The K944565 reference device includes traditionally manufactured (subtractive, milled) titanium plates of various sizes and shapes that range in thickness from 0.3 - 0.6 mm and are intended {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while the word "martin" is in gray. The letters "L.P." are in black and are smaller than the other letters in the logo. for the stabilization of bone in oral-maxillo-craniofacial surgery. Comparative performance testing and additional bench analysis has addressed any minor differences between the subject, primary predicate, and reference devices. The KLS Martin IPS Planning System (K182789) has been included as a reference device to support compatibility with the subject device. A device comparison table of the subject, predicate, and reference devices is provided below. {7}------------------------------------------------ ## kts martın LP. | | KLS Martin Individual<br>Patient Solutions | KLS Martin Individual<br>Patient Solutions<br>K163579 | TruMatch CMF<br>Titanium 3D Printed<br>Implant<br>K173039 | KLS Martin Micro<br>Osteosynthesis System<br>(1.5mm)<br>K944565 | KLS Martin IPS<br>Planning System<br>K182789 | |----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | | Indications for Use | KLS Martin Individual<br>Patient Solutions implant<br>devices are intended for<br>use in the stabilization,<br>fixation, and reconstruction<br>of the maxillofacial /<br>midface and mandibular<br>skeletal regions. | KLS Martin Individual<br>Patient Solutions implant<br>devices are intended for<br>use in the stabilization and<br>fixation of mandibular<br>fractures and mandibular<br>reconstruction. | The TruMatch CMF<br>Titanium 3D Printed<br>Implant is a patient specific<br>implant and is intended for<br>bone fixation and<br>reconstruction, restoration<br>of bone defects and<br>intended to provide<br>continuity in regions where<br>the bone is missing and/or<br>to augment the bone by<br>means of an onlay device<br>in the maxillofacial<br>skeleton, midface and chin. | The KLS Martin Micro<br>Osteosynthesis System<br>(1.5mm) is used in<br>oral-maxillo-cranio-facial<br>surgery to stabilize<br>fractured bone structures.<br>The bone segments are<br>attached to the plate with<br>screws to prevent<br>movement of the screws. | The KLS Martin Individual<br>Patient Solutions (IPS)<br>Planning System is<br>intended for use as a<br>software system and image<br>segmentation system for<br>the transfer of imaging<br>information from a medical<br>scanner such as a CT based<br>system. The input data file<br>is processed by the IPS<br>Planning System and the<br>result is an output data file<br>that may then be provided<br>as digital models or used as<br>input to a rapid prototyping<br>portion of the system that<br>produces physical outputs<br>including anatomical<br>models, guides, splints, and<br>case reports for use in<br>maxillofacial surgery. The<br>IPS Planning System is<br>also intended as a<br>pre-operative software tool<br>for simulating/ evaluating<br>surgical treatment options. | | | KLS Martin Individual<br>Patient Solutions | KLS Martin Individual<br>Patient Solutions<br>K163579 | TruMatch CMF<br>Titanium 3D Printed<br>Implant<br>K173039 | KLS Martin Micro<br>Osteosynthesis System<br>(1.5mm)<br>K944565 | KLS Martin IPS<br>Planning System<br>K182789 | | | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | | Patient-specific? | Yes. Devices are<br>manufactured<br>patient-specific, based on a<br>CT scan of the patient. | Yes. Devices are<br>manufactured<br>patient-specific, based on a<br>CT scan of the patient. | Yes. Devices are<br>manufactured<br>patient-specific, based on a<br>CT scan of the patient. | No. | Yes. Devices are<br>manufactured<br>patient-specific, based on a<br>CT scan of the patient. | | Classification | 21 CFR 872.4760.<br>Class II | 21 CFR 872.4760.<br>Class II | 21 CFR 872.4760.<br>Class II | 21 CFR 872.4760,<br>Class II | 21 CFR 872.4120, Class II<br>21 CFR 892.2050, Class II | | Product Code | JEY | JEY | JEY | JEY | DZJ, LLZ | | Material | Anatomical Models:<br>Epoxy/Acrylic Resins<br>Implants: CP Titanium or<br>Ti-6Al-4V | Anatomical Models:<br>Epoxy/Acrylic Resins<br>Implants: CP Titanium or<br>Ti-6Al-4V | Implants: CP Titanium | Implants: CP Titanium &<br>Ti-6Al-4V | Anatomical Models:<br>Epoxy/Acrylic Resins<br>Cutting/Marking Guides:<br>Polyamide, Ti-6Al-4V, CP<br>Titanium<br>Splints: acrylic/methacrylic<br>resins | | Manufacturing<br>Method | Epoxy/Acrylic Resins:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: Traditional<br>and 3D (Additive;<br>Selective Laser Melting) | Epoxy/Acrylic Resins:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive:<br>Selective Laser Melting) | CP Titanium: 3D Additive | Subtractive (traditional) -<br>Milling | Epoxy/Acrylic Resins:<br>Stereolithography (SLA)<br>CP Titanium: Traditional<br>(Subtractive)<br>Ti-6Al-4V: 3D (Additive;<br>Selective Laser Melting)<br>Polyamide: 3D (Additive;<br>Selective Laser Sintering) | | Sterilization | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | | Anatomical Sites | Maxillofacial / Midface &<br>Mandible | Mandible | Maxillofacial / Midface &<br>Chin | Craniomaxillofacial | Maxillofacial / Mandible | | | KLS Martin Individual<br>Patient Solutions | KLS Martin Individual<br>Patient Solutions<br>K163579 | TruMatch CMF<br>Titanium 3D Printed<br>Implant<br>K173039 | KLS Martin Micro<br>Osteosynthesis System<br>(1.5mm)<br>K944565 | KLS Martin IPS<br>Planning System<br>K182789 | | | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | | Plate Specifications | | | | | | | Thickness | • Orbital: 0.3 mm – 1.0 mm<br>• Maxillofacial / midface reconstruction:<br>o 0.6 mm – 10 mm<br>• Mandibular reconstruction: 1.0 mm – 3.0 mm | Mandibular reconstruction:<br>1.0 mm – 3.0 mm | • Orbital: 0.4 mm – 1.5 mm<br>• Midface reconstruction:<br>o 0.8 mm – 10 mm<br>• Mandibular reconstruction: 1.2 mm – 10 mm | 0.3 mm – 0.6 mm | Not applicable | | Style | Non-compression<br>Compression<br>Threaded | Non-compression<br>Compression<br>Threaded | Mesh-Shaped | Standard<br>Non-locking | Not applicable | | Width<br>(screw-hole dependent) | Maxillofacial / midface:<br>Min: ≥ 4.5 mm (around screw holes)<br>Min: ≥ 3 mm (not around screw hole)<br>Max: Dependent on screw-hole<br>Mandibular:<br>Same as predicate | Min: 7 mm<br>Max: 8.5 mm | Unknown | Not applicable | Not applicable | | Length | Maxillofacial / midface:<br>Min: 18 mm<br>Max: 350 mm<br>Mandibular:<br>Same as predicate | Min: 31 mm<br>Max: 320 mm | Min: 10 mm<br>Max: 294 mm | Not applicable | Not applicable | | Degree of curvature<br>(in-plane) | Maxillofacial / midface | Min: 90°<br>Max: 180° | Unknown | Not applicable | Not applicable | | | KLS Martin Individual<br>Patient Solutions | KLS Martin Individual<br>Patient Solutions<br>K163579 | TruMatch CMF<br>Titanium 3D Printed<br>Implant<br>K173039 | KLS Martin Micro<br>Osteosynthesis System<br>(1.5mm)<br>K944565 | KLS Martin IPS<br>Planning System<br>K182789 | | | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | | | Min: 30°<br>Max: 180°<br>Mandibular:<br>Same as predicate | | | | | | Degree of curvature<br>(out-of- plane) | Maxillofacial / midface:<br>Min: 15°<br>Max: 180°<br>Mandibular:<br>Same as predicate | Min: 60°<br>Max: 180° | Unknown | Not applicable | Not applicable | | Hole spacing | Orbital & Maxillofacial /<br>midface:<br>≥4.5 mm<br>Mandibular:<br>Same as predicate | ≥8 mm | Unknown | Not applicable | Not applicable | | Number of Holes | Orbital & Maxillofacial /<br>midface:<br>Min: ≥ 2 per side of defect<br>Max: Dependent on length<br>& hole spacing<br>Mandibular:<br>Same as predicate | Min: 4<br>Max: Dependent on length<br>& hole spacing | Unknown | Not applicable | Not applicable | | Screw Diameter | Orbital<br>1.5 mm<br>Maxillofacial / midface<br>1.5 mm - 2.3 mm<br>Mandibular:<br>Same as predicate | 2.0 mm - 3.2 mm | Unknown | 1.5 mm - 1.8 mm | 1.5 mm - 2.7 mm | | Screw Length | Orbital & Maxillofacial /<br>midface | 5 mm - 22 mm | Unknown | 3.5 mm - 15 mm | 4 mm - 22 mm | | | KLS Martin Individual<br>Patient Solutions | KLS Martin Individual<br>Patient Solutions<br>K163579 | TruMatch CMF<br>Titanium 3D Printed<br>Implant<br>K173039 | KLS Martin Micro<br>Osteosynthesis System<br>(1.5mm)<br>K944565 | KLS Martin IPS<br>Planning System<br>K182789 | | | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | | | 3.5 mm – 22 mm<br>Mandibular:<br>Same as predicate | | | | | | Screw Style | maxDrive & crossDrive<br>(Drill-Free, locking<br>[ThreadLock Taper Screw<br>-TLTS], standard) | maxDrive & crossDrive<br>(Drill-Free, locking<br>[ThreadLock Taper Screw<br>-TLTS], standard) | Unknown | Centre-Drive | maxDrive & crossDrive<br>(Drill-Free, locking<br>[ThreadLock Taper Screw<br>-TLTS]) | {8}------------------------------------------------ # kts martın L.P. {9}------------------------------------------------ # kts martın L.P. {10}------------------------------------------------ # kts martın L.P. {11}------------------------------------------------ ## kls martın LP. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while "marti…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...