The TruMatch CMF Titanium 3D Printed Implant is a patient specific implant and is intended for bone fixation and reconstruction, restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton, midface and chin.
Device Story
Patient-specific implant; manufactured via additive manufacturing using commercially pure titanium; features mesh-like structure for osseointegration. Input: patient CT imaging data. Output: custom-fit titanium implant. Used in plastic and reconstructive surgery; operated by surgeons. Integrated screw holes allow fixation using compatible Synthes screw systems. Guides aid in positioning. Device provides structural continuity or augmentation in maxillofacial regions. Benefits: patient-specific anatomical fit; eliminates manual contouring by surgeon.
Clinical Evidence
Bench testing only. Mechanical performance tested per ASTM F382; results showed equivalent static bending and non-inferior fatigue properties. Sterilization validated per ISO 17665-1, ISO 14161, and ISO 11737-2 to SAL 10^-6. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, chemical characterization, and leachable substances assessment) with all tests passing. Packaging validated per ISTA 2A.
Indicated for bone fixation, reconstruction, restoration of bone defects, and bone augmentation in the maxillofacial skeleton, midface, and chin.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
TruMatch CMF Titanium 3D Printed Implant System (K170272)
{0}------------------------------------------------
July 10, 2018
Materialise NV Lina Ramirez Regulatory Officer Technologielaan 15 Leuven, 3001 BELGIUM
Re: K173039
Trade/Device Name: TruMatch CMF Titanium 3D Printed Implant Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY Dated: June 8, 2018 Received: June 11, 2018
Dear Lina Ramirez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173039
Device Name TruMatch CMF Titanium 3D Printed Implant
#### Indications for Use (Describe)
The TruMatch CMF Titanium 3D Printed Implant is a patient specific implant and is intended for bone fixation and reconstruction, restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton, midface and chin.
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K173039- 510(k) SUMMARY
The following section is included as required by the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92
## Submitter information
| Company name | Materialise N.V. |
|-----------------------------------|-----------------------------------|
| Establishment registration number | 3003998208 |
| Street Address | Technologielaan 15 |
| City | Leuven |
| Zip code | 3001 |
| Country | Belgium |
| Phone number | +32 16 74 49 56 |
| Fax number | +32 16 39 66 00 |
| Principal contact person | Lina Ramirez |
| Contact title | Regulatory Officer |
| Contact e-mail address | regulatory.affairs@materialise.be |
| Additional contact person | Mieke Janssen |
| Contact title | Senior Regulatory Officer |
| Contact e-mail address | mieke.janssen@materialise.be |
## Date of preparation
This summary was last updated on July 9, 2018.
## Submission information
| | TruMatch CMF Titanium 3D Printed Implant |
|------------------------------------------|--------------------------------------------------------------------------------------|
| Trade name(s) | TruMatch CMF Ti 3D-Printed Plates and TruMatch CMF Ti 3D-<br>Printed Surgical Guides |
| | TruMatch CMF Ti 3D-Printed Implants. |
| Common or Usual name | Bone plate |
| Classification name | Bone plate |
| Product code (classification regulation) | JEY |
| Classification Panel | Dental |
| Device class | Class II (21 CFR 872.4760) |
{4}------------------------------------------------
# Predicate device
The predicate device to which substantial equivalence is claimed to:
| TruMatch CMF Titanium 3D Printed Implant System | |
|-------------------------------------------------|-------------------------------------------------|
| Bone plate and screw | |
| Trade or proprietary or model name | TruMatch CMF Titanium 3D Printed Implant System |
| 510(k) number | K170272 |
| Decision date | 08/08/2017 |
| Product code | JEY (21 CFR 872.4760) |
| Manufacturer | Materialise N.V. |
# Reference devices:
| OsteoFab Patient Specific Facial Device - OPSFD | |
|-------------------------------------------------|-----------------------------------------|
| Bone plate and screw | |
| Trade or proprietary or model name | OsteoFab patient Specific facial Device |
| 510(k) number | K133809 |
| Decision date | 07/28/2014 |
| Product code | KKY (21 CFR 878.3500) |
| Manufacturer | Oxford Performance Materials, Inc. |
| Synthes MatrixMANDIBLE Plate and Screw System | |
|-----------------------------------------------|-----------------------------------------------|
| Bone plate and screw | |
| Trade or proprietary or model name | Synthes MatrixMANDIBLE Plate and Screw System |
| 510(k) number | K063790 |
| Decision date | 04/16/2007 |
| Product code | JEY (21 CFR 872.4760) |
| Manufacturer | Synthes (USA) |
{5}------------------------------------------------
| Synthes Craniofacial Plate and Screw System | | | | |
|---------------------------------------------|---------------------------------------------|--|--|--|
| Bone plate and screw | | | | |
| Trade or proprietary or model name | Synthes Craniofacial Plate and Screw System | | | |
| 510(k) number | K080331 | | | |
| Decision date | 04/30/2008 | | | |
| Product code | JEY (21 CFR 872.4760) | | | |
| Manufacturer | Synthes (USA) | | | |
| Synthes Matrixorthognathis Fixation System | |
|--------------------------------------------|--------------------------------------------|
| Bone plate and screw | |
| Trade or proprietary or model name | Synthes Matrixorthognathis Fixation System |
| 510(k) number | K083388 |
| Decision date | 03/12/2009 |
| Product code | JEY (21 CFR 872.4760) |
| Manufacturer | Synthes (USA) |
| The Synthes (USA) Neuro Plate and Screw System | |
|------------------------------------------------|------------------------------------------------|
| Bone plate and screw | |
| Trade or proprietary or model name | The Synthes (USA) Neuro Plate and Screw System |
| 510(k) number | K042365 |
| Decision date | 11/18/2004 |
| Product code | JEY |
| Manufacturer | Synthes (USA) |
| | |
{6}------------------------------------------------
| Synthes Patient Specific Cranial/Craniofacial Implant (PSCI) | |
|--------------------------------------------------------------|--------------------------------------------------------------|
| Plate, Cranioplasty, preformed, non-alterable | |
| Trade or proprietary or model name | Synthes Patient Specific Cranial/Craniofacial Implant (PSCI) |
| 510(k) number | K053199 |
| Decision date | 12/14/2005 |
| Product code | GXN (21 CRF) 882.5330 |
| Manufacturer | Synthes (USA) |
### Device Information
### Indications for use
The TruMatch CMF Titanium 3D Printed Implant is a patient specific implant and is intended for bone fixation and reconstruction, restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton, midface and chin.
### Device Description
The TruMatch CMF Titanium 3D Printed Implant is a patient specific implant and is intended for bone fixation and reconstruction, restoration of bone defects and intended to provide continuity in regions where the bone is missing and/or to augment the bone by means of an onlay device in the maxillofacial skeleton, midface and chin.
The implants feature a mesh-like structure. The mesh-like structures are designed with the same elementary pattern. This pattern was designed to obtain implants with mechanical properties close to those of bone and to allow for osseointegration.
The TruMatch CMF Titanium 3D Printed Implant can be used in combination with TruMatch CMF Titanium 3D Printed Accessories (patient-specific guides), cleared as SurgiCase guides (K103136) and TruMatch CMF Titanium 3D Printed Implant System (K170272). The guides are intended to aid with implant positioning.
The TruMatch CMF Titanium 3D Printed Implant provides surgeons with a patient-specific implant solution for plastic and reconstructive surgery. The device is constructed based on the patient's CT imaging data.
The TruMatch CMF Titanium 3D Printed Implant is designed to fit the patient's anatomy and is not contoured manually by the surgeon. The TruMatch CMF Titanium 3D Printed Implant is designed and manufactured with integrated screw holes to fixate the bone using: MatrixMIDFACE (K050608), MatrixMANDIBLE (K063790, K121574), MatrixORTHOGNATHIC (K083388), MatrixNEURO screws (K123723, K042365), and Synthes Craniofacial Screw System (K050608).
{7}------------------------------------------------
The TruMatch CMF Titanium 3D Printed Implant contains the following applications:
| Reconstruction applications: Orbital | | |
|--------------------------------------|-------------------------------------|----------------------------------------|
| Brand name | TruMatch CMF Ti 3D-Printed Implants | |
| Material | Commercially pure titanium | |
| Type of design | Patient Specific | |
| Type of application | Implant thickness | Patient specific associated instrument |
| Orbital implants | 0.4-1.5mm | Orbital guide |
| Reconstruction applications: Mandible, midface | | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|----------------------------------------|
| Brand name | TruMatch CMF Ti 3D-Printed Implant | | |
| Material | Commercially pure titanium | | |
| Type of design | Patient Specific | | |
| Range of length | 10-294mm | | |
| Curvature | 0°-12°/mm length | | |
| Type of application | Range of shapes | Implant thickness | Patient specific associated instrument |
| Midface reconstruction | Mesh-shaped, contoured to the patient's anatomy One/multi piece One/multi layered | 0.8-10mm | Midface guide |
| Mandibular reconstruction | Mesh-shaped, contoured to patient's anatomy Single/double strut Straight Curved/crescent Subcondylar Plated extensions One/multi layered Combinations of the above | 1.2-10mm | Mandibular guide |
{8}------------------------------------------------
# Comparison to the Predicate Device
Intended use:
| | Subject device:<br>TruMatch CMF 3D Printed Implants<br>(K173039) | Predicate device:<br>TruMatch CMF 3D Printed Implant System<br>(K170272) |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code | JEY | JEY |
| Classification | Class II | Class II |
| Intended Use | The TruMatch CMF Titanium 3D Printed<br>Implant is a patient specific Implant and<br>is intended for bone fixation and<br>reconstruction, restoration of bone<br>defects and intended to provide<br>continuity in regions where the bone is<br>missing and/or to augment the bone by<br>means of an onlay device in the<br>maxillofacial skeleton, midface and chin. | The TruMatch CMF 3D Printed Implant<br>System is intended for bone repositioning,<br>fixation and reconstruction of the<br>maxillofacial skeleton, midface, mandible<br>and chin in adolescents (greater than 12 to<br>21 years of age) and adults.<br><br>Specific indications for use:<br>● Orthognathic surgery.<br>● Reconstructive mandible and<br>maxillofacial surgery.<br>● Mandible and maxillofacial trauma<br>surgery. |
Both the subject device and the predicate device have the same indications for use: bone fixation and reconstruction in the maxillofacial skeleton, midface and chin. Both devices are used in the same anatomical regions. The subject device and the reference device (OsteoFab Patient Specific Facial Device – OPSFD), are both intended for bone augmentation/enhancement.
| | Subject device: | Predicate device: |
|----------------------------|----------------------------------------|----------------------------------------------|
| | TruMatch CMF Titanium 3D Printed | 3D<br>TruMatch CMF<br>Titanium<br>Printed |
| | Implants | Implant System |
| | | |
| | (K173039) | (K170272) |
| Product code | JEY | JEY |
| Classification | Class II | Class II |
| Technical<br>specification | Bone plate | Bone plate |
| Fixation method | Synthes screw system | Synthes screw system |
| Material(s) | Commercially pure titanium | Commercially pure titanium |
| Manufacturing<br>method | Additive manufacturing | Additive manufacturing |
| Provided<br>sterile? | No | No |
| Sterilization<br>method | Moist heat | Moist heat |
| Device<br>thickness | 0.4 mm – 10 mm | 0.8 mm - 3.0 mm |
| Patient-specific | Yes. Devices are manufactured patient- | Yes. Devices are manufactured patient- |
| configuration? | specific, based on a CT scan of the | specific, based on a CT scan of the patient. |
| | patient. | |
{9}------------------------------------------------
Both the subject and the predicate device are manufactured in commercially pure titanium. The subject device and the predicate device are fixed with screws.
Both the subject device and the predicate device are not provided sterile and have the same sterilization method, moist heat.
The subject device and the predicate device use an additive manufacturing technique to manufacture the implants. The subject device and the predicate are manufactured from commercially pure titanium in a mesh-like structure, and can be used with accessories (patientspecific guides).
The subject device and predicate device are made patient specific, manufactured from patient CT scan data.
The thickness range of the subject device falls into the thickness range of the predicate and the reference devices.
### Performance data
The following non-clinical testing was conducted as a basis for the determination of substantial equivalence:
| Test | Test method summary | Results |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanical testing | ASTM F382: Standard<br>Specification and Test Method<br>for Metallic Bone Plates | The results of this test indicate<br>that the subject device has<br>equivalent static bending<br>properties and has non-inferior<br>fatigue bending properties<br>compared to the reference<br>devices. |
| Sterilization testing | Steam sterilization validation<br>according to ISO 17665-1 and<br>ISO 14161, ISO11737-2:2009 | The provided sterilization<br>instructions effectively steam<br>sterilize the subject device to a<br>SAL of 10-6. |
| Environmental conditioning<br>and simulated shipping<br>testing | According to ISTA 2A | The packaging specifications are<br>found to be adequate to protect<br>the device from damage during<br>shipment. |
| Compatibility testing | Combination of user need<br>validation lab and engineering<br>rationale | The subject device is compatible<br>with the Synthes fixation<br>systems. |
{10}------------------------------------------------
| Biocompatibility test overview | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Test/assessment description | Test report conclusion |
| • Cytotoxicity:<br>ISO 10993-5: Tests for in-vitro cytotoxicity -<br>L929 Neutral red uptake Cytotoxicity | - no cytotoxic effect |
| • Sensitization:<br>ISO 10993-10: Tests for irritation and<br>delayed-type hypersensitivity - Kligman<br>Maximization test | - in compliance with requirements of the ISO<br>10993-10 guidelines |
| • Intra-cutaneous reactivity:<br>ISO 10993-10: Tests for irritation and<br>delayed-type hypersensitivity - Intra-<br>cutaneous Injection Test | - in compliance with requirements of the ISO<br>10993-10 guidelines |
| • Systemic toxicity:<br>ISO 10993-11: Tests for systemic toxicity -<br>Systemic injection test | - test passed and is considered negative based<br>on standards set by ISO 10993-11 |
| • Chemical characterization:<br>ISO10993-18: Biological Evaluation of<br>Medical Devices - Part 18: Chemical<br>Characterization of Materials (2005) | - chemical characterization as per report |
| • Assessment of allowable limits for leachable<br>substances:<br>ISO 10993-17 (2009) "Biological evaluation<br>of medical devices — Part 17: Establishment<br>of allowable limits for leachable substances" | - no toxicological concern remains<br>- further biological testing are considered not<br>justified |
The following nonclinical tests were conducted on the TruMatch CMF Titanium 3D Printed Implant as a basis for the determination of substantial equivalence: Performance testing of the TruMatch CMF Titanium 3D Printed implants and accessories, including
- Mechanical performance of the subject device compared to the predicate and reference devices
- Compatibility testing
- . Biocompatibility testing
- Sterilization testing
- . Environmental conditioning and simulated shipping testing
### Conclusion
Non-clinical tests demonstrate that the TruMatch CMF Titanium 3D Printed Implant is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.